MSO3000HD Series High Definition Oscilloscopes
“
Specifications
- Model: MSO3000HD Series
- MSO3054HD:
- Analog channel number: 4
- Analog bandwidth: 500 MHz
- MSO3034HD:
- Analog channel number: 4
- Analog bandwidth: 350 MHz
- MSO3024HD:
- Analog channel number: 4
- Analog bandwidth: 200 MHz
Product Usage Instructions
1. General Inspection
Before using the MSO3000HD series oscilloscope for the first
time, it is recommended to follow these steps:
- Check for damages caused by transport.
2. Before Use
To perform a quick verification of the instrument’s normal
operations:
- Connecting to the Power Supply:
- Frequency: 50 Hz/60 Hz or 400 Hz
- Boot Check:
- Press the power soft switch key and ensure the indicator
changes from red to green. - The oscilloscope will display a boot animation and then enter
the normal interface.
- Press the power soft switch key and ensure the indicator
- Connecting Probe:
- This oscilloscope provides 2 pieces of compensating signal
probe. - Connect the BNC of the probe to the BNC of oscilloscope’s
CH1. - Connect the probe to the probe compensating signal connection
clip. - Connect the ground alligator clip of the probe with the ground
terminal of compensating signal connection clip. - The output of compensating signal connection clip: amplitude is
approximately 3 Vpp, with a default frequency of 1 kHz.
- This oscilloscope provides 2 pieces of compensating signal
- Function Check:
- Press the AUTO key, a square wave (amplitude 3 Vpp, frequency 1
kHz) should appear on the screen. - Repeat the step to check all channels.
- If the displayed waveform does not match the expected square
waveform, perform the ‘Probe Compensation’ procedure.
- Press the AUTO key, a square wave (amplitude 3 Vpp, frequency 1
FAQ
Q: What should I do if the displayed waveform does not match
the expected square waveform during function check?
A: If the displayed waveform is incorrect, perform the ‘Probe
Compensation’ procedure described in the user manual. Adjust the
probe’s variable capacitance until the display matches the correct
compensation waveform.
“`
MSO3000HD Series High-Definition Oscilloscopes
Quick Start Guide
This manual applies to model: MSO3000HD Series
V1.2 May, 2025
1 / 25
Quick Start Guide
MSO3000HD Series
Limited Warranty and Liability
UNI-T guarantees that the Instrument product is free from any defect in material and workmanship within three years from the purchase date. This warranty does not apply to damages caused by accident, negligence, misuse, modification, contamination, or improper handling. If you need a warranty service within the warranty period, please contact your seller directly. UNI-T will not be responsible for any special, indirect, incidental, or subsequent damage or loss caused by using this device. For the probes and accessories, the warranty period is one year. Visit instrument.uni-trend.com for full warranty information.
Scan to Download relevant document, software, firmware and more.
Register your product to confirm your ownership. You will also get product notifications, update alerts, exclusive offers and all the latest information you need to know.
is the licensed trademark of UNI-TREND TECHNOLOGY (CHINA) CO., Ltd. UNI-T products are protected under patent laws in China and internationally, covering both granted and pending patents. Licensed software products are the properties of UNI-Trend and its subsidiaries or suppliers, all rights reserved. This manual contains information that replaces all earlier published versions. The product information in this document subject to update without notice. For more information on UNI-T Test & Measure Instrument products, applications, or service, please contact UNI-T instrument for support, the support center is available on www.uni-trend.com ->instruments.uni-trend.com https://instruments.uni-trend.com/ContactForm/
Headquarter
UNI-TREND TECHNOLOGY (CHINA) CO., Ltd. Address: No.6, Industrial North 1st Road, Songshan Lake Park, Dongguan City, Guangdong Province, China Tel: (86-769) 8572 3888
Europe
UNI-TREND TECHNOLOGY EU GmbH Address: Affinger Str. 12 86167 Augsburg Germany Tel: +49 (0)821 8879980
North America
UNI-TREND TECHNOLOGY US INC. Address: 3171 Mercer Ave STE 104, Bellingham, WA 98225 Tel: +1-888-668-8648
Copyright © 2025 by UNI-Trend Technology (China) Co., Ltd. All rights reserved.
