Rotronic HL-1D

Rotronic Hygrolog HL-1D Data Logger User Manual

Model: HL-1D | Brand: Rotronic

1. Introduction

The Rotronic Hygrolog HL-1D is a high-precision temperature and humidity data logger designed for reliable and continuous monitoring of environmental conditions. This device is equipped with multilevel security features and is compliant with 21 CFR Part 11 and GAMP 5 regulations, making it ideal for sensitive applications such as vaccine storage, pharmaceutical environments, server rooms, and other critical areas requiring stringent data integrity.

This manual provides essential information for the proper setup, operation, maintenance, and troubleshooting of your HL-1D data logger to ensure optimal performance and longevity.

2. Safety Information

Please read and understand all safety instructions before operating the device. Failure to follow these guidelines may result in damage to the device or personal injury.

3. Product Overview

The Rotronic Hygrolog HL-1D features a clear display and intuitive controls for easy operation. Familiarize yourself with the main components:

Front view of Rotronic Hygrolog HL-1D data logger

Figure 3.1: Front view of the Rotronic Hygrolog HL-1D data logger, showing the display and control buttons.

Front and back view of Rotronic Hygrolog HL-1D data logger

Figure 3.2: Front and back view of the HL-1D, illustrating the USB port and battery compartment on the rear.

4. Setup

4.1. Initial Inspection

Upon unpacking, inspect the device for any visible damage. If any damage is found, contact your supplier immediately.

4.2. Battery Installation

The HL-1D requires one CR2 battery for operation. To install or replace the battery:

  1. Locate the battery compartment cover on the back of the device.
  2. Use a coin or a flat-head screwdriver to gently twist the cover counter-clockwise to unlock it.
  3. Insert the CR2 battery, ensuring the correct polarity (+/-) as indicated inside the compartment.
  4. Replace the cover and twist clockwise to lock it securely.

4.3. Software Installation

To configure the data logger, download data, and analyze measurements, you will need the Rotronic HW4-lite software. This software is essential for full functionality, including setting logging intervals, alarm thresholds, and managing data files.

4.4. Connecting to a Computer

Use the provided USB cable to connect the HL-1D data logger to your computer. The USB port is located on the back of the device. Once connected, the HW4-lite software should detect the device, allowing you to proceed with configuration or data download.

5. Operation

5.1. Power On/Off

The device powers on automatically when a battery is inserted. It remains in a low-power state until logging is initiated or a button is pressed. There is no manual power-off button; the device enters a sleep mode when inactive or logging is stopped.

5.2. Display Modes

Press the MODE button to cycle through different display views:

5.3. Starting and Stopping Data Logging

To begin or end a data logging session:

  1. To Start Logging: Press and hold the START/STOP button for approximately 3 seconds until the display indicates that logging has begun (e.g., a blinking icon or specific message).
  2. To Stop Logging: Press and hold the START/STOP button again for approximately 3 seconds until the display indicates that logging has stopped.

The logging interval can be configured using the HW4-lite software, ranging from 30 seconds to 24 hours. The device has a storage capacity of 64,000 data points.

5.4. LED Indicators

The HL-1D features two LED indicators that provide status information:

6. Data Management

Data recorded by the HL-1D can be downloaded and analyzed using the Rotronic HW4-lite software.

  1. Connect the HL-1D to your computer using the USB cable.
  2. Launch the HW4-lite software.
  3. The software will detect the connected device. Follow the software's instructions to download the logged data.
  4. Once downloaded, you can view, graph, export, and manage your data. The software supports features for 21 CFR Part 11 and GAMP 5 compliance, ensuring data integrity and audit trails.

7. Maintenance

7.1. Cleaning

Clean the exterior of the device with a soft, damp cloth. Do not use abrasive cleaners, solvents, or immerse the device in water. Ensure the ventilation slots are free from dust and debris.

7.2. Storage

When not in use for extended periods, store the HL-1D in a cool, dry place, away from direct sunlight and extreme temperatures. It is recommended to remove the battery if storing for several months to prevent leakage.

7.3. Calibration

For applications requiring high accuracy and compliance, periodic calibration of the HL-1D is recommended. Refer to the Rotronic HW4-lite software for calibration options or contact Rotronic support for professional calibration services.

