StreamTeck-logo

StreamTeck T60 Wireless Physiological Transmitter and Central Monitoring Software

StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software-product

Product Information

Product Specifications

  • Material: ABS
  • Power: 5V/2A
  • Operating time: Continuous-time
  • RF range: Wireless sensor: 60GHz ~ 64GHz, Wireless transmission: 2.4/5GHz
  • Operating environment temperature: 15~40°C
  • Operating environment humidity: 30~80%, no condensation
  • Operation/storage/transportation pressure: 700~1060 hPa
  • Storage/transportation environment temperature: 0~50°C
  • Storage/transportation environment humidity: 30~80%, no condensation
  • Size: 190*190*60mm
  • Weight: 250g

StreamTeck Wireless Physiological

Transmitter and Central Monitoring Software

Caution! Please read the user manual and follow the instructions before use.
The information provided by the product is assisted by medical care personnel under professional judgment.
This product is not suitable for use as a basis for diagnosis.

Product Description and Usage

  • Product composition: This product includes the wireless physiological transmitter, central monitoring software and other compatible hardware like computer, keyboard, mouse and monitor.
  • Product features: StreamtTeck central monitoring software is connected to a single or multiple wireless physiological transmitters through the computer’s internet. The monitoring information is used to display, record, trace the physiological parameter. The software can alarm notification function, including alarm setting and centralized monitoring and management of a single or multiple beds, suitable for the nursing station of long-term care institutions or the general wards of medical institutions. The system provides continuous real-time display and storage of physiological parameters and vacancy bed detection and can trace back the stored data. The central monitoring software should be operated by professional medical and nursing staff. It can only be used as assistance for medical and nursing staff instead of clinical treatment and diagnosis.

Product applicable functions

  1. View the continuous monitoring physiological parameter, including heart rate, breathing rate, body temperature and bed occupation transmit by the wireless physiological transmitter.
  2. Warning to medical units using the product (system visual warning).
  3. Observe and review the patients’ physiological data in detail.

Detailed Product Specifications

Wireless Physiological Transmitter (T60) StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (1)
Material ABS
Power 5V/2A
Operating time Continuous-time
RF range Wireless sensor: 60GHz ~ 64GHz

Wireless transmission: 2.4/5GHz

Operating environment temperature 15~40℃
Operating environment humidity 30~80%, no condensation
Operation/storage/transportation pressure 700~1060 hPa
Storage/transportation environment

temperature

0~50℃
Storage/transportation environment humidity 30~80%, no condensation
Size 190*190*60mm
Weight 250g
Medical Power Supply StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (2)
Brand Adapter Tech. ATM012T-W050U
Output voltage 5V
Output current 2A
Input voltage range 100~240 VAC
Size 80.2*40.0*40.5±1mm (L*W*H)
Central monitoring

software

Central monitoring software with compatible hardware

including computer, keyboard, mouse and monitor

Install environment Minimum demand
Processor X86 3.0GHz four cores
Operating system/software Construct with LAMP (Linux, Apache, MySQL, PHP)
RAM More than 4G RAM
Hard drive space RAID1 or RAID5,no less than 1TB
Internet require A fixed IP and two-way 10MB bandwidth
Future Function
The number of supported

bed

Receive physiological parameters from 32 beds at most
Wireless communication Receive physiological parameter data from T60
Auto-connect and reconnect
 

 

 

 

 

 

 

 

System                              function management

Home page (login page)
Back: Back to the login page
Overview: List all the T60 devices in use
“Overview” page list all the information of the T60 device and patient data, including floor, room, bed number, patient name, age, gender, group, real-time heart rate, breathing rate, body temperature and vacancy bed detection.
Device setting: Add new T60 device to the device list, and setting corresponding location floor, room number and group, and can be removed from the device list.
Group setting: Adding group and group description, and can be deleted from the group list.
Account and user authority setting: Set account

authority to be user or administrator. The account and password can be changed.

 

 

Patient basic information management

Patient data setting: Set patient medical record number,

name, gender and birthday.

