Milwaukee PORTABLE pH/EC/TDS METER
USER MANUAL
MODEL: MW801, MW802
Smart Combined Meter
WARRANTY:
This instrument is warranted from all defects in materials and manufacturing for a period of two years from the date of purchase. The probe is warranted for a period of six months. If during this period the repair or replacement of parts is required, where the damage is not due to negligence or erroneous operation by the user, please return the parts to either distributor or Milwaukee office and the repair will be effected free of charge.
Note: Milwaukee reserves the right to modify the design, construction and appearance of its products without advance notice.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is found, notify your Dealer. Each meter is supplied complete with:
- MA850DIN preamplified non refillable double junction combination pH electrode with built-in EC/TDS probe and temperature sensor; 1m (3.3') cable.
- pH7.01, 1413 μS/cm & 1382 ppm (for MW801) or 1500 ppm (for MW802) solutions, 20 mL each.
- Instruction Manual.
- 1 x 9V battery.
OPERATION
Slide the battery compartment cover located on the back of the meter off, and install the battery into the battery clip connector while observing polarity. Connect the probe to the meter securely by aligning the pins with the socket and pushing the plug in. Always remove the electrode protective cap before use. Make sure that the meter has been calibrated before taking any measurements (see calibration procedure). Turn the instrument on by pressing the ON/OFF key.
pH MEASUREMENTS
If the electrode has been left dry, soak the tip in a pH 7 (M10007) buffer solution for a few minutes to reactivate it. Submerge the tip of the electrode into the sample to be measured, select the pH mode and allow the reading to stabilize before taking measurements.
EC/TDS MEASUREMENTS
Immerse the tip of the electrode into the sample to be tested. Select the measurement range (EC or TDS) and wait for the temperature sensor to reach thermal equilibrium with the samples before taking measurements. AFTER MEASUREMENTS the instrument should be switched off and the probe cleaned and stored with a few drops of storage solution in the protective cap.
CALIBRATION PROCEDURE:
Remove the protective cap from the electrode. Switch the meter on. Select the range pressing the appropriate key for pH, EC or TDS.
pH Calibration
A) PREPARATION
Single point manual calibration. Choose buffer. If you are going to measure samples near pH 7 use pH 7.01 buffer solution (M10007B) for calibration; use pH 4.01 buffer solution (M10004B) for acidic measurements, or pH 10.01 (M10010B) for alkaline measurements.
B) PROCEDURE
Rinse the tip of the electrode and immerse it in the calibration solution. Wait a couple of minutes for the reading to stabilize. Measure the temperature of the buffer solution with a thermometer. Adjust the calibration knob until the LCD shows the pH value of the buffer at the measured temperature (see pH versus temperature chart).
The calibration is now complete and the meter is ready for use.
EC/TDS calibration:
A) PREPARATION
Single point EC calibration. Choose appropriate standard: Model MW801: use 1382 ppm (M10032B) for TDS or 1.413 μS/cm (M10031B) for EC calibration. Model MW802: use 1500 ppm (M10442B) for TDS or 1.413 mS/cm (M10031B) for EC calibration.
B) PROCEDURE
Immerse the tip of the electrode in the standard solution. Wait for the reading to stabilize, and thermal equilibrium be reached. Turn the EC/TDS calibration knob until the display shows the EC or TDS reading at 25°C. The calibration is now complete and the meter is ready for use. The instrument should be re-calibrated at least once a month, or when the probe or battery is replaced. Note: The conversion between EC and TDS is made by a built-in circuit. If calibration is made in EC range, TDS range is automatically calibrated (or vice versa).
BATTERY REPLACEMENT:
The meters are powered by a 9V battery that is in a compartment located in the rear of the instrument. When the battery becomes weak, the instrument automatically switches off. Replacement must only take place in a non-hazardous area using an alkaline 9V battery. Slide off the battery compartment cover and replace the old battery with a new one. Make sure that the battery contacts are fully engaged in the connector. Seat the battery in its compartment and replace the cover.
pH VERSUS TEMPERATURE CHART
TEMP | pH VALUES | ||||
°C | °F | MA9004 | MA9007 | MA9010 | |
0 | 32 | 4.01 | 7.13 | 10.32 | |
5 | 41 | 4.00 | 7.10 | 10.24 | |
10 | 50 | 4.00 | 7.07 | 10.18 | |
15 | 59 | 4.00 | 7.04 | 10.12 | |
20 | 68 | 4.00 | 7.03 | 10.06 | |
25 | 77 | 4.01 | 7.01 | 10.01 | |
30 | 86 | 4.02 | 7.00 | 9.96 | |
35 | 95 | 4.03 | 6.99 | 9.92 | |
40 | 104 | 4.04 | 6.98 | 9.85 | |
45 | 113 | 4.05 | 6.98 | 9.85 | |
50 | 122 | 4.06 | 6.98 | 9.82 | |
55 | 131 | 4.07 | 6.98 | 9.79 | |
60 | 140 | 4.09 | 6.98 | 9.77 | |
65 | 149 | 4.11 | 6.99 | 9.76 | |
70 | 158 | 4.12 | 6.99 | 9.75 |
OPTIONAL ACCESSORIES
- M10007B pH 7.01 buffer solution, 25x20 mL sachet
- M10004B pH 4.01 buffer solution, 25x20 mL sachet
- M10010B pH 10.01 buffer solution, 25x20 mL sachet
- M10031B 1413 μS/cm calibration solution, 25x20 mL sachet
- M10032B 1382 ppm calibration solution, 25x20 mL sachet
- M10442B 1500 ppm calibration solution, 25x20 mL sachet
- MA9015 Electrode storage solution, 220 mL bottle
- MA850DIN pH/EC/TDS spare probe with 1m cable
SPECIFICATIONS:
MW801 | MW802 | |
RANGE | 0.0 to 14.0 pH 0 to 1990 ppm 0 to 1990 μS/cm | 0.00 to 14.00 pH 0 to 4000 ppm 0.00 to 6.00 mS/cm |
RESOLUTION | 0.1 pH 10 ppm 10 μS/cm | 0.10 pH 10 ppm 0.01 mS/cm |
ACCURACY (@ 20°C) | ±0.2 pH ±2 % Full scale | ±0.20 pH ±2 % Full scale |
CALIBRATION SOLUTIONS | M10007B (pH 7.01) M10031B (1.413 mS/cm) M10032B (1382ppm) M10442B (1500ppm) | |
CONVERSION FACTOR | 0.5 | 0.68 |
CALIBRATION | Manual, at 1 point | |
TEMPERATURE COMPENSATION | Automatic from 0 to 50°C | |
PROBE | MA850DIN combination pH/EC/TDS probe | |
ENVIRONMENT | 0 to 50°C, 95% RH max | |
BATTERY LIFE /TYPE | 1x9V alkaline approx. 150 hours of use | |
DIMENSIONS | 143 x 80 x 32 mm | |
WEIGHT | 220 g (with battery) |