TREE REMO-R Universal Indicator
Specifications
- Product Name: REMO Indicators
- Type: Universal Indicator
- Models: REMO, REMO-R
- Manufacturer: LW Measurements, LLC
- Standards: EN 61326-1: 2006
Declaration of Conformity
Declaration of conformity for apparatus with CE mark
- We hereby declare that the product to which this declaration refers conforms to the following standards.
- Electronic scale: REMO Indicators
Available Models:
- REMO
- REMO-R
| Mark Applied | EU Directive | Standards |
| CE |
2004/108/EC |
EN 61326-1: 2006 |
Signature:
- Boon Lim, R & D Manager
- Date: 01/01/2024
- LW Measurements, LLC, 620 Carlson Court, Rohnert Park, CA 94928
Customer Service
USA
- LW Measurements LLC,
- 620 Carlson Court
- Rohnert Park, CA 94928
- USA
- Tel: 707-542-2185
- http://www.lwmeasurements.com

Introduction
What you should know about this Operation Manual:
- Tree® Professional Weighing Equipment products are simple to operate. Nevertheless, you should read through these operating instructions in their entirety, so that you can make optimum use of the full potential of the weighing scale in your daily work activities.
- These operating instructions contain guidance in the form of pictograms and keyboard diagrams, which should help in finding the required information:
- For the labeling of potential hazards, please refer to the Safety proportion of this operation manual.
Safety
Representations & Symbols
- Note: Important instructions which involve safety are highlighted with the appropriate mark:
DANGER
Safety Recommendations
- When using the weighing equipment in surroundings with increased safety requirements, the corresponding regulations must be observed.
- The weighing scale may only be used with the power adapter supplied.
- Before connecting the power adapter to the scale, the user must ensure that the operating voltage stated on the power adapter is compliant with the mains voltage. If not, please contact Customer Service at the address above.
- If the power adapter or its cable is damaged, the weighing scale must immediately be disconnected from the electricity supply (unplug the power adapter).
- If there should be any reason to believe that it is no longer safe to operate the scale, it should be immediately unplugged from the electricity supply (unplug power adapter) and secured against inadvertent operation.
- During maintenance work, it is essential to follow the recommendations set out in maintenance and servicing.
- The weighing scale must not be operated in an area subject to explosion risks.
- Care must be taken when weighing liquids to ensure that no liquid is spilled inside the scale, into connections on the rear of the equipment, or the power adapter. If liquid is spilled on the scale, it must immediately be unplugged from the main electricity supply (unplug power adapter).
- The weighing scale may be operated after it has first been inspected by a service technician.
- These operating instructions must be read by each user and should always be easily accessible at the workplace.
Indicator
The weighing scale consists of certain parts.
- Indicator
- Mounting Bracket
- Load Cell Wiring
- AC Adapter
- Operation Manual
- Mounting Screws

Features
- Auto back light (selectable)
- Auto off (selectable)
- Animal weighing
- Hold function
- Peak hold function
- Any Cal calibration software
- RS232 function
- To work either as an indicator, or a display
- Three modes: Normal mode / Setting mode /Internal Auto Calibration mode
- Zero range: + 10% of full capacity
- Zero Range (0 to 10% of full capacity)
- Tare Range (10% to full capacity)
- 4.5Ah rechargeable battery (REMO-R only)
- Display options:5 digital LCD
- Error message indication:
Overload
Unstable when power on
Calibration code is not enough or operation error
Low battery indication
Interface Data:
- 1 signal interface:(microphone socket)
- Red——-E+(positive source) / Blue ——- S+(positive signal)
- Black——-E-(negative source) / Green——-S-(negative signal) No sign (yellow) ——- GND
- 2 recharge battery socket:
- AC/DC 10V――――――external positive, inner negative
- RS232 interface:DB9
RS232 TRANSMIT FUNCTION
- 1. RS232 interface:DB9

RS232 Transmission Agreement:
- Mode: Simplex Asynchronous Serial
- Baud rate: 9600 Data Bit:8
- Parity Bit:NONE
- Stop Bit:1
- Data Format:ASCII
- Transmission Information Format: 20 Byte, blank=20H

