ROCKER VFC 47 Continuous Filter Holder and Filtration System
Important Notice
This instrument is designed for laboratory usage only. Please read this manual carefully before installing and operating. The instrument shall not be modified in any way. Any modification will void the warranty and may result in potential hazard. We are not responsible for any injury or damage caused by any non-intended purposes and modifying the instrument without authorization.
- Check the voltage specified on the nameplate and ensure it matches the line voltage in your location.
- Install the instrument in a clean, dust-less, and ventilated area under 40°C.
- Never use the pump with any corrosives, flammable, or toxic material.
- The pump is not designed to start against the applied vacuum. To prevent damage, verify that the inlet of the pump is at atmospheric pressure before each start. If necessary, partially unscrew the vacuum regulator knob to vent to the atmosphere.
- This pump has a thermal protection device that automatically shuts off when overheated.
- The temperature of the pump surface is very high after use or during work, please don’t touch it to avoid being burnt.
- The pump must not be used to create vacuum and pressure simultaneously.
- Do not use any lubricant, which may damage the pump.
- For any instrument issues, please contact the manufacturer or service agent for assistance. Do NOT disassemble it improperly.
- Please discard packing materials according to local regulations.
- Visit the official website and the latest product guide for detailed information.
- Operating condition
- Ambient temperature: 5 ~ 40°C
- Relative humidity: 80% RH Max.
- Power supply: 100-120V~, 50/60Hz or 200-240V~, 50/60Hz
- Altitude: up to 2000 m
- Pollution degree: II
- Indoor use
Caution: Hot surface
- Before the operation, check the compatibility of the pump (including pump head, valve plates, and diaphragm) and filtration apparatus materials with the used medium.
- Warning: The PTFE-coated diaphragm and pump head can offer good resistance to a wide range of chemicals except for strong acids and alkalis.
Unpacking
Please check if the package is complete without any damage before unpacking. When unpacking, please make sure you have all accessories indicated on the list. If there is any problem, please keep the serial number along with the packing case and contact your local distributor immediately for assistance.
VFC 47, Continuous Filter Holder
Standard Package Includes: |
Glass Cover |
Sintered Glass Support Base, 47 mm |
Silicone Stopper (No. 8) |
Aluminum Clamp, 47 mm |
Silicone Tube Ø 6 x 12 mm, 100 cm |
PTFE Sinker |
Instruction Manual |
Rocker 300C – VFC 47, Continuous Solvent Purification System
Standard Package Includes: |
Rocker 300C, PTFE Coated Chemical Resistant Vacuum Pump (with Hoop holder) |
VFC 47, Continuous Filter Holder |
GL45 Filtration Adaptor |
Laboratory Glass Bottle, 1000 mL |
Instruction Manual |
- The sinker and silicone tube are included in the VFC 47 Continuous Filter Holder.
- An optional 500 mL laboratory glass bottle (Cat. No. 197000-58) is available.
Main Part Diagram
- VFC 47, Continuous Filter Holder
Item | Designation | Item | Designation |
1a | Cover | 4a | Silicone Stopper (No. 8) |
2a | Aluminum Clamp | 5a | Silicone Tube Ø 6 x 12 mm |
3a | Sintered Glass Support Base | 6a | Sinker |
- Rocker 300C, PTFE Coated Chemical Resistant Vacuum Pump
Item | Designation | Item | Designation |
1b | Power Switch | 4b | Vacuum Gauge |
2b | Air Outlet | 5b | Moisture Trap |
3b | Air Inlet | 6b | Vacuum Regulator |
Parts Material List
- VFC 47, Continuous Filter Holder
Part | Material | Part | Material |
Filter Holder | Borosilicate glass | Tube | Silicone |
Inlet Hose Barb | PP, Silicone | (opt.) Bottle | Borosilicate Glass |
Sinker | PTFE |
- Rocker 300C, PTFE Coated Chemical Resistant Vacuum Pump
Part | Material | Part | Material |
Pump Head | PTFE coated | Diaphragm | PTFE coated |
Valve Plate(s) | FKM |
- If compatibility issues arise, replace the part with a suitable material, such as PTFE tubing.
- Listed materials exclude parts not in direct contact with the filtrate, refer to the official website for the latest information.
Installation and Operation
Installation
- Assemble the VFC 47 Continuous Filter Holder as shown below.
- The GL45 filtration adapter and bottle are not included with the purchase of a VFC 47.
Operation
- Turn on the pump, block the inlet (3b), and regulate the desired vacuum by vacuum regulator (6b).
- A trap bottle is recommended to connect before the pump when handling moisture-rich samples.
- Screw the GL45 filtration adaptor onto a laboratory glass bottle, and insert the filter holder on it using a silicone stopper (4a).
- Place a membrane on the support base (3a), assemble the cover (1a) onto the support base, and clamp them securely.
- Connect the GL45 filtration adaptor to the pump inlet (3b) using a silicone tube.
- If the silicone tube is incompatible with the medium, replace it with a suitable material, such as PTFE.
