STIL Ref. 4817 Digital Refractometer

Instrument structure

Instrument measurement
Before the testing:
The instrument and liquid test sample should be placed in same environment temperature. (The temperature difference between the environment, the instrument, and the liquid sample must not exceed 2°C). Otherwise it will affect the accuracy of the testing.
Shake the liquid sample thoroughly before testing..
If the instrument has not been used for a long time, resetting the”Zero point”before using.
Check the battery level. Low battery power may affect measurement accuracy.
Power on :
Press and hold the”ON/OFF/START”for one second until the screen display.
Power off:
Press and hold the”ON/OFF/START”for at least one second until the screen display”Power off”.
Operating steps:
Open the shading plate counterclockwise.
Use pure water and a soft cloth to clean the detection area, then remove the residual water.
Add 3~5 drops of liquid test sample to the detection slot.
Close the shading plate clockwise.
Press the”ON/OFF/START”shortly, or touch the “touch instruction point”.
The screen shows the value of BRIX and Value of Refractive index.
Zero point calibration:
Generally the instrument can be used directly. Zero point calibration need to be performed when the instrument is long time unused or there’s a zero offset.
The method is as follows:
Use the pure water to clean the detection slot with soft cloth repeatedly.
Add 3~5 drops of pure water to the detection slot.
Press and hold the”ZERO”until the screen shows” zeroing”, release the button and start.
When the Screen shows”Zero Success”, it means the Zero point calibration is finished. When the screen shows”Zero Failure”, it means the Zero point calibration is failed.
Press the”ON/OFF/START”. If the displayed value is 0.0%, the zero-point calibration is successful.
Note
Make sure the clean pure water is used for Zero Point
Calibration.
Make sure to clean the detection slot after testing.
The instrument should be placed in the environment of dry and normal temperature. Avoid exposing the instrument to high temperatures or direct sunlight to protect measurement accuracy and the OLED display.
Don’t wash the instrument immersion into the water to avoid the leaking.
Digital Refractometer is a precision optical instrum-ent, please using and maintaining carefully and softly. Avoid the damage to the prism in detection slot from cut and bump.
Avoid sample evaporation, as it may affect the measurement results, please operate fast.
Low battery power may cause incorrect measurement data.
Technical parameters
| Parameter | Value |
| Measuring range | 0.0–95 Brix (%) |
| Resolution | 0.1 Brix (%) / 0.1°C |
| Accuracy | ±0.3 Brix (%) / 1°C |
| Temperature compensation | 10–80°C |
| Operating temperature | 10–40°C |
| Sample volume | ~0.5 mL (3–5 drops) |
| Measure time | ~3s |
| Power supply | 2 x AAA |
| Dimension | 154 × 52 × 44 mm |
| Weight | 140g |
Common trouble shooting
- Concentration is too high or too low :
Situation Paraphrase Cause To do Concentration is too high or too low. Outside the measurement range The sugar content of the tested sample is higher than the detection range Dilute the testing sample. Influence of external light Close the shading plate properly The tested liquid contains particles Filter out the particles Situation Cause To do Concentration is too high or too low. Liquid temperature is lower or higher than the instrument range The temperature of the liquid should be the same as the temperature of the instrument and the environment Pas de mesure Missing sample solution Add the test solution to be tested. - Zero failed :
Cause To do Detection slot is not clean Cleaning the detection slot. The calibration solution is not correct Change the calibration solution. - Test data is incorrect :
Cause To do The surface of the prism is uncleaned Wipe or clean prism Improper operation Please follow the instructions Photometric effect Operation after covering the shading plate Operation at high temperature Operation at normal temperature Low battery power The power is displayed as empty, which will affect the test data. - Mauvaise interprétation de la qualité :
Cause To do Misjudgment of product quality due to improper operation Please follow the instructions