8. Troubleshooting

If you encounter issues with your HL-1D data logger, refer to the following common problems and solutions:

ProblemPossible CauseSolution
Device does not turn on/display is blank.Dead or incorrectly installed battery.Check battery polarity. Replace with a new CR2 battery.
Cannot connect to computer/software.USB cable issue, driver not installed, software not running.Ensure USB cable is securely connected. Reinstall HW4-lite software and drivers. Try a different USB port.
Data logging not starting.START/STOP button not held long enough, device memory full.Press and hold START/STOP button for 3 seconds. Download data and clear device memory using HW4-lite.
Readings appear inaccurate.Sensor obstruction, device placed in unsuitable environment, calibration needed.Ensure ventilation slots are clear. Move device to a stable environment. Consider professional calibration.
Red LED flashing.Alarm limit exceeded or low battery.Check current readings against set alarms. Replace battery if low.

9. Specifications

Detailed technical specifications for the Rotronic Hygrolog HL-1D:

Technical specifications sheet for Rotronic HL-1D, BL-1D, TL-1D data loggers

Figure 9.1: Technical data sheet for the Rotronic HL-1D, including measurement ranges, accuracy, and other features.

ParameterValue
Measurement MethodROTRONIC HYGROMER® IN-1
Humidity Measurement Range0...100 %RH (non-condensing)
Humidity Accuracy (23 °C ±5 °C)±3.0 %RH
Humidity Resolution0.1 %RH
Temperature Measurement MethodNTC
Temperature Measurement Range-30...70 °C
Temperature Accuracy (23 °C ±5 °C)±0.3 °C
Temperature Resolution0.1 °C
Logging Interval30 s...24 h (configurable)
Storage Capacity64,000 measured values
Display Refresh Rate5 sec. (standard)
LED Indicators2 LEDs (Green: logging, Red: alarm/low battery)
Communication SoftwareHW4-lite (V3.9 or later)
Battery Type1 x CR2
Battery Life~3 years (at 10 min logging interval, room temperature)
IP Protection ClassIP67
Weight85 g (3 ounces)
Dimensions (W x H x D)60 x 90 x 23 mm (2.36"W x 3.54"H)
FDA/GAMP CompatibilityFDA 21 CFR Part 11 / GAMP 5-compliant HW4 software
StandardsEN 12830
AccessoriesMini USB cable AC0003 (contained in HL-1D set)
Dimensions of Rotronic Hygrolog HL-1D data logger

Figure 9.2: Physical dimensions of the Rotronic Hygrolog HL-1D data logger.

10. Warranty and Support

10.1. Warranty Information

Rotronic products are manufactured to the highest quality standards. This product comes with a standard manufacturer's warranty covering defects in materials and workmanship. Please refer to the warranty card included with your product or visit the official Rotronic website for detailed warranty terms and conditions.

10.2. Technical Support

For technical assistance, troubleshooting beyond this manual, or inquiries regarding calibration and service, please contact Rotronic customer support:

Related Documents - HL-1D

Preview Rotronic TL-CC1 Cold Chain Temperature Logger Datasheet and Specifications
Explore the Rotronic TL-CC1, a single-use cold chain temperature logger designed for reliable monitoring of sensitive goods like pharmaceuticals and food. This datasheet details its quick setup, automatic PDF reporting, high storage capacity, manipulation-proof design, and compliance with GxP, EN 12830, FDA 21 CFR Part 11, and GAMP 5 standards. Includes technical specifications and quick start guide.
Preview Rotronic RMS: Compliant Medication Storage Monitoring for Healthcare Facilities
Ensure compliant medication storage in retirement homes and healthcare facilities with Rotronic's RMS (Rotronic Monitoring System). This guide details temperature monitoring, regulatory requirements (GxP, GDP), and the features of Rotronic's data logging solutions for pharmaceutical quality management.
Preview HygroPalm HP31/32/GTS Short Instruction Manual
This manual provides a concise overview of the HygroPalm HP31, HP32, and HP-GTS handheld instruments, covering their general description, device overview, operation, settings, data saving, viewing saved data, alarms, adjustments, and technical specifications. It details how to use the device for measuring and logging temperature and humidity, including software installation, battery management, display functions, and data logging capabilities.
Preview HygroFlex Humidity Temperature Transmitter Instruction Manual v4
Comprehensive instruction manual for the Rotronic HygroFlex series of humidity and temperature transmitters, covering models HygroFlex 1, 2, and 3. Details installation, configuration, operation, and maintenance for industrial and scientific applications.
Preview Rotronic PC-Series Humidity Probes: Short Instruction Manual
A concise guide to the installation, wiring, and technical specifications of Rotronic's PC-Series humidity and temperature probes, covering models PC33, PC52, PC62, PC62V, and PCMini52.
Preview Rotronic HF3A Humidity and Temperature Transmitter Datasheet
Comprehensive technical datasheet for the Rotronic HF3A Humidity and Temperature Transmitter, detailing its features, applications, housing options, technical specifications, and performance characteristics for HVAC and industrial use.