1. Introduction
The CEM DT-9054 is a digital RCD (Residual Current Device) tester designed to verify the operational integrity of leakage switches. This device provides accurate measurements of trip time and trip current, ensuring electrical safety in various installations. It features a clear digital display and intuitive controls for ease of use.

2. Setup
Before using the DT-9054, ensure the area is safe and the power supply to the circuit under test is off, if required for connection. The device is designed for direct plug-in operation.
- Inspection: Unpack the DT-9054 and inspect it for any signs of damage. Do not use the device if it appears damaged.
- Power Connection: Plug the DT-9054 directly into the electrical outlet or circuit you intend to test. Ensure a secure connection.
- Familiarization: Take a moment to identify the ON/OFF switch, TEST button, RANGE button, and 180° button on the device.

3. Operating Instructions
The DT-9054 is used to test RCDs by simulating a leakage current and measuring the time it takes for the RCD to trip.
- Power On: Ensure the DT-9054 is securely plugged into the circuit to be tested. Turn the device ON using the ON/OFF switch. The digital display should illuminate.
- Select Trip Current: Use the RANGE button to cycle through the available trip current settings (e.g., 10mA, 30mA, 100mA, 300mA, 500mA, 650mA). Select the appropriate current for the RCD being tested.
- Select Phase Angle (if applicable): The 180° button allows you to switch between 0° and 180° phase angles for testing. This is important for comprehensive RCD testing as some RCDs may react differently depending on the phase of the leakage current.
- Initiate Test: Press the TEST button. The device will inject a simulated leakage current into the circuit. Observe the RCD under test; it should trip.
- Read Results: After the RCD trips, the DT-9054's digital display will show the trip time in milliseconds (ms). Note this value.
- Reset and Re-test: Reset the RCD. You may perform additional tests at different trip current settings or phase angles as required by safety standards or your testing protocol.
- Power Off: Once testing is complete, turn the DT-9054 OFF and unplug it from the circuit.

4. Maintenance
Proper maintenance ensures the longevity and accuracy of your DT-9054 RCD Tester.
- Cleaning: Wipe the device with a soft, dry cloth. Do not use abrasive cleaners or solvents.
- Storage: Store the tester in a cool, dry place, away from direct sunlight and extreme temperatures.
- Inspection: Periodically check the plug and casing for any signs of wear, cracks, or damage. If damage is found, discontinue use and contact support.
- Calibration: For professional use, periodic calibration by a qualified service center is recommended to maintain accuracy.
5. Troubleshooting
If you encounter issues with your DT-9054, refer to the following common problems and solutions:
- Device does not power on:
- Ensure the device is securely plugged into a live electrical outlet.
- Check the ON/OFF switch position.
- RCD does not trip during test:
- Verify that the selected trip current setting is appropriate for the RCD.
- Ensure the RCD itself is functional. Test with a known good RCD if possible.
- Check the supply voltage to ensure it is within the operational range of the tester (195V - 253V).
- Inaccurate readings:
- Ensure the device is used within its specified operating temperature range.
- Consider professional calibration if consistent inaccuracies are observed.
6. Specifications
| Attribute | Value |
|---|---|
| Model Number | DT-9054 |
| Brand Name | CEM |
| Place of Origin | China |
| Display Type | Digital Only |
| Supply Voltage | 195V - 253V / 45 - 65Hz |
| Operating Temperature | -20°C to 50°C |
| Dimensions (HxWxD) | 200mm x 70mm x 45mm |
| Weight | 260g |
| Trip Current Settings | 10mA, 30mA, 100mA, 300mA, 500mA, 650mA |
| Trip Current Measurement Accuracy | ±10% ±2digits |
| Trip Time Ranges | 0ms to 300ms |
| Trip Time Accuracy | ±5% ±3digits |
| Ramp Test | x0.4 to 1.1 times nominal rated current |
7. User Tips
- Always ensure the RCD being tested is easily accessible for resetting after a trip.
- Perform tests at both 0° and 180° phase angles for a comprehensive evaluation of the RCD's performance.
- Keep a log of your RCD test results for compliance and maintenance records.
- If testing multiple RCDs, allow a short cool-down period between tests to prevent overheating, especially at higher current settings.
8. Warranty & Support
For warranty information, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.
Seller: ruilande shangdian Store