Threshold setting for alarm events: Set alarm trigger threshold of heart rate, breathing rate and body

temperature.

Physiological information tracking Physiological data record
Data search (base on year/ month/ date/ hour)
Display history physiological data
System self-protection If the system fails, restart itself and restore it to the

original setting.

 

 

Authority management

Not login: Can not use SmartCaring
Ward management: Execute and set all operation and

system parameter

Normal user: Set patient basic information and alarm

trigger threshold

Patient basic information

display

Floor, room, bed number, patient medical record

number, name, gender, age, group

 

 

Alert status display

Physiological parameters exceed the threshold.
T60 device disconnected hint.
Bed vacancy and timer display.
Patient motion display: T60 transmitter stop detecting

and transmit the last data before patient motion

 

Alarm color

Red: Alarm (immediately call a doctor)
Green: normal
Orange: bed vacancy
 

Default alarm threshold

Heart rate: >120bpm                     <50bpm
Breath rate: >25/per minute          <8/per minute
Body temperature: >37.5℃          <36℃
Alarm threshold setting Heart rate/ breathing rate/ body temperature
StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (3) IEC 60417-5032 : Alternating Current.
StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (4) IEC 60417-5031 : Direct Current
StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (5) ISO 7010-M002: Follow instructions for use

Usage Scenarios

This product is suitable for long-term care institutions and the nursing station in the general wards of medical institutions.

Contraindications

  1. Severe physical disability prevents safe and effective measurement.
  2. Without the patient’s consent.
  3. Do not use for the fetus.

Side Effect
According to Federal Communications Commission (FCC) requirements, the specific absorption rate (SAR) should be lower than 10w/m2. The passed measurement result of the product is 5.33w/m2 at a distance of 20cm. If the distance is less than 20cm, it may cause some electromagnetic wave damage to the human body, such as a mobile phone.

Caution

  1. The data gain by the wireless physiological transmitter and the central monitoring software can only be used as assistance instead of the clinical treatment and diagnosis.
  2. Avoid the wireless physiological transmitter, central monitoring software and the product accessories from sunlight and high temperature and water.
  3. Do not disassemble or repair the wireless physiological transmitter by yourself. It may affect the accuracy of the measurement result and the safety of the device.
  4. This device complies with Part 15 of the FCC Rules and ISED’s licence-exempt RSSs. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation
  5. This device has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures:
    • Reorient or relocate the receiving antenna.
    • Increase the separation between the equipment and receiver.
    • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
    • Consult the dealer or an experienced radio/TV technician for help.
  6. Any changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate this equipment.
  7. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
  8. This device complies with FCC and ISED radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator & your body.
  9. The wireless physiological transmitter complies with the electromagnetic compatibility (EMC) test standard but is still recommended to stay away from other microwave devices, RF devices, and radar devices, to avoid influencing the abovementioned devices.
  10. Data may lose if the wireless physiological transmitter is powered off or abnormally turned off. Do not arbitrarily turn off the device. If the device is accidentally turned off, immediately reconnect the device to the power source.
  11. The installation and the setting of the wireless physiological transmitter can only be carried out by manufacture trained and authorized personnel.
  12. If the server PC of the central monitoring software is powered off or abnormally turned off, data may lose. Do not arbitrarily turn off the computer. If the computer is accidentally turned off, immediately turn it on and start the software.
  13. For the installation and setting of the central monitoring software, please refer to the user manual, or operate by the staff of Streamteck Inc.
  14. Remove the monitoring device on the software will remove the setting of the device. Please make sure and check that the bed related to the device no longer needs to be monitored.
  15. It is important to choose a proper value when setting the threshold of the physiological parameters. The extreme or improper value may cause alarm invalidation.
  16. The computer operating the central monitoring software can not be moved or run another software.
  17. In case of serious incident that has occurred, please contact the manufacturer and local authorities immediately.
  18. WARNING – Do not modify this equipment without authorization of the manufacturer.
  19. Do not place the power cord where it is difficult to disconnect and may be stepped by other persons.
  20. Disconnect device: disconnect the power cord to fully power off the device
  21. CLASSIFICATION:
    • Supply Class II adapter
    • No applied part
    • Continuous Operation
    • Not AP or APG category
  22. Disconnect this equipment from AC outlet before cleaning. Do not use liquid or spray detergents for cleaning.
  23. CAUTION! This adapter Adaptor Tech/ ATM012T-W050U is a forming part of the device
  24. Only use the power cord with following specification: 18AWG min., type SJT, 125V/10A, UL listed, 3m max, hospital grade if for USA market
  25. Use a power cord that matches the voltage of the power outlet, which has been approved and complies with the safety standard of your particular country.
  26. Do not dispose this device in the bin, dispose and recycle it according to national regulation

