TSI-LOGO

TSI 5200 Series Gas Flow Multi Mete

TSI-5200-Series-Gas-Flow-Multi-Meter-PRODUCT

 

Instrument Kit Components

5200 Series Low Flow Gas Multi-Meter

Qty Item Description Part/Model
1 Low Flow Gas Flow Multi-Meter, 3/8-inch tube ends Measures Flow, Temperature, Absolute Pressure 5200
Low Flow Gas Flow Multi-Meter, 3/8-inch tube ends

Measures Flow, Volume, Temperature, Absolute Pressure, Differential Pressure

5210
Low Flow Gas Flow Multi-Meter, 3/8-inch tube ends

Measures Flow, Volume, Temperature, Absolute Pressure, Differential Pressure, Relative Humidity

5220
Low Flow Gas Flow Multi-Meter (High Accuracy), 3/8-inch tube ends, includes Tabletop Stand

Measures Flow, Volume, Temperature, Absolute Pressure, Differential Pressure, Relative Humidity

5230
Low Flow Gas Flow Multi-Meter (Wide Accuracy Spec), 3/8-inch tube ends

Measures Flow, Temperature, Absolute Pressure

5203
1 Calibration Certificate N/A
1 5200 Series Standard Accessory Kit 52000
1 Low Pressure Measurement Kit (5210, 5220, 5230 models only) 5200-LPMK
1 License key for FLO-Sight™ PC Software (Includes a free trial of the advanced version) 5000-PC

*FLO-Sight software can be downloaded from www.tsi.com.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-1download Operator’s Manual at TSI.com

Instrument Overview

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-2

1. Power Button (On/Off) 5. USB-C Power/ Communications Port
2. Color Touchscreen Display* 6. USB-A Communications Ports (2)
3. Interchangeable Tube Ends 7. Low Differential Pressure Ports (5210/5220/5230 models only)
4. Mounting Inserts (M3 thread) 8. Collars

dimensions shown in inches [mm]

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-3

Setup Basics

Supplying Power
Connect the USB cable from the USB-C port on the meter to the USB-A port of: 1) the 5000 Series power supply, 2) a computer, or 3) other USB sources that can provide 5 VDC.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-4

USB Hub Cable

Some computer USB ports are not capable of sufficiently powering up the Meter. If you are unable to power your Meter from the computer, connect the USBC-A power cable to the provided USB hub cable and plug the hub cable into the computer.

Turning Meter ON/OFF

The Meter may power itself on when connected to a power source. If not, press the power button to turn the Meter on. To turn the meter off, press and hold the power button for 5 seconds.

Connect Inlet Filter

TSI® recommends an inlet filter be used at all times while operating the Flow Multi-Meter.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-5

Changing Tube Ends

  1. Unscrew collars and remove tube ends.TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-6
  2. Insert replacement tube ends into the meter while aligning the anti-rotation locking notch on the tube end with the tab on the meter.TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-7
  3. Screw collars back onto the meter.

Touchscreen Display Overview

Meter Home Screen

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-8

1.     Meter Home Tab 6. Volume and Triggers Tab
2.     Measurement Readings 7. Log Data Button
3.     Type of Gas 8. Menu Dropdown Screen
4.     Flow Directional Indicator 9. Device Name and Information
5.     Pause Display Button 10. Alerts Indicator

*Available on 5210, 5220, and 5230 models only

Pause Display Button
The Pause Button is used to pause the display from updating. The current measurement values will remain fixed on the screen. The Meter will continue to make measurements while paused, and pausing the display does not affect any active datalog sessions.

  • When paused, the button’s icon will change to a play iconTSI-5200-Series-Gas-Flow-Multi-Meter-FIG-9
  • Press the button again to resume display updates.

Menu Screen

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-10

1.     Go to Meter Home Screen 5. Rotate Display 180º
2.     Set Logging Parameters * 6. Settings Menu
3.     Data Management * 7. Press anywhere outside menu to close menu screen
4.     Reset Device Settings  

*Available on 5210, 5220, and 5230 models only

Touchscreen Display
The touchscreen display is resistive and responds to pressure from your finger or other objects. You can also operate the Meter while wearing gloves.

Configuring the Meter

Measurement Parameters
Press a value field, use the scroll arrows to select a parameter, and click Save.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-11

Units of Measurement
Press a units field, use the scroll arrows to select the desired units, and click Save.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-12

Type of Gas
Press the type of gas field, use the scroll arrows to select the type of gas, and click Save.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-13

Volume and Triggers

(Models 5210, 5220, and 5230 only)

Volume is a calculated measurement that integrates flow over time and is controlled through triggers. You can set begin and end triggers for volume measurements using flow rate, absolute pressure, or low differential pressure values.

Setting Triggers
The meter must be stopped to edit trigger settings. On the Volume tab, confirm that the Meter is stopped (shown below left) and press on the field you would like to edit.
If a volume measurement is active (shown below right), then press the Stop button to cancel the measurement and allow edits to be made to trigger settings.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-14

In the Select Mode screen, you can choose: 

Trigger Parameter Flow, Absolute Pressure, or Low Pressure
Test Mode Single or Continuous

Use the arrows to scroll through the selection options and click Save to complete.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-15

In the Select Trigger screens, you can choose: 

Begin / End Trigger Values Select from available values
Begin / End Trigger Slopes Positive or Negative

Note trigger options for the volume begin and end triggers. These are set as the default options and cannot be changed by the user.

Trigger Units Trigger Values
Flow Std L/min +/– 0.1, 0.5, 1, 2, 3, 5, 7.5, 10, 20
Absolute Pressure kPa 90, 95, 98, 100, 102, 104, 107, 110
Low Differential Pressure cmH2O +/– 0.5, 1, 2, 3, 4, 5, 10

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-16

Making Volume Measurements

Once all settings have been saved, press the Play button to lock editing and begin making volume measurements.
The Multi-Meter will notify you on screen when the begin trigger, and then end trigger, is reached. Once the end trigger has been reached, the Meter will deliver the volume measurement.

TSI-5200-Series-Gas-Flow-Multi-Meter-FIG-17

TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations.

CONTACT

TSI Incorporated – Visit our website www.tsi.com for more information.

  • USA Tel: +1 800 680 1220
  • UK Tel: +44 149 4 459200
  • France Tel: +33 1 41 19 21 99
  • Germany Tel: +49 241 523030
  • India Tel: +91 80 67877200
  • China Tel: +86 10 8219 7688
  • Singapore Tel: +65 6595 6388

P/N 6012621 Rev D ©2024 TSI Incorporated Printed in U.S.A.