TiePie-engineering-LOGO

TiePie engineering HS3 Handy Scope Channel USB Oscilloscope with Function Generator

TiePie-engineering-HS3-Handy-Scope-Channel-USB-Oscilloscope-with-Function-Generator-PRODUCT

Product Usage Instructions

  • Safety
    • It is crucial to follow safety guidelines when using the Handyscope HS3. Avoid measuring directly on the line voltage to prevent potential hazards. Always use proper isolation transformers or differential probes when working with line voltage or grounded power supplies.
  • Driver Installation
    • Before using the Handyscope HS3, ensure that the drivers are correctly installed on your computer. Follow the instructions provided in the manual to install the drivers.
  • Hardware Installation
    • Power the instrument using the recommended method. Connect the Handyscope HS3 to your computer via USB. If issues arise, try plugging into a different USB port.
  • Front Panel
    • The front panel includes input connectors for CH1 and CH2, a GENERATOR output connector, and a power indicator. Familiarize yourself with these components before use.
  • Rear Panel
    • The rear panel features power options such as USB power cable and a power adapter. Additionally, there is an Extension Connector for further connectivity.

FAQs

Q: What should I do if the Handyscope HS3 is not recognized by my computer?

A: Try plugging the device into a different USB port on your computer. If the issue persists, check if the drivers are correctly installed.

Q: Can I measure line voltage directly with the Handyscope HS3?

A: It is highly advised against measuring line voltage directly as it can be dangerous. Always use isolation transformers or differential probes for such measurements.

“`

Handyscope HS3
User manual
TiePie engineering

ATTENTION! Measuring directly on the line voltage can be very dangerous. The outside of the BNC connectors at the Handyscope HS3 are connected with the ground of the computer. Use a good isolation transformer or a differential probe when measuring at the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS3 is connected to a positive voltage. This short-circuit current can damage both the Handyscope HS3 and the computer.
Copyright ©2025 TiePie engineering. All rights reserved. Revision 2.55, October 2025 This information is subject to change without notice. Despite the care taken for the compilation of this user manual, TiePie engineering can not be held responsible for any damage resulting from errors that may appear in this manual.

Safety

1

When working with electricity, no instrument can guarantee complete safety. It is the responsibility of the person who works with the instrument to operate it in a safe way. Maximum security is achieved by selecting the proper instruments and following safe working procedures. Safe working tips are given below:

· Always work according (local) regulations.
· Work on installations with voltages higher than 25 VAC or 60 VDC should only be performed by qualified personnel.
· Avoid working alone.
· Observe all indications on the Handyscope HS3 before connecting any wiring
· Check the probes/test leads for damages. Do not use them if they are damaged
· Take care when measuring at voltages higher than 25 VAC or 60 VDC. · Do not operate the equipment in an explosive atmosphere or in the pres-
ence of flammable gases or fumes.
· Do not use the equipment if it does not operate properly. Have the equipment inspected by qualified service personal. If necessary, return the equipment to TiePie engineering for service and repair to ensure that safety features are maintained.
· Measuring directly on the line voltage can be very dangerous. The outside of the BNC connectors at the Handyscope HS3 are connected with the ground of the computer. Use a good isolation transformer or a differential probe when measuring at the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS3 is connected to a positive voltage. This short-circuit current can damage both the Handyscope HS3 and the computer.

Safety 1

Declaration of conformity

TiePie engineering Koperslagersstraat 37 8601 WL Sneek The Netherlands

2

EC Declaration of conformity
We declare, on our own responsibility, that the product
Handyscope HS3-10MHz Handyscope HS3-25MHz Handyscope HS3-50MHz Handyscope HS3-100MHz

for which this declaration is valid, is in compliance with
EC directive 2011/65/EU (the RoHS directive) including up to amendment 2021/1980,
EC regulation 1907/2006 (REACH) including up to amendment 2021/2045,
and with

