UNI-T UPO1000HD Series High Resolution Oscilloscopes User Guide

UPO1000HD Series High Resolution Oscilloscopes

Product Information

Specifications:

  • Model: UPO1000HD Series
  • Analog channel number: 2 or 4
  • Analog bandwidth: 80MHz or 150MHz

Product Usage Instructions

1. General Inspection

Before using the UPO1000HD series oscilloscopes for the first
time, it is recommended to inspect the instrument for any damages
caused by transport.

2. Getting Started Guide

This chapter is designed to introduce the usage of the
oscilloscope for the first time, including the front and rear
panels, user interface, and touch screen function.

2.1. Before Use

To perform a quick verification of the instrument’s normal
operations:

  • Adjust the supporting feet properly for stable placement and
    operation.
  • Connect the oscilloscope to a DC power supply of at least
    12V/3A using the provided power adaptor.
  • Press the power soft switch key to boot up the
    oscilloscope.
  • Connect the provided compensating signal probe to CH1 using the
    BNC connector.
Supporting Feet

Unfold the supporting feet for stable placement and fold them
for storage or transport.

Power Adaptor Specifications
Input Output
100V – 240V, 50Hz – 60Hz, 1.6A Max DC, 12V, 3A, 65W
Connect Probe

Connect the probe to CH1 and the ground alligator clip to the
ground terminal of the compensating signal connection clip. The
output signal should have an amplitude of approximately 3Vpp with a
default frequency of 1kHz.

FAQ

What should I do if the oscilloscope does not power on?

If the oscilloscope does not power on, please check the power
connection and ensure that the power adaptor is correctly plugged
into a power source within the specified voltage range.

How can I access software updates for my UPO1000HD Series
oscilloscope?

To access software updates for your oscilloscope, you can scan
the QR code in the manual or visit the manufacturer’s website to
download the relevant software/firmware updates.

“`

UPO1000HD Series High-Resolution Oscilloscopes
Quick Start Guide
This manual applies to model: UPO1000HD Series
V1.0 April 2025

Limited Warranty and Liability
UNI-T guarantees that the Instrument product is free from any defect in material and workmanship within three years from the purchase date. This warranty does not apply to damages caused by accident, negligence, misuse, modification, contamination, or improper handling. If you need a warranty service within the warranty period, please contact your seller directly. UNI-T will not be responsible for any special, indirect, incidental, or subsequent damage or loss caused by using this device. For the probes and accessories, the warranty period is one year. Visit instrument.uni-trend.com for full warranty information.

Scan to Download relevant document, software, firmware and more.

Register your product to confirm your ownership. You will also get product notifications, update alerts, exclusive offers and all the latest information you need to know.

is the licensed trademark of UNI-TREND TECHNOLOGY (CHINA) CO., Ltd. UNI-T products are protected under patent laws in China and internationally, covering both granted and pending patents. Licensed software products are the properties of UNI-Trend and its subsidiaries or suppliers, all rights reserved. This manual contains information that replaces all earlier published versions. The product information in this document subject to update without notice. For more information on UNI-T Test & Measure Instrument products, applications, or service, please contact UNI-T instrument for support, the support center is available on www.uni-trend.com ->instruments.uni-trend.com https://instruments.uni-trend.com/ContactForm/

Headquarter
UNI-TREND TECHNOLOGY (CHINA) CO., Ltd. Address: No.6, Industrial North 1st Road, Songshan Lake Park, Dongguan City, Guangdong Province, China Tel: (86-769) 8572 3888

Europe UNI-TREND TECHNOLOGY EU GmbH Address: Affinger Str. 12 86167 Augsburg Germany Tel: +49 (0)821 8879980

North America
UNI-TREND TECHNOLOGY US INC. Address: 3171 Mercer Ave STE 104, Bellingham, WA 98225 Tel: +1-888-668-8648

Copyright © 2025 by UNI-Trend Technology (China) Co., Ltd. All rights reserved.

Quick Start Guide
1. UPO1000HD Series Oscilloscopes

UPO1000HD Series

UPO1000HD series digital oscilloscopes includes the following models.

Model

Analog channel number

Analog bandwidth

UPO1152HD

2

150MHz

UPO1154HD

4

150MHz

UPO1082HD

2

80MHz

UPO1084HD

4

80MHz

Instruments.uni-trend.com

4 / 24

Quick Start Guide
2. Getting Started Guide

UPO1000HD Series

This chapter is to introduce on using the oscilloscope for the first time, the front and rear panels, the user interface, as well as touch screen function.

