kuman P4100
kuman P4100 Universal 100:1 High Voltage Oscilloscope Probe User Manual
Model: P4100
1. Introduction
This user manual provides comprehensive instructions for the safe and effective use of the kuman P4100 Universal 100:1 High Voltage Oscilloscope Probe. Designed for use with various oscilloscopes including Rigol, Atten, Owon, and Siglent, this probe allows for accurate measurement of high voltage signals up to 2000V (Vp-p) with a 100:1 attenuation ratio and a bandwidth of 100 MHz. Please read this manual thoroughly before operating the device.
2. Safety Information
WARNING: This probe is designed for high voltage measurements. Improper use can result in electric shock, injury, or damage to equipment. Always adhere to the following safety precautions:
- Ensure the oscilloscope and the probe are properly grounded before making any connections.
- Do not exceed the maximum input voltage of 2000V (Vp-p).
- Always connect the probe's ground lead to the circuit's ground before connecting the probe tip to the signal source.
- Disconnect the probe tip from the signal source before disconnecting the ground lead.
- Inspect the probe cable and connectors for any damage before each use. Do not use if damaged.
- Use only in dry conditions. Avoid exposure to moisture or extreme temperatures.
- Do not attempt to disassemble or modify the probe. Refer servicing to qualified personnel.
3. Package Contents
Verify that all items listed below are present in your package:
- 1 x kuman P4100 Oscilloscope Probe
- 1 x Ground Lead with Alligator Clip
- 1 x Hook Tip (removable)
- 1 x Adjustment Tool
- 4 x Color Rings (for identification)
- 1 x User Manual (this document)

Figure 3.1: Complete package contents of the kuman P4100 probe.
4. Product Overview and Features
The kuman P4100 is a passive high voltage probe designed for general purpose oscilloscope measurements. Its key features include:
- Universal 100:1 High Voltage Probe: Provides a 100:1 attenuation ratio, allowing oscilloscopes to safely measure higher voltages.
- Wide Compatibility: Usable with oscilloscopes from various manufacturers such as Rigol, Atten, Owon, and Siglent.
- High Bandwidth: Supports measurements up to 100 MHz, suitable for a wide range of signals.
- Robust Design: Constructed with durable ABS material for reliable performance.

Figure 4.1: Overview of the kuman P4100 High Voltage Probe.
5. Setup
5.1 Connecting the Probe to the Oscilloscope
- Ensure your oscilloscope is powered off.
- Connect the BNC connector of the P4100 probe to an input channel on your oscilloscope.
- Attach the ground lead's alligator clip to the ground terminal of the oscilloscope.
- Power on your oscilloscope.
5.2 Probe Compensation Adjustment
Proper probe compensation is crucial for accurate measurements. This adjusts the probe's capacitance to match the oscilloscope's input capacitance, ensuring a flat frequency response.
- Connect the probe tip to the oscilloscope's probe compensation output (usually a square wave signal, often 1kHz, 5Vp-p).
- Connect the probe's ground clip to the oscilloscope's ground terminal.
- Observe the square wave on the oscilloscope screen.
- Using the provided adjustment tool, carefully turn the trimmer screw on the probe body until the square wave corners are flat and sharp, without overshoot or undershoot.

Figure 5.1: Probe tip with retractable hook and BNC connector detail.
6. Operating Instructions
The P4100 probe has a 100:1 attenuation ratio. This means the voltage displayed on the oscilloscope screen will be 1/100th of the actual voltage at the probe tip. Remember to multiply the oscilloscope's reading by 100 to get the true voltage.
- Ensure the probe is correctly connected and compensated as described in the Setup section.
- Always connect the probe's ground clip to the circuit's ground point first.
- Carefully touch the probe tip to the test point where you wish to measure the high voltage signal.
- Observe the waveform on the oscilloscope screen. Adjust the oscilloscope's vertical scale (Volts/Div) and horizontal scale (Time/Div) as needed to get a clear display.
- To calculate the actual voltage, multiply the voltage reading on the oscilloscope by 100. For example, if the oscilloscope reads 10V, the actual voltage is 10V * 100 = 1000V.
- Always disconnect the probe tip from the test point before disconnecting the ground clip.