EN 55011:2016/A1:2017 EN 55022:2011/C1:2011

IEC 61000-6-1:2019 EN IEC 61000-6-3:2007/A1:2011/C11:2012 EN

according the conditions of the EMC directive 2014/30/EU

also with Canada: ICES-001:2004 and IEC 61010-1:2010/A1:2019

Australia/New Zealand: AS/NZS CISPR 11:2011 USA: UL 61010-1, Edition 3

and is categorized as 30 VRMS, 42 Vpk, 60 VDC

Sneek, 1-9-2022 ir. A.P.W.M. Poelsma

Declaration of conformity

3

Environmental considerations
This section provides information about the environmental impact of the Handyscope HS3.
End-of-life handling
Production of the Handyscope HS3 required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the Handyscope HS3’s end of life.
In order to avoid release of such substances into the environment and to reduce the use of natural resources, recycle the Handyscope HS3 in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
The shown symbol indicates that the Handyscope HS3 complies with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE).

4

Chapter 2

Introduction

3

Before using the Handyscope HS3 first read chapter 1 about safety.

Many technicians investigate electrical signals. Though the measurement may not be electrical, the physical variable is often converted to an electrical signal, with a special transducer. Common transducers are accelerometers, pressure probes, current clamps and temperature probes. The advantages of converting the physical parameters to electrical signals are large, since many instruments for examining electrical signals are available.
The Handyscope HS3 is a portable two channel measuring instrument with Arbitrary Waveform Generator. The Handyscope HS3 is available in several models with different maximum sampling rates: 10 MSa/s, 25 MSa/s, 50 MSa/s or 100 MSa/s. The native resolution is 12 bits, but user selectable resolutions of 8, 14 and 16 bits are available too, with adjusted maximum sampling rate:

Resolution HS3-100

HS3-50

HS3-25

HS3-10

8 bit 100 MSa/s 50 MSa/s 25 MSa/s 10 MSa/s

12 bit 50 MSa/s 50 MSa/s 25 MSa/s 10 MSa/s

14 bit 3.125 MSa/s 3.125 MSa/s 3.125 MSa/s 3.125 MSa/s

16 bit 195 kSa/s 195 kSa/s 195 kSa/s 195 kSa/s

Table 3.1: Maximum sampling rates

With the accompanying software the Handyscope HS3 can be used as an oscilloscope, a spectrum analyzer, a true RMS voltmeter or a transient recorder. All instruments measure by sampling the input signals, digitizing the values, process them, save them and display them.
3.1 Sampling
When sampling the input signal, samples are taken at fixed intervals. At these intervals, the size of the input signal is converted to a number. The accuracy of this number depends on the resolution of the instrument. The higher the resolution, the smaller the voltage steps in which the input range of the instrument is divided. The acquired numbers can be used for various purposes, e.g. to create a graph.TiePie-engineering-HS3-Handy-Scope-Channel-USB-Oscilloscope-with-Function-Generator-FIG- (1)

Introduction

5

Figure 3.1: Sampling
The sine wave in figure 3.1 is sampled at the dot positions. By connecting the adjacent samples, the original signal can be reconstructed from the samples. You can see the result in figure 3.2.TiePie-engineering-HS3-Handy-Scope-Channel-USB-Oscilloscope-with-Function-Generator-FIG- (2)

Figure 3.2: “connecting” the samples
3.2 Sampling rate
The rate at which the samples are taken is called the sampling rate, the number of samples per second. A higher sampling rate corresponds to a shorter interval between the samples. As is visible in figure 3.3, with a higher sampling rate, the original signal can be reconstructed much better from the measured samples.TiePie-engineering-HS3-Handy-Scope-Channel-USB-Oscilloscope-with-Function-Generator-FIG- (3)

6

Chapter 3

Figure 3.3: The effect of the sampling rate

3.2.1

The sampling rate must be higher than 2 times the highest frequency in the input signal. This is called the Nyquist frequency. Theoretically it is possible to reconstruct the input signal with more than 2 samples per period. In practice, 10 to 20 samples per period are recommended to be able to examine the signal thoroughly.
Aliasing
When sampling an analog signal with a certain sampling rate, signals appear in the output with frequencies equal to the sum and difference of the signal frequency and multiples of the sampling rate. For example, when the sampling rate is 1000 Sa/s and the signal frequency is 1250 Hz, the following signal frequencies will be present in the output data:

Multiple of sampling rate …
-1000 0
1000 2000

1250 Hz signal
-1000 + 1250 = 250 0 + 1250 = 1250
1000 + 1250 = 2250 2000 + 1250 = 3250

-1250 Hz signal
-1000 – 1250 = -2250 0 – 1250 = -1250
1000 – 1250 = -250 2000 – 1250 = 750

Table 3.2: Aliasing

As stated before, when sampling a signal, only frequencies lower than half the sampling rate can be reconstructed. In this case the sampling rate is 1000 Sa/s, so we can we only observe signals with a frequency ranging from 0 to 500 Hz. This means that from the resulting frequencies in the table, we can only see the 250 Hz signal in the sampled data. This signal is called an alias of the original signal.
If the sampling rate is lower than twice the frequency of the input signal, aliasing will occur. The following illustration shows what happens.TiePie-engineering-HS3-Handy-Scope-Channel-USB-Oscilloscope-with-Function-Generator-FIG- (4)

Introduction

7

Figure 3.4: Aliasing
In figure 3.4, the green input signal (top) is a triangular signal with a frequency of 1.25 kHz. The signal is sampled with a rate of 1 kSa/s. The corresponding sampling interval is 1/1000Hz = 1ms. The positions at which the signal is sampled are depicted with the blue dots. The red dotted signal (bottom) is the result of the reconstruction. The period time of this triangular signal appears to be 4 ms, which corresponds to an apparent frequency (alias) of 250 Hz (1.25 kHz – 1 kHz).
To avoid aliasing, always start measuring at the highest sampling rate and lower the sampling rate if required.
3.3 Digitizing
When digitizing the samples, the voltage at each sample time is converted to a number. This is done by comparing the voltage with a number of levels. The resulting number is the number corresponding to the level that is closest to the voltage. The number of levels is determined by the resolution, according to the following relation: LevelCount = 2Resolution. The higher the resolution, the more levels are available and the more accurate the input signal can be reconstructed. In figure 3.5, the same signal is digitized, using two different amounts of levels: 16 (4-bit) and 64 (6-bit).TiePie-engineering-HS3-Handy-Scope-Channel-USB-Oscilloscope-with-Function-Generator-FIG- (5)

8

Chapter 3

Figure 3.5: The effect of the resolution
The Handyscope HS3 measures at e.g. 12 bit resolution (212=4096 levels). The smallest detectable voltage step depends on the input range. This voltage can be calculated as:
V oltageStep = F ullInputRange/LevelCount
For example, the 200 mV range ranges from -200 mV to +200 mV, therefore the full range is 400 mV. This results in a smallest detectable voltage step of 0.400 V / 4096 = 97.65 µV.
3.4 Signal coupling
The Handyscope HS3 has two different settings for the signal coupling: AC and DC. In the setting DC, the signal is directly coupled to the input circuit. All signal components available in the input signal will arrive at the input circuit and will be measured. In the setting AC, a capacitor will be placed between the input connector and the input circuit. This capacitor will block all DC components of the input signal and let all AC components pass through. This can be used to remove a large DC component of the input signal, to be able to measure a small AC component at high resolution.
When measuring DC signals, make sure to set the signal coupling of the input to DC.
3.5 Probe compensation
The Handyscope HS3 is shipped with a probe for each input channel. These are 1x/10x selectable passive probes. This means that the input signal is passed through directly or 10 times attenuated.

Introduction

9

When using an oscilloscope probe in 1:1 the setting, the bandwidth of the probe is only 6 MHz. The full bandwidth of the probe is only obtained in the 1:10 setting
The x10 attenuation is achieved by means of an attenuation network. This attenuation network has to be adjusted to the oscilloscope input circuitry, to guarantee frequency independency. This is called the low frequency compensation. Each time a probe is used on an other channel or an other oscilloscope, the probe must be adjusted. Therefore the probe is equiped with a setscrew, with which the parallel capacity of the attenuation network can be altered. To adjust the probe, switch the probe to the x10 and attach the probe to a 1 kHz square wave signal. Then adjust the probe for a square front corner on the square wave displayed. See also the following illustrations.
Figure 3.6: correct

Figure 3.7: under compensated

10 Chapter 3

Figure 3.8: over compensated

Driver installation

4

Before connecting the Handyscope HS3 to the computer, the drivers need to be installed.

