solventum logo3M™ Attest™ Auto-reader 490 System 
Quick Reference Guidesolventum 490 System Attest Auto Reader

VH2O2 BI PCD use

solventum 490 System Attest Auto Reader - VH2O2 BI PCD use
Remove
Remove 1295PCD from pouch.
Remove only the number needed.
Leave remaining in pouch.
Confirm and verify
Confirm process indicator is unprocessed. Verify media ampoule is intact.
Place
DO NOT open prior to sterilization. Place in sterilizer; top rack, towards front of chamber.
solventum 490 System Attest Auto Reader - VH2O2 BI PCD use 1
Don
After cycle completion, don
safety glasses and gloves.
Verify
Verify 1348 CI is ACCEPT.
Verify media ampoule is intact.
Remove, verify, identify
Remove 1295 BI from pack.
Verify process indicator is exposed. Identify 1295 BI by writing on label.

STEAM and VH2O2 BI activation and incubation

solventum 490 System Attest Auto Reader - activation
Crush
Squeeze BI in activator to close cap and crush media ampoule.
Flick
Remove BI and flick wrist to ensure media flows to growth chamber.
solventum 490 System Attest Auto Reader - Inspect
Inspect
Verify media has transferred to BI growth chamber at the vial bottom.
Incubate
Place activated BI in any well. Display shows remaining minutes of incubation or test results. DO NOT REMOVE BI until the (+) or (-) symbol indicates the test is complete.

solventum 490 System Attest Auto Reader - Positive control BINOTES:

  • Activate and incubate 1295 BI within 1 hour of completion of the VH2O2 sterilization cycle
  • Gloves are recommended when activating 1295 BI
  • Allow 1491 and 1492V BIs to cool 10 minutes prior to activation

Interpretation of Solventum sterilization assurance core products

STEAM Indicator Unexposed Acceptable Exposed
3M Attest BI 1491 and 1492V solventum 490 System Attest Auto Reader - Unexposed solventum 490 System Attest Auto Reader - Acceptable Exposed
After processing, indicator on cap turns light brown or darker.
3M Comply Indicator Tape 1322 solventum 490 System Attest Auto Reader - darker solventum 490 System Attest Auto Reader - darker 1
After processing, indicator stripes turn dark-brown/black.
3M Attest Indicator Tape 1355 solventum 490 System Attest Auto Reader - darker 2 solventum 490 System Attest Auto Reader - darker 3
After processing, indicator stripes turn dark-brown/black.
VH2O2 Indicator Unexposed Acceptable Exposed Color Range
3M Attest BI 1295 solventum 490 System Attest Auto Reader - Unexposed 1 solventum 490 System Attest Auto Reader - Unexposed 2
After processing, indicator stripes turn toward pink.
3M Comply Indicator Tape 1228 solventum 490 System Attest Auto Reader - Unexposed 3solventum 490 System Attest Auto Reader - pink
After processing, indicator stripes turn toward pink.
3M Attest Tri-Metric Chemical Indicator 1348/1348E solventum 490 System Attest Auto Reader - Chemical Indicator solventum 490 System Attest Auto Reader - stripes
After processing, a color change towards pink should have entered the ACCEPT window.
STEAM Indicator Unprocessed Acceptable Processed PASS Unacceptable Processed FAIL
3M Attest Steam Chemical Integrator 1243 solventum 490 System Attest Auto Reader - Unprocessed solventum 490 System Attest Auto Reader - Unprocessed 1 solventum 490 System Attest Auto Reader - Unprocessed 2 solventum 490 System Attest Auto Reader - Unacceptable solventum 490 System Attest Auto Reader - Unacceptable 1
After processing, indicator bar crosses into the ACCEPT region.   If the color bar is in the REJECT region or on the line, it is considered a fail.
3M Comply Bowie-Dick Plus Test Pack 00135LF 3M Comply Early Warning Test Sheet (above) 3M Comply Bowie-Dick Test Sheet (below) solventum 490 System Attest Auto Reader - Unprocessed 3

solventum 490 System Attest Auto Reader - Unprocessed 4

solventum 490 System Attest Auto Reader - FAIL solventum 490 System Attest Auto Reader - REJECT solventum 490 System Attest Auto Reader - REJECT 1
When indicator ink circle on Early Warning Test sheet is lighter than the color standard, this may indicate that sterilizer maintenance should be scheduled.
solventum 490 System Attest Auto Reader - Unprocessed 5 solventum 490 System Attest Auto Reader - PASS solventum 490 System Attest Auto Reader - PASS 1 solventum 490 System Attest Auto Reader - REJECT 2 solventum 490 System Attest Auto Reader - REJECT 3
After processing, a uniform color change to dark brown/black is a pass. Fail result is indicated by a non-uniform color development,with a lighter colored area, usually near the center.
STEAM Indicator Unprocessed Processed Processed PASS Processed FAIL
3M Attest eBowie-Dick Test System 3M Attest eBowie-Dick Test Card (left) 3M Attest eBowie-Dick Auto-reader (right)

solventum 490 System Attest Auto Reader - Unprocessed 8

solventum 490 System Attest Auto Reader - Processed

solventum 490 System Attest Auto Reader - Processed PASS solventum 490 System Attest Auto Reader - Processed FALL
After processing, indicator ink circle on test card darkens. After processing, a light-up green check mark () is a pass. After processing, a light-up red X is a fail.

© Solventum 2025. Solventum, the S logo are trademarks of Solventum or its affiliates. 3M, the 3M logo, Attest and Comply are trademarks of 3M. Other trademarks are the property of their respective owners. 70-2011-7096-9 Rev 5/25
*Refer to 3M™ Attest Auto-reader 490 Operator Manuals and 3M™ Attest Biological Indicator 1295, 1491 and 1492V Instructions for Use for complete instructions and other important information. Refer to the indications for use provided in the biological indicator Instructions for Use. The indications for use should align with the sterilization cycle to be monitored. This Quick Reference Guide applies to 3M™ Attest” Auto-reader 490 having software version 4.2.7 or greater. Refer to the Instructions for Use for complete instructions and other important information related to the use of all other products included in this Guide.

solventum logoHealth Care Help Line 1-800-228-3957
Learn more at go.Solventum.com/sterilization
Solventum Medical Surgical
3M Center, Building 275
2510 Conway Avenue East
St. Paul, MN 55144 USA