Figure 6.1: Detail of the probe's ground lead connection.
7. Maintenance
To ensure the longevity and accuracy of your kuman P4100 probe, follow these maintenance guidelines:
- Cleaning: Use a soft, dry cloth to clean the probe body and cable. Do not use abrasive cleaners or solvents. Ensure the probe is disconnected from all power sources before cleaning.
- Storage: Store the probe in a clean, dry environment, away from direct sunlight and extreme temperatures. Keep it in its original packaging or a protective case to prevent damage to the tip and cable.
- Inspection: Regularly inspect the probe cable for cuts, kinks, or insulation damage. Check the connectors for corrosion or bending. If any damage is found, discontinue use and consider replacement.
- Recalibration: Periodically check the probe compensation, especially if you notice discrepancies in measurements or if the probe has been subjected to significant temperature changes.
8. Troubleshooting
If you encounter issues with your kuman P4100 probe, refer to the following common problems and solutions:
| Problem | Possible Cause | Solution |
|---|---|---|
| Distorted waveform (overshoot/undershoot) | Incorrect probe compensation. | Perform probe compensation adjustment (Section 5.2). |
| No signal or very weak signal | Probe not properly connected; oscilloscope settings incorrect; faulty probe. | Check BNC connection and ground lead. Adjust oscilloscope's Volts/Div setting. Test with a known good signal source. |
| Excessive noise in waveform | Poor grounding; electromagnetic interference. | Ensure ground lead is securely connected to circuit ground. Minimize cable length. Move away from strong electromagnetic fields. |
| Inaccurate voltage readings | Incorrect oscilloscope scaling; probe compensation off. | Remember to multiply oscilloscope reading by 100. Re-check probe compensation. |
9. Specifications
| Parameter | Value |
|---|---|
| Attenuation Ratio | 100:1 |
| Input Resistance | 100 MΩ |
| Input Capacitance | 5 pF |
| Compensation Range | 10 pF - 35 pF |
| Bandwidth | 100 MHz |
| Rise Time | 3.5 ns |
| Maximum Operating Voltage | 2000V Working Voltage (Vp-p) |
| Cable Length | 1.1 m |
| Material | ABS |
| Item Weight | 2.39 ounces (approx. 67.7 grams) |
10. Warranty and Support
For warranty information or technical support regarding your kuman P4100 Oscilloscope Probe, please contact kuman customer service through their official website or the retailer from whom you purchased the product. Please have your product model number (P4100) and purchase details ready when contacting support.
You can visit the kuman store for more information: kuman Official Store
Related Documents - P4100
![]() |
RIGOL PHA High-Voltage Differential Probe User Guide User guide for RIGOL PHA0150 and PHA1150 high-voltage differential probes, covering features, safety, usage, specifications, and troubleshooting for accurate high-voltage measurements. |
![]() |
FNIRSI P6100/P4100 Passive Oscilloscope Probe User Manual and Specifications This document provides comprehensive instructions, specifications, maintenance guidelines, and assembly details for the FNIRSI P6100 and P4100 series X1&X10 passive oscilloscope probes, designed for accurate signal measurement. |
![]() |
MICSIG Oscilloscopes and Probes: Comprehensive Product Catalog Explore the full range of MICSIG oscilloscopes and probes, including high-resolution, tablet, modular, and automotive diagnostic models, along with advanced current and isolation probes. Discover cutting-edge technology for precise signal measurement. |
![]() |
Micsig Product Selection Guide: Oscilloscopes and Test Probes Explore the comprehensive range of Micsig oscilloscopes, including high-resolution, tablet, and automotive models, along with various test probes like differential, current, and optical-fiber isolated probes. Find the perfect solution for your electrical testing needs. |
![]() |
Tektronix P5102 100 MHz High-Voltage Probe Instruction Manual Comprehensive instruction manual for the Tektronix P5102 100 MHz High-Voltage Probe, covering safety, operation, adjustment, verification, maintenance, and specifications. Includes detailed guidance for high-voltage measurements. |
![]() |
UNI-T UT-P20 Oscilloscope Probe Kit - Specifications and Safety Detailed specifications, safety instructions, and accessory list for the UNI-T UT-P20 high-voltage oscilloscope probe kit, including its voltage derating curve and compensation adjustment guide. |
Ask a question about this manual
Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.