4.1
4.1.1 4.1.2

Introduction
To operate a Handyscope HS3, a driver is required to interface between the measurement software and the instrument. This driver takes care of the low level communication between the computer and the instrument, through USB. When the driver is not installed, or an old, no longer compatible version of the driver is installed, the software will not be able to operate the Handyscope HS3 properly or even detect it at all.
The installation of the USB driver is done in a few steps. Firstly, the driver has to be pre-installed by the driver setup program. This makes sure that all required files are located where Windows can find them. When the instrument is plugged in, Windows will detect new hardware and install the required drivers.
Where to find the driver setup
The driver setup program and measurement software can be found in the download section on TiePie engineering’s website. It is recommended to install the latest version of the software and USB driver from the website. This will guarantee the latest features are included.
Executing the installation utility
To start the driver installation, execute the downloaded driver setup program. The driver install utility can be used for a first time installation of a driver on a system and also to update an existing driver.
The screen shots in this description may differ from the ones displayed on your computer, depending on the Windows version.

Driver installation 11

Figure 4.1: Driver install: step 1 When drivers were already installed, the install utility will remove them before installing the new driver. To remove the old driver successfully, it is essential that the Handyscope HS3 is disconnected from the computer prior to starting the driver install utility. When the Handyscope HS3 is used with an external power supply, this must be disconnected too. Clicking “Install” will remove existing drivers and install the new driver. A remove entry for the new driver is added to the software applet in the Windows control panel.
Figure 4.2: Driver install: Copying files
12 Chapter 4

Figure 4.3: Driver install: Finished
Driver installation 13

14 Chapter 4

Hardware installation

5

Drivers have to be installed before the Handyscope HS3 is connected to the computer for the first time. See chapter 4 for more information.

5.1 5.1.1
5.2 5.3

Power the instrument
The Handyscope HS3 is powered by the USB, no external power supply is required. Only connect the Handyscope HS3 to a bus powered USB port, otherwise it may not get enough power to operate properly.
External power
In certain cases, the Handyscope HS3 cannot get enough power from the USB port. When a Handyscope HS3 is connected to a USB port, powering the hardware will result in an inrush current higher than the nominal current. After the inrush current, the current will stabilize at the nominal current.
USB ports have a maximum limit for both the inrush current peak and the nominal current. When either of them is exceeded, the USB port will be switched off. As a result, the connection to the Handyscope HS3 will be lost.
Most USB ports can supply enough current for the Handyscope HS3 to work without an external power supply, but this is not always the case. Some (battery operated) portable computers or (bus powered) USB hubs do not supply enough current. The exact value at which the power is switched off, varies per USB controller, so it is possible that the Handyscope HS3 functions properly on one computer, but does not on another.
In order to power the Handyscope HS3 externally, an external power input is provided for. It is located at the rear of the Handyscope HS3. Refer to paragraph 7.1 for specifications of the external power intput.
Connect the instrument to the computer
After the new driver has been pre-installed (see chapter 4), the Handyscope HS3 can be connected to the computer. When the Handyscope HS3 is connected to a USB port of the computer, Windows will detect new hardware.
Depending on the Windows version, a notification can be shown that new hardware is found and that drivers will be installed. Once ready, Windows will report that the driver is installed.
When the driver is installed, the measurement software can be installed and the Handyscope HS3 can be used.
Plug into a different USB port
When the Handyscope HS3 is plugged into a different USB port, some Windows versions will treat the Handyscope HS3 as different hardware and will install the drivers again for that port. This is controlled by Microsoft Windows and is not caused by TiePie engineering.

