Victor 01S

Victor 01S Temperature Source Signal Generator User Manual

Model: 01S

1. Introduction

The Victor 01S Temperature Source Signal Generator is a battery-powered, handheld portable instrument designed for precise output of electrical and physical parameters. It is capable of analog output for DC voltage, resistance, thermocouple (TC), and thermal resistance (RTD) signals, making it an essential tool for calibration and testing in various industrial and laboratory settings.

This manual provides detailed instructions for the safe and efficient use of your Victor 01S signal generator. Please read it thoroughly before operation and keep it for future reference.

2. Safety Information

To ensure safe operation and service of the instrument, follow these safety precautions:

  • Always inspect the instrument and test leads for damage before use. Do not use if damaged.
  • Do not apply more than the rated voltage between terminals or between any terminal and ground.
  • Use caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC. These voltages pose a shock hazard.
  • Ensure the battery compartment is securely closed before operation.
  • Do not operate the instrument in explosive atmospheres.
  • Refer to the specifications section for maximum input/output limits.

3. Package Contents

Upon unpacking, please verify that all items listed below are present and undamaged:

Packaging accessories including the calibrator, test leads, and manual.

Figure 3.1: Packaging Accessories

  1. Packing Box
  2. Victor 01S Temperature Signal Generator
  3. Alligator Clips
  4. Industrial Test Leads (Red and Black)
  5. Latch Key
  6. Thermocouple Adapter
  7. User Manual
  8. Product Qualification Certificate

4. Product Overview

The Victor 01S features a compact design with a high-definition large LCD screen for clear readings. It includes various functions such as flashlight lighting, low battery pressure tips, automatic shut-down, and a clear zero function. The device supports both Thermocouple (TC) and Thermal Resistance (RTD) functions, along with unit conversion for temperature.

Front view of Victor 01S showing LCD, buttons, and input/output terminals.

Figure 4.1: Victor 01S Front View with Key Features

4.1 Panel Introduction

The front panel of the Victor 01S is designed for intuitive operation. Refer to the diagram below for component identification:

Detailed diagram of the Victor 01S panel with numbered labels for each button and terminal.

Figure 4.2: Victor 01S Panel Layout

  1. LCD display screen
  2. Turn temperature compensation on/off (RJ-ON button)
  3. Power on/off button
  4. Switch output function (FUNC button)
  5. Select output range (RANGE button)
  6. Output setting button (Up/Down/Left/Right arrows)
  7. Analog output resistance low-end terminal (3W)
  8. 3W: 3-wire resistance measurement SENSE terminal
  9. 4W: 4-wire resistance measurement SENSE terminal
  10. Analog output resistance high-end terminal (4W)
  11. Select temperature unit (°C/°F button)
  12. Backlight/flashlight button
  13. ZERO button
  14. Configuration button (CONFIG)
  15. Voltage output (+) output terminal (V)
  16. Voltage output common (-) (return) terminal (COM)

5. Setup

5.1 Battery Installation

The Victor 01S is powered by three 1.5V AA batteries. To install or replace batteries:

  1. Ensure the device is powered off.
  2. Locate the battery compartment on the back of the instrument.
  3. Use a suitable tool (e.g., a coin) to unlock and open the battery compartment cover.
  4. Insert three 1.5V AA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
  5. Close the battery compartment cover and secure it.
Back view of Victor 01S showing the battery compartment for three AA batteries.

Figure 5.1: Battery Compartment

Note: The instrument features low pressure tips to indicate when batteries need replacement.

5.2 Initial Power On

Press the Power On/Off button (3) to turn on the device. The LCD screen (1) will illuminate, displaying the current settings.

6. Operation

The Victor 01S can output various signals. Use the FUNC button (4) to cycle through different output modes (DCV, Ohm, TC, RTD). Use the RANGE button (5) to select the appropriate measurement range within each mode. The arrow buttons (6) are used to adjust the output value.

6.1 DC Voltage Output

To output a DC voltage signal:

  1. Press FUNC (4) until 'DCV' mode is selected.
  2. Connect the test leads to the V (+) terminal (15) and COM (-) terminal (16).
  3. Use the RANGE button (5) to select the desired voltage range.
  4. Adjust the output voltage using the arrow buttons (6).

6.2 Resistance (Ohm) Output

To output a resistance signal:

  1. Press FUNC (4) until 'Ohm' mode is selected.
  2. Connect the test leads to the appropriate resistance terminals (3W, 4W, or Ω/COM) based on the required measurement type (2-wire, 3-wire, or 4-wire).
  3. Use the RANGE button (5) to select the desired resistance range.
  4. Adjust the output resistance using the arrow buttons (6).

