Apera Instruments PH700, LLC-AI501, LLC-AI1109

Apera Instruments PH700 Benchtop Lab pH Meter and Calibration Buffer Solution Kit User Manual

Models: PH700, LLC-AI501, LLC-AI1109

1. Introduction

This user manual provides detailed instructions for the proper setup, operation, maintenance, and troubleshooting of your Apera Instruments PH700 Benchtop Lab pH Meter and the accompanying pH Calibration Buffer Solution Kit. This system is designed for accurate and reliable pH measurements in laboratory environments.

2. Safety Precautions

3. Package Contents

Verify that all items are present and undamaged upon unpacking:

Apera Instruments PH700 Benchtop Lab pH Meter with electrode holder and various pH buffer solutions.

This image displays the complete Apera Instruments PH700 Benchtop Lab pH Meter setup, including the meter unit, electrode, electrode holder, and the accompanying pH calibration buffer solutions (pH 4.00, 7.00, 10.01) and 3M KCL storage solution.

4. Setup

  1. Unpack Components: Carefully remove all components from their packaging.
  2. Assemble Electrode Holder: Attach the electrode holder to the meter unit according to the provided diagram in the quick start guide (if applicable).
  3. Connect Electrode: Connect the 201T-F combination pH electrode to the designated BNC connector on the back of the PH700 meter. Ensure a secure connection.
  4. Connect Power: Plug the power adapter into the meter's power input and then into a standard electrical outlet.
  5. Initial Electrode Preparation: Before first use, remove the protective cap from the electrode. Rinse the electrode thoroughly with distilled water. If the electrode has been stored dry, soak it in 3M KCL storage solution for at least 30 minutes to rehydrate the glass membrane.

5. Operating Instructions

5.1. Calibration

The PH700 meter features advanced auto-calibration with a self-diagnosis function. Regular calibration is crucial for accurate measurements.

  1. Prepare Buffer Solutions: Pour small amounts of pH 4.00, pH 7.00, and pH 10.01 buffer solutions into separate, clean beakers. Ensure the solutions are at room temperature.
  2. Rinse Electrode: Rinse the electrode with distilled water and gently blot dry with a lint-free tissue.
  3. Start Calibration: Turn on the meter. Follow the on-screen prompts or refer to the meter's specific button functions to enter calibration mode.
  4. Calibrate with pH 7.00: Immerse the electrode into the pH 7.00 buffer solution. Stir gently and wait for the reading to stabilize. The meter will automatically recognize the buffer and complete the calibration point.
  5. Calibrate with pH 4.00 or pH 10.01: Rinse the electrode with distilled water. Immerse it into either the pH 4.00 or pH 10.01 buffer solution (depending on your expected sample range). The meter will guide you through the process. For optimal accuracy, perform a 3-point calibration (4.00, 7.00, 10.01).
  6. Exit Calibration: Once all desired points are calibrated, exit calibration mode. The meter is now ready for measurement.
Three bottles of Apera Instruments pH calibration buffer solutions: pH 4.00 (red), pH 7.00 (green), and pH 10.01 (blue).

A close-up view of the three primary pH calibration buffer solutions provided with the kit, color-coded for easy identification and labeled with their respective pH values.

5.2. Measurement

  1. Prepare Sample: Pour your sample solution into a clean beaker.
  2. Rinse Electrode: Rinse the electrode with distilled water and gently blot dry.
  3. Immerse Electrode: Immerse the electrode into the sample solution, ensuring the sensing bulb and reference junction are fully submerged.
  4. Stir and Stabilize: Gently stir the solution to ensure homogeneity and wait for the reading on the meter to stabilize. The PH700's advanced digital processing technology helps generate quick and stable readings.
  5. Record Reading: Once stable, record the pH and temperature readings displayed on the meter.
  6. Clean After Use: After each measurement, rinse the electrode thoroughly with distilled water.

