Milwaukee PORTABLE pH/EC/TDS METER

USER MANUAL

MODEL: MW801, MW802

Smart Combined Meter

WARRANTY:

This instrument is warranted from all defects in materials and manufacturing for a period of two years from the date of purchase. The probe is warranted for a period of six months. If during this period the repair or replacement of parts is required, where the damage is not due to negligence or erroneous operation by the user, please return the parts to either distributor or Milwaukee office and the repair will be effected free of charge.

Note: Milwaukee reserves the right to modify the design, construction and appearance of its products without advance notice.

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is found, notify your Dealer. Each meter is supplied complete with:

OPERATION

Slide the battery compartment cover located on the back of the meter off, and install the battery into the battery clip connector while observing polarity. Connect the probe to the meter securely by aligning the pins with the socket and pushing the plug in. Always remove the electrode protective cap before use. Make sure that the meter has been calibrated before taking any measurements (see calibration procedure). Turn the instrument on by pressing the ON/OFF key.

pH MEASUREMENTS

If the electrode has been left dry, soak the tip in a pH 7 (M10007) buffer solution for a few minutes to reactivate it. Submerge the tip of the electrode into the sample to be measured, select the pH mode and allow the reading to stabilize before taking measurements.

EC/TDS MEASUREMENTS

Immerse the tip of the electrode into the sample to be tested. Select the measurement range (EC or TDS) and wait for the temperature sensor to reach thermal equilibrium with the samples before taking measurements. AFTER MEASUREMENTS the instrument should be switched off and the probe cleaned and stored with a few drops of storage solution in the protective cap.

CALIBRATION PROCEDURE:

Remove the protective cap from the electrode. Switch the meter on. Select the range pressing the appropriate key for pH, EC or TDS.

pH Calibration

A) PREPARATION

Single point manual calibration. Choose buffer. If you are going to measure samples near pH 7 use pH 7.01 buffer solution (M10007B) for calibration; use pH 4.01 buffer solution (M10004B) for acidic measurements, or pH 10.01 (M10010B) for alkaline measurements.

B) PROCEDURE

Rinse the tip of the electrode and immerse it in the calibration solution. Wait a couple of minutes for the reading to stabilize. Measure the temperature of the buffer solution with a thermometer. Adjust the calibration knob until the LCD shows the pH value of the buffer at the measured temperature (see pH versus temperature chart).

The calibration is now complete and the meter is ready for use.

EC/TDS calibration:

A) PREPARATION

Single point EC calibration. Choose appropriate standard: Model MW801: use 1382 ppm (M10032B) for TDS or 1.413 μS/cm (M10031B) for EC calibration. Model MW802: use 1500 ppm (M10442B) for TDS or 1.413 mS/cm (M10031B) for EC calibration.

B) PROCEDURE

Immerse the tip of the electrode in the standard solution. Wait for the reading to stabilize, and thermal equilibrium be reached. Turn the EC/TDS calibration knob until the display shows the EC or TDS reading at 25°C. The calibration is now complete and the meter is ready for use. The instrument should be re-calibrated at least once a month, or when the probe or battery is replaced. Note: The conversion between EC and TDS is made by a built-in circuit. If calibration is made in EC range, TDS range is automatically calibrated (or vice versa).

BATTERY REPLACEMENT:

The meters are powered by a 9V battery that is in a compartment located in the rear of the instrument. When the battery becomes weak, the instrument automatically switches off. Replacement must only take place in a non-hazardous area using an alkaline 9V battery. Slide off the battery compartment cover and replace the old battery with a new one. Make sure that the battery contacts are fully engaged in the connector. Seat the battery in its compartment and replace the cover.

pH VERSUS TEMPERATURE CHART

TEMPpH VALUES
°C°FMA9004MA9007MA9010
0324.017.1310.32
5414.007.1010.24
10504.007.0710.18
15594.007.0410.12
20684.007.0310.06
25774.017.0110.01
30864.027.009.96
35954.036.999.92
401044.046.989.85
451134.056.989.85
501224.066.989.82
551314.076.989.79
601404.096.989.77
651494.116.999.76
701584.126.999.75