2.1. General Inspection
It is recommended to inspect the instrument follow the steps below before using the UPO1000HD series oscilloscopes for the first time. (1) Check for Damages caused by Transport
If the packaging carton or the foam plastic cushions are severely damaged, please contact the UNI-T distributor of this product immediately. (2) Check Accessories Please check appendix for the list of accessories. If any of the accessories are missing or damaged, please contact UNI-T or local distributors of this product. (3) Machine Inspection If the instrument appears to be damaged, not working properly, or has failed the functionality test, please contact UNI-T or local distributors of this product. If the equipment is damaged due to shipping, please keep the packaging and notify both the transportation department and UNI-T distributors, UNI-T will arrange maintenance or replacement.

2.2.Before Use
To perform a quick verification of the instrument’s normal operations, please follow the steps below. (1) Adjust the Supporting Feet
Adjust the supporting feet properly to use them as stands, tilting the oscilloscope upward for stable placement, improved operation, and better observation. You can also fold the supporting feet when the instrument is not in use for easier storage or transport.

Instruments.uni-trend.com

5 / 24

Quick Start Guide

UPO1000HD Series

Supporting Feet Unfold the supporting feet

Fold the supporting Feet

(2) Connect to Power Supply The power requirements of the oscilloscope are DC, at least 12 V/3 A. Please use the power adaptor provided in the accessories to connect the oscilloscope to the AC power source (100 V – 240 V, 50 Hz – 60 Hz).

Power Connector

Power Adaptor Specifications Terminal Input output

Specification 100V – 240 V, 50 Hz – 60 Hz, 1.6 A Max
DC, 12 V, 3 A, 65 W

(3) Boot Check Press the power soft switch key and the indicator should change from red to green. The oscilloscope will show a boot animation, and then enter the normal interface.
(4) Connect Probe This oscilloscope provides 1 pieces of compensating signal probe. Connect the BNC of the probe to the BNC of oscilloscope’s CH1 and connect the probe to the “Probe Compensating Signal Connection Clip”, and then connect the ground alligator clip of the probe with the ground

Instruments.uni-trend.com

6 / 24

Quick Start Guide

UPO1000HD Series

terminal of compensating signal connection clip. The output of compensating signal connection clip: amplitude is approximately 3 Vpp, with a default frequency of 1 kHz.

Probe compensating signal connection clip

Ground terminal
Probe Compensating Signal Connection Clip and Ground Terminal
(5) Function Check Press the AUTO key, a square wave (amplitude 3 Vpp, frequency 1 kHz) should appear on the screen. Repeat the step 4 to check all channels.
(6) Probe Compensation When the probe is connected to any input channel for the first time, this step might be adjusted to match the probe and the input channel. Probes that are not compensated may lead to measurement errors or mistake. Please follow the following steps to adjust the probe compensation. Set the attenuation coefficient in the probe menu to 10x and set the probe switch to 10x. Connect the oscilloscope probe to CH1. If using the probe’s hook tip, ensure it makes stable contact. Connecting the probe to the “Probe Compensating Signal Connection Clip” of the oscilloscope and connect the ground alligator clip to the ground terminal of probe compensating signal connection clip. Open CH1 and press the AUTO key. View the displayed waveform, as shown in the following figure.

Excessive Compensation Correct Compensation Insufficient Compensation Probe Compensation Calibration
If the displayed waveform resembles the above Insufficient Compensation or Excessive Compensation, use a non-metallic screwdriver to adjust the probe’s variable capacitance until the display matches the Correct Compensation waveform.
WarningTo avoid electric shock when using the probe to measure high voltage, please ensure that the probe insulation is in good condition and avoid physical contact with any metallic part of the probe.

Instruments.uni-trend.com

7 / 24

Quick Start Guide
2.3.Front Panel
1

23

UPO1000HD Series

4
5 7 8 6

12 11

10 9

Front Panel

Table 1 Button and Terminal on Front Panel

No.

Description

1

Display area

2

Measurement analysis area

3

Multi-function area

4

Control area

5

Horizontal control area

6

Trigger control area

No.

