LabCon Map PC Recording 4 Channel Mapping PC Software
This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net
OPERATOR PAGE PARAMETERS
Parameters | Settings (Default Value) |
Time Start Command >>
Time Abort Command >> |
Yes No
(Default : No) |
Time Interval (H:M) >> | 0.00 to 500.00 (HH:MM)
(Default : 0.10) |
Ctrl Set Value >> | Setpoint LO limit to Setpoint HI limit (Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple) (Default : 25.0) |
Ctrl Lo Deviation >> | For RTD & DC Linear : 0.2 to 99.9 For Thermocouple : 2 to 99
(Default : 2.0) |
Ctrl Hi Deviation >> | For RTD & DC Linear : 0.2 to 99.9 For Thermocouple : 2 to 99
(Default : 2.0) |
Change Password >> | 1 to 100
(Default : 0) |
SUPERVISORY
SUPERVISORY > SENSOR INPUT
Parameters | Settings (Default Value) |
Ctrl Zero Offset >> | -50 to 50
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple) (Default : 0.0) |
Map-1 Zero Offset >> Map-2 Zero Offset >> Map-3 Zero Offset >> Map-4 Zero Offset >> | -50 to 50
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple) (Default : 0.0) |
SUPERVISORY > RECORDING
Parameters | Settings (Default Value) |
Rec.
Interval (min) >> |
0 to 250 Min (Default : 5 Min) |
Alarm
Toggle Record >> |
Enable Disable
(Default : Enable) |
Del PC Records >> | Yes No
(Default : No) |
Del
Display Record >> |
Yes No
(Default : No) |
GSM Machine ID >>
(refer note) |
1 to 128
(Default : 1) |
RTC Time >> | 0 to 23 for Hour
0 to 59 for Min (Default : NA) |
RTC Date >> | 1 to 31 (Default : NA) |
RTC Month >> | 1 to 12 (Default : NA) |
RTC Year >> | 2000 to 2099
(Default : NA) |
Device Id >> | 1 to 127
(Default : 1) |
SUPERVISORY > CONTROL
Parameters | Settings (Default Value) |
Tune >> | Yes No
(Default : No) |
Setpoint LO Limit >> | Min Range for the Selected Input Type to Setpoint HI Limit (Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple) (Default : 0.0) |
Setpoint | Setpoint LO Limit to |
HI Limit >> | Max Range for the Selected
Input Type |
(Resolution 0.1°C for RTD / | |
DC Linear & 1°C for Thermocouple) | |
(Default : 600.0) | |
Compressor | 0 to 100 |
Setpoint | (Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple) |
(Default : 45.0) | |
Compressor Hyst >> | 0.1 to 99.9
(Default : 2.0) |
Heat Ctrl Action >> | ON-OFF PID
(Default : PID) |
Heat Hyst >> | 0.1 to 99.9
(Default : 0.2) |
Heat Only Control | Heat + Cool Control Zone : Single | Heat + Cool Control Zone : Dual |
Proportional Band >>
0.1 to 999.9 (Default : 50.0) |
Proportional Band >>
0.1 to 999.9 (Default : 50.0) |
Cz Prop Band >> Proportional Band for Cool Pre-dominant zone 0.1 to 999.9
(Default : 50.0) |
Integral Time >> 0 to 3600 sec (Default : 100 sec) | Integral Time >> 0 to 3600 sec (Default : 100 sec) | Cz Integral Time >>
Integral Time for Cool Pre-dominant zone 0 to 3600 sec (Default : 100 sec) |
Derivative Time >>
0 to 600 sec (Default : 16 sec) |
Derivative Time >>
0 to 600 sec (Default : 16 sec) |
Cz Derivative Time >> Derivative Time for Cool Pre-dominant zone
0 to 600 sec (Default : 16 sec) |
Cycle Time >> 0.5 to 100.0 sec (Default : 10.0 sec) |
Cycle Time >> 0.5 to 100.0 sec (Default : 10.0 sec) |
Hz Prop Band >> Proportional Band for Heat Pre-dominant zone 0.1 to 999.9
(Default : 50.0) |
Overshoot Inhibit >> Enable Disable
(Default : Disable) |
Overshoot Inhibit >> Enable Disable
(Default : Disable) |
Hz Integral Time >>
Integral Time for Heat Pre-dominant zone 0 to 3600 sec (Default : 100 sec) |
Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec) |
Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec) |
Hz Derivative Time >> Derivative Time for Heat Pre-dominant zone
0 to 600 sec (Default : 16 sec) |
Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec) |
||
Overshoot Inhibit >> Enable Disable
(Default : Disable) |
||
Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec) |
SUPERVISORY > PASSWORD
Parameters | Settings (Default Value) |
Change Password >> | 1000 to 1999
(Default : 123) |
SUPERVISORY > EXI
Parameters | Settings (Default Value) |
Exit Setup Mode >> | Yes No
(Default : No) |
FACTORY
FACTORY > CONTROL SENSOR INPUT
Parameters | Settings (Default Value) |
Input Type >> | Refer Table 1 (Default : RTD Pt100) |
Signal LO >> | |
Signal HI >> | |
Range LO >> | -199.9 to RANGE HI
(Default : 0.0) |
Range HI >> | RANGE LO to 999.9
(Default : 100.0) |
FACTORY > MAPPING PARAMETERS
Parameters | Settings (Default Value) | ||||||||||||||||||
Mapping Points >> | 0 to 4
(Default : 0) |
||||||||||||||||||
Input Type >> | Refer Table 1 (Default : RTD Pt100) | ||||||||||||||||||
