UB-PT-N1 PT100 Temperature Sensor

Specifications:

  • Model: Micro USB/Audio
  • Power Supply: 237mA(@5V), 129mA(@12V)
  • Measuring Range: -30~200
  • Measuring Accuracy: RS485 Modbus RTU Protocol
  • Connector: 5
  • Max Current: 3m
  • Dimensions: 85*33*60mm
  • Sensor Length: 200mm
  • Sensor Diameter: 5mm
  • Cable Length: 3m
  • Tail Line Working Temperature: -30~200
  • Communication Protocol: RS485 Modbus RTU Protocol
  • RS485 Address: 0xC2
  • Baud Rate: 1200 bit/s, 2400 bit/s, 4800 bit/s, 9600 bit/s
    (default), 19200 bit/s

Product Usage Instructions:

Wiring Instruction:

RS485 Micro USB Audio
VCC Red Red
A Green White
B White Green
GND Black Black

Communication Protocols:

1. Communication Basic Parameters:

  • Coding System: Modbus-RTU
  • Data Bit: N/A
  • Parity Checking Bit: N/A
  • Stop Bit: N/A
  • Error Checking: 16-bit CRC code
  • Baud Rate: 1200 bit/s, 2400 bit/s, 4800 bit/s, 9600 bit/s
    (default), 19200 bit/s

2. Data Frame Format:

The Modbus-RTU communication protocol is used in the following
format:

  • Initial Structure: 4 bytes in time.
  • Address code: 1 byte, default 0xC2.
  • Function code: 1 byte, support function code 0x03 (read only)
    and 0x06 (read/write).
  • Data area: N bytes, 16-bit data, high byte comes first.
  • Error check: 16-bit CRC code.
  • End Structure: 4 bytes of time.

Product Application:

  1. Do not directly place the sensor in a high-temperature
    environment or the environment of steam, water mist, water curtain,
    or condensation for a long time.
  2. The temperature tolerance of the sensor tail line is 200°C.
    Exceeding this temperature will lead to the failure of normal
    temperature measurement and even permanent damage.

Frequently Asked Questions (FAQ):

Q: What is the recommended working temperature range for the
sensor?

A: The sensor has a working temperature range of -30°C to
200°C.

Q: What communication protocol does the device support?

A: The device supports the standard MODBUS RTU protocol.

Q: How long is the sensor’s cable length?

A: The sensor’s cable length is 3 meters.

“`

Introduction
The PT100 sensor is made of specially treated platinum wire wrapped around the inner surface of the sensor. Compared with the general axial thermal resistance, it can reflect the actual temperature of the measured object quickly and accurately. The sensor chip imported from Germany with high stability, and adopts a more stable three-wire connection mode to ensure the accuracy of temperature measurement and anti-interference ability of the system.The probe shell is made of 304 stainless steel, which has good resistance to acid and alkali corrosion. The shell is completely wrapped and can reach IP68 waterproof grade, which means the sensor can be continuously measured in water.

Use Case Scenarios Widely used in chemical plants, power plants, oil refineries, cold storage, sewage treatment plants, steel plants, food plants and other industrial temperature measurement sites.

Features High precision, wide range, good consistency. Super stability and anti-interference. Standard MODBUS RTU protocol.

Specification

Model Power Supply Measuring Range Measuring Accuracy
Connector Max Current Dimensions Sensor Length Sensor Diameter Cable Length Tail Line Working Temperature Communication Protocol RS485 Address
Baud Rate

Specification UB-PT-N1 DC 5V~12V -200~600 ±(2%+1)
Micro USB/Audio 237mA(@5V), 129mA(@12V)
85*33*60mm 200mm 5 3m
-30~200 RS485 Modbus RTU Protocol
0xC2 1200 bit/s,2400 bit/s, 4800 bit/s, 9600 bit/s (default), 19200 bit/s

Wiring Instruction

RS485 Micro USB
Audio

VCC Red Red

B White Green

A Green White

GND Black Black

Communication protocols

1. Communication basic parameters

Coding System Data Bit
Parity Checking Bit Stop Bit
Error Checking Baud Rate

Communication Basic Parameter 8­bit binary 8 bits none 1 bit CRC Check
1200 bit/s, 2400 bit/s, 4800 bit/s, 9600 bit/s (default), 19200 bit/s

2. Data Frame Format

The Modbus-RTU communication protocol is used in the following format:

Initial structure 4 bytes in time.

Address code: 1 byte, default 0xC2.

Function code: 1 byte, support function code 0x03 (read only) and 0x06 (read/write).

Data area: N bytes, 16-bit data, high byte comes first.

Error check: 16-bit CRC code.

End structure 4 bytes of time.

Request

Slave Address

Function Code Register Address No. of Registers

CRC LSB

CRC MSB

1 byte

1 byte

2 bytes

2 bytes

1 byte

1 byte

Response

Slave Address Function Code No. of Bytes Content 1 Content 1

Content n

CRC

1 byte

1 byte

1 byte

2 bytes

2 bytes

2 bytes

2 bytes

3. Register Address

Address 0x0000 0x0064
0x0065

Content Temperature
Address
Baud Rate

Register Address

Register Length Function Code

Description of definitions

1

03

Signed integer data, divided by 10, in []

1

03/06

1255

2:1200, 3:2400, 4:4800, 5:9600, 6:19200,

1

03/06

7:38400, 8:43000, 9:56000, 10:57600, 11:115200

Product Application 1. Do not directly place the sensor in a high temperature environment or the environment of steam, water mist, water curtain or condensation for a long time. 2. The temperature tolerance of the sensor tail line is 200 . Exceeding this temperature will lead to the failure of normal temperature measurement and even permanent damage.

Documents / Resources

UbiBot UB-PT-N1 PT100 Temperature Sensor [pdf] User Guide
WS1, WS1 Pro, UB-PT-N1 PT100 Temperature Sensor, UB-PT-N1, PT100 Temperature Sensor, Temperature Sensor, Sensor

References

Leave a comment

Your email address will not be published. Required fields are marked *