LS171 Provide Color Comparison
“
Specifications
- Illumination geometry: D/8, specular component
include (SCI) - Illumination light source: Full spectrum LED
light source - Measuring aperture: 8mm
- Measurement conditions: About 1s
- Color space: CIE Lab, Luv, LCh, Yxy, CMYK,
RGB, Hex - Color difference formula: E*ab, E*uv, E*94,
E*cmc(2:1), E*cmc(1:1), E*cmc(1.4:1), E*00 - Repeatability: Standard deviation E*ab is
within 0.03 (after calibration, the average value of 30
measurements on the whiteboard at an interval of 3s) - Dimension: Weight: About 126g
- Battery power: Rechargeable lithium battery
3.7V@600mAh, can be used continuously for 10,000 times under full
charge - Display: OLED screen
- Charging port: USB (Type-C)
- Data transmission: Bluetooth
- Operating temperature range: 0~45°C, 0~85%RH
(no condensation) - Storage temperature range: -25~55°C, 0~85%RH
(no condensation) - Supply Voltage: DC5V
- Operating Current: 10mA
- Operating Power Consumption: 50mW
Product Usage Instructions
I. Mobile App Installation
To install the colorimeter APP, follow these steps:
- Scan the QR code on the cover with a mobile browser or WeChat
to download and install the LScolor application. - Alternatively, search for LScolor on the iOS App Store or
Google Play App market and follow the prompts to download and
install the application. - During installation, ensure to set permissions to “Allow all”
to enable device search and app usage.
II. Product Introduction
The colorimeter complies with various standards and provides
accurate color measurement using different color spaces and
formulas.
III. Product Features
The product features include an OLED screen for standalone use
or pairing with a mobile app, multiple color charts for accurate
color matching, QC testing function with customizable color
difference thresholds, various color spaces for diverse needs,
color data import and sharing capabilities, stable values with
built-in compensation optical path, and a rechargeable lithium
battery for long-term use.
Frequently Asked Questions (FAQ)
Q: What are the supported operating systems for the colorimeter
APP?
A: The colorimeter APP supports Android operating system 7.0 and
above or iOS 7.0 and above.
Q: How long can the colorimeter’s rechargeable battery last
under full charge?
A: The rechargeable lithium battery can be used continuously for
up to 10,000 times under full charge.
“`
Shenzhen Linshang Technology Co., Ltd.
Colorimeter
Model: LS171 User Manual V2.15
Please read this manual carefully before using and reserve it for reference.
2024-08-06 Page 1 of 13
I Mobile App Installation
Shenzhen Linshang Technology Co., Ltd.
The colorimeter APP supports Android operating system 7.0 and above or iOS 7.0 and above. Choose one of the following methods to install:
(1) Scan the QR code on the cover with a mobile browser or WeChat, follow the prompts to download and install LScolor application.
(2) Search for “LScolor” from the iOS App Store or Google Play App market and follow the prompts to download and install LScolor application.
APP icon Precautions: Open the APP during the installation process or after the first installation, the mobile phone will prompt for permission settings. The customer needs to set “Allow all”, otherwise it will appear that the APP can’t search for the device and the APP can’t be used.
II Product Introduction
LS171 is a portable App smart colorimeter. It greatly improve and expand the function of the colorimeter using the mobile phone App’s large memory and convenient operation. The mobile APP comes with multiple electronic color charts to quickly match the measured color to the closest color card number and support measured color sharing through WeChat, WhatsApp, Skype, email, etc. The colorimeter also has a color comparison function, the color difference threshold can be set and a variety of color difference formulas can be selected, which can realize fast QC testing. It is equipped with an OLED display, which can be used separately from the mobile phone APP.
Standards for the product JJG 595-2002 Colorimeters and Color Difference Meters GB/T 3978-2008 Standard illuminants and geometric conditions GBT 7921-2008 Uniform color space and color difference formula GB/T 11186.1-1989 Methods for measuring the colour of paint films–Part 1: Principles GB/T 11186.2-1989 Methods for measuring the colour of paint films–Part 2: Colour measurement GB/T 11186.3-1989 Methods for measuring the colour of paint films–Part 3: Calculation of colour differences
2024-08-06 Page 2 of 13
III Parameters
Shenzhen Linshang Technology Co., Ltd.