Industry Canada statement
This device complies with ISED’s licence‐exempt RSSs. Operation is subject to the following two conditions:

  1. This device may not cause harmful interference, and
  2. This device must accept any interference received, including interference that may cause undesired operation.

Caution :

  1. The device for operation in the band 5150‐5250 MHz is only for indoor use to reduce the potential for harmful interference to co‐channel mobile satellite systems; FOR MOBILE DEVICE USAGE (>20cm/low power)

Radiation Exposure Statement
This  equipment  complies  with  ISED  radiation  exposure  limits  set forth for an  uncontrolled environment. This equipment should be installed and operated with greater than 20 cm between the radiator & your body.

Guidance and manufacturer’s declaration – electromagnetic emissions
The model T60 is intended for use in the electromagnetic environment specified below. The customer or the user of the model HA2402 SERIES should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment guidance
RF emissions CISPR 11 Group 1 The model T60 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The model T60 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/

flicker emissions IEC 61000-3-3

Not applicable
Recommended separation distances between

portable and mobile RF communications equipment and the model T60

The model T60 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the model T60 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the model T60 as recommended below, according to the maximum output power of the communications

equipment.

Rated maximum output power of transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

d = 1,2 StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (6)

80 MHz to 800 MHz

d = 1,2StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (6)

800 MHz to 2,5 GHz

d = 2,3 StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (6)

0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Guidance and manufacturer’s declaration – electromagnetic immunity
The model T60 is intended for use in the electromagnetic environment specified below. The customer or the user of the model T60 should assure that it is used in such an environment.
Immunity test IEC 60601test level Compliance level Electromagnetic environmentguidance
Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidityshould be at least 30 %.
Electrical fast transient/burst ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality should be that of a typical commercial or hospital environment.
IEC 61000-4-4 ±1 kV for input/output ±1 kV for input/output
lines lines
SurgeIEC 61000-4-5 ±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital environment.
interruptions and <5 % UT <5 % UT Mains power quality should be that of a typical commercial or hospital environment. If the user of the model HA2402 SERIES requires continued operation during powermains interruptions, it is recommended that the model HA2402 SERIES be powered from an uninterruptible power supply or a battery.
voltage variations (>95 % dip in UT) (>95 % dip in UT)
on power supply for 0,5 cycle for 0,5 cycle
input lines
40 % UT 40 % UT
IEC 61000-4-11 (60 % dip in UT) (60 % dip in UT)
for 5 cycles for 5 cycles
70 % UT 70 % UT
(30 % dip in UT) (30 % dip in UT)
for 25 cycles for 25 cycles
<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
for 5 sec for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The model HA2402 SERIES is intended for use in the electromagnetic environment specified below. The customer or the user of the model HA2402 SERIES should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment

guidance

Portable and mobile RF communications equipment should be used no closer to any part of the model T60, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1,2 StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (6)
d = 1,2 StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (6) 80 MHz to 800 MHz
 

Conducted RF IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz

3 Vrms d = 2,3 StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (6) 800 MHz to 2,5 GHz
 

Radiated RF IEC 61000-4-3

 

3 V/m

80 MHz to 2,5 GHz

 

 

3 V/m

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (7)
  • NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
  • NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
  • Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the model T60 is used exceeds the applicable RF compliance level above, the model T60 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model T60.
  • Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Detail Instructions

The wireless physiological transmitters are installed and set by the engineer of StreamTeck Inc. The wireless sensor detects physiological signals, including heart rate, breathing rate and body temperature, in a non-contact manner of the static, lying patient. The physiological parameters can be transmitted to the central monitoring software at a nursing station in long-term institutions or the general wards of medical institutions via a wireless network for integration.