Quick Start Guide
2. MSO3000HD Series Oscilloscopes
MSO3000HD Series
MSO3000HD series high-definition oscilloscopes includes three models.
Model MSO3054HD
Analog channel number 4
Analog bandwidth 500 MHz
Digital
MSO3034HD
4
350 MHz
MSO3024HD
4
200 MHz
Standard Option ×Not support
Gen
Instruments.uni-trend.com
4 / 24
Quick Start Guide
3. Getting Started Guide
MSO3000HD Series
This chapter is to introduce on using the MSO3000HD series oscilloscope for the first time, the front and rear panels, the user interface, as well as touch screen function.
3.1. General Inspection
It is recommended to inspect the instrument follow the steps below before using the MSO3000HD series oscilloscopes for the first time. (1) Check for Damages caused by Transport
If the packaging carton or the foam plastic cushions are severely damaged, please contact the UNI-T distributor of this product immediately. (2) Check Accessories Please check appendix for the list of accessories. If any of the accessories are missing or damaged, please contact UNI-T or local distributors of this product. (3) Machine Inspection If the instrument appears to be damaged, not working properly, or has failed the functionality test, please contact UNI-T or local distributors of this product. If the equipment is damaged due to shipping, please keep the packaging and notify both the transportation department and UNI-T distributors, UNI-T will arrange maintenance or replacement.
3.2.Before Use
To perform a quick verification of the instrument’s normal operations, please follow the steps below. (1) Connecting to the Power Supply
Refer to Connecting Power Supply, use the assembled power line or other power line that meets the local country standards to connect the oscilloscope. When the power switch on the rear panel is not opened, the power soft indicator in the left bottom on the rear panel is extinguished, which indicates this soft switch key is no-effect. When the power switch on the rear panel is opened, the power soft indicator in the left bottom on the rear panel is illuminated with red, and then press the soft switch key to enable the oscilloscope.
Instruments.uni-trend.com
5 / 24
Quick Start Guide
Voltage Range 100 V-240 VAC (fluctuant±10%) 100 V-120 VAC (fluctuant±10%)
Frequency 50 Hz/60 Hz
400 Hz
MSO3000HD Series
(2) Boot Check Press the power soft switch key and the indicator should change from red to green. The oscilloscope will show a boot animation, and then enter the normal interface.
(3) Connecting Probe This oscilloscope provides 2 pieces of compensating signal probe. Connect the BNC of the probe to the BNC of oscilloscope’s CH1 and connect the probe to the “probe compensating signal connection clip”, and then connect the ground alligator clip of the probe with the ground terminal of compensating signal connection clip. The output of compensating signal connection clip: amplitude is approximately 3 Vpp, with a default frequency of 1 kHz.
Probe compensating signal connection clip 1, 2
Ground terminal Probe Compensating Signal Connection Clip and Ground Terminal
(4) Function Check Press the AUTO key, a square wave (amplitude 3 Vpp, frequency 1 kHz) should appear on the screen. Repeat the step 3 to check all channels. If the displayed square waveform does not match the one shown above, please perform the ‘Probe Compensation’ procedure described in the next section.
(5) Probe Compensation When the probe is connected to any input channel for the first time, this step might be adjusted to match the probe and the input channel. Probes that are not compensated may lead to measurement errors or mistake. Please follow the following steps to adjust the probe compensation. Set the attenuation coefficient in the probe menu to 10x and set the probe switch to 10x. Connect the oscilloscope probe to CH1. If using the probe’s hook tip, ensure it makes stable contact. Connecting the probe to the “probe compensation signal connection clip” of the oscilloscope and connect the ground alligator clip to the ground terminal of probe compensating signal connection clip. Open CH1 and press the AUTO key.
Instruments.uni-trend.com
6 / 24
Quick Start Guide
View the displayed waveform, as shown in the following figure 2.