Description

7

Shortcut keys and search navigation

8

Vertical control area

Probe compensating signal connection 9
clip and ground terminal

10

Analog channel input terminal

11

USB HOST port

12

Power soft switch key

1 Probe compensation signal connection terminal and ground terminalConnect the BNC end of the probe to the BNC connector on Channel 1 of the oscilloscope. Attach the probe tip to the “Probe Compensation Signal Terminal,” and connect the probe’s ground alligator clip to the “Ground Terminal” beneath the probe compensation signal terminal. This setup will output the oscilloscope’s internal signal. Please refer to the “Before Use – Connect Probe” section.
2 Analog channel input terminal: Connect the oscilloscope probe or BNC cable to these BNC

Instruments.uni-trend.com

8 / 24

Quick Start Guide
connectors to input a signal into the oscilloscope.
2.4. Rear Panel
1

UPO1000HD Series
3 4 5

Table 2 Terminal on Rear Panel

No.

Description

1

AUX Out

2

USB Device

3

LAN port

Rear Panel

65

No.

Description

4

USB Type-C Power Connector

5

Safety lock

6

Ground terminal

2.5.Operation Panel

(1) Vertical Control 1 , 2 , 3 , 4 Analog channel setting key respectively represents CH1, CH2, CH3, and CH4. Four channel’s tab are identified by different colors and it also corresponding to the colors of waveforms on the screen and the channel input connectors. Press any keys to enter the related channel menu (activate or disable the channel). Math Press this key to open the mathematical operation menu
to perform math operation (add, subtract, multiply, divide), digital filter and advanced operation. Ref Loading the reference waveform from “Local or USB”, so the measured waveform can

Instruments.uni-trend.com

9 / 24

Quick Start Guide

UPO1000HD Series

compare with the reference waveform. Scale Vertical scale rotary knob is used to adjust the vertical scale in the current channel.
Turn clockwise to decrease the scale, turn counterclockwise to increase the scale. The amplitude of waveform will increase or decrease with the adjustment and the scale at the

bottom of screen

will change in real-time. The vertical scale is step with 1-2-5,

press this rotary knob to adjust the vertical scale between coarse tuning and fine tuning.

PositionVertical position rotary knob is used to move the vertical position of the waveform

in the current channel. Press this rotary knob to move the channel position back to the

vertical midpoint.

(2) Horizontal Control

Scale Horizontal scale rotary knob is used to adjust all channel time base. During the adjustment, the waveform

is compressed or extended in horizontal show on the

screen and the horizontal scale value will change in real-time. The time base is step with 1-2-5, press this rotary knob to adjust the horizontal scale between coarse tuning and fine tuning. PositionHorizontal position rotary knob is used to move the trigger point to left or right side that relative to the center of the screen. During the adjustment, all channel waveforms move to the left or right side and the horizontal shift value on the top of the screen

will change in real-time. Press this rotary knob to move the current position back to the horizontal midpoint. (3) Trigger Control
SlopePress the Slope key to switch the trigger edge: rising edge, falling edge, or any edge, The corresponding edge’s indicator light will illuminate when switching the edge.
Menu Press the Menu key to enter the trigger menu. SourcePress the Source key to switch the trigger source.
The panel numbers 1 to 4 correspond to channels CH1 to CH4. When switching the source, the corresponding number’s indicator light will illuminate. Position Trigger level rotary knob, turn clockwise to increase the level, turn counterclockwise to decrease the level. During the adjustment, the trigger level

on the top right will change in real-time. When the trigger is single

Instruments.uni-trend.com

10 / 24

Quick Start Guide

UPO1000HD Series

level, press this rotary knob to turn the trigger level to the trigger signal and quickly turn to 50%. When the trigger status is READY or TRIGED, the corresponding Ready or Trig’d indicator light in the trigger control area will illuminate.

(4) Control Area

Auto/NormalPress the Auto/Normal key to switch the

oscilloscope trigger mode between Auto and Normal. Run/StopThis key is used to set the operating mode of

the oscilloscope to Run or Stop. In the RUN state, the key illuminates green. In the STOP state, the key illuminates red. ForceWhen the trigger mode is Normal or Single, press the force trigger key to generate

a trigger. AutosetAfter pressing the Autoset key, the oscilloscope automatically adjusts the vertical

scale, time base, and trigger mode based on the input signal to display the optimal waveform. SinglePress the Single key to set the oscilloscope’s trigger mode to Single, the key

illuminates orange.