- Connect the filter holder’s inlet hose barb to the sinker using tubing, then place the sinker (6a) into the sample container for continuous filtration.
- Power on the pump to start filtration.
Maintenance
- The vacuum pump is not autoclavable. Please clean the surface with pure water or 75% ethanol.
- After finishing the experiment, please remove the liquid in the moisture trap and keep pumping the air for at least 2 minutes to withdraw the residual steam.
- Please remove the liquid in the moisture trap after every use. If the liquid in the moisture trap is more than 70%, please stop the experiment immediately to empty the liquid.
- Filtration apparatus and accessories are autoclavable (121°C, 1 bar, 20 min). Please make sure to rinse the apparatus with pure water and separate each part before autoclaving.
- Valve plate(s), O-ring(s), silicone tube(s), and diaphragm are consumables, it is recommended to replace them every year or as needed to ensure good operation.
How to clean the VFC 47 Continuous Filter Holder
- Disassemble the filter holder and vacuum bottle completely.
- Soak all parts in clean water. Avoid abrasive cleansers. If necessary, use a soft bristle brush or sponge to remove residues.
- Rinse all parts thoroughly. The final rinse should be done with distilled or deionized water.
- Store in a clean, dry place.
Troubleshooting
Problem | Reason and Solution |
Failed to start |
→ Loose plug à Reconnect plug to power supply.
→ Wrong voltage à Reconnect to the power specified on the thenameplatee. → Vacuum exists à Release vacuum and restart. → The switch is broken à Contact the distributor for assistance. |
Failed to adjust the vacuum |
→ Loose regulator à Tighten the regulator by yourself or contact the distributor for assistance.
→ Faulty vacuum regulator à Contact distributor for assistance. → Faulty vacuum gauge à Contact distributor for assistance. |
Poor vacuum |
→ Improperly vacuum setting à Adjust the vacuum regulator.
→ Air leakage à Replace or tighten the tubing and accessories that possible leaks. |
Low filtration speed |
→ Improperly vacuum setting à Adjust the vacuum regulator.
→ Air leakage à Replace or tighten the tubing and accessories that possible leaks. |
Others |
→ Contact the distributor for assistance. |
Ordering information
- 189303-11(22) Rocker 300C – VFC 47, Continuous Solvent Purification System, AC110V, 60Hz (AC220V, 50Hz)
- 167200-S4 VFC 47, Continuous Filter Holder (47 mm)
- 167210-36-05 PTFE sinker
- 197100-65 GL45 filtration adaptor
- 167210-01 Glass cover
- 167230-03 Sintered glass support base, 47 mm (including No.8 silicone stopper)
- 167230-07 Glass support base, 47 mm (to be used with SS support screen and PTFE gasket)
- 167230-31 SS support screen, 47 mm
- 167230-41 PTFE gasket
- 167240-01 Aluminum clamp, 47 mm
- 167100-16-1 Silicone stopper (No.8)
- 167200-38 Silicone tube, Ø 6 x 12 mm, 100 cm
- 197000-58 Laboratory glass bottle, 500 mL
- 197000-59 Laboratory glass bottle, 1000 mL
- 167100-52-1 Whatman ME 25 (MCE) membrane, 47 mm, 0.45 μm, 100/PK
- 167100-20 PALL A/E glass fiber membrane filter, 47 mm, 1 μm, 100/PK
Vacuum Pump
- 189300-11(22) Rocker 300C, PTFE Coated Chemical Resistant Vacuum Pump, AC110V,60Hz (AC 220V,50Hz)
- 189400-11(22) Rocker 400C, PTFE Coated Chemical Resistant Vacuum Pump, AC110V,60Hz (AC 220V,50Hz)
- 189410-11(22) Rocker 410C, PTFE Coated Chemical Resistant Vacuum Pump, AC110V,60Hz (AC 220V,50Hz)
- 167300-11(22) Rocker 300 Vacuum Pump, AC110V,60Hz (AC 220V,50Hz)
- 167400-11(22) Rocker 400 Vacuum Pump, AC110V,60Hz (AC 220V,50Hz)
- 167410-11(22) Rocker 410 Vacuum Pump, AC110V,60Hz (AC 220V,50Hz)
- 169300-11(22) Chemker 300 Vacuum pump, AC110V,60Hz (AC 220V,50Hz)
Rocker Scientific Co., Ltd.
- Tel: +886-2-26033311 Fax: +886-2-26036622
- E-mail: export@rocker.com.tw
- https://www.rocker.com.tw
Documents / Resources
![]() |
ROCKER VFC 47 Continuous Filter Holder and Filtration System [pdf] Instruction Manual VFC 47, VFC 47 Continuous Filter Holder and Filtration System, Continuous Filter Holder and Filtration System, Filter Holder and Filtration System, Holder and Filtration System, Filtration System, System |
![]() |
ROCKER VFC 47 Continuous Filter Holder and Filtration System [pdf] Instruction Manual VFC 47 Continuous Filter Holder and Filtration System, VFC 47, Continuous Filter Holder and Filtration System, Filter Holder and Filtration System, Filtration System, System |