These release notes contain important information about version 1.2.0 of the AFG1062 software.
Introduction
This document provides supplemental information regarding the behavior of the AFG1062 software. This information is grouped into six categories:
| Revision history | Lists the version of software, the document version, and the date of the software release. |
| New features/enhancements | Summary of each significant new feature included. |
|
Problem fixes |
Summary of each significant software/firmware bug fix. |
|
Known problems |
Description of each significant known problem and ways to work around it. |
| Installation instructions | Detailed instructions describing how to install the software. |
| Appendix A – Previous versions | Information about previous versions of the software. |
Revision history
This document is periodically updated and distributed with releases and service packs to provide the most up-to-date information. This revision history is included below.
| Date | Software version | Document number |
| 9/1/2025 | V1.2.0 | 077187500 |
| 8/8/2024 | V1.1.0 | – |
| 10/29/2019 | V1.0.9 | – |
| 5/27/2018 | V1.0.6
V1.0.5 |
–
– |
| 4/23/2017 | V1.0.4
V1.0.3
V1.0.2 |
–
–
– |
VERSION 1.2.0
New features/enhancements
| Issue number | CAS-255025-Q1G0R5 |
| Models affected | AFG1062 |
| Enhancement | Five new SCPI commands added. Details below. |
| Issue number | CES2T-100 |
| Models affected | AFG1062 |
| Enhancement | Upon firmware upgrade, acceptance of the End User License Agreement (EULA) on the front panel of the instrument is required before use. |
[SOURce[1|2]]:BURSt:PERiod
This command sets or queries the burst period (trigger interval) in burst mode.
- Conditions: None
- Group: Source
- Syntax: [SOURce[1|2]]:BURSt:PERiod {<seconds> | MINimum | MAXimum}
- [SOURce[1|2]]:BURSt:PERiod? {MINimum | MAXimum}
- Arguments: <seconds>::=<NRf>[<units>] where:
- <NRf> is the burst period in seconds.
- <units>::=[ns | μs | ms | s]
- Returns: <seconds>
- Example: SOURce1:BURSt:PERiod 100ms
Sets the channel 1 burst period to 100 ms.
[SOURce[1|2]]:BURSt:PHASe
This command sets or queries the phase of the output waveform in burst mode for the specified channel. Set the value in radians or degrees. If no units are specified, the default is RAD. The query returns the value in RAD.
- Conditions: None
- Group: Source
- Syntax: [SOURce[1|2]]:BURSt:PHASe {<phase> | MINimum | MAXimum}
- [SOURce[1|2]]:BURSt:PHASe? {MINimum | MAXimum}
- Arguments: <phase>::=<NR3>[<units>]
- where:
- <NR3> is the phase of the output waveform.
- <units>::=[RAD | DEG]
- If units are omitted, RAD is selected. The setting ranges are:
- RAD: -2 Pl to +2 Pl, relative to phase value.
- DEG: -360 to +360, relative to phase value.
- Returns: <phase>
- Example: SOURce1:BURSt:PHASe MAXimum
Sets the maximum value for the burst mode phase of the output waveform for CH 1.
[SOURce[1|2]]:BURSt:POLarity
This command sets or queries the trigger gate polarity in burst gated mode.
- Conditions: None
- Group: Source
- Syntax: [SOURce[1|2]]:BURSt:POLarity {POSitive|NEGative}
- [SOURce[1|2]]:BURSt:POLarity?
- Arguments: POSitive specifies a positive trigger gate polarity.
- NEGative specifies a negative trigger gate polarity.
- Returns: POS|NEG
- Example: [SOURce[1|2]]:BURSt:POLarity POSitive
- Sets the trigger gate polarity of CH 1 to positive.
SYSTem:CLKSrc
This command sets or queries the system’s clock source.
- Group: System
- Syntax: SYSTem:CLKSrc {INTernal|EXTernal}
- SYSTem:CLKSrc?
- Arguments: INTernal specifies the clock source as the internal clock signal.
- EXTernal specifies the clock source as an external clock signal.
- Returns: INT|EXT
- Example: SYSTem:CLKSrc INTernal
- Sets the system clock source to the internal clock.
TRIGger:SLOPe
This command sets or queries the slope edge the external trigger event occurs on.
- Group: Trigger
- Syntax: TRIGger:SLOPe {POSitive|NEGative}
- TRIGger:SLOPe?
- Arguments: POSitive specifies that the trigger event occurs on the rising edge of the externaltrigger signal.
- NEGative specifies that the trigger event occurs on the falling edge of the externaltrigger signal.
- Returns: POS|NEG,POS|NEG,POS|NEG,POS|NEG where the first and third return value is the external trigger slope of sweep mode for channel 1 and 2, and the second and fourth return value is the external trigger slope of burst mode for channel 1 and 2.
- Example: TRIGger:SLOPe POSitive Sets the trigger slope to positive, which triggers on the rising edge of the signal.