(5) Measurement Analysis Area

MeasurePress the Measure key to enter the

measurement menu and set the counter, voltmeter, parameter snapshot, measurement statistics, add measurement, clear measurement, and global settings. CursorsPress the Cursor key to enter the cursor measurement menu and set time, voltage,

and screen measurement for each source. AnalyzePress the Analyze to open the analysis module menu, where the user can access

functions such as the voltmeter, counter, power analysis, waveform recording, and pass/fail test.

(6) Multi-function Area

Multipurpose AMulti-function knob A. When a numerical menu is selected in the function popup window, allowing the user to adjust values using the knob.
Multipurpose BMulti-function knob B. When setting

Instruments.uni-trend.com

11 / 24

Quick Start Guide

UPO1000HD Series

numerical parameters in a text box, the user can rotate this knob to shift the selected digit.

(7) Shortcut key area

Quick: Press the Quick key to quickly save the screen

waveform as a PNG bitmap to the default or a custom image

path.

: Press the Clear key to clear all recalled waveforms and parameter measurement

statistics from the screen. Touch/Lock: Press the Touch/Lock key to disable the touchscreen; the backlight will turn

on. To enable the touchscreen again, press the key once more, and the backlight will turn

off. Default: Restore factory settings. Press the Default key to reset all oscilloscope settings to

their default values.

(8) Home Menu

Press the Home icon at the top-right of the screen to open “Home” quick menu, including: Measure, Cursor, DVM, Counter, Math, Ref, AreaHist, Search, Decode, Navigation, Record, Mark, FFT, PwrAnaly, Pass/Fail, Storage, Display, Utility, Help, and Area draw. The quick menu key allows the user to quickly enter the corresponding function menu.
Home Menu

Instruments.uni-trend.com

12 / 24

Quick Start Guide

2.6. User Interface

12

3

4

5

UPO1000HD Series

6

7

16

15

14

Table 3 User Interface

13

12

11 10

98

No.

Description

No.

Description

1

Waveform display window

9

Volts/div signal bar

2

Trigger state icon

10

Multi-window display area

3

Horizontal base label

11

Ref label

Sampling rate and memory

4

12

depth label

FFT label

5

Trigger info label

13

Math label

6

Function toolbar

14 Measured result display window

7

Home menu

15

Channel label

8

Notification

16

Analog channel label

2.7.Help System
The help system provides information about the function keys (including menu keys) on the front panel. The user can access the help system by following these steps. In Home menu, click on the help icon to open the help menu.

Instruments.uni-trend.com

13 / 24

Quick Start Guide

UPO1000HD Series

In each function menu popup, click on the help icon on the top right to open the relevant help menu.
The help screen is divided into two parts, the left side is Help Options, and the right side is Help Display Area. By selecting a help option, the user can view all related help content on the right.

Instruments.uni-trend.com

14 / 24

Quick Start Guide
3. Parameter Setting

UPO1000HD Series

UPO1000HD series supports use the Multipurpose A (Multipurpose rotary knob) and touch screen to set the parameter. The setting steps as follows. (1) Multipurpose rotary knob
For the parameter of time and voltage, once the parameter is selected, rotating the Multipurpose rotary knob on the front panel to enter the parameter value. (2) Touch screen Once the parameter or text field has been selected, double-click to pop up the virtual keyboard to enter the parameter value, label name, or file name. a. Enter character string
When naming a file or folder, use the character keyboard to enter a string. b. Text field
Enter text: input letters, numbers, and special characters, with a maximum length of 16 characters. c. Clear key Press the Clear key to delete all contents in the text field. d. Caps key Press the Caps key to switch between the upper and lower case. e. Tab key Press the Tab key to enter 2 spaces at a time. f. Shift key Press the Shift key to switch among number, special character, upper and lower case. g. Arrow keys (left, right) If part of the content needs to be changed, press the , key to move the cursor to left or right and then to edit the content. h. Space key Press the Space key to enter one space in the text field. i. Backspace key Press the Backspace key to delete a single character. This key is used to delete a character when the text field contains a large amount of content. j. Enter key Once the content has been entered, press the Enter key to confirm the setting and close the virtual keyboard.

