TSI 5200 Series Gas Flow Multi Mete
Instrument Kit Components
5200 Series Low Flow Gas Multi-Meter
Qty | Item Description | Part/Model |
1 | Low Flow Gas Flow Multi-Meter, 3/8-inch tube ends Measures Flow, Temperature, Absolute Pressure | 5200 |
Low Flow Gas Flow Multi-Meter, 3/8-inch tube ends
Measures Flow, Volume, Temperature, Absolute Pressure, Differential Pressure |
5210 | |
Low Flow Gas Flow Multi-Meter, 3/8-inch tube ends
Measures Flow, Volume, Temperature, Absolute Pressure, Differential Pressure, Relative Humidity |
5220 | |
Low Flow Gas Flow Multi-Meter (High Accuracy), 3/8-inch tube ends, includes Tabletop Stand
Measures Flow, Volume, Temperature, Absolute Pressure, Differential Pressure, Relative Humidity |
5230 | |
Low Flow Gas Flow Multi-Meter (Wide Accuracy Spec), 3/8-inch tube ends
Measures Flow, Temperature, Absolute Pressure |
5203 | |
1 | Calibration Certificate | N/A |
1 | 5200 Series Standard Accessory Kit | 52000 |
1 | Low Pressure Measurement Kit (5210, 5220, 5230 models only) | 5200-LPMK |
1 | License key for FLO-Sight™ PC Software (Includes a free trial of the advanced version) | 5000-PC |
*FLO-Sight software can be downloaded from www.tsi.com.
download Operator’s Manual at TSI.com
Instrument Overview
1. Power Button (On/Off) | 5. USB-C Power/ Communications Port |
2. Color Touchscreen Display* | 6. USB-A Communications Ports (2) |
3. Interchangeable Tube Ends | 7. Low Differential Pressure Ports (5210/5220/5230 models only) |
4. Mounting Inserts (M3 thread) | 8. Collars |
dimensions shown in inches [mm]
Setup Basics
Supplying Power
Connect the USB cable from the USB-C port on the meter to the USB-A port of: 1) the 5000 Series power supply, 2) a computer, or 3) other USB sources that can provide 5 VDC.
USB Hub Cable
Some computer USB ports are not capable of sufficiently powering up the Meter. If you are unable to power your Meter from the computer, connect the USBC-A power cable to the provided USB hub cable and plug the hub cable into the computer.
Turning Meter ON/OFF
The Meter may power itself on when connected to a power source. If not, press the power button to turn the Meter on. To turn the meter off, press and hold the power button for 5 seconds.
Connect Inlet Filter
TSI® recommends an inlet filter be used at all times while operating the Flow Multi-Meter.
Changing Tube Ends
- Unscrew collars and remove tube ends.
- Insert replacement tube ends into the meter while aligning the anti-rotation locking notch on the tube end with the tab on the meter.
- Screw collars back onto the meter.
Touchscreen Display Overview
Meter Home Screen
1. Meter Home Tab | 6. Volume and Triggers Tab |
2. Measurement Readings | 7. Log Data Button |
3. Type of Gas | 8. Menu Dropdown Screen |
4. Flow Directional Indicator | 9. Device Name and Information |
5. Pause Display Button | 10. Alerts Indicator |
*Available on 5210, 5220, and 5230 models only
Pause Display Button
The Pause Button is used to pause the display from updating. The current measurement values will remain fixed on the screen. The Meter will continue to make measurements while paused, and pausing the display does not affect any active datalog sessions.
- When paused, the button’s icon will change to a play icon
- Press the button again to resume display updates.
Menu Screen
1. Go to Meter Home Screen | 5. Rotate Display 180º |
2. Set Logging Parameters * | 6. Settings Menu |
3. Data Management * | 7. Press anywhere outside menu to close menu screen |
4. Reset Device Settings |
*Available on 5210, 5220, and 5230 models only
Touchscreen Display
The touchscreen display is resistive and responds to pressure from your finger or other objects. You can also operate the Meter while wearing gloves.
Configuring the Meter
Measurement Parameters
Press a value field, use the scroll arrows to select a parameter, and click Save.
Units of Measurement
Press a units field, use the scroll arrows to select the desired units, and click Save.
Type of Gas
Press the type of gas field, use the scroll arrows to select the type of gas, and click Save.
Volume and Triggers
(Models 5210, 5220, and 5230 only)
Volume is a calculated measurement that integrates flow over time and is controlled through triggers. You can set begin and end triggers for volume measurements using flow rate, absolute pressure, or low differential pressure values.
Setting Triggers
The meter must be stopped to edit trigger settings. On the Volume tab, confirm that the Meter is stopped (shown below left) and press on the field you would like to edit.
If a volume measurement is active (shown below right), then press the Stop button to cancel the measurement and allow edits to be made to trigger settings.
In the Select Mode screen, you can choose:
Trigger Parameter | Flow, Absolute Pressure, or Low Pressure |
Test Mode | Single or Continuous |
Use the arrows to scroll through the selection options and click Save to complete.
In the Select Trigger screens, you can choose:
Begin / End Trigger Values | Select from available values |
Begin / End Trigger Slopes | Positive or Negative |
Note trigger options for the volume begin and end triggers. These are set as the default options and cannot be changed by the user.
Trigger | Units | Trigger Values |
Flow | Std L/min | +/– 0.1, 0.5, 1, 2, 3, 5, 7.5, 10, 20 |
Absolute Pressure | kPa | 90, 95, 98, 100, 102, 104, 107, 110 |
Low Differential Pressure | cmH2O | +/– 0.5, 1, 2, 3, 4, 5, 10 |
Making Volume Measurements
Once all settings have been saved, press the Play button to lock editing and begin making volume measurements.
The Multi-Meter will notify you on screen when the begin trigger, and then end trigger, is reached. Once the end trigger has been reached, the Meter will deliver the volume measurement.
TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations.
CONTACT
TSI Incorporated – Visit our website www.tsi.com for more information.
- USA Tel: +1 800 680 1220
- UK Tel: +44 149 4 459200
- France Tel: +33 1 41 19 21 99
- Germany Tel: +49 241 523030
- India Tel: +91 80 67877200
- China Tel: +86 10 8219 7688
- Singapore Tel: +65 6595 6388
P/N 6012621 Rev D ©2024 TSI Incorporated Printed in U.S.A.