- Example: 50kg

Application & Conformity
- The following are instructions on how to correctly use the indicator:
- The indicator may only be used for the weighing of solid materials and of liquids contained in secure containers.
- The maximum capacity load of the weighing scale must be correctly entered in the indicators program, otherwise the weighing scale may be damaged.
- When using the indicator in combination with other devices including those produced by other manufacturers, the appropriate regulations for the safe use of the relevant attachments and their application in accordance with instructions must be observed.
- The indicator has been manufactured and tested in accordance with the standards and recommendations set out in the declaration of conformity.
- The power adapter supplied for the indicator complies with the appropriate electrical protection class.
The following applies to all REMO series indicators.
Power supply:
- Input: 110V AC 50Hz to 60Hz
- REMO Output: 9v 400mAh
- REMO-R Output: 12v DC 1000mAh
- Allowable ambient conditions:
- Temperature: 32ºF – 104ºF / 5°C. 40°C
Relative humidity:
- 25% – 85%, non-condensing
- NOTE: If you have any questions on the technical data or require detailed technical information on your balance, please contact your technical representative.
Getting Started
- The indicator is packaged in an environmentally friendly carton, which provides optimum protection for the indicator during transportation.
- We suggest that you keep the original packaging to avoid damage if you are shipping or transporting the scale to a different location. It is also the best way to keep it in the best condition if it will not be used for an extended period.
- To avoid damage, please follow the instructions provided below, when unpacking the scale:
- Unpack the indicator carefully.
- When outside temperatures are very low, the indicator should be stored for a couple hours in its box in a dry room at normal temperature. This prevents any condensation from settling on the unit when opening the box.
- Check the indicator immediately after unpacking for any external visual damage. If there is any damage on the indicator, contact customer service immediately.
- If the indicator is not to be used immediately after purchase, it should be stored in a dry place where fluctuations in temperature are low. (reference pg.18).
- Read through these operating instructions before you work with the unit and pay attention to the safety recommendations.
Delivery
- Inspect delivery for completeness immediately upon unpacking all components.
Checklist for complete delivery
| Component | Present Yes / No |
| Indicator unit | |
| Power adapter | |
| Operating manual | |
| Mounting bracket | |
| Load cell wiring |
Assembly & Installation
The indicator is delivered in a partly dismantled condition. Assemble the individual components in the following sequence:
- Mount your machine on a clean, dry surface to ensure optimal visibility of your readings.
Connecting the AC Adapter
- The following safety recommendations must be observed when connecting the balance:
DANGER
- The scale should only be connected to the mains voltage socket with the power adapter supplied. Before connecting the power adapter to the mains socket, check to see that the operating voltage stated on the power adapter complies with the local mains voltage.
- If the operating voltage is not the same as the mains voltage, the power adapter must not be connected to the mains socket and contact customer service.
Placement of Scale
- The location in which the scale is placed is very important for the scale to work to its full potential. Certain conditions can affect the capabilities of the scale, such as presence of air flow, variations in temperature, and direct sunlight. Please follow the recommendations given below when choosing a location to place your scale.
- Place the indicator on a solid, firm and preferably vibration-proof surface.
- Make sure that the indicator cannot be shaken or knocked over.
- Protect from direct solar radiation.
- Avoid drafts and excessive temperature fluctuations.
- Avoid placing the indicator near or on any magnetic surfaces.
Remo Indicator Mounting Drawing

