STIL Ref. 4817 Digital Refractometer User Manual

STIL Ref. 4817 Digital Refractometer

STIL Ref. 4817 Digital Refractometer

Instrument structure

Instrument structure

Instrument measurement

Before the testing:
The instrument and liquid test sample should be placed in same environment temperature. (The temperature difference between the environment, the instrument, and the liquid sample must not exceed 2°C). Otherwise it will affect the accuracy of the testing.
Shake the liquid sample thoroughly before testing..
If the instrument has not been used for a long time, resetting the”Zero point”before using.
Check the battery level. Low battery power may affect measurement accuracy.

Power on :
Press and hold the”ON/OFF/START”for one second until the screen display.

Power off:
Press and hold the”ON/OFF/START”for at least one second until the screen display”Power off”.

Operating steps:
Open the shading plate counterclockwise.
Use pure water and a soft cloth to clean the detection area, then remove the residual water.
Add 3~5 drops of liquid test sample to the detection slot.
Close the shading plate clockwise.
Press the”ON/OFF/START”shortly, or touch the “touch instruction point”.
The screen shows the value of BRIX and Value of Refractive index.

Zero point calibration:
Generally the instrument can be used directly. Zero point calibration need to be performed when the instrument is long time unused or there’s a zero offset.

The method is as follows:
Use the pure water to clean the detection slot with soft cloth repeatedly.
Add 3~5 drops of pure water to the detection slot.
Press and hold the”ZERO”until the screen shows” zeroing”, release the button and start.
When the Screen shows”Zero Success”, it means the Zero point calibration is finished. When the screen shows”Zero Failure”, it means the Zero point calibration is failed.
Press the”ON/OFF/START”. If the displayed value is 0.0%, the zero-point calibration is successful.

Note
Make sure the clean pure water is used for Zero Point
Calibration.
Make sure to clean the detection slot after testing.
The instrument should be placed in the environment of dry and normal temperature. Avoid exposing the instrument to high temperatures or direct sunlight to protect measurement accuracy and the OLED display.
Don’t wash the instrument immersion into the water to avoid the leaking.
Digital Refractometer is a precision optical instrum-ent, please using and maintaining carefully and softly. Avoid the damage to the prism in detection slot from cut and bump.
Avoid sample evaporation, as it may affect the measurement results, please operate fast.
Low battery power may cause incorrect measurement data.

Technical parameters

Parameter Value
Measuring range 0.0–95 Brix (%)
Resolution 0.1 Brix (%) / 0.1°C
Accuracy ±0.3 Brix (%) / 1°C
Temperature compensation 10–80°C
Operating temperature 10–40°C
Sample volume ~0.5 mL (3–5 drops)
Measure time ~3s
Power supply 2 x AAA
Dimension 154 × 52 × 44 mm
Weight 140g

Common trouble shooting

  • Concentration is too high or too low :
    Situation Paraphrase Cause To do
    Concentration is too high or too low. Outside the measurement range The sugar content of the tested sample is higher than the detection range Dilute the testing sample.
    Influence of external light Close the shading plate properly
    The tested liquid contains particles Filter out the particles
    Situation Cause To do
    Concentration is too high or too low. Liquid temperature is lower or higher than the instrument range The temperature of the liquid should be the same as the temperature of the instrument and the environment
    Pas de mesure Missing sample solution Add the test solution to be tested.
  • Zero failed :
    Cause To do
    Detection slot is not clean Cleaning the detection slot.
    The calibration solution is not correct Change the calibration solution.
  • Test data is incorrect :
    Cause To do
    The surface of the prism is uncleaned Wipe or clean prism
    Improper operation Please follow the instructions
    Photometric effect Operation after covering the shading plate
    Operation at high temperature Operation at normal temperature
    Low battery power The power is displayed as empty, which will affect the test data.
  • Mauvaise interprétation de la qualité :
    Cause To do
    Misjudgment of product quality due to improper operation Please follow the instructions

Symbol

Logo

Documents / Resources

PDF thumbnailRef. 4817 Digital Refractometer
User Manual · Refractometer, 4817, Digital

References

Ask a Question

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

Ask a Question

Ask about setup, compatibility, troubleshooting, or anything missing from this manual. Name and email are optional.