Signal LO >> |
|
||||||||||||||||||
Signal HI >> |
|
||||||||||||||||||
Range LO >> | -199.9 to RANGE HI
(Default : 0.0) |
||||||||||||||||||
Range HI >> | RANGE LO to 999.9
(Default : 100.0) |
FACTORY > ALARM PARAMETERS
Parameters | Settings (Default Value) |
Hysteresis >> | 0.1 to 99.9
(Default : 0.2) |
Inhibit >> | Yes No
(Default : Yes) |
FACTORY > HEAT COOL SELECT
Parameters | Settings (Default Value) |
Control Strategy >> | Heat Only Cool Only Heat + Cool
(Default : Heat + Cool) |
Control Strategy : Cool Only | |
Time Delay (sec) >> | 0 to 1000 Sec (Default : 200 Sec) |
Control Strategy : Heat + Cool | |
Compressor Strategy >> | CONT. OFF CONT. ON SP BASED PV BASED
(Default : CONT. ON) |
CONT. ON | SP BASED | PV BASED | ||
Time
Delay (sec) >> 0 to 1000 Sec (Default : 200 Sec) |
Boundary Set Value >>
0 to 100 (Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
|
Time Delay (sec) >> 0 to 1000 Sec (Default : 200 Sec) |
FACTORY > SOAK TIMER PARAMETERS
Parameters | Settings (Default Value) |
Enable >> | No Yes
(Default : No) |
Start Band >> | 0 to 999.9
(Default : 0.5) |
Holdback Strategy >> | None Up Down Both
(Default : None) |
Hold Band >> | 0.1 to 999.9
(Default : 0.5) |
Heat Off >> | No Yes
(Default : No) |
Cool Off >> | No Yes
(Default : No) |
Power Recovery >> | Abort Restart Continuous
(Default : Restart) |
FACTORY > DOOR OPEN
Parameters | Settings (Default Value) |
Enable >> | No Yes
(Default : No) |
Start Band >> | 0 to 999.9
(Default : 0.5) |
Holdback Strategy >> | None Up Down Both
(Default : None) |
Hold Band >> | 0.1 to 999.9
(Default : 0.5) |
Heat Off >> | No Yes
(Default : No) |
Cool Off >> | No Yes
(Default : No) |
Power Recovery >> | Abort Restart Continuous
(Default : Restart) |
FACTORY > MAINS FAILURE
Parameters | Settings (Default Value) |
Enable >> | No Yes
(Default : No) |
Start Band >> | 0 to 999.9
(Default : 0.5) |
Holdback Strategy >> | None Up Down Both
(Default : None) |
Hold Band >> | 0.1 to 999.9
(Default : 0.5) |
Heat Off >> | No Yes
(Default : No) |
Cool Off >> | No Yes
(Default : No) |
Power Recovery >> | Abort Restart Continuous
(Default : Restart) |
FACTORY > SMS ALERT
Parameters | Settings (Default Value) |
Enable >> | No Yes
(Default : No) |
Start Band >> | 0 to 999.9
(Default : 0.5) |
Holdback Strategy >> | None Up Down Both
(Default : None) |
Hold Band >> | 0.1 to 999.9
(Default : 0.5) |
Heat Off >> | No Yes
(Default : No) |
Cool Off >> | No Yes
(Default : No) |
Power Recovery >> | Abort Restart Continuous
(Default : Restart) |
FACTORY > FACTORY DEFAULT
Parameters | Settings (Default Value) |
Set To Default >> | Yes No
(Default : No) |
FACTORY > PASSWORD
Parameters | Settings (Default Value) |
Set To Default >> | Yes No
(Default : No) |
FACTORY > EXIT
Parameters | Settings (Default Value) |
Exit Setup Mode >> | Yes No
(Default : No) |
TABLE 1
TABLE 1 | ||
What it means | Range (Min. to Max.) | Resolution |
Type J Thermocouple | 0 to +960°C |
Fixed 1°C |
Type K Thermocouple | -200 to +1376°C | |
Type T Thermocouple | -200 to +385°C | |
Type R Thermocouple | 0 to +1770°C | |
Type S Thermocouple | 0 to +1765°C | |
Type B Thermocouple | 0 to +1825°C | |
Type N Thermocouple | 0 to +1300°C | |
Reserve |
Reserved for customer specific Thermocouple type not listed above. The type shall be specified in accordance with the ordered (optional on request) Thermocouple type. | |
3-wire, RTD Pt100 | -199.9 to 600.0°C | Fixed 0.1°C |
0 to 20mA DC current |
-199.9 to 999.9 units |
Fixed 0.1 unit |
4 to 20mA DC current | ||
0 to 5.0V DC voltage | ||
0 to 10.0V DC voltage | ||
1 to 5.0V DC voltage |
FRONT PANEL KEYS
Symbol | Key | Function |
![]() |
Scroll |
Press to scroll through various Process Information Screens in Normal Operation Mode. |
![]() |
Alarm Acknowledge |
Press to acknowledge and mute (if active) alarm output. |
![]() |
DOWN |
Press to decrease the parameter value. Pressing once decreases the value by one count; keeping pressed speeds up the change. |
![]() |
UP |
Press to increase the parameter value. Pressing once increases the value by one count; keeping pressed speeds up the change. |
![]() |
SET-UP |
Press to enter or exit set-up mode. |
![]() |
ENTER |
Press to store the set parameter value and to scroll to the next parameter. |
PV ERROR INDICATIONS
Message | Error Type | Cause |
OPEN | Sensor Open | Sensor (RTD Pt100) Broken / Open |
OVR | Over-range | Temperature above Max. Specified Range |
UNR | Under-range | Temperature below Min. Specified Range |
ELECTRICAL CONNECTIONS
4 Mapping Channels