Display Messages
- The STABLE icon indicates when the readings stabled.
- The ZERO icon indicates when the weights tared to zero.
- The LO-BAT icon indicates when running out of battery.
- The kg – Kilograms (1000 grams) is a unit of measure.
- The lb. – Pounds (16 ounces) is unit of measure.
Key Functions
(icon) key turns the indicator on or off.
key is used to hold the weight value.
key sets display to zero or subtracts the container weight.
key changes the weight units, kg, lb.
key is used to send data to an external device through RS232 connection.
Function Settings
Auto Shutoff
-
- Press and hold the
key then press
key to turn on the scale, the display will show A-ON or A-OFF, press
key to select (A-ON means auto power off function activate, A-OFF means auto power off function inactivate). Continue to backlight.
- Press and hold the
Backlight
- Press key
, the display will show L-ON, L-OFF, or L-Au, press
key to select (L-ON means auto backlight function is active, L-OFF means auto backlight function is inactive, L-Au means backlight under auto mode). Continue to print function.
Print Function
- Press
key, the display will show P-XX. Press
key to select OFF, key, con, or Stb. (OFF means RS232 function is inactive; Key means that you can output the data by pressing key. Con means that the data outputs continually. Stb means that the data will output when the reading is stable). Continue to power-on mode .
Power-On Mode
- Press
key, the display will show J-ON or J-OFF, press
key to select (JON means showing zero when turning on the power, J-OFF means showing weight placed when turning on the power.) Continue to weighing function.
Weighing Function
- Press
key, the display will show F-Nor or F-PEA or F-Ani. Press
key to select (F-Nor means normal weighting function is activated. F-PEA means peak hold weighing function is activated. F-Ani means animal weighing function is activated). Continue to display mode.
Display Mode
- Press
key, the display will show N-dsp or N-Ind. Press
key to select (Ndsp means it can display synchronously as a separate display for another indicator through the RS232 output. N-Ind means it works only as an indicator for the scale.) Continue to peak hold mode.
Peak Hold Mode
- In FUNCTION SETTING 5, please choose F-PEA. F-PEA means peak hold weighing function is active, before weighing, please press key
. Continue animal anti shock range.
Animal Anti-Shock Range (F-Ani only)
- Press
key, the display will show Kd 50, Kd 100, Kd 200, or Kd 400. Press
key to select animal scale anti-shock range among 50, 100, 200, and 400. Continue to filter range.
Filter Range (F-Ani only)
- • Press
key, the display will show Fd 0.5, Fd 3, Fd 5, Fd 10. Press key to select animal scale filter range among 0.5, 3, 5, or 10. Continue to animal scale filter time.
Animal Scale Filter Time (F-Ani only)
- • Press
key, the display will show Ft 1 or Ft 2 or Ft 3 or Ft 4. Press
key to select animal scale filter time among 1, 2, 3, or 4. Continue to unit select function.
Unit Select Function
- Press
key, the display will show U-on or U-off, press
key to select U-on or U-off. (U-on means you can switch unit between kg and lb., U-off means lb. unit will be disabled.)
Turn indicator off to save selections.
- ***NOTE: Kd, Fd, and Ft work in Animal Weighing Mode (F-Ani) only. ***
HOLD Function
- Place the item on the scale platform. Wait for the stable indicator to be displayed.
- Press the
key. The “ . “ (hold indicator) will turn ON. - Remove the item from the scale platform. The item’s weight reading will remain on the display. Now you can also select Unit (lb. or kg) by pressing
key. (lb. cannot be selected if lb. is disabled in Function Setting)
Quit HOLD function:
- Once an item is weighed and the hold function is enabled, and you want to quit the HOLD function.
- Press the
key again to cancel the “. “ (hold indicator) and return the scale to weighing mode.
Calibration
What you need:
- Using a calibration or test weight that is 2/3 of the max capacity is strongly recommended for recalibration.
- If you do not have a test weight or calibration weight. You can use any object of “known weight” to calibrate with our scales AnyCal Calibration software.
- NOTE: when calibrating even if you use a “known weight” the more exact the weight is the more precise the calibration will be.
Prepare for calibration by following these steps:
- While the indicator is OFF, press and hold key
then press key
. The display will show CAPU=. - Press
key to select capacity unit kg or lb.(unit lb. cannot be selected if the lb. is disabled in FUNCTION SETTING). - Press
key to confirm the selection. - Press
key again to enter the next menu, the display will show CAP. - Press
key to set capacity, press
key to move the flashing digit to right, press
key to increase the flashing digit value. - Press key to confirm the selection.
- Press
key again to enter the next menu, the display will show CALU=. - Press
key to select calibration unit kg or lb. (unit lb. cannot be selected if the lb. is disabled in Function Settings). - Press
key to confirm the selection. - Press
key again to enter next menu, the display will show CAL. - Press
key to confirm the selection (the calibration weight should be more than 2/3 of full capacity). - Press
s key again to enter next menu, the display will show DIV, - Press
key to select division. - Press
key to confirm the selection. - Press
key again to enter next menu, the display will show Adb, press key to select (19.20). - Press