Illumination geometry
D/8, specular component include (SCI)
Illumination light source
Full spectrum LED light source
Measuring aperture
8mm
Measurement conditions
Light source D65, field of view 10°
Measuring time
About 1s
Color space
CIE Lab, Luv, LCh, Yxy, CMYK, RGB, Hex
Color difference formula
E*ab, E*uv, E*94, E*cmc(2:1), E*cmc(1:1), E*cmc(1.4:1), E*00
Repeatability
Standard deviation E*ab is within 0.03 (Measurement conditions: after calibration, the average value of 30 measurements on the whiteboard at an interval of 3s)
Dimension
57mm×57mm×89.4mm
Weight
About 126g
Battery power
Rechargeable lithium battery 3.7V@600mAh, can be used continuously for 10,000 times under full charge
Display
128×64 dot matrix OLED
Charging port
USB (Type-C)
Data transmission
Bluetooth
Operating temperature range 0~45, 0~85%RH (no condensation)
Storage temperature range
-25~55, 0~85%RH (no condensation)
Supply Voltage
DC5V
Operating Current
10mA
Operating Power Consumption 50mW
IV Features
1. With OLED screen, it can be used alone, or used with mobile phone App.
2024-08-06 Page 3 of 13
Shenzhen Linshang Technology Co., Ltd. 2. Built in multiple electronic color charts, quickly match the closest color code, the color code
matching accuracy rate is greater than 90%. 3. Equipped with QC testing function, color difference threshold (tolerance value) can be set and a
variety of color difference formulas can be selected. 4. A variety of color spaces are available to facilitate color measurement for various needs. 5. Support color data import and sharing to achieve multi-device data sharing.Built-in compensation
optical path makes the values stable, allowing long-period calibration. 6. Built-in rechargeable lithium battery, can be used continuously for 10,000 times under full charge.
V Operation
1. Power on/off
Power on: Short press the power button to turn on the colorimeter, after that, the indicator light will flash. Shut down: Long press the power button to turn off the colorimeter; when the mobile phone is not connected to Bluetooth, it will automatically turn off without any operation in 3 minutes; when the mobile phone is connected to Bluetooth and it will automatically turn off without any operation in 30 minutes.
2. Bluetooth connection
Turn on the colorimeter, open the App and first display the LOGO. After the LOGO is displayed for 3s, enter the “Connect Device” interface and automatically start searching for Bluetooth devices. The searched devices will be displayed at the bottom of the interface. Click the device that matches the SN of the colorimeter and wait for the device to connect. After the connection is successful, it will jump to “Calibration interface”.
Precautions: If there is a Bluetooth connection error or connection failure, you can restart the App and the
colorimeter, or turn off/start Bluetooth in the setting interface and then reconnect. Click the “Browse” button, you can enter the App to view historical data without connecting to the
colorimeter.
3. Calibration
After the App is connected to the colorimeter, enter the “equipment calibration” interface. The user can perform the calibration operation or skip the calibration. Equipment calibration After closing the bottom cap of the colorimeter, click the “Calibration” button. If the calibration is successful, the colorimeter will directly enter the measurement interface; if the calibration is unsuccessful, it will prompt “Calibration failed”.
2024-08-06 Page 4 of 13
Shenzhen Linshang Technology Co., Ltd. If it prompts that the calibration failed, the possible reasons are as follows: (1) The bottom cap is not closed properly; (2) The standard tile on the bottom cap is too dirty and needs to be cleaned; (3) The attenuation of the light source leads to abnormal use, the colorimeter needs to be returned to
the factory for inspection and repair. Skip calibration The colorimeter skips the calibration operation and enters the measurement interface. When the colorimeter has not been used for a long time, it is recommended to calibrate.
4. Measurement
After the App is successfully connected to the colorimeter, enter the main measurement interface. There are two measurement modes: “Color Scan” and “Color Compare”. After the App is opened, the default mode is “Color Scan”, there are two measurement switch buttons at the bottom of the main interface. The font of the button becomes blue when selected. The current measurement mode, battery power, historical records, model and connection status are displayed on the top information bar. Battery life
Display the current power of the colorimeter. When the battery is low, charge it in time. History records
Click the history icon, if it is the color measurement interface, enter the “My Colors” interface; if it is the color compare interface and enter the “Compare History” interface. Connection status Display the connection status of the colorimeter. If the App is connected to the colorimeter, the corresponding model will be displayed; if not, it will prompt that the device is disconnected and the model display will be blank.
4.1. Color measurement
After the colorimeter is connected, it will enter the color measurement interface. Click the “Scan” button on the App or short press the button and the colorimeter will show the data to the App after measuring. The display interface after color measurement including the following parameters: Color of the test sample: After the measurement, the measurement button area will display the color value of the test sample and the value type can be set by the user. Matching colors: According to the color chart selected by the user, the closest three color codes are matched and displayed.
You can continue other operations as follows:
Enter name
2024-08-06 Page 5 of 13
Shenzhen Linshang Technology Co., Ltd. After measuring the color, you can enter a name for saving or sharing. Save The user can enter name and click the “Save” icon to save the measured color. The color is saved in “My Colors”. Share The user can enter name and click the “Share” icon to share the measurement results in image format or data format (the sharing methods are: WeChat, WhatsApp, Skype, email, etc.) with others. The shared data can be used for App data import. Select color chart Choose a set of color chart or multiple sets of color charts for color matching in color measurement. After the color chart is selected, only the colors in the color chart are calculated and matched. The three color codes that are closest to the measured color are displayed.
