Oakton PC700
Oakton PC700 Meter User Manual
Model: AO-35413-00
1. Introduction
The Oakton PC700 Meter is a compact benchtop instrument designed for precise measurement of pH, conductivity, and temperature. It features a large, easy-to-read dual display that shows both the measured parameter and temperature simultaneously. The meter incorporates advanced functionalities such as automatic pH buffer recognition for USA and NIST buffer sets, automatic or manual temperature compensation, and the ability to store up to 100 data sets. Its intuitive membrane keypad and quick reference card facilitate ease of use in laboratory environments.
2. Package Contents
Verify that all items are present and undamaged upon unpacking:
- Oakton PC700 Meter Unit
- pH Electrode
- Conductivity/Temperature Probe
- Electrode Stand
- Power Adapter (if included)
- Quick Reference Card
3. Setup
Follow these steps to set up your Oakton PC700 Meter:
- Position the Meter: Place the PC700 meter on a stable, level surface in your laboratory.
- Assemble the Electrode Stand: Attach the electrode stand components as per the included instructions, ensuring it is stable and can securely hold the probes.
- Connect the pH Electrode: Connect the pH electrode to the BNC connector on the back or side of the meter. Ensure a secure connection.
- Connect the Conductivity/Temperature Probe: Connect the conductivity/temperature probe to the 2.5 mm jack designated for temperature and conductivity measurements.
- Power Connection: Connect the power adapter to the meter and then plug it into a suitable electrical outlet.

4. Operating Instructions
This section outlines the basic operation of the PC700 Meter.
4.1 Powering On/Off
- Press the POWER button to turn the meter on.
- Press and hold the POWER button to turn the meter off.
4.2 Mode Selection
The PC700 meter can measure pH, mV, conductivity, TDS, and temperature. Use the dedicated mode buttons (e.g., pH/mV, COND/TDS) to switch between measurement parameters.
4.3 Calibration
Accurate measurements require proper calibration. The PC700 supports up to 6-point pH calibration and conductivity cell constant entry.
- pH Calibration: The meter features auto-buffer recognition for USA (1.68, 4.01, 7.00, 10.01, 12.45) and NIST (1.68, 4.01, 6.18, 9.86, 12.45) buffer sets. Refer to the quick reference card for detailed calibration steps.
- Conductivity Calibration: Calibrate using standard solutions. The meter stores conductivity/TDS cell constants (0.1, 1.0, and 10).
4.4 Temperature Compensation
The meter offers automatic (ATC) or manual temperature compensation (MTC). Temperature can be displayed in either degrees Fahrenheit (°F) or Celsius (°C).
4.5 Data Storage and Retrieval
The PC700 can store up to 100 data sets. Use the MEM button to store readings and the RECALL function to retrieve stored data.
4.6 Display Hold
Press the HOLD button to freeze the current reading on the display for easier recording.
5. Maintenance
Proper maintenance ensures the longevity and accuracy of your PC700 Meter.
- Cleaning the Meter: Wipe the meter's exterior with a damp cloth. Avoid abrasive cleaners or solvents. Ensure the membrane keypad is kept clean.
- Electrode Care:
- Always keep the pH electrode bulb hydrated in storage solution when not in use. Never store in distilled water.
- Rinse probes with distilled water after each use to prevent sample contamination and buildup.
- Inspect probes regularly for damage or buildup. Clean as necessary using appropriate cleaning solutions recommended by the manufacturer.
- Storage: Store the meter and probes in a clean, dry environment away from extreme temperatures and direct sunlight.
6. Troubleshooting
Refer to the following table for common issues and their solutions:
| Problem | Possible Cause | Solution |
|---|---|---|
| Inaccurate pH readings | Electrode not calibrated; dirty electrode; expired buffer solutions; damaged electrode | Calibrate the pH electrode; clean the electrode; use fresh buffer solutions; replace the electrode if damaged |
| Unstable conductivity readings | Dirty conductivity cell; air bubbles in cell; incorrect cell constant | Clean the conductivity cell; gently tap the probe to dislodge bubbles; verify or recalibrate cell constant |
| Meter does not power on | No power supply; faulty power adapter | Check power connection; ensure outlet is functional; test with a different power adapter if available |
| Display shows 'Error' message | Calibration error; probe malfunction | Refer to the quick reference card for specific error codes; re-calibrate or check probe connections |
For issues not listed, consult the quick reference card or contact customer support.
7. Specifications
Key technical specifications for the Oakton PC700 Meter:
- Model Number: AO-35413-00
- Buffer Recognition: DIN, NIST, USA
- pH Calibration: Up to 6 points
- Conductivity Accuracy: 1% Full Scale (FS)
- pH Connection: BNC
- Temperature Connection: 2.5 mm jack
- Display Type: LCD (Dual-display)
- Data Storage: Up to 100 data sets
- Temperature Compensation: Automatic or Manual, selectable in °F or °C
- Package Dimensions: 12.5 x 9.25 x 5 inches
- Weight: 3.85 Pounds
- Manufacturer: COLE-PARMER
8. Warranty and Support
For warranty information, please refer to the documentation provided with your purchase or contact the manufacturer, COLE-PARMER, directly. Technical support and service inquiries should also be directed to the manufacturer or your authorized distributor.
Manufacturer: COLE-PARMER
For further assistance, visit the Oakton brand page or contact your supplier.