Problem fixes
| Issue number | CAS-290934-H5T6R9 |
| Models affected | AFG1062 |
| Symptom | Pulse waveform output in continuous mode goes to pulse high level for an extended time before going to 0 V when the output is turned off. |
| Resolution | This issue has been corrected. |
| Issue number | CAS-296793-Y7Y0W4 |
| Models affected | AFG1062 |
| Symptom | Pulse waveform output incorrectly inverted when an offset other than 0 V was used. |
| Resolution | This issue has been corrected. |
| Issue number | CAS-333592-P5X5K9 |
| Models affected | AFG1062 |
| Symptom | The [SOURce[1|2]]:PULSe:DCYCle command generates an error. |
| Resolution | This issue has been corrected. |
| Issue number | CAS-362740-R9L3J4 |
| Models affected | AFG1062 |
| Symptom | Ref Clk/Counter could not synchronize to an external reference clock or counter input. |
| Resolution | This issue has been corrected. |
| Issue number | CAS-40294-GS62Q8 |
| Models affected | AFG1062 |
| Symptom | When using pulse waveform in burst mode, changing the pulse period incorrectly changed the pulse width. |
| Resolution | This issue has been corrected. |
Known problems
| Issue number | CAS-215793-Z6H6W9 |
| Models affected | AFG1062 |
| Symptom | Ref Clk Out cannot output an external reference clock or counter it receives
through its Ref Clk/Counter input. |
| Workaround | When synchronizing the clocks of more than two instruments, place the AFG as either the origin of the clock signal or the last instrument being synchronized. |
Installation instructions
CAUTION. Updating your instrument firmware is a sensitive operation; it is critical that you follow the instructions below. If you do not, you may cause damage to your instrument. For example, to prevent damage to the instrument, do not remove the USB flash drive at any time while updating the firmware, and do not power off the instrument during the update process.
To update your instrument firmware:
- Visit tek.com and search for the AFG1062 firmware.
- Download the compressed .zip file to your computer.
- Unzip the downloaded file and copy the .tfb file to the USB flash drive root directory.
- Insert the USB flash drive into the instrument front panel.
- On the front panel, press the Utility button. The Utility menu is displayed.
- From the instrument Utility menu, select System.
- Press Next Page until the Update firmware option is displayed. This option will be disabled if you do not have a USB flash drive inserted or Access Protection is turned on.
- Select Update firmware.
- With the navigation wheel, select USBDEVICE, then select Enter.
- With the navigation wheel, select the .tfb file, then select Execute.
- Reading file… will be displayed, followed by Updating file system… with a progress bar.
CAUTION. Updating the firmware takes approximately two minutes. Do not remove the USB flash drive during the update process.
If you accidentally remove the USB drive during the update, do not power off the instrument. Restart the upgrade process from the Utility menu. - When the upgrade is finished, the instrument will display “Firmware upgrade success!”
- The instrument will automatically reboot.
- When the instrument has rebooted, the instrument End-User License Agreement (EULA) is displayed.You must accept the EULA before you can use the instrument.
- Remove the USB drive. You can verify that the firmware has been updated by going to the Utility menuand viewing the version.
Firmware downgrades
If your instrument does not function properly after a firmware upgrade, you can downgrade by following the steps under Installation instructions with an earlier version of the firmware file. However, please note the following conditions before attempting a downgrade.
- If your instrument serial number is 2231001 or higher, you can only downgrade to Version 1.1.0 ornewer
- If your instrument serial number is 2231000 or lower, you can only downgrade to Version 1.1.0 ornewer once the instrument is upgraded to Version 1.1.0 or newer
If you attempt a downgrade and it is not successful due to the specified conditions, you will receive an error.
Appendix A – Previous versions
VERSION 1.1.0
This release does not include any critical fixes or enhancements.
VERSION 1.0.9
Problem fixes
| Issue number | AR-62341 |
| Models affected | AFG1062 |
| Symptom | Pulse duty cycle stuck at 50% in burst mode, even when user attempts to change it. |
| Resolution | This issue has been corrected. |
| Issue number | AR-62464 |
| Models affected | AFG1062 |
| Symptom | Cycle counts larger than 65000 in burst mode is incorrect. |
| Resolution | This issue has been corrected. |
VERSION 1.0.6
Problem fixes
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | Output signal requires optimization when trigger source is manual or external in
burst mode. |
| Resolution | This issue has been corrected. |
VERSION 1.0.5
Problem fixes
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | The arbitrary waveform does not output in burst mode. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | Phase align malfunctions in burst mode. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | DC offset is inconsistent with over waveforms. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | The HFR malfunctions. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | Abnormal arbitrary waveform output under certain conditions. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | Thumbnail does not update when changes to the settings are made. |
| Resolution | This issue has been corrected. |
VERSION 1.0.4
Problem fixes
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | The [SOURce[1|2]]:PHASe[:ADJust] command does not work in burst mode. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | The trigger level is not correct in sweep mode. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | The arbitrary waveform quality requires optimization. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | The counter function requires optimization. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | More PI commands require support. |
| Resolution | This issue has been corrected. |
VERSION 1.0.3
New features/enhancements
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | Added support for the DATA:DATA? EMEM command. |
| Resolution | This issue has been corrected. |
VERSION 1.0.2
New features/enhancements
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | Align phase function upgraded. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | User settings can be saved to the instrument (.tfs format). |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | User waveforms can be moved from a USB drive to the instrument (.tfw
format). |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | FAT16 file system has been upgraded to support USB drives. |
| Resolution | This issue has been corrected. |
| Issue number | Not applicable |
| Models affected | AFG1062 |
| Symptom | Additional SCPI commands added. |
| Resolution | This issue has been corrected. |
Frequently Asked Questions
Q: How can I update the firmware of AFG1062?
A: To update the firmware, follow the installation instructions provided in the user manual for Version 1.2.0.
Q: What should I do if I encounter a known problem with the software?
A: If you encounter a known problem, refer to the problem fixes section in the release notes for possible workarounds.