4.2. Color compare
Click the “Color Compare” button at the bottom to switch to the color compare interface Standard color measurement
Click the “Scan Standard” button and the current measured value will be used as the standard color. Sample color measurement
Click the “Scan Sample” button and the current measured value is used as the sample color. The display interface after color difference comparison is as follows:
When the standard color is measured or added and the sample color is measured or added, then a
2024-08-06 Page 6 of 13
Shenzhen Linshang Technology Co., Ltd. comparison measurement is completed (a comparison record is automatically stored). The color difference calculation is performed on the two colors. The color difference comparison results generally have the following three judgments:
LabCh value
The LabCh value is obtained by the sample color data minus the standard color data and the color cast direction between the two is judged by L, a and b. The value of E The value of E is calculated according to the selected color difference formula. QC testing
QC testing is suitable for color difference detection in the production process. Users can set whether this function is needed according to actual needs. If the QC testing function is set (QC testing is ticked) and the E threshold is set, if the calculated E is greater than the set threshold, “NG” is displayed, and the judgment area is displayed in red; if the calculated E is less than or equal to the set threshold, it will display “PASS” and the judgment area will be green. E setting Set the formula of E calculation and threshold of QC testing. Save standard
Users can enter the name of the color and save the standard color in “My Colors”. Input standard
Users can input Lab or LCh values as the standard color. Add standard color
Standard color addition is to add the measured color in “My Colors” or the color in the standard color charts as the standard color. After clicking the “Add Standard” button, jump to the “add color” interface, open the folder, click the color you want to add and then use the selected color as the standard color and return to the measurement interface. Add sample color
Sample color addition is to add the measured color in the history or the color in the standard color chart as the sample color. The operation of adding sample color is the same as adding standard color above.
5. Menu bar
In the measurement interface, click the menu bar ” ” in the upper left corner, the menu bar as shown in the figure below appears, corresponding to different parameter settings.
5.1. Parameters settings
2024-08-06 Page 7 of 13
Shenzhen Linshang Technology Co., Ltd. “Parameter setting” can set the color format, E formula, QC mode and matching color chart. Color format
The color format corresponding to the color space includes: Lab, Luv, LCh, Yxy, CMYK, RGB and Hex. Users can select different format to tick according to their needs. Up to 3 items can be selected. Lab is ticked by default. Options cannot be changed, so you can choose up to 2 items.
After this option is changed, the color format in the “Color Scan” interface will change accordingly. The color parameters in “My Color” will also change. E formula The E formula corresponding to the type of color difference formula includes E*ab, E*uv, E*94, E*cmc(2:1), E*cmc(1:1), E*00, a total of 6 color difference formulas. Users can choose one of them as the color difference calculation standard according to their needs.
After this option is changed, the E calculation of the “Color Compare” interface will be changed accordingly. The results displayed in the “Compare History” should be recalculated and determined according to the selected formula (if QC testing is selected, QC judgment must be recalculated based on the calculated E). QC mode
The QC mode is the parameter setting for QC testing. The QC testing function can be selected according to the needs. The threshold can be set in the range of 0.1-50. After the QC testing function is turned on, the QC judgment result will appear in the “Color Compare” interface after each color comparison. If the input threshold exceeds the range, it will automatically return to the default threshold. Select color chart
The purpose of selecting color chart is to select a set of color chart or multiple sets of color charts for color matching in color measurement. After the color chart is selected, three closest color codes in the color chart are matched and displayed. Reset to defaults
The parameter settings are restored to the factory options.
5.2. My Colors
“My Colors” is used to save the color measured by the user. The user can add up to 5 folders to save the measured color. Folders can be created, deleted and modified. The measured color can also be deleted, modified name and shared.
Create folder
In the “My Colors” interface, click “Create Folder”, enter the name and click “Confirm”.
Delete folder
To delete a single folder, click the box on the left side of the folder to select the folder (the box is ticked)
2024-08-06 Page 8 of 13
Shenzhen Linshang Technology Co., Ltd. and click the “Delete” icon in the upper right corner to delete. If you need to delete all, first click the “select all” icon to select all the folders (the boxes in front of all folders are ticked) and then click the “delete” icon to delete all folders.
Modify the folder name
After opening the folder, click on the name and the folder naming box will pop up, you can re-enter the name and click “Confirm”.
View color data
The user can click on any folder to view the saved color and view the name of the saved color, measurement time and measurement data. Single deletion or all deletion can be performed (the deletion operation is the same as the folder deletion operation). Click on a single color to view the measurement data and matching color codes and you can share and modify the name.