Hardware installation 15

16 Chapter 5

Front panel

6

Figure 6.1: Front panel
6.1 CH1 and CH2 input connectors
The CH1 and CH2 BNC connectors are the main inputs of the acquisition system. The outside of the BNC connectors is connected to the ground of the Handyscope HS3. Connecting the outside of the BNC connector to a potential other than ground will result in a short circuit that may damage the device under test, the Handyscope HS3 and the computer.
6.2 GENERATOR output connector
The OUT BNC connector is the output of the internal Arbitrary Waveform Generator. The outside of this BNC connector is connected to the ground of the Handyscope HS3.
When the generator is switched off in the software, the generator output is switched to a high impedance, floating state, the output voltage is then undefined.
When the generator is switched on in the software and set to pause, the generator output is switched to a low impedance (50 ), the output voltage is at the selected offset level.
6.3 Power indicator
A power indicator is situated at the top cover of the instrument. It is lit when the Handyscope HS3 is powered.

Front panel 17

18 Chapter 6

Rear panel

7

Figure 7.1: Rear panel
7.1 Power
The Handyscope HS3 is powered through the USB. If the USB cannot deliver enough power, it is possible to power the instrument externally. The Handyscope HS3 has two external power inputs located at the rear of the instrument: the dedicated power input and a pin of the extension connector.

Handyscope HS3s with SN# 11832 and lower do not have a dedicated power input at the rear, they only have an external power input on the extension connector.
The specifications of the dedicated power connector are:

Pin Center pin Outside bushing

Dimension Ø1.3 mm Ø3.5 mm

Description ground positive

Figure 7.2: Power connector
Besides the external power input, it is also possible to power the instrument through the extension connector, the 25 pin D-sub connector at the rear of the instrument. The power has to be applied to pin 3 of the extension connector. Pin 4 can be used as ground.
The following minimum and maximum voltages apply to both power inputs:

SN# <12941 SN# >12941

Minimum
4.5 VDC 4.5 VDC

Maximum
6 VDC 12 VDC

Table 7.1: Maximum voltages

Rear panel 19

7.1.1

Note that the externally applied voltage should be higher than the USB voltage to relieve the USB port.
USB power cable
The Handyscope HS3 is delivered with a special USB external power cable.

Figure 7.3: USB power cable
One end of this cable can be connected to a second USB port on the computer, the other end can be plugged in the external power input at the rear of the instrument. The power for the instrument will be taken from two USB ports of the computer.

The outside of the external power connector is connected to +5 V. In order to avoid shortage, first connect the cable to the Handyscope HS3 and then to the USB port.

7.1.2 7.2

Power adapter
In case a second USB port is not available, or the computer still can’t provide enough power for the instrument, an external power adapter can be used. When using an external power adapter, make sure that:
· the polarity is set correctly · the voltage is set to a valid value for the instrument and higher than the USB
voltage · the adapter can supply enough current (preferably >1 A) · the plug has the correct dimensions for the external power input of the in-
strument
USB
The Handyscope HS3 is equipped with a USB 2.0 High speed (480 Mbit/s) interface with a fixed cable with type A plug. It will also work on a computer with a USB 1.1 interface, but will then operate at 12 Mbit/s.

20 Chapter 7

7.3 Extension Connector

Figure 7.4: Extension connector

To connect to the Handyscope HS3 a 25 pin female D-sub connector is available, containing the following signals:

Pin Description
1 Ground 2 Reserved 3 External Power in DC 4 Ground 5 +5V out, 10 mA max. 6 Ext. sampling clock in (TTL) 7 Ground 8 Ext. trigger in (TTL) 9 Data OK out (TTL) 10 Ground 11 Trigger out (TTL) 12 Reserved 13 Ext. sampling clock out (TTL)

Pin Description
14 Ground 15 Ground 16 Reserved 17 Ground 18 Reserved 19 Reserved 20 Reserved 21 Generator Ext Trig in (TTL) 22 Ground 23 I2C SDA 24 I2C SCL 25 Ground

Table 7.2: Pin description Extension connector

All TTL signals are 3.3 V TTL signals which are 5 V tolerant, so they can be connected to 5 V TTL systems.
For instruments with serial number 14266 and higher, pins 9, 11, 12, 13 are open collector outputs. Connect a pull-up resistor of 1 kOhm to pin 5 when using one of these signals. For older instruments, the outputs are standard TTL outputs and no pull-up is required.

