Labconco CApture Portable Fuming System Instruction Manual
Model: 3420022
Brand: Labconco
Product Overview
The Labconco CApture Portable Fuming System is designed for quickly fixing latent fingerprints at crime scenes using microburst or traditional cyanoacrylate (CA) fuming methods. This system protects the user from hazardous fumes by combining all necessary components into one compact unit, eliminating the need for external ducting.
An image showing the Labconco CApture Portable Fuming System. The unit features a clear, hinged polycarbonate chamber on the left, revealing an internal stainless steel manifold with a shelf and hook. The right side houses the main unit with controls and the 'CApture Labconco' logo. A 'Caution Hot Surface Inside Door' warning label is visible.
Key Features:
Patented collapsible design for portability.
Clear, polycarbonate chamber for observation.
Microprocessor-controlled heater with a temperature range of 120° to 350° F (50° to 177° C).
Stainless steel support manifold with control slide slots, shelf, and hook.
LCD control panel for temperature display and programming.
Setup Instructions
Unpacking: Carefully remove the fuming system and all accessories from its packaging. Retain packaging for future transport or storage.
Placement: Place the unit on a stable, level surface. Adjust the leveling foot with its locking pin to ensure stability.
Chamber Assembly: If collapsed for transport, extend and secure the clear polycarbonate chamber. Ensure it is properly latched.
Accessory Installation: Install the stainless steel support manifold, shelf, and hook as needed for your specific fuming application. These components slide into the designated slots and perforations within the chamber.
Heater Pan Placement: Place one or more of the ten included aluminum heater pans into the two heater receptacles located within the unit. Close the heater door.
Power Connection: Connect the 6-foot (1.8 meters) power cord to the unit and then to a suitable 230 Volts, 50/60 Hz power outlet. Ensure the main power switch/circuit breaker located at the rear of the unit is in the OFF position before connecting.
Filter Check: Verify that the particulate/activated carbon filter cartridge is correctly installed. This filter is crucial for safe operation and fume containment.
Operating Instructions
Power On: Flip the rear-mounted main power switch/circuit breaker to the ON position.
Temperature Setting: Use the "Up" and "Down" buttons on the control panel to set the desired fuming temperature. The LCD will display the set point. Press "Set" to confirm. The heater can be set between 120° F (50° C) and 350° F (177° C).
Fuming Agent Placement: Place the cyanoacrylate (Super Glue) or other fuming agent into the heater pan(s) inside the chamber.
Sample Placement: Carefully place the items to be fumed onto the shelf or hang them from the hook within the chamber.
Chamber Closure: Securely close the clear polycarbonate chamber door.
Operation Mode: Use the three-way heater/off/vent switch on the control panel:
Heater: Activates the heater to begin the fuming process. The LCD will display the actual heater temperature.
Off: Turns off the heater and exhaust blower.
Vent: Activates the built-in exhaust blower to clear fumes from the chamber after the fuming cycle is complete. Ensure the chamber is clear before opening.
Monitoring: Monitor the fuming process through the clear chamber. Once latent prints are developed, switch the unit to "Vent" mode to clear the chamber.
Completion: After ventilation, carefully open the chamber and remove the fumed items.
Maintenance
Filter Replacement: The particulate/activated carbon filter cartridge has a capacity for approximately 100 development cycles. Replace the filter regularly to ensure effective fume removal and user safety. Refer to the filter replacement guide for specific instructions.
Chamber Cleaning: Clean the polycarbonate chamber regularly with a mild, non-abrasive cleaner and a soft cloth to maintain visibility and prevent residue buildup. Do not use harsh chemicals that may damage the polycarbonate.
Heater Pan Disposal: Dispose of used aluminum heater pans according to local regulations for chemical waste.
General Cleaning: Wipe down the exterior of the unit with a damp cloth as needed. Ensure the unit is unplugged before cleaning.
Troubleshooting
Problem
Possible Cause
Solution
Unit does not power on.
Power cord not securely plugged in; Main power switch/circuit breaker is OFF or tripped.
Ensure power cord is fully inserted. Check that the main power switch/circuit breaker is ON. Reset the circuit breaker if tripped. Verify power outlet functionality.
Heater not reaching set temperature.
Heater switch not in "Heater" position; Heater pan not properly seated; Malfunctioning heater element.
Confirm the three-way switch is set to "Heater". Ensure heater pan is correctly placed. If problem persists, contact Labconco support.
Fumes not clearing after fuming cycle.
Switch not in "Vent" position; Filter cartridge is saturated or improperly installed; Blower malfunction.
Set the three-way switch to "Vent". Replace the filter cartridge if it has reached its capacity (approx. 100 cycles). Ensure filter is correctly installed. If issue continues, contact Labconco support.
Poor fingerprint development.
Incorrect temperature setting; Insufficient fuming agent; Chamber not sealed properly.
Verify the temperature setting is appropriate for the fuming agent used. Ensure sufficient fuming agent is present. Check that the chamber door is securely closed and latched.
Specifications
Model Number: 3420022
Brand: Labconco
Item Weight: 24 lbs (10.9 kg)
Shipping Weight: 30 lbs (13.6 kg)
Power Requirements: 230 Volts, 50/60 Hz
Heater Temperature Range: 120° to 350° F (50° to 177° C)
Filter Capacity: Approximately 100 development cycles
Power Cord: 6' (1.8 meters) with Schuko plug
Certifications: ETL listed (Models 3432000), CE conformity marking (Models 3420020, 3420022 and 3420024)
First Available: March 19, 2013
Warranty and Support
For specific warranty information regarding your Labconco CApture Portable Fuming System, please refer to the documentation included with your purchase or visit the official Labconco website. Warranty terms may vary based on region and purchase date.
For technical support, service, or to order replacement parts, please contact Labconco customer service directly:
Refer to the website for contact numbers and email support.
When contacting support, please have your model number (3420022) and serial number ready for faster assistance.
Ask a question about this manual
Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.