Color data sharing
In any folder, click the “Share” icon in the upper right corner and all the measured colors under the folder will be exported in data format.
5.3. Compare History
“Comparison History” saves the result of color difference comparison. Display the measurement time, E value and QC testing result of each comparison result. Single item deletion or all deletion can be performed and the comparison result can also be shared in data format. Delete To delete a single record, click the box on the left of the single record to select it (the box is ticked) and click the “delete” icon in the upper right corner to delete it. If you need to delete all, first click the “select all” icon to select all folders (the boxes in front of all records are ticked) and then click the “delete” icon to delete all folders. Share Click the “Share” icon in the upper right corner to export all the color difference comparison records.
5.4. Standard Color Chart
“Standard Colors” is the recorded standard color chart. Users can view the color information of the standard color chart, use the fuzzy search function to quickly search the color code or save the color code in “My Colors”. Fuzzy search After entering the characters, click the “Search” button to match all the colors containing the entered characters, which is convenient for users to find standard colors.
2024-08-06 Page 9 of 13
Save color
Shenzhen Linshang Technology Co., Ltd.
After opening any color card, click the “Save” button to save the standard color code to “My Colors”.
5.5. Device Information
“Device Info” displays the basic information of the connected colorimeter, including: device type, serial number, production date, calibration date, battery life, device status, firmware version, App version, manufacturer and service call.
5.6. Calibration
The colorimeter enters the calibration interface and calibrate it according to the interface prompts.
5.7. Disconnect Bluetooth
When the colorimeter is connected and enters the App, the item displays “Disconnect BT”. After clicking “Disconnect BT”, the colorimeter is disconnected and jumps to the “Connect Device” interface to search for the colorimeter again.
When you choose to browse and enter the App, the item displays “Connect”. After clicking “Connect”, it will jump to the “Connect Device” interface for colorimeter search.
6. Color import
Users can import the colors shared by others into their own App, but the colors must be tested by the same model of colorimeter and the file shared by selecting “Export Data” can be imported. The specific method of importing colors is as follows:
Take WeChat sharing and import as an example, find the file to be imported on WeChat, the file name is “import_mycolor” and click Open.
Figure 1
Figure 2
2024-08-06 Page 10 of 13
Shenzhen Linshang Technology Co., Ltd. (1) If the Figure 1 interface appears, click “Open with other app” and select the App “Colorimeter”, after
confirmation, jump to the import interface and select the file to complete the import. (2) If the Figure 2 interface appears, click “…” in the upper right corner, select “Open with other app”, then
select the App “Colorimeter”, after confirmation, jump to the import interface, select the folder to complete the import.
(3) If the import is successful, it will prompt “Import successful”. After the import is successful, the last record in the folder is the imported data.
(4) If the model does not match, it will prompt “Unmatched device model”. You need to connect to the corresponding model of the instrument before importing.
VI Precaution
1. When the colorimeter has not been used for a long time, it is recommended to perform calibration before using it.
2. Please ensure that the sample is uniform and the surface is flat and clean, otherwise it will affect the measurement accuracy.
3. After the base is removed from the host, it should be stored in a clean place to avoid contamination of the standard tile.
4. Do not insert any objects into the colorimeter to clean it, as this will damage it and affect measurement accuracy and operational safety.
5. The indicator light of the colorimeter is flashing red and it needs to be charged in time. During the charging process, the red light will be on. After charging, the light will turn to blue from red.
6. If not use the colorimeter for more than half a year, charge it regularly to prevent the battery from over-discharge and damage.
2024-08-06 Page 11 of 13
Shenzhen Linshang Technology Co., Ltd. 7. The recommended calibration cycle is one year, and the company provides calibration services. 8. Due to the different mobile phone displays, there may be a slight color difference between the actual
color and the color displayed on the mobile phone App.
VII Locating Plate
In order to meet the requirements for precise alignment of some measurement areas, the colorimeter is equipped with a locating plate.
Locating plate
Measurement area
VIII Packing List
No.
Description
1
Colorimeter
2
USB data cable
3
Cleaning cloth
4
Locating plate
5
User manual
6
Calibration Report
Place the meter on the positioning hole
Quantity
Unit
1
Set
1
pcs
1
pcs
1
pcs
1
pcs
1
pcs
IX Service
1. The colorimeter has one-year warranty. If the meter works abnormally, please send the whole colorimeter to our company for maintenance.
2. Provide users with spare parts and lifelong maintenance services. 3. Provide the users with the gauge calibration service. 4. Free technical support for long term.
2024-08-06 Page 12 of 13
Manufacturer: Shenzhen Linshang Technology Co., Ltd. Website: www.linshangtech.com Service hotline: 086-755-86263411 Email: sales21@linshangtech.com
Shenzhen Linshang Technology Co., Ltd.
2024-