Rear panel 21

22 Chapter 7

Specifications

8

The accuracy of a channel is defined as a percentage of the Full Scale range. The Full Scale range runs from -range to range and is effectively 2 * range. When the input range is set to 4 V, the Full Scale range is -4 V to 4 V = 8 V. Additionally a number of Least Significant Bits is incorporated. The acuracy is determined in the highest resolution.
When the accuracy is specified as ±0.2% of the Full Scale range ± 1 LSB, and the input range is 4 V, the maximum deviation the measured value can have is ±0.2% of 8 V = ±16 mV. ±1 LSB equals 8 V / 65536 (= number of LSB at 16 bit) = ± 122 µV. Therefore the measured value will be between 16.122 mV lower and 16.122 mV higher than the actual value. When e.g. applying a 3.75 V signal and measuring it in the 4 V range, the measured value will be between 3.76612 V and 3.73388 V.
8.1 Acquisition system

Number of input channels

2 analog

CH1, CH2

BNC, female

Type

Single ended

Resolution

8, 12, 14, 16 bit user selectable

Accuracy

0.2% of full scale ± 1 LSB

Ranges (full scale)

±200 mV ±2 V ±20 V

±400 mV ±4 V ±40 V

Coupling

AC/DC

Impedance

1 M / 30 pF

Noise Maximum voltage

210 µVRMS (200 mV range, 12 bit, 50 MSa/s) 50 µVRMS (200 mV range, 16 bit, 195 kSa/s)
200 V (DC + AC peak <10 kHz)

Bandwidth (-3dB)

50 MHz

AC coupling cut off frequency (-3dB)±1.5 Hz

Maximum sampling rate

HS3-100

HS3-50

HS3-25

HS3-10

8 bit

100 MSa/s 50 MSa/s

25 MSa/s

10MSa/s

12 bit

50 MSa/s

50 MSa/s

25 MSa/s

10MSa/s

14 bit

3.125 MSa/s 3.125 MSa/s 3.125 MSa/s 3.125 MSa/s

16 bit

195 kSa/s 195 kSa/s 195 kSa/s 195 kSa/s

Maximum streaming rate

10 kSa/s

Sampling source

internal quartz, external

Internal

Quartz

Accuracy Stability

±0.01% ±100 ppm over -40C to +85C

Time base aging

±5 ppm/year

External

On extension connector

Voltage

3.3 V TTL, 5 V TTL tolerant

Frequency range

10 MHz – 100 MHz

Memory

128 kSamples per channel (256 kSamples with disabled generator)

±800 mV ±8 V ±80 V
HS3-5 * 5 MSa/s 5 MSa/s 3.125 MSa/s 195 kSa/s

* The HS3-5 model is no longer available, its specs remain available here for reference

Specifications 23

8.2 8.3

Trigger system
Triggering is only availabe when the Handyscope HS3 operates in block mode, not when operating in streaming mode

System Source
Trigger modes
Level adjustment Hysteresis adjustment Resolution Pre trigger Post trigger Digital external trigger
Input Range Coupling

digital, 2 levels CH1, CH2, digital external, AND, OR, AWG Start, AWG Stop, AWG New period rising slope, falling slope, inside window, outside window 0 to 100% of full scale 0 to 100% of full scale 0.024 % (12 bits) 0 to 128 ksamples (0 to 100%, one sample resolution) 0 to 128 ksamples (0 to 100%, one sample resolution)
extension connector 0 to 5 V (TTL) DC

Arbitrary Waveform Generator

Output channel DAC Resolution Output range Amplitude Range Resolution Accuracy Noise
DC offset Range Resolution Accuracy
Coupling Impedance Bandwidth System Memory Operating mode Maximum sampling rate Sampling source Accuracy Stability Time base aging Waveforms Symmetry