MSO3000HD Series
Excessive Compensation Correct Compensation Insufficient Compensation Probe Compensation Calibration
If the displayed waveform resembles the above “Insufficient Compensation” or “Excessive Compensation”, use a non-metallic screwdriver to adjust the probe’s variable capacitance until the display matches the “Correct compensation” waveform.
WarningTo avoid electric shock when using the probe to measure high voltage, please ensure that the probe insulation is in good condition and avoid physical contact with any metallic part of the probe.
3.3.Front Panel
1
23
4
5 6 7
14
13
12
11
10
9
8
Front Panel
Instruments.uni-trend.com
7 / 24
Quick Start Guide
MSO3000HD Series
Table 1 Front Panel Description
No.
Description
No.
Description
Probe compensating signal connection
1
Display area
8
clip and ground terminal
2
Measurement analysis area
9
Vertical control area
3
Multi-function area
10
Analog channel input terminal
4
Control area
11
Digital channel input terminal
5
Horizontal control area
12
Gen output port
6
Trigger control area
13
USB HOST port
7
Signal output area
14
Power soft switch key
Probe compensation signal connection terminal and ground terminalConnect the BNC end of
the probe to the BNC connector on Channel 1 of the oscilloscope. Attach the probe tip to the
“Probe Compensation Signal Terminal,” and connect the probe’s ground alligator clip to the
“Ground Terminal” beneath the probe compensation signal terminal. This setup will output the
oscilloscope’s internal signal. Please refer to the ” Before Use – Connecting the Probe” section.
Analog channel input terminal: Connect the oscilloscope probe or BNC cable to these BNC
connectors to input a signal into the oscilloscope.
Digital channel input port: Use the UT-M15 logic probe, provided as an accessory, to connect
the oscilloscope to the device under test for digital channel usage. Please refer to the “Digital
Channel” section.
Signal Source Output Interface G1, G2: These BNC connections can output signals such as
continuous wave, amplitude modulation, frequency modulation, amplitude-shift keying,
frequency-shift keying, and sweep signals. Please refer to the “Function/Arbitrary Waveform
Generator (Gen)” section.
Instruments.uni-trend.com
8 / 24
Quick Start Guide
3.4. Rear Panel
MSO3000HD Series
1
4
2
5
6
3
7
10
Table 2 Rear Panel Description
No.
Description
1
EXT Trig
2
AUX Out
3
10MHz REF
4
USB HOST port
5
HDMI port
Rear Panel
No. 6 7 8 9 10
9
8
Description USB Device
LAN port Power switch AC power input socket
Safety lock
3.5.Operation Panel
(1) Vertical Control
1 , 2 , 3 , 4 Analog channel setting key respectively represents CH1, CH2, CH3, and CH4. Four channel’s tab are identified by different colors and it also corresponding to the colors of waveforms on the screen and the channel input connectors. Press any keys to enter the related channel menu (activate or disable the channel).
Instruments.uni-trend.com
9 / 24
Quick Start Guide
MSO3000HD Series
Math Press this key to open the mathematical operation menu to perform math operation (add, subtract, multiply, divide), digital filter and advanced operation.
Ref Loading the reference waveform from `local or USB”, so the measured waveform can compare with the reference waveform.
BusPress this key to enter protocol decoding setting, to set the decoding of RS232, I2C, SPI, CAN, CAN-FD, LIN, FlexRay, Audio, 1553B, Manchester, SENT, and ARINC429.
DigitalPress this key to enter Digital setting, to set basics, grouping, threshold, bus and label.
Scale Vertical scale rotary knob is used to adjust the vertical scale in the current channel. Turn clockwise to decrease the scale, turn counterclockwise to increase the scale. The amplitude of waveform will increase or decrease with the adjustment and the scale at the
bottom of screen
will change in real-time. The vertical scale is step with1-2-5,
press this rotary knob to adjust the vertical scale between coarse tuning and fine tuning.
PositionVertical position rotary knob is used to move the vertical position of the waveform
in the current channel. Press this rotary knob to move the channel position back to the
vertical midpoint.