OPTIONAL ACCESSORIES

SPECIFICATIONS:

MW801MW802
RANGE0.0 to 14.0 pH
0 to 1990 ppm
0 to 1990 μS/cm
0.00 to 14.00 pH
0 to 4000 ppm
0.00 to 6.00 mS/cm
RESOLUTION0.1 pH
10 ppm
10 μS/cm
0.10 pH
10 ppm
0.01 mS/cm
ACCURACY (@ 20°C)±0.2 pH
±2 % Full scale
±0.20 pH
±2 % Full scale
CALIBRATION SOLUTIONSM10007B (pH 7.01)
M10031B (1.413 mS/cm)
M10032B (1382ppm) M10442B (1500ppm)
CONVERSION FACTOR0.50.68
CALIBRATIONManual, at 1 point
TEMPERATURE COMPENSATIONAutomatic from 0 to 50°C
PROBEMA850DIN combination pH/EC/TDS probe
ENVIRONMENT0 to 50°C, 95% RH max
BATTERY LIFE /TYPE1x9V alkaline approx. 150 hours of use
DIMENSIONS143 x 80 x 32 mm
WEIGHT220 g (with battery)
Models: MW801-MW802 Portable pH-EC-TDS Meter, MW801-MW802, Portable pH-EC-TDS Meter, pH-EC-TDS Meter, Meter

PDF preview unavailable. Download the PDF instead.

manMW801 MW802 ENG Acrobat Distiller 8.1.0 (Windows)

Related Documents

Preview Milwaukee MW805 MAX Portable pH EC TDS Temperature Meter Instruction Manual
Instruction manual for the Milwaukee MW805 MAX portable meter, covering pH, EC, TDS, and temperature measurements. Learn about features, operation, calibration, and maintenance.
Preview Milwaukee MW805 MAX Portable Meter Instruction Manual
Comprehensive instruction manual for the Milwaukee MW805 MAX portable meter, detailing its features, specifications, operation, calibration, maintenance, and troubleshooting for pH, EC, TDS, and temperature measurements.
Preview Milwaukee MW100/MW101/MW102/MW500 pH/ORP/Temperature Meters
Milwaukee's MW100, MW101, MW102, and MW500 are compact, portable pH, ORP, and temperature meters designed for ease of use and reliable measurements in educational, agricultural, horticultural, and environmental applications. These meters feature a larger LCD display, ergonomic design, and long battery life.
Preview Milwaukee MW180 MAX Bench Meter Instruction Manual
Comprehensive instruction manual for the Milwaukee MW180 MAX Bench Meter, detailing its features, specifications, operation, calibration, and maintenance for pH, mV, EC, TDS, NaCl, and temperature measurements.
Preview Milwaukee MC311 Conductivity Controller User Manual
User manual for the Milwaukee MC311 Conductivity Controller, detailing specifications, operation, calibration, and installation procedures for accurate conductivity measurements.
Preview Milwaukee LDM 50 Laser Distance Meter User Manual and Specifications
Comprehensive user manual and technical specifications for the Milwaukee LDM 50 Laser Distance Meter, detailing its functions, safety instructions, and operational procedures for accurate distance, area, and volume measurements.
Preview Milwaukee LDM 50 Laser Distance Meter Quickstart Guide
This guide provides essential information for operating the Milwaukee LDM 50 Laser Distance Meter, covering its primary functions, operational steps, and key features for efficient use.
Preview Milwaukee Auto Voltage/Continuity Tester W/ Resistance 2213-20
Milwaukee Auto Voltage/Continuity Tester W/ Resistance (2213-20) offers accurate measurements, a backlit LCD, LED worklight, and auto-ranging for efficient electrical testing. Includes belt clip, test leads, batteries, and manual.