The monitoring information is used to display, record, trace physiological parameters. The software is capable of alarm notification function, including alarm setting and centralized monitoring and management of a single or multiple beds, which is suitable for the nursing station of the long-term care institutions and the general wards of medical institutions. The system provides continuous real-time display and storage of physiological parameters and vacancy bed detection and can trace back the stored data. The central monitoring software should be operated by professional medical and nursing staff. It can only be used as assistance for medical and nursing staff instead of clinical treatment and diagnosis.

Wireless Physiological Transmitter

  1. Install the wireless physiological transmitter at a distance of 50 cm above the hospital bed (show as the picture). After the device is set up, connect the device to the power supply.StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (8)
  2. The device will detect the physiological signals via RF radio and infrared rays and transmit them to the receiver (server), the server computer at the nursing station of long-term care institution, nursing home or the general ward of a medical institution for monitoring.

Central Monitoring Software

  1. Open the browser, enter the central monitoring software URL and enter the homepage.StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (9)
  2. Enter account and password to log in. StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (10)
  3. Enter the main screen of the monitoring system. StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (11)
  4. Device setting
    Press the left side button “device setting.” StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (12) Press the “+create” button, jump to the next page, enter device number, MAC address and the location of the device, including the group, floor, room and bed number, then press “save,” go back to the previous page. StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (13)
  5. Patient data setting:
    Press the left side button “Patient List.”StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (14) Press the “+create” button, jump to the next page, enter patient No., name, birthday, physiological alarm threshold (heart rate, breathing rate and body temperature), and the related device, then press
    “SAVE.” StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (15)
  6. Group setting:
    Press the left side button “Group Setting.”StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (16) Press the “+create” button, jump to the next page, enter the group name and the group description, then press “SAVE.” StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (17) 7. History physiological data record and trace Press “Overview” and see the monitoring data of all patients. Click “>” unfold the patient’s personal information.StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (18) The left side shows the patient’s personal information, including name, age and gender. The right side shows every history data of each physiological parameter. StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (19) Choose “Start Time” and “End Time,” then press the button StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (20) to  adjust the observation time. Click the below button like heart rate, breathing rate and body temperature to choose the physiological data you want to see. Press the left side button “V” to fold the history data. StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (21)8. Alarm status: when detecting that the physiological is higher or lower the set threshold, the status bar will turn red, and the physiological parameter will flash a red warning.StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (22) The movement of the patient, like turning over, will affect physiological signal detection accuracy. Therefore, any motion of the patient will show on the central monitoring software, and the showing parameter number will be the last data before the patient moves. StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (23)

Troubleshooting

Wireless Physiological Transmitter (T60)
If there is an unexpected situation or other concerns about the wireless physiological signal transmitter, please remove the power supply of the device and reconnect it after five seconds. If there is still a problem after reconnecting the power supply to the device, please record the current situation and notify the agent or our company for assistance.

  1. Physiological signal measurement
    Abnormal state: heart rate, breathing rate, and body temperature of the central monitoring software show “Error” text.
    Handling: remove the power of the T60 transmitter and reconnect. Once the monitor still shows “Error” text, please notify the agent or our company for assistance.StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (24)

Central Monitoring Software
If there is an unexpected situation or other concerns about the central monitoring software, please close the software and re-start the software, or turn off and turn on the server computer and start the software. If there is still a problem, please record the current situation and notify the agent or our company for assistance.
When the software is abnormal, it may be a problem with the network or server of the computer using the software.

  1. Check if the computer is connected to the intranet, then open Google chrome browser to connect the software URL. If you cannot connect to the management console correctly, please check the server status.
  2. Check if the server is connected to the intranet, then reboot the server. If you cannot connect to the management console correctly, please notify the agent or our company for help.

StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (25)

Regular Maintenance
Wireless physiological transmitter:
The wireless physiological transmitter is a fixed type. In order to ensure the safety of use, please regularly check whether the fixing crews and brackets are rusted and loose or not. Hence wipe the device with a damp fabric to clean up the dust. After a disaster such as an earthquake, please check the device is not loose.

Central monitoring software:
Regularly back up the patient’s physiological signal history stored by the software to avoid data loss caused by the malfunction of the server computer.

Authorized maintenance service
This product does not provide users with self-maintenance. If you need any maintenance, please contact the agent or our company.

  • Distributor name: StreamTeck Scientific Inc.
  • Distributor address: No. 174, Huamei St., West Dist., Taichung City 403, Taiwan
  • Manufacturer name: WelFare Technology Inc.
  • Manufacturer address: No. 130-8, Waixizhou, Shuishang Township, Chiayi County 608006, Taiwan
  • Marketing time:

Operational Principle

  1. The principle and process of the wireless sensor detecting heart rate and breathing rate
    【Symbol Description】
    • 100:Wireless physiological signal transmitter
    • 110:Processor
    • 120:Storage media
    • 121:Communication module
    • 122:Signal processing module
    • 130:Transceiver
    • 31、32: Wireless signal
    • 41、42: Reflected signal
    • 51、52、53、54:Time period
      S201、S202、S203、S204、S205、S206、S207、S208、S209、S210、S211、 S212、S213、S214、S401、S402、S403、S404、S405、S406:Steps StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (26) StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (27) StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (28)
    • 【0001】 The monitoring device 100 (wireless physiological signal transmitter) in Fig. 1 is suitable for monitoring people’s status in a specific area. The monitoring device 100 may include a processor 110, a storage medium 120, and a transceiver 130.
    • 【0002】 The processor 110 such as a central processing unit (CPU), other programmable micro control units (MCUs) of the general-purpose or special-purpose, microprocessor, digital signal processor (DSP), programmable controller, application-specific integrated circuit (ASIC), graphics processing unit (GPU), image signal processor (ISP), an image processing unit (IPU), an arithmetic logic unit (ALU), a complex programmable logic device (CPLD), field-programmable gate array (FPGA) or other similar components or a combination of the above components. The processor 110 may be coupled to the storage media 120, the transceiver 130 and access and execute multiple modules and various application programs stored in the storage media 120.
    • 【0003】 The storage media 120 such as any fixed or removable random access memory (RAM), read-only memory (ROM), flash memory, hard disk drive (HDD), solid-state drive (SSD) or similar components or a combination of the above components, and used to store multiple modules or various applications that can be executed by the processor 110. In this example, the storage media 120 can store multiple modules, including a communication module 121 and a signal processing module 122, the functions described later.
    • 【0004】 In step S201, the communication module 121 can transmit multiple wireless signals to specific areas through the transceiver 130 at a plurality of different time points. The specific area is that the patient is stationary and lying on the bed. 【0005】 In step S202, the communication module 121 receives a mixed reflection signal corresponding to wireless signals’ plurality from a specific area through the transceiver 130. The mixed reflection signal includes corresponding to wireless signals’ plurality, respectively of multiple reflection signals. Fig. 3 illustrates a schematic diagram of multiple wireless signals and mixed reflection signals according to an instance of this invention. Suppose the plurality of wireless signal comprising a wireless signal 31 and a radio signal 32; the mixed reflection signal includes at least a reflection signal 41 corresponds to the wireless signal 31, and a reflection signal 42 corresponds to the wireless signal 32. In Fig. 3, the time period 51 represents the time it takes from transmitting the infinite signal 31 to receiving the reflected signal 41, the time period 52 represents the time it takes from transmitting the infinite signal 32 to receiving the reflected signal 42, the time period 53 represents the time difference between the transmission of the wireless signal 31 and the transmission of the wireless signal 32. Period 54 represents the time difference between receiving the reflected signal 41 and receiving the reflected signal 42.
    • 【0006】 In Fig. 2, in step S203, the signal processing module 122 performs a one-dimensional Fourier transform on the mixed reflection signal to generate a one-dimensional signal. The one-dimensional signal includes points corresponding to at least one object of the reflected wireless signal cloud data.
    • 【0007】 In step S204, the signal processing module 122 can determine whether the one-dimensional signal corresponds to a living creature. If the one-dimensional signal corresponds to a biological signal, step S205 is entered. If the one-dimensional signal is not a biological signal, go back to step S201. For example, the signal processing module 122 can determine whether the mixed reflection signal contains the person’s information to be monitored based on the value of each carrier in the one-dimensional signal. If multiple values of multiple carriers that can correspond to vital signs or posture changes in the one-dimensional signal are all zero, it means that the mixed reflection signal does not contain the information of the person to be monitored (or that multiple wireless signals have not scanned creatures in a specific area). Therefore, the signal processing module 122 can determine that the one-dimensional signal is not a signal corresponding to a creature. On the other hand, if at least one of the multiple values of multiple carriers that can correspond to the vital signs in the one-dimensional signal is not zero, it means that the mixed reflection signal contains information about the person to be monitored (or represents multiple wireless signals have scanned creatures in a specific area). Therefore, the signal processing module 122 can determine that the one-dimensional signal corresponds to a creature.
    • 【0008】 In step S205, the signal processing module 122 generates a filtered signal based on the one-dimensional signal. Specifically, the signal processing module 122 determines that the specific carrier corresponds to a non-living creature according to the value of the specific carrier of the one-dimensional signal, and in response to the specific carrier corresponding to the non-living creature, convert the specific carrier from the one-dimensional Signal filtering to generate filtered signals.
    • 【0009】 In step S206, the signal processing module 122 may perform band-pass filtering on the filtered signal to generate a first processed signal. The purpose of this step is to filter out signals that are not vital signs signals. Generally speaking, the human breathing signal frequency is approximately between 0.1 Hz and 0.6 Hz, and the frequency of the human heartbeat signal is approximately between 0.8 Hz and 4 Hz. Therefore, the signal processing module 122 can be configured to filter signals below 0.1 Hz and above 4 Hz, thereby generating the first processed signal.
      In step S207, the signal processing module 122 can determine the monitored person’s vital signs according to the first processed signal, where the vital signs include breathing rate and heart rate.

