Yooning MP-2500 Microplate Centrifuge Instruction Manual
Yooning MP-2500 Microplate Centrifuge

General Information

Thank you for purchasing our Product.
This Manual for users contains function and operation of the Instrument. In order to use the instrument properly, please read this manual carefully before operation.

Please check the instrument and accessories with the packing list at the first time you open the instrument packing case.

Introduce

MP-2500 is a transient centrifuge designed for microplates, which is convenient for the wall drops. It is suitable for 96-well, 384 and small-capacity micro plates, as well as skirts and various standard PCR microplates.

Operation Guide

The Structure

The Structure

Fane Plate and Display

Fane Plate and Display

Key functions

Buttons Press the up and down key to set the time.
Buttons The Short button
Buttons The start/stop button

The Operation

Timer Setting

a) When the Instrument powers on, it starts self-checking with the sound of de…
b) About 1s later, 10:00 in the time display is the last set time.
c) Press Buttons or Buttons to set the timer. Long press to set the time quickly.

The Touch 

Press Buttons, it will run with the max. speed. When it reaches the maximum speed, the timer begins to work. It decelerates to stop when. releasing button

The Start/Stop

Close the lid after setting the time. Then press Buttons to run. When it reaches the maximum speed, the timing begins. Press the button during the running, it will decelerate to stop with 3 sound of de…

Note: 

  1. It doesn’t run with 2 sounds of de… and E1 is displayed if the cover not closing The time will be displayed after closing the cover.
  2. Do not open the cover before the rotor stops.
  3. Samples should be placed symmetrically in equal quantities

Safety

  1. place on a stable work platform
  2. Don’t move the instrument during operation

Clean

Clean the instrument with a dry cloth and a non-corrosive cleaning solution with a pH of 7 ± 1; Do not immerse the instrument in a liquid or use a liquid to wet the instrument; Ensure that all parts of the instrument are completely dry before running the instrument.

The Parameters

VoltageDC24V, 2APower48W
Speed2800rpmNoise≤60dB
Max. RCF600 xgSpeed Deviation≤10%
Time1~99sDeceleration rate≤6s
Dimension (mm)220x245x200Weight2.5kg

Troubles and Shootings

TroublesCausesShootings
The rotor not

working

1. The power cord not plugged in.Check the power cord
2. No power supply to the socketCheck the socket
3. The switch not turning onCheck the switch
4. The lid Not closedCheck the lid
Noise, severe vibration or abnormal noise after starting up1. Not placed on a horizontal platformPlace it on a horizontal platform
2. Loose screws 

Tighten the screws

3. Samples not placed symmetricallyPlace the sample symmetrically
E1 errorLid cant closeContact supplier
E2 errorOver speedContact supplier
E3 errorMotor stallingContact supplier

Performance Test

ItemMicroplate CentrifugeModelMP-2500
DateSerial No.
No.ContentMethodsStandardsResults
1Basic

Function

Visual checkValid 

□Qualified

2AppearanceVisual checkPass□Qualified
3LabelVisual checkPass□Qualified
4SpeedTachometer2800±10%rpm 

□Qualified

5Deceleration

Time

CalculagraphT≤6s 

□Qualified

 

6

Noise LevelSound Level Meter≤60dB□Qualified
7AmplitudeDial Gauge≤0.3mm 

□Qualified

8 

Continuous

Running Test

Operation30 min for 5 times without problem 

□Qualified

 

Result

 

Remarks:

 

QC:                                                                                       Confirmer:

Packing List

 

No.

 

Item

 

Model/Type

 

Quantity

 

Confirm

 

1

Micro Plate

Centrifuge

 

MP-2500

 

1

 

2

Adapters for

tubes

 

2

 

3

 

Power Line

 

DC24V

 

1

 

4

 

QC PASS Card

 

1

 

5

 

Manual

 

1

 

QC:                                                                           Date:

Hangzhou Yooning Instrument Co., Ltd.
Address: 4 floor, #6 Building, No.3 Shengang Road,
Yuhang District,Hangzhou, China
Customer Service:+86-571-86907383 / 87613616
Website: www.yooning.com

Company Logo

Documents / Resources

Yooning MP-2500 Microplate Centrifuge [pdf] Instruction Manual
MP-2500, MP-2500 Microplate Centrifuge, Microplate Centrifuge, Centrifuge

References

Leave a comment

Your email address will not be published. Required fields are marked *