SWEVY Digital Refractometer Installation Guide

SWEVY Digital Refractometer Installation Guide
SWEVY Digital Refractometer Installation Guid

Instructions for the Digital Refractometer

Specifications

Resolution 0.1% 0.01%
Accuracy ±0.2% ±0.15%
Temperaturecompensation 10-40 °C
Battery 2*AAA
Size and weight 117(L) * 55(H) * 29(H) * mm, 113g

Status Notice

Error Messages Reason
Excessive light External light is too strong to measure.
High TempLow Temp The temperature in the measuring slot is out of range.
Re zero Please follow the zero-clearing step again.
Less solution Too little solution in measuring slot.
Less water Too little clean water in measuring slot.
Out of range Excessive concentration
Low voltage Low voltage
Fault Possible damage to components.

Warning Icon Warning

  • This instrument is a precise optical instru-ment, Prevent it from collision and fall.
  • Forbid hard objects to scratch the optical prism in measuring slot.
  • When it is out of use, please store it in a place as dry as possible.

Change battery

This instrument is powered by two AAA batteries. Remove the battery or battery cover in the following order.
Change battery

Parts
Overview

Language
Language

Cautions

  1. Cold or hot measuring solution
    Solution: Drop in the measuring solution and wait 20 seconds before measuring. Because the instrument needs to do temperature supplement, this requires the instrument and the measuring solution to maintain the same temperature ,and this takes more than 20 seconds.
    Wait 20 seconds
  2. Measuring solution containing suspended substances
    Solution (Step 1):
    Remove suspended solids from measuring solution with filter spoon or gauze as much as possible.
    Remove suspended solids
    Solution (Step 2):
    Mix measuring solution evenly (do not use metal and hard objects, and avoid scratching the optical prism). Because uneven measuring solutions or suspended solids can lead to instrumental uncertainty in measurement.
    Measuring solution containing suspended substances

Cleaning

If there is dirt in the measuring slot or the instrument is not used for a long time, It should be cleaned first.
Cleaning

Zero-setting

Measure whether the clean water or pure water is 0.0% before using, and if it is not 0.0%, it will be zerged first.
Zero-setting

Measurement

Measure again after zero clearing.
Measurement

SWEVY Logo

Documents / Resources

PDF thumbnailDigital Refractometer
Installation Guide · r fractom tre, num rique, taux de sucre, brix, Digital Refractometer, Refractometer

References

Ask a Question

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

Ask a Question

Ask about setup, compatibility, troubleshooting, or anything missing from this manual. Name and email are optional.