VORTEX B0C7PG8H72 One Handed Operation Aid
Product Specifications:
- Product Name: Vortex
- Intended Use: Spacer/valved holding chamber for delivery of aerosol medication with metered dose inhalers
- Manufacturer: PARI Respiratory Equipment, Inc.
- Contact Information: Tel: 1-804-253-7274, Email: product.info@pari.com
- Address: 2412 PARI Way, Midlothian, VA 23112
Product Usage Instructions
Assembling the Parts:
- Step 1:
- Step 4:
- Vigorously shake the inhaler before every use.
- Connect the inhaler to the connecting ring on the inhalation chamber.
- Step 5 (only if using a mask): Attach the mask or mouthpiece securely.
Instructions for Cleaning:
Washing by Hand:
- Clean the parts before first use and then clean and disinfect once a week thereafter.
- Remove the one-handed operation aid from the chamber.
- A. Hand Washing:
- Soak the part for 15 minutes in lukewarm water with liquid detergent. Agitate gently, then rinse in clear water.
- Rinse the part in clean water and shake out any excess water.
- Air dry thoroughly on a clean, lint-free cloth or drying rack before storing for future use.
- B. Dishwasher Cleaning:
- Place the part in the top shelf of the dishwasher for cleaning.
- Remove from the dishwasher, shake out excess water, and air dry as instructed above.
Disinfection: After cleaning, disinfect all parts by placing them in boiling water for 5 minutes. Dry as instructed above.
Frequently Asked Questions
Can I use Vortex with any metered dose inhaler?
Yes, Vortex is designed to work with most metered dose inhalers for aerosol medication delivery.
How often should I clean and disinfect Vortex?
You should clean and disinfect Vortex once a week after first use to maintain proper hygiene and functionality.
Product is non-sterile, not made with BPA or latex.
For use only with PARI Vortex® Non Electrostatic Holding Chamber.
READ ALL INSTRUCTIONS BEFORE USE.
Vortex is indicated for use as a spacer/valved holding chamber for use in delivery of aerosol medication with metered dose inhalers.
ASSEMBLING THE PARTS
Step 1:
- Push the one-handed operation aid onto the inhalation chamber from the back.
Step 2:
- Remove the cap on the metered dose inhaler.
- Shake the metered dose inhaler vigorously before every use.
Step 3:
- Insert the metered dose inhaler in the connecting ring on the inhalation chamber.
Step 4:
- Position the one-handed operation aid so that its round end encloses the metered dose inhaler. To do this, you may have to push the one-handed operation aid farther forwards or backwards.
Step 5 (only if using a mask):
Attach the mask (pediatric mask)
- Take the protective cap off the mouthpiece of the VORTEX.
- Attach the mask to the VORTEX mouthpiece.
OR
Attach the mask (adult mask)
- Take the protective cap off the mouthpiece of the VORTEX.
- Fit the mask connector into the opening of the PARI adult mask soft.
- Attach the mask to the mouthpiece of the VORTEX with the mask connector.
INSTRUCTIONS FOR CLEANING
DISINFECTION: To disinfect after cleaning, place all parts in boiling water for 5 minutes. Dry as instructed above.
PARI Respiratory Equipment, Inc.
Tel: 1-804-253-7274
www.pari.com | product.info@pari.com
2412 PARI Way, Midlothian, VA 23112
© 2024 PARI Respiratory Equipment, Inc.
051D7019 Rev A 01-24