Instruments.uni-trend.com

15 / 24

Quick Start Guide
Keyboard name
Tab key Caps key
Shift key

Input area

UPO1000HD Series
Clear key Backspace key Enter key

Virtual keyboard Space Key

Arrow keys: left, right

(3) Enter numeric value When setting or editing a parameter, use the numeric keyboard to enter the numeric value. 1. Click the number or unit to enter

Arrow keys: left, right

Input area

Backspace key

Clear Maximum key Default key Minimum key Enter key

Numeric

Unit

After entering all the values and selecting the desired units, the numeric keypad will automatically close, completing the parameter setting. Additionally, the user can manually close the numeric keypad by clicking the confirm key, in which case the unit will default to the preset unit. On the numeric keypad, the user can also perform the following operations:

Instruments.uni-trend.com

16 / 24

Quick Start Guide

UPO1000HD Series

a. Delete the entered parameter value. b. Set the parameter to the maximum or minimum value (sometimes specifically the
maximum or minimum value for the current state). c. Set the parameter to the default value. d. Clear the parameter input field. e. Move the cursor to modify the parameter value. 2. Enter binary, hexadecimal system value During the decoding trigger, use the numeric keypad to enter binary or hexadecimal values for data and address settings. Enter Method: Tap to select the number or text field to be edited, and then use the numeric keypad to enter the desired numeric or letter values.

Binary input

Hexadecimal input

Default key Maximum key Minimum key Ok key

Numeric keyboard

Arrow keys

(4) After entering all the values and pressing the OK button, the numeric keypad will automatically close, completing the parameter setting. Additionally, on the numeric keypad, you can perform the following operations: a. Move the cursor to modify the parameter value. b. Set the parameter to the maximum or minimum value (sometimes specifically for the current state). c. Set the parameter to the default value. d. Clear the parameter input field. e. Delete the entered parameter value.

Instruments.uni-trend.com

17 / 24

Quick Start Guide
4. Touch Screen

UPO1000HD Series

UPO1000HD series provides 7-inch super capacitive touch screen, multiple point touch control and gesture control. UPO1000HD has easily operating system with flexible and high-sensitive touch screen features for great waveform display and excellent user experience. Touch control function includes tap, squeeze, drag, and rectangle drawing. NoteThe menu displayed on the screen of the oscilloscope can all use the touch control function. (1) Tap
Use one finger to slightly tap on an icon or a word on the screen as shown in the following figure. Tap gestures can be used for: Tap the menu display on the screen and then to setup. Tap the function icon on the top right corner to open the corresponding function. Tap the pop-up numeric keyboard to set the parameter. Tap the virtual keyboard to set the label name and file name.
Tap a message to pop up a × button on the top right corner to close the pop-up window.
Tap another window displayed on the screen and then set up. Tap a message to reveal help key in the top right corner. Clicking this key will open the
corresponding function’s help menu, providing detailed guidance on the selected feature.

Tap Gestures
(2) Pinch Squeeze two fingers together or separate. Pinch gesture can zoom out or zoom in the waveform. To zoom out the waveform, pinch two fingers together and then slide them apart; to zoom in, spread two fingers apart and then pinch them together, as shown in the following figure. Pinch gestures can be used for: Adjust the horizontal time base of waveform by squeezing in the horizontal direction. Adjust the vertical time base of waveform by squeezing in the vertical direction.

Instruments.uni-trend.com

Pinch Gestures

18 / 24

Quick Start Guide

UPO1000HD Series

(3) Drag
Use one finger to press and drag the selected item to the aimed position as shown in the following figure. Drag gestures can be used: Drag the waveform to change the waveform position. Drag the window to change the window position.
Drag the cursor to change the cursor position.

Drag Gestures
(4) Rectangle Drawing Open the HOME menu and click the icon Rectangle Drawing to enable the function, drag your finger to draw a rectangle on the screen as shown in Figure (a), (b), move the finger, a menu will appear on the screen, at this point, Region A, Region B, Intersection, Non-intersect can be selected. Drag your finger from bottom right to the top left on the screen to draw the trigger area.

(a)

(b)

Rectangle Drawing Gesture

Select Zone A

Draw the trigger zone A

Open the trigger zone A

Open zone trigger menu

Select Zone B

Draw the trigger zone B

Open the trigger zone B

Open one trigger menu

Note: Click on Rectangle Drawing to step through rectangle drawing and operating waveform

mode. Click on rectangle drawing, if the icon shows , it indicates that rectangle drawing

mode is enabled; if the icon shows , it indicates that operating waveform mode is enabled.