key to confirm the selection.
Start the calibration by following these steps:
- Press
key, the display will show CAL, then AD value. - Press
key when the stable indicator lights, the display will show the flashing calibration weight, place the shown weight on the platform. - Press
key when the stable indicator lights, the display will show “—–“, and AD value. - Turn OFF the scale.
- Calibration is Complete.
- Turn ON the scale, place tested weight to check whether it is correct.
- If not, please repeat the above instructions.
If this device has been previously programmed, please see “Start the Calibration” steps above for a simplified procedure.
Maintenance & Service
DANGER
- For maintenance work, the balance must be disconnected from the power supply (remove power adapter plug from socket). Also, ensure that the balance cannot be connected to the power supply during the work by a third party.
- Make sure that no liquid spills into the scale while performing maintenance work. If liquid is spilled on the scale, it must be inspected by a service technician.
- Regularly maintain the weighing pan and the weighing pan holder by removing any dirt or dust from under the weighing pan and on the weighing scale housing. Use a soft brush or a soft, lint-free cloth moistened with a mild soap solution.
- Never use solvents, acids, alkalis, paint thinners, scouring powders, or other aggressive or corrosive chemicals for cleaning; these substances can cause damage to the surfaces of the scale housing.
Transportation & Storage
- Your weighing machine is a precision instrument, treat it carefully.
- Avoid shaking, severe impacts, and vibration during transportation.
- Make sure that there are no marked temperature fluctuations during transportation and that the weighing machine does not become damp (condensation).
- If you would like to take the weighing machine out of service for an extended period, disconnect it from the electricity supply, clean it thoroughly (refer to Maintenance & Service) and store it in a place that meets the following conditions:
- No violent shaking or vibrations
- Minimum temperature fluctuations
- No direct solar radiation
- Minimum moisture
- The weighing machine should preferably be dispatched and transported in the original packaging to avoid transportation damage.
CAUTIONS
- When the recharging sign “
” indicates, please charge the battery for 8 to 10 hours. When the charging is done, the orange light will turn green. - Only use the charger (adapter) supplied with the indicator.
- After recharging, if the battery does not last long, replace it with a new battery.
Troubleshooting
| Error Code | Issue | Troubleshoot | Solution |
|
ERR-O/0 |
Overload error |
Turn scale off, then back on, and try weighing using weight under max capacity to verify the loadcell still functions. | If scale still registers accurate weight refrain from continued over loading. If error continues loadcell may be damaged. |
|
ERR-Z/2 |
Exceeded zero tracing range when powered on. |
Turn scale off and clear weighing platform. Turn scale back on and attempt recalibration. |
If the scale was recalibrated successfully then the error is resolved. If error persists, scale may need a linearity calibration or has a damaged loadcell. |
|
ERR-S/5 |
Scale not stable when powered on. |
Make sure scale is stable when powered on. Do not touch, shift, or place anything on the weighing platform until after the
countdown. |
If error persists, scale may need recalibration. If error continues after recalibration, then loadcell may be damaged. |
|
ERR-C |
Calibration weight was not entered during calibration | Enter calibration weight during calibration process. |
If error persists, call support. |
| ERR-P | Counting set error | Place sample before entering counting settings. | If error persists, call support. |
| ERR-L | Calibration weight was not placed during calibration. | Be sure to place weight when recalibrating. | If error persists, call support. |
| ERR-E | Software issue | If error persists, call support. | If error persists, call support. |
|
ERR-B/6 |
Low battery error |
Attempt to charge battery. |
If error persists, battery will need to be replaced. Make sure to not leave scale with rechargeable battery plugged in constantly. |
| For any error that is not shown on this list or that persists after troubleshooting you will need to contact support. | |||
- We are genuinely grateful for your recent purchase. Your decision to choose TREE by LW Measurements means a lot to us.
- Your support enables us to continue creating and providing high-quality products and services. We hope you enjoy your new scale and that it brings you as much satisfaction as it brings us to serve customers like you.
- We would love to know what you think of our TREE products and your experience with us. If you have a chance, please visit www.lwmeasurements.com and leave a review.
- Thank you for being a part of the LW Measurements family. We look forward to serving you again in the future.
All TREE products are designed in the U.S.
- Scan to Contact Us OR
- Call:707-542-2185

FAQs
Q: What should I do if liquid is spilled on the scale?
A: If liquid is spilled on the scale, immediately unplug it from the main electricity supply and ensure that no liquid enters the connections on the equipment or power adapter.
Documents / Resources
![]() |
TREE REMO-R Universal Indicator [pdf] User Manual REMO, REMO-R, REMO-R Universal Indicator, REMO-R, Universal Indicator, Indicator |