(2) Horizontal Control Menu Horizontal menu key is used to display the horizontal scale, time base mode (XY/YT), horizontal, auto roll, quick roll time base, horizontal position, time base extension and time base selection.
Scale Horizontal scale rotary knob is used to adjust all channel time base. During the adjustment, the waveform is compressed or extended in horizontal show on the screen and
the horizontal scale value
will change in real-time. The time base is
step with 1-2-5, press this rotary knob to adjust the horizontal scale between coarse tuning
and fine tuning. PositionHorizontal position rotary knob is used to move the trigger point to left or right
side that relative to the center of the screen. During the adjustment, all channel waveforms
move to left or right side and the horizontal shift value on the top of the screen
will change in real-time. Press this rotary knob to move the current
Instruments.uni-trend.com
10 / 24
Quick Start Guide
MSO3000HD Series
position back to the horizontal midpoint. (3) Trigger Control
SlopePress the Slope key to switch the trigger edge: rising edge, falling edge, or any edge, The corresponding edge’s indicator light will illuminate when switching the edge.
Menu Press the Menu key to enter the trigger menu. SourcePress the Source key to switch the trigger source. The panel numbers 1 to 4 correspond to channels C1 to C4. When switching the source, the corresponding number’s indicator light will illuminate. Position Trigger level rotary knob, turn clockwise to increase the level, turn counterclockwise to decrease the level. During the adjustment, the trigger level
on the top right will change in real-time. When the trigger is single level, press this rotary knob to turn the trigger level to the trigger signal and quickly turn to 50%. When the trigger status is READY or TRIGED, the corresponding Ready or Trig’d indicator light in the trigger control area will illuminate. (4) Control Area
Auto/NormalPress the Auto/Normal key to switch the oscilloscope trigger mode between Auto and Normal.
Run/StopThis key is used to set the operating mode of the oscilloscope to “Run” or “Stop”. In the “RUN” state, the key is illuminated in green. In the “STOP” state, the key is illuminated in red.
ForceWhen the trigger mode is Normal or Single, press the force trigger key to generate a trigger.
AutosetAfter pressing the Autoset key, the oscilloscope automatically adjusts the vertical scale, time base, and trigger mode based on the input signal to display the optimal waveform.
SinglePress the Single key to set the oscilloscope’s trigger mode to “Single”; the key will illuminate in orange.
(5) Measurement Analysis Area MeasurePress the Measure key to enter the measurement menu, to set the counter, voltmeter,
Instruments.uni-trend.com
11 / 24
Quick Start Guide
MSO3000HD Series
parameter snapshot, measurement statistics, add measurement, clear measurement,
and global settings. CursorPress the Cursor key to enter the cursor measurement menu, to set time, voltage,
and screen measurement for each source. AnalyzePress the Analyze to open the analysis module menu, where the user can access
functions such as the voltmeter, counter, power analysis, waveform recording, and pass/fail
test.
(6) Multi-function Area
QuickPress the Quick key to quickly save the screen
waveform as a PNG bitmap to the default or a custom image
path.
Press the Clear key to clear all recalled waveforms and
parameter measurement statistics from the screen. Touch/Lock Press the Touch/Lock key to disable the touchscreen function; the
background light will illuminate. To enable the touchscreen again, press this key once more,
and the background light will turn off. DefaultRestore factory settings. Press the Default key to reset all oscilloscope settings to
their default values. Multipurpose AMulti-function knob A. When a numerical menu is selected in the function
popup window, the LED on the multi-function knob will light up, allowing the user to adjust
values using the knob. Multipurpose BMulti-function knob B. When setting numerical parameters in a text box,
the user can rotate this knob to shift the selected digit.
(7) Signal Output Area
G1Press the G1 key to open the signal source Gen
settings menu and select G1. G2Press the G2 key to open the signal source Gen
settings menu and select G2.
(8) Home Menu
Press the Home icon on the top right corner to pop up “Home” quick menu, including the quick
menuvoltmeter, FFT, signal source, Math, reference, help, cursor, Bode diagram, storage,
counter, measurement, regional drawing, display, auxiliary, decoding, search, regional diagram,
navigation, waveform recording, power analysis, and Pass/Fail. Press the quick menu to enter
the corresponding function module.