6.3 Thermocouple (TC) Output

The Victor 01S supports 11 types of thermocouples (R, S, K, E, J, T, N, B, A, C, D).

Diagram showing 11 types of thermocouples (R, S, K, E, J, T, N, B, A, C, D) and their temperature ranges.

Figure 6.1: Supported Thermocouple Types and Ranges

  1. Press FUNC (4) until 'TC' mode is selected.
  2. Connect the thermocouple adapter to the V (+) and COM (-) terminals.
  3. Connect the thermocouple to the adapter.
  4. Use the CONFIG button (14) to select the desired thermocouple type (e.g., K, J, T).
  5. Use the RANGE button (5) to select the appropriate temperature range.
  6. Adjust the output temperature using the arrow buttons (6).
  7. Use the °C/°F button (11) to switch between Celsius and Fahrenheit units.
  8. The RJ-ON button (2) can be used to turn temperature compensation on/off.

6.4 Thermal Resistance (RTD) Output

The Victor 01S can simulate 5 types of thermal resistance (Pt100, Pt200, Pt500, Pt1000, Cu50).

Diagram showing 5 types of thermal resistance (RTD) sensors (Pt100, Pt200, Pt500, Pt1000, Cu50) and their temperature ranges.

Figure 6.2: Supported RTD Types and Ranges

  1. Press FUNC (4) until 'RTD' mode is selected.
  2. Connect the test leads to the appropriate resistance terminals (3W or 4W) based on the RTD type.
  3. Use the CONFIG button (14) to select the desired RTD type (e.g., Pt100, Cu50).
  4. Use the RANGE button (5) to select the appropriate temperature range.
  5. Adjust the output temperature using the arrow buttons (6).
  6. Use the °C/°F button (11) to switch between Celsius and Fahrenheit units.

6.5 Other Functions

  • ZERO Function (13): Press the ZERO button to clear the current reading or set a reference point.
  • Backlight/Flashlight (12): Press the backlight/flashlight button to illuminate the screen or use the built-in flashlight for working in low-light conditions.
  • Automatic Shut-down: The instrument features an automatic shut-down function to conserve battery life.

7. Specifications

The Victor 01S Temperature Source Signal Generator offers high precision and a wide range of measurement capabilities. All specifications are based on a calibration performed within one year, at 23°C ±5°C, 20-70% RH, with an accuracy of ± (% set value + reading).

Table of detailed specifications for DCV, Ohm, Thermocouple, and RTD output functions, including measurement range, output range, resolution, and accuracy.

Figure 7.1: Detailed Specifications Table

Analog Output Function Specifications
Output functionMeasurement rangeOutput rangeResolutionAccuracy
DCV100mV-10.000mV ~ 110.00mV0.01mV0.05%+0.03mV
1V-0.1000V ~ 1.1000V0.0001V0.05%+0.3mV
OMH400Ω0 ~ 400.0Ω0.1Ω0.05%+0.2Ω
4KΩ0 ~ 4.000KΩ0.05%+2Ω
TCR0°C ~ 1767°C1°C0°C-100°C: 1.5°C
100°C-1767°C: 1.2°C
S0°C ~ 1767°C1°C0°C-100°C: 0.9°C
100°C-1767°C: 0.9°C
K-200.0°C ~ 1372.0°C0.1°C-200°C-0°C: 1.0°C
0°C-1300°C: 0.8°C
1300°C-1372°C: 1.0°C
E-200.0°C ~ 1000.0°C0.1°C-200°C-0°C: 1.0°C
0°C-800°C: 0.8°C
800°C-1000°C: 1.0°C
J-200.0°C ~ 1200.0°C0.1°C-200°C-0°C: 1.0°C
0°C-1200°C: 0.8°C
T-250.0°C ~ 400.0°C0.1°C-250°C-0°C: 1.0°C
0°C-400°C: 0.6°C
N-200.0°C ~ 1300.0°C0.1°C-200°C-0°C: 1.0°C
0°C-1300°C: 0.8°C
B600°C ~ 1820°C1°C600°C-800°C: 1.0°C
800°C-1820°C: 0.9°C
A0°C ~ 2500°C1°C0°C-1500°C: 2.0°C
1500°C-2500°C: 2.0°C
C0°C ~ 2310°C1°C0°C-1600°C: 2.0°C
1600°C-2310°C: 2.0°C
D0°C ~ 2310°C1°C0°C-270°C: 2.0°C
270°C-2310°C: 2.0°C
RTDPt100(385)-200°C ~ 800°C0.1°C-200°C-0°C: 0.3°C
0°C-600°C: 0.2°C
600°C-800°C: 0.3°C
Pt200(385)-200°C ~ 630°C0.1°C-200°C-0°C: 0.3°C
0°C-630°C: 0.2°C
Pt500(385)-200°C ~ 630°C0.1°C-200°C-0°C: 0.3°C
0°C-630°C: 0.2°C
Pt1000(385)-200°C ~ 630°C0.1°C-200°C-0°C: 0.3°C
0°C-630°C: 0.2°C
Cu50-50°C ~ 150°C0.1°C-50°C-0°C: 0.8°C
0°C-150°C: 0.6°C

