WHALETEQ logoWECG400
Parallel Testing ECG Tester
for wearable device production lines
WHALETEQ WECG400 Parallel Testing ECG TesterQuick Start Guide

Hardware Overview

WHALETEQ WECG400 Parallel Testing ECG Tester - Hardware OverviewDimensions (W x L x H): 270 x 225 x 50 mm
Note: The WECG400 is rack-mountable with the 1U height (excluding the foot pads) design.

Installation

WHALETEQ WECG400 Parallel Testing ECG Tester - Installation

  1. (Optional) Ground the WECG400 by connecting one end of a grounding wire to the GND terminal and the other end to a suitable grounded object1.
  2. Connect the RA, LA, and RLD electrodes of DUTs respectively to the RA, LA, and RLD electrode terminals on the WECG400.
  3. Connect a computer to the WECG400 using a USB Type-A to USB Type-B cable.
  4. Plug a power adapter into an AC power source, then plug the power adapter cable into the WECG400’s power jack. Now the WECG400 is powered on.
  5. Activate the installed WECG400 software to conduct parallel or sequential testing2.
    Note: 1. For more information on grounding, please refer to Notes for Installation – Reduce environmental noise interference.
    2. For more information on testing, please refer to Operation.

Operation

Before starting the testing:

  • The WECG400 supports both parallel and sequential testing, and users can develop programs by using SDK to set up suitable testing methods.
  • The WECG400 is not a standalone model. Please connect the WECG400 to a PC/laptop to conduct testing.
  • Parallel testing
    1. Set the input impedance of the WECG400 to 10MΩ through PC.
    2. After connecting 4 DUTs to the WECG400, switch the WECG400 to the desired input impedance value and choose the test item.
    3. The WECG400 will simultaneously send signals to the 4 DUTs. After the test is complete, switch the input impedance of the WECG400 back to 10MΩ.
    4. Disconnect the tested DUTs and repeat the above steps for further testing.
    Note: Set the input impedance of the WECG400 to 10MΩ before each test can minimize the impact of electrostatic discharge (ESD) and surges from DUTs on the WECG400 and test results.
  • Sequential testing
    1. Connect the first DUT, then choose the test items and the DUT through PC to send signals to it.
    2. Connect the second and third DUTs, then choose the DUTs through PC while the first DUT is under testing.
    3. After completing testing on the first DUT, the WECG400 tests other DUTs one at a time by users’ designated order.
    4. Disconnect the tested DUTs, then connect new DUTs to the WECG400 and repeat the above steps.
    Note: If there is excessive noise during sequential testing, parallel testing is recommended.

Notes for Installation – Reduce environmental noise interference
WHALETEQ WECG400 Parallel Testing ECG Tester - interference

Note:

  1. Shorter cables are recommended by bringing a closer connection between the WECG400 and DUTs to reduce possible interference.
  2. The metal plate’s size should cover the WECG400 and the PC at least. It’s unnecessary to ground the metal plate itself.

LED Description

WHALETEQ WECG400 Parallel Testing ECG Tester - Description

Performance Test Software – Manual Mode

WHALETEQ WECG400 Parallel Testing ECG Tester - Manual Mode

  1. Click the ”Manual” tab.
  2. Select the desired waveform.
  3. Set the frequency.
  4. Set the amplitude.
  5. Set the pulse width*.
  6. Set the DC offset.
  7. Select desired input impedance.
  8. (Optional) Connect the signal ground inside the WECG400 to the chassis ground.
  9. Check RA or LA for testing.
  10. Click ”Play” to output configured signals.
    *Note: Pulse width is only available when Rectangle Pulse or Triangle Pulse is selected in the Waveform section.

Performance Test Software – Quick Mode: IEC 60601-2-47 test items

WHALETEQ WECG400 Parallel Testing ECG Tester - test items

  1. Click the ”Quick ” tab.
  2. Select the desired test item (dynamic range, frequency response, or input impedance).
  3. Set the values on the selected test item*.
  4. Check RA or LA for testing.
  5. Click ”Play” to output configured signals.
    *Note: Dropdown lists showing in light gray provide selectable values.

Performance Test Software – Quick Mode: Play raw data

WHALETEQ WECG400 Parallel Testing ECG Tester - raw data

  1. Click the ”Quick” tab.
  2. Check ”Play Raw Test”.
  3. Click the ”…” button to select a raw data file.
  4. (Optional) Check ”Loop” to play the raw data in a loop.
  5. Check RA or LA for testing.
  6. Click ”Play” to output raw data signals.

WHALETEQ WECG400 Parallel Testing ECG Tester - qr codewww.whaleteq.com

service@whaleteq.com
WHALETEQ WECG400 Parallel Testing ECG Tester - home icon8F., No. 125 Songjiang Rd., Zhongshan Dist., Taipei City 104474, Taiwan
WHALETEQ WECG400 Parallel Testing ECG Tester - callicon+886-2-2517-6255
WHALETEQ WECG400 Parallel Testing ECG Tester - icon1+886-2-2596-0702
Copyright © 2013-2024, All Rights Reserved.
WhaleTeq Co. LTD

WHALETEQ logo

WHALETEQ  is a registered trademark of WhaleTeq Co. LTD.
All other trademarks or trade names are the property of their respective holders.
All information, documentation, firmware, software utilities, and specifications are subject to change without prior notification by the manufacturer.

Documents / Resources

WHALETEQ WECG400 Parallel Testing ECG Tester [pdf] User Guide
WECG400 Parallel Testing ECG Tester, WECG400, Parallel Testing ECG Tester, Testing ECG Tester, ECG Tester