Instruments.uni-trend.com
12 / 24
Quick Start Guide
MSO3000HD Series
Home Menu
3.6. User Interface
12
3
4
5
6
7
17
16
14
13
12
11
10
98
Instruments.uni-trend.com
13 / 24
Quick Start Guide
MSO3000HD Series
Table 3 User Interface Description
No.
Description
No.
Description
1
Waveform display window
10
Multi-window display area
2
Trigger state icon
11
Digital channel label
3
Horizontal base label
12
Ref label
Sampling rate and storage depth
4
13
label
FFT label
5
Trigger info label
14
Math label
6
Function toolbar
15 Measured result display window
7
Home menu
16
Channel label
8
Notification
17
Analog channel label
9
Volts/div signal bar
3.7.Help System
The help system provides information about the function keys (including menu keys) on the front panel. The user can access the help system by following these steps. In Home menu, click on the help icon ” ” to open the help menu. In each function menu popup, click on the help icon ” ” on the top right to open the relevant
help menu. The help screen is divided into two parts, the left side is “Help Options”, and the right side is “Help Display Area”. By selecting a help option, the user can view all related help content on the right.
Instruments.uni-trend.com
14 / 24
Quick Start Guide
MSO3000HD Series
Instruments.uni-trend.com
15 / 24
Quick Start Guide
4. Parameter Setting
MSO3000HD Series
MSO3000HD series supports use the Multipurpose A (Multipurpose rotary knob) and touch screen to set the parameter. The setting steps as follows. (1) Multipurpose rotary knob
For the parameter of time and voltage, once the parameter is selected, rotating the Multipurpose rotary knob on the front panel to enter the parameter value. (2) Touch screen Once the parameter or text field has been selected, double-click to pop up the virtual keyboard to enter the parameter value, label name, or file name. a. Enter character string
When naming a file or folder, use the character keyboard to enter a string. b. Text field
Enter text: input letters, numbers, and special characters, with a maximum length of 16 characters. c. Clear key Press the “Clear” key to delete all contents in the text field. d. Caps key Press the “Caps” key to switch between upper and lower case. e. Tab key Press the “Tab” key to enter 2 spaces at a time. f. Shift key Press the “Shift” key to switch among number, special character, upper and lower case. g. Arrow keys (left, right) If part of the content needs to be changed, press the “, ” key to move the cursor to left or right and then to edit the content. h. Space key Press the “Space” key to enter one space in the text field. i. Backspace key Press the “Backspace” key to delete a single character. This key is used to delete a character when the text field contains a large amount of content. j. Enter key Once the content has been entered, press the “Enter” key to confirm the setting and close the virtual keyboard.
Instruments.uni-trend.com
16 / 24
Quick Start Guide
Keyboard
Input area
MSO3000HD Series
Tab Caps key
Shift
Clear key Backsap ce key Enter
(3) Enter numeric value
Virtual
Space Key
Arrow keys: left, right
When setting or editing a parameter, use the numeric keyboard to enter the numeric value.
1. Click the number or unit to enter
Arrow keys: left, right
Input area
Backspace key
Clear Maximum key Default key Minimum key Enter key
Numeric
Unit
After entering all the values and selecting the desired units, the numeric keypad will automatically close, completing the parameter setting. Additionally, the user can manually close the numeric keypad by clicking the confirm key, in which case the unit will default to the preset unit. On the numeric keypad, the user can also perform the following operations:
Instruments.uni-trend.com
17 / 24
Quick Start Guide
MSO3000HD Series
a. Delete the entered parameter value. b. Set the parameter to the maximum or minimum value (sometimes specifically the
maximum or minimum value for the current state). c. Set the parameter to the default value. d. Clear the parameter input field. e. Move the cursor to modify the parameter value. 2. Enter binary, hexadecimal system value During the decoding trigger, use the numeric keypad to enter binary or hexadecimal values for data and address settings. Enter Method: Tap to select the number or text field to be edited, and then use the numeric keypad to enter the desired numeric or letter values.