Note: Temperature coefficient: 0.1 × Basic accuracy/°C (temperature range <18°C or >28°C).

Maximum applied voltage at output terminal: approximately 30Vpk.

Maximum applied current at output terminal: approximately 25mA.

Load characteristics: capacitive load ≥0.01uF (DCV/OHM/TC/RTD).

8. Maintenance

8.1 Cleaning

To clean the instrument, wipe the case with a damp cloth and a mild detergent. Do not use abrasives or solvents. Ensure the instrument is completely dry before use.

8.2 Battery Replacement

Replace batteries when the low battery indicator appears on the display to ensure accurate measurements and proper functionality. Refer to Section 5.1 for battery installation instructions.

8.3 Storage

When not in use for extended periods, remove the batteries to prevent leakage and store the instrument in a cool, dry place away from direct sunlight and extreme temperatures.

9. Troubleshooting

  • No display or erratic readings: Check battery levels and replace if necessary. Ensure test leads are securely connected.
  • Inaccurate readings: Verify the correct output function and range are selected. Ensure proper connection of sensors/probes. Consider recalibration if accuracy issues persist after checking connections and settings.
  • No output: Confirm the device is powered on and the correct output function is selected. Check for any error messages on the display.

10. User Tips

  • The versatility of this signal generator means its 'capacity' depends entirely on the specific measurement type you are performing (e.g., resistance, thermocouple, or thermal resistance).
  • Remember that the device supports a wide array of thermocouple types (R, S, K, E, J, T, N, B, A, C, D) and RTD types (Pt100, Pt200, Pt500, Pt1000, Cu50). Always select the correct type in the settings for accurate simulation.
  • For precise measurements, especially with resistance, utilize the 4-wire connection method when possible to minimize lead resistance errors.

11. Manufacturing Process

The Victor 01S signal generator is manufactured using advanced techniques, including automated circuit board assembly to ensure high quality and reliability.

Video 11.1: Manufacturing process of circuit boards for the Victor 01S.

12. Support

For technical assistance, troubleshooting beyond this manual, or inquiries regarding your Victor 01S Temperature Source Signal Generator, please contact your local distributor or the manufacturer's customer support. Refer to the product packaging or qualification certificate for contact details.

Related Documents - 01S

Preview VICTOR 4105A Digital Grounding Resistance Meter Instruction Manual
Comprehensive instruction manual for the VICTOR 4105A Digital Earth Tester, covering overview, safety precautions, work principle, technical specifications, measurement methods, battery installation, and troubleshooting.
Preview VICTOR Electronic Testing and Measuring Instruments Catalog | Digital Multimeters, Oscilloscopes, and More
Explore the comprehensive range of VICTOR electronic testing and measuring instruments, including digital multimeters, clamp meters, oscilloscopes, function generators, and more. High-quality, reliable equipment for professional use.
Preview Victor 1226 Calculator Quick Start Guide
A quick start guide for the Victor 1226 calculator, covering setup, paper loading, time and date setting, troubleshooting, and available accessories.
Preview Victor Personal Massager User Manual and Joyhub App Guide
Comprehensive user manual for the Victor personal massager, detailing product features, app connectivity with Joyhub, usage instructions, care, warnings, and warranty information. Supports multiple languages.
Preview Victor 1226 Thermal Printing Calculator Instruction Manual
This manual provides instructions for operating the Victor 1226 Thermal Printing Calculator, covering its features, functions, and specifications. It details preparation, key identification, date and time functions, battery replacement, and various calculation examples.
Preview Victor Digital Anemometer Instruction Manual
Instruction manual for the Victor Digital Anemometer, detailing its functions, LCD display, operation, and specifications. Covers air velocity and temperature measurements, data hold, unit selection, and power management.