Instruments.uni-trend.com

19 / 24

Quick Start Guide
5. Remote Control

UPO1000HD Series

UPO1000HD series digital oscilloscopes can communicate with a PC via USB and LAN port for remote control. Remote control is implemented using SCPI (Standard Commands for Programmable Instruments). UPO1000HD series has three methods for remote control. (1) Custom Programming
The user can perform the programming control on the oscilloscope through SCPI (Standard Commands for Programmable Instruments). For detailed descriptions on command and programming, please refer to UPO1000HD Series High-Resolution Oscilloscopes-Programming Manual. (2) PC Software Control (Instrument manager) The user can use a PC software to remotely control the oscilloscope. The instrument manager can display the oscilloscope screen in real time and control the operation with the mouse. It is recommended to use the PC software provided by UNI-T. It can be downloaded from UNI-T official website (https://www.uni-trend.com). Operating steps: Setup the communication between the instrument and a PC. Open the instrument manager software and search the instrument source. Right-click to open the oscilloscope, operate the instrument manager to remotely control
the oscilloscope (refer to Instrument Manager-User’s Manual for more details). This device supports remote control through communication with a computer via USB, LAN interfaces. Remote control is implemented using the SCPI command set, allowing users to operate and configure the device remotely. (3) Web Control Once the network is connected, users can access a web page via the IP address. After logging in with the username and password, they can control the device. The Web Control feature displays the instrument’s screen interface in real-time. It supports web access from PCs, smartphones, and iPads, and allows for both internal and external remote control of the device.

Instruments.uni-trend.com

20 / 24

Quick Start Guide
6. Troubleshooting

UPO1000HD Series

(1) If the oscilloscope remains on a black screen without any display when pressing the power soft key. a. Check whether the power plug is properly connected, and the power supply is normal. b. Check whether the power switch is turned on. If the power switch is turned on, the power soft key on the front panel should be green. When the power soft key is enabled, the power soft key should be blue, and the oscilloscope will make an active sound. There should be a normal relay rattle when the soft switch key is pressed. c. If the relay has sound, it indicates that the oscilloscope is normal boot-up. Press the Default key and press the Yes key, if the oscilloscope returns to normal, indicating that the backlight brightness is set too low. d. Restart the oscilloscope after completing the above steps. e. If the product still does not work properly, contact the UNI-T Service Center for assistance.
(2) After signal acquisition, the waveform of the signal does not appear on the screen. a. Check whether probe and DUT are connected properly. b. Check whether the signal output channel is open. c. Check whether the signal connecting line is connected to analog channel. d. Check whether the signal source has DC offset. e. Plug out the connected signal, to check whether the base line is within the screen range (If not, please perform self-calibration). f. If the product still does not work properly, contact the UNI-T Service Center for assistance.
(3) The measured voltage amplitude value is 10 times larger or 10 times smaller than the actual value. Check whether the channel probe attenuation coefficient settings are consistent with the used probe attenuation rate.
(4) There is a waveform display but not stable. a. Check the trigger settings in trigger menu whether is consistent with the actual signal input channel. b. Check the trigger type: the general signals should use “Edge” trigger. The waveform can only be displayed stably if the trigger mode is set correctly. c. Try to change trigger coupling to HF rejection or LF rejection, to filter out the high-frequency or low-frequency noise that interfere with the trigger.
(5) No waveform is displayed after pressing the Run/Stop key.

Instruments.uni-trend.com

21 / 24

Quick Start Guide

UPO1000HD Series

a. Check whether the trigger mode is set to Normal or Single and verify if the trigger level exceeds the waveform range.
b. If the trigger mode is normal or single and the trigger level is in the center, set the trigger mode to Auto.

c. Press the Autoset key to automatically complete the above settings.

(6) Waveform refresh is very slow. a. Check whether the acquisition method is set to Average and if the average times are large. b. Check whether the memory depth is maximum c. Check whether the trigger holdoff is large. d. Check whether it is normal trigger and is slow time base. All the above will lead to slow waveform refresh, it is recommended to restore the factory settings, then the waveform can be refreshed normally.

Instruments.uni-trend.com

22 / 24

110401113347X

1

2

3

4

5

6

7

DWH

CHK
APPRO

:148×210±1mm. 128G 60g ,…

REV.0
MODEL : UPO1000HD

Part NO. 1 1 0 4 0 1 1 1 3 3 4 7 X
()
UNI-TREND TECHNOLOGY (CHINA) CO.,LTD