Binary input
Hexadecimal input
Numeric keyboard
Arrow keys
Default key Maximum Minimum key Ok key
(4) After entering all the values and pressing the “Ok” button, the numeric keypad will automatically close, completing the parameter setting. Additionally, on the numeric keypad, you can perform the following operations: a. Move the cursor to modify the parameter value. b. Set the parameter to the maximum or minimum value (sometimes specifically for the current state). c. Set the parameter to the default value. d. Clear the parameter input field. e. Delete the entered parameter value.
Instruments.uni-trend.com
18 / 24
Quick Start Guide
5. Touch Screen
MSO3000HD Series
MSO3000HD series provides 10.1-inch super capacitive touch screen, multiple point touch control and gesture control. MSO3000HD has easily operating system with flexible and high-sensitive touch screen features for great waveform display and excellent user experience. Touch control function includes tap, squeeze, drag, and rectangle drawing. NoteThe menu displayed on the screen of the oscilloscope can all use the touch control function. (1) Tap
Use one finger to slightly tap on an icon or a word on the screen as shown in the following figure. Tap gesture can use for: Tap the menu display on the screen and then to setup Tap the function icon on the top right corner to open the corresponding function Tap the pop-up numeric keyboard to set the parameter Tap the virtual keyboard to set the label name and file name
Tap a message to pop up a × button on the top right corner to close the pop-up window.
Tap other window displayed on the screen and then to setup Tap a message to reveal a help key in the top right corner. Clicking this key will open the
corresponding function’s help menu, providing detailed guidance on the selected feature
Tap Gesture
(2) Pinch Squeeze two fingers together or separate. Pinch gesture can zoom out or zoom in the waveform. To zoom out the waveform, pinch two fingers together and then slide them apart; to zoom in, spread two fingers apart and then pinch them together, as shown in the following figure. Adjust the horizontal time base of waveform by squeezing on the horizontal direction. Adjust the vertical time base of waveform by squeezing on the vertical direction.
Instruments.uni-trend.com
Pinch Gesture
19 / 24
Quick Start Guide
MSO3000HD Series
(3) Drag Use one finger to press and drag the selected item to the aimed position as shown in the following figure. Drag gesture can use for: Drag the waveform to change the waveform position. Drag the window to change the window position. Drag the cursor to change the cursor position.
Drag Gesture
(4) Rectangle Drawing Open the HOME menu and click the icon “Rectangle Drawing” to enable the function, drag your finger to draw a rectangle on the screen as shown in Figure (a), (b), move the finger, a menu will appear on the screen, at this point, “Region A”, “Region B”, “Intersection”, “Non-intersect” can be selected. Drag your finger from bottom right to the top left on the screen to draw the trigger area.
(a)
(b)
Rectangle Drawing Gesture
Select “Zone A”
Draw the trigger zone A
Open the trigger zone A
Open “Zone Trigger” menu
Select “Zone B” Draw the trigger zone B Open the trigger zone B Open “Zone Trigger” menu
Note: Click on “rectangle drawing” to step through rectangle drawing and operating waveform
mode. Click on “rectangle drawing”, if the icon shows , it indicates that “rectangle drawing”
mode is enabled; if the icon shows , it indicates that “operating waveform” mode is enabled.