1 analog, female BNC 14 bit @ 50 MSa/s -12 V to 12 V (open circuit)
±0.12 V, ±1.2 V, ±12 V (open circuit) 13 bit 0.4 % of range 0.12 V range: 900 µVRMS 1.2 V range: 1.3 mVRMS 12 V range: 1.5 mVRMS
-12 V to 12 V (open circuit) 13 bit 0.4 % of range DC 50 2 MHz DDS 256k points Continuous, triggered, gated 50 MSa/s internal quartz ±0.01% ±100 ppm over -40C to +85C ±5 ppm/year sine, square, triangle, noise, DC and user defined 0 to 100%

24 Chapter 8

8.4 Interface

Interface
8.5 Power

USB 2.0 High Speed (480 Mbit/s) (USB 1.1 Full Speed (12 Mbit/s) and USB 3.0 compatible)

Input Consumption
8.6 Physical

from USB or external input 500 mA max

Instrument height Instrument length Instrument width Weight USB cord length
8.7 I/O connectors

25 mm / 1.0″ 170 mm / 6.7″ 140 mm / 5.2″ 480 gram / 17 ounce 1.8 m / 70″

8.8

CH1, CH2 Generator out Power Extension connector USB

BNC, female BNC, female 3.5 mm power socket D-sub 25 pins female Fixed cable with type A plug

System requirements

PC I/O connection Operating System

USB 2.0 High Speed (480 Mbit/s) (USB 1.1 and USB 3.0 compatible)
Windows 10 / 11, 64 bits

8.9 Environmental conditions

Operating Ambient temperature Relative humidity
Storage Ambient temperature Relative humidity

0C to 55C 10 to 90% non condensing
-20C to 70C 5 to 95% non condensing

Specifications 25

8.10 Certifications and Compliances

CE mark compliance

Yes

RoHS

Yes

REACH

Yes

EN 55011:2016/A1:2017

Yes

EN 55022:2011/C1:2011

Yes

IEC 61000-6-1:2019 EN

Yes

IEC 61000-6-3:2007/A1:2011/C11:2012

Yes

ICES-001:2004

Yes

AS/NZS CISPR 11:2011

Yes

IEC 61010-1:2010/A1:2019

Yes

UL 61010-1, Edition 3

Yes

8.11 Probes

Model
Bandwidth Rise time Input impedance Input capacitance Compensation range Working voltage (DC + AC peak)
8.12 Package contents

HP-3250I
X1
6 MHz
58 ns
1 M
oscilloscope impedance
56 pF
+ oscilloscope capacitance

300 V 150 V CAT II

X10 250 MHz 1.4 ns 10 M
incl. 1 M oscilloscope impedance
13 pF
10 to 30 pF 600 V 300 V CAT II

Instrument Probes Accessories Software Drivers Software Development Kit Manual

Handyscope HS3 2 x HP-3250I X1 / X10 switchable USB power cable Windows 10 / 11, 64 bits, via website Windows 10 / 11, 64 bits, via website Windows 10 / 11 (64 bits) and Linux, via website Instrument manual and software manual

26 Chapter 8

If you have any suggestions and/or remarks regarding this manual, please contact:

TiePie engineering Koperslagersstraat 37 8601 WL Sneek The Netherlands

Tel.: Fax: E-mail: Site:

+31 515 415 416 +31 515 418 819 support@tiepie.nl www.tiepie.com

TiePie engineering Handyscope HS3 instrument manual revision 2.55, October 2025

Documents / Resources

TiePie engineering HS3 Handy Scope Channel USB Oscilloscope with Function Generator [pdf] User Manual
100 MS-s 12-bit 2-kanaals, USB-oscilloscoop met functiegenerator, HS3 Handy Scope Channel USB Oscilloscope with Function Generator, HS3, Handy Scope Channel USB Oscilloscope with Function Generator, USB Oscilloscope with Function Generator, with Function Generator, Function Generator

References

Leave a comment

Your email address will not be published. Required fields are marked *