6. Maintenance

6.1. Electrode Care and Storage

Apera Instruments 3M KCL Storage Solution for pH/ORP Electrodes, 4 oz bottle.

This image shows the 4 oz bottle of 3M KCL storage solution, essential for maintaining the pH electrode's performance and extending its lifespan.

6.2. Buffer Solution Storage

6.3. Meter Cleaning

7. Troubleshooting

ProblemPossible CauseSolution
Unstable readings / Drifting pHDirty electrode; Electrode not properly hydrated; Contaminated buffer solutions; Air bubbles on electrode bulb.Clean electrode; Soak electrode in 3M KCL solution; Use fresh buffer solutions; Gently tap electrode to remove bubbles.
Calibration error / Fails to calibrateExpired or contaminated buffer solutions; Damaged electrode; Incorrect calibration procedure.Use fresh, unexpired buffer solutions; Inspect electrode for damage; Review calibration steps in manual.
No display / Meter not turning onPower adapter issue; Loose connection.Check power adapter and cable connections; Try a different power outlet.
Slow response timeElectrode aging; Electrode not clean.Clean electrode; Rehydrate electrode; Consider electrode replacement if issue persists.

8. Specifications

FeatureDescription
pH Accuracy+/- 0.01 pH at 25 °C (traceable to NIST SRMs)
Electrode201T-F 3-in-1 Combination pH Electrode (measures pH and temperature simultaneously)
Temperature CompensationAutomatic Temperature Compensation (ATC)
CalibrationAdvanced Auto. calibration mode with calibration guide and self-diagnosis (1-3 points)
TechnologyAdvanced digital processing technology for quick and stable readings
Buffer SolutionspH 4.00, 7.00, 10.01 (8oz each), color-coded, double-sealed, 2-year expiration from manufacture for unopened bottle
Storage Solution3M KCL Storage Solution (4oz) for pH/ORP Electrodes
ManufacturingProduced in an ISO 9001:2015 certified facility using high-purity reagents and deionized water

9. Warranty and Support

Apera Instruments products are designed for reliability and performance. For specific warranty information, please refer to the warranty card included with your product or visit the official Apera Instruments website. For technical support, troubleshooting assistance, or to inquire about replacement parts, please contact Apera Instruments customer service directly.

Apera Instruments, LLC
Website: www.aperainst.com
Contact Information: Refer to product packaging or website for current contact details.

Related Documents - PH700, LLC-AI501, LLC-AI1109

Preview Apera PH700 Benchtop pH Meter Instruction Manual
Comprehensive instruction manual for the Apera PH700 Benchtop pH Meter, covering features, specifications, operation, calibration, maintenance, and troubleshooting.
Preview Apera PH700 Benchtop pH Meter Instruction Manual
Comprehensive instructions for operating, calibrating, and maintaining the Apera PH700 Benchtop pH Meter. It covers technical specifications, instrument description, measurement procedures for pH and ORP, parameter settings, electrode care, and troubleshooting.
Preview APERA Instruments pH 4.00 Buffer Solution Safety Data Sheet
Safety Data Sheet (SDS) for APERA Instruments pH 4.00 Buffer Solution, providing essential information on identification, hazards, first aid, handling, storage, and regulatory compliance.
Preview PH60Z-HT Smart pH Meter User Manual - Apera Instruments
User manual for the Apera Instruments PH60Z-HT Smart pH Meter, detailing its features, operation, calibration, maintenance, and troubleshooting for high temperature and caustic solutions.
Preview Apera EC20 Pocket Conductivity Tester: Instruction Manual & Guide
Comprehensive instruction manual for the Apera EC20 Pocket Conductivity Tester. Learn about setup, calibration, conductivity measurement, technical specifications, and warranty information.
Preview Apera SX811-SS Portable pH Meter for Food Testing: Instruction Manual & Specifications
Comprehensive instruction manual and technical specifications for the Apera SX811-SS portable pH meter, designed for accurate food testing. Learn about features, calibration, setup, USB communication, and troubleshooting.