Instruments.uni-trend.com
20 / 24
Quick Start Guide
6. Remote Control
MSO3000HD Series
MSO3000HD series high-definition oscilloscopes can communicate with a PC via USB and LAN port for remote control. Remote control is implemented using SCPI (Standard Commands for Programmable Instruments). MSO3000HD series has three methods for remote control. (1) Custom Programming
The user can perform the programming control on the oscilloscope through SCPI (Standard Commands for Programmable Instruments). For detailed descriptions on command and programming, please refer to MSO3000HD Series High-Definition Oscilloscopes-Programming Manual. (2) PC Software Control (Instrument manager) The user can use a PC software to remotely control the oscilloscope. The instrument manager can display the oscilloscope screen in real time and control the operation with the mouse. It is recommended to use the PC software provided by UNI-T. It can be downloaded from UNI-T official website (https://www.uni-trend.com). Operating steps: Setup the communication between the instrument and a PC Open the instrument manager software and search the instrument source Right-click to open the oscilloscope, operate the instrument manager to remotely control
the oscilloscope (refer to Instrument Manager-User’s Manual for more details) .This device supports remote control through communication with a computer via USB, LAN, and WiFi interfaces. Remote control is implemented using the SCPI command set, allowing users to operate and configure the device remotely. (3) Web Control Once the network is connected, users can access a web page via the IP address. After logging in with the username and password, they can control the device. The Web Control feature displays the instrument’s screen interface in real-time. It supports web access from PCs, smartphones, and iPads, and allows for both internal and external remote control of the device.
Instruments.uni-trend.com
21 / 24
Quick Start Guide
7. Troubleshooting
MSO3000HD Series
(1) If the oscilloscope remains black screen without any display when press the power soft key. a. Check if the power plug is properly connected and the power supply is normal. b. Check if the power switch is turned on. If the power switch is turned on, the power soft key on the front panel should be green. When the power soft key is enabled, the power soft key should be blue, and the oscilloscope will make active sound. There should be a normal relay rattle when the soft switch key is pressed. c. If the relay has sound, it indicates that the oscilloscope is normal boot-up. Press the Default key and press the “Yes” key, if the oscilloscope returns to normal, indicating that the backlight brightness is set too low. d. Restart the oscilloscope after completing the above steps. e. If the product still does not work properly, contact the UNI-T Service Center for assistance.
(2) After signal acquisition, the waveform of the signal does not appear on the screen. a. Check whether probe and DUT are connected properly. b. Check whether the signal output channel is open. c. Check whether the signal connecting line is connect to analog channel. d. Check whether the signal source has DC offset. e. Plug out the connected signal, to check whether the base line is within the screen range (If not, please perform self-calibration). f. If the product still does not work properly, contact the UNI-T Service Center for assistance.
(3) The measured voltage amplitude value is 10 times larger or 10 times smaller than the actual value. Check whether the channel probe attenuation coefficient settings are consistent with the used probe attenuation rate.
(4) There is a waveform display but not stable. a. Check the trigger settings in trigger menu whether is consistent with the actual signal input channel. b. Check the trigger type: the general signals should use “Edge” trigger. The waveform can only be displayed stably if the trigger mode is set correctly. c. Try to change trigger coupling to HF rejection or LF rejection, to filter out the high-frequency or low-frequency noise that interfere the trigger.
(5) No waveform is displayed after pressing the Run/Stop key. a. Check whether the trigger mode is set to “Normal” or “Single” and verify if the trigger level
Instruments.uni-trend.com
22 / 24
Quick Start Guide
MSO3000HD Series
exceeds the waveform range. b. If the trigger mode is in normal or single and the trigger level is in the center, set the trigger
mode to Auto. c. Press the Autoset key to automatically complete the above settings. (6) Waveform refresh is very slow. a. Check whether the acquisition method is set to “Average” and if the average times are
large. b. Check whether the storage depth is maximum c. Check whether the trigger holdoff is large. d. Check whether it is normal trigger and is slow time base. e. All the above will lead to slow waveform refresh, it is recommended to restore the factory
settings, then the waveform can be refreshed normally.
Instruments.uni-trend.com
23 / 24
110401112667X
1
2
3
4
5
6
7
DWH
CHK APPRO
:148×210±1mm. 128G 60g ,…
REV.0
MODEL : MSO3000HD
art NO. 110401112667X
()
UNI-TREND TECHNOLOGY (CHINA) CO.,LTD
Documents / Resources
![]() |
UNI T MSO3000HD Series High Definition Oscilloscopes [pdf] User Guide MSO3000HD Series, MSO3000HD Series High Definition Oscilloscopes, MSO3000HD Series, High Definition Oscilloscopes, Definition Oscilloscopes, Oscilloscopes |