IR Sensor temprature algorithm

StreamTeck-T60-Wireless-Physiological-Transmitter-and-Central-Monitoring-Software- (29)

Professional installation instruction

  1. Installation personal
    This product is designed for specific application and needs to be installed by a qualified personal who has RF and related rule knowledge. The general user shall not attempt to install or change the setting.
  2. Installation location
    The product shall be installed at a location where the radiating antenna can be kept 20 cm from nearby person in normal operation condition to meet regulatory RF exposure requirement.
  3. External antenna
    Use only the antennas which have been approved by the applicant. The non-approved antenna(s) may produce unwanted spurious or excessive RF transmitting power which may lead to the violation of FCC/IC limit and is prohibited.
  4. Installation procedure
    Please refer to user’s manual for the detail.
  5. Warning
    Please carefully select the installation position and make sure that the final output power does not exceed the limit set force in relevant rules. The violation of the rule could lead to serious federal penalty.

FAQ

Q: Is the product suitable for diagnostic purposes?
A: No, this product is not suitable for use as a basis for diagnosis. It is meant to assist medical care personnel under professional judgment.

Documents / Resources

StreamTeck T60 Wireless Physiological Transmitter and Central Monitoring Software [pdf] User Manual
STKT60, 2AY6FSTKT60, T60 Wireless Physiological Transmitter and Central Monitoring Software, T60, Wireless Physiological Transmitter and Central Monitoring Software, Physiological Transmitter and Central Monitoring Software, Transmitter and Central Monitoring Software, Central Monitoring Software, Monitoring Software, Software

References

Leave a comment

Your email address will not be published. Required fields are marked *