Milwaukee MW101

Milwaukee MW101 PH Meter Instruction Manual

Model: MW101

1. Introduction

The Milwaukee MW101 PH Meter is a portable device designed for accurate pH measurements with a 0.01 pH resolution. It features manual temperature compensation and a compact, ergonomic design, making it suitable for various applications including educational, agricultural, horticultural, and environmental analysis. This manual provides detailed instructions for the proper setup, operation, and maintenance of your MW101 PH Meter to ensure reliable and consistent performance.

Milwaukee MW101 PH Meter main unit

Figure 1: Milwaukee MW101 PH Meter main unit.

2. Safety Information

Please read all safety information before operating the device. Failure to follow these instructions may result in injury or damage to the meter.

  • Always handle the pH electrode with care. It is a delicate component.
  • Do not immerse the meter body in water or other liquids. The meter is not waterproof.
  • Keep calibration solutions and cleaning solutions out of reach of children and pets. Follow safety data sheet (SDS) instructions for chemical handling.
  • Ensure the battery compartment is securely closed to prevent moisture ingress.
  • Dispose of batteries and the meter according to local regulations.

3. Setup

3.1 Unpacking and Inspection

Carefully unpack the MW101 PH Meter and all accessories. Verify that the following items are included:

  • MW101 PH Meter unit
  • MA911B/1 pH electrode
  • pH 7.01 calibration solution sachet (20 ml)
  • Calibration screwdriver
  • 9V battery
  • Instruction manual (this document)

Inspect all components for any signs of damage. If any items are missing or damaged, contact your supplier immediately.

3.2 Battery Installation

  1. Locate the battery compartment cover on the back of the meter.
  2. Slide or unclip the cover to open the compartment.
  3. Connect the 9V battery to the battery clip, ensuring correct polarity.
  4. Place the battery inside the compartment and close the cover securely.

3.3 Electrode Connection and Preparation

  1. Connect the MA911B/1 pH electrode BNC connector to the BNC socket on the top of the meter. Ensure a firm connection.
  2. Remove the protective cap from the electrode tip. The cap contains a storage solution to keep the electrode hydrated.
  3. Rinse the electrode tip thoroughly with distilled or deionized water. Do not wipe the electrode tip, as this can create static charges and damage the sensitive glass bulb.
  4. The electrode is now ready for calibration and measurement.

4. Operating Instructions

4.1 Powering On/Off

Press the ON/OFF button located on the front panel to turn the meter on or off.

4.2 Calibration

For accurate measurements, the MW101 PH Meter requires regular calibration. It is recommended to calibrate the meter before each use or at least once a day.

  1. Turn on the meter.
  2. Rinse the electrode with distilled or deionized water and blot dry.
  3. Open the pH 7.01 calibration solution sachet and immerse the electrode tip into the solution.
  4. Allow the reading to stabilize on the display.
  5. Using the supplied calibration screwdriver, adjust the pH7 trimmer (located on the front panel) until the display shows "7.01".
  6. Rinse the electrode again with distilled or deionized water.
  7. If a two-point calibration is desired for greater accuracy, repeat steps 3-5 using a pH 4.01 or pH 10.01 buffer solution and adjust the corresponding pH4/10 trimmer.
  8. The meter is now calibrated and ready for use.

4.3 Manual Temperature Compensation

The MW101 features manual temperature compensation. pH measurements are temperature-dependent. To ensure accuracy, measure the temperature of your sample and adjust the Temperature Compensation knob on the front panel to match the sample's temperature.

4.4 Taking Measurements

  1. Ensure the meter is calibrated and the temperature compensation is set to the sample's temperature.
  2. Rinse the electrode with distilled or deionized water and blot dry.
  3. Immerse the electrode tip into the sample solution, ensuring the sensing bulb is fully submerged.
  4. Gently stir the electrode in the sample to ensure homogeneity and allow the reading to stabilize on the display.
  5. Record the pH value.
  6. After measurement, rinse the electrode thoroughly with distilled or deionized water before storing.

5. Maintenance

5.1 Electrode Storage

Proper electrode storage is critical for its longevity and performance.

  • Always keep the electrode tip hydrated. Never allow the electrode to dry out.
  • Store the electrode with its protective cap filled with a proper electrode storage solution (e.g., Milwaukee MA9015). Do not use distilled or deionized water for long-term storage, as this can deplete the electrolyte.
  • If the electrode has dried out, soak it in storage solution for at least 1 hour before use.

5.2 Electrode Cleaning

Regular cleaning helps maintain electrode accuracy.

  • For general cleaning, rinse the electrode with distilled or deionized water.
  • For stubborn deposits or buildup, use a specialized electrode cleaning solution (e.g., Milwaukee MA9016). Follow the cleaning solution's instructions.
  • After cleaning, rinse thoroughly and re-calibrate the meter.

5.3 Battery Replacement

When the display becomes dim or the meter shows erratic readings, it may be time to replace the 9V battery. Follow the battery installation steps in Section 3.2.

5.4 General Unit Cleaning

Wipe the meter's exterior with a damp cloth. Do not use abrasive cleaners or solvents.

6. Troubleshooting

ProblemPossible CauseSolution
Meter does not turn on.Dead or improperly installed battery.Check battery connection; replace 9V battery.
Erratic or unstable readings.
  • Dirty or dry electrode.
  • Improper calibration.
  • Air bubbles around electrode tip.
  • Electrode nearing end of life.
  • Clean and rehydrate electrode (soak in storage solution).
  • Perform a fresh calibration.
  • Gently tap electrode to dislodge bubbles.
  • Consider replacing the electrode.
Slow response time.
  • Dirty or aged electrode.
  • Low battery.
  • Clean and rehydrate electrode.
  • Replace 9V battery.
Inaccurate readings after calibration.
  • Expired or contaminated calibration solutions.
  • Incorrect temperature compensation setting.
  • Use fresh calibration solutions.
  • Verify and adjust the temperature compensation knob to the sample temperature.

7. Specifications

Milwaukee MW101 PH Meter with dimensions

Figure 2: Milwaukee MW101 PH Meter with approximate dimensions (6.4 inches / 16 cm height).

ModelMW101
Measurement Range0.00 to 14.00 pH (typical for pH meters)
Resolution0.01 pH
Accuracy±0.02 pH (typical, after calibration)
Temperature CompensationManual, 0 to 50°C (32 to 122°F)
ElectrodeMA911B/1 (included)
Power Supply1 x 9V battery (included)
Battery LifeApproximately 300 hours of continuous use
DimensionsApprox. 160 x 80 x 35 mm (6.3 x 3.1 x 1.4 inches)
WeightApprox. 220 g (8 ounces)
ManufacturerMilwaukee
ASINB0009YH06Y

8. Warranty

Specific warranty terms and conditions for the Milwaukee MW101 PH Meter are provided by the manufacturer. Please refer to the warranty card included with your product or visit the official Milwaukee Instruments website for detailed information regarding warranty coverage, duration, and claims procedures. Keep your proof of purchase for warranty purposes.

9. Support

For technical assistance, troubleshooting beyond this manual, or to inquire about replacement parts and accessories, please contact Milwaukee Instruments customer support.

Manufacturer: Milwaukee Instruments
Website: www.milwaukeeinstruments.com
Please refer to the manufacturer's website for the most current contact information, including phone numbers and email addresses.

Related Documents - MW101

Preview Milwaukee MW100/MW101/MW102/MW500 pH/ORP/Temperature Meters
Milwaukee's MW100, MW101, MW102, and MW500 are compact, portable pH, ORP, and temperature meters designed for ease of use and reliable measurements in educational, agricultural, horticultural, and environmental applications. These meters feature a larger LCD display, ergonomic design, and long battery life.
Preview Milwaukee MW805 MAX Portable Meter Instruction Manual
Comprehensive instruction manual for the Milwaukee MW805 MAX portable meter, detailing its features, specifications, operation, calibration, maintenance, and troubleshooting for pH, EC, TDS, and temperature measurements.
Preview Milwaukee MW801 & MW802 Portable pH/EC/TDS Meter User Manual
User manual for Milwaukee portable pH/EC/TDS meters, models MW801 and MW802. Includes operation, calibration, maintenance, and specifications.
Preview Milwaukee MW805 MAX Portable pH EC TDS Temperature Meter Instruction Manual
Instruction manual for the Milwaukee MW805 MAX portable meter, covering pH, EC, TDS, and temperature measurements. Learn about features, operation, calibration, and maintenance.
Preview Milwaukee MW801 MW802 Portable pH EC TDS Meter User Manual and Specifications
User manual for the Milwaukee MW801 and MW802 portable pH, EC, and TDS meters. Includes instructions for preliminary examination, operation, pH and EC/TDS measurements, calibration procedures, battery replacement, and technical specifications.
Preview Milwaukee LDM 50 Laser Distance Meter User Manual and Specifications
Comprehensive user manual and technical specifications for the Milwaukee LDM 50 Laser Distance Meter, detailing its functions, safety instructions, and operational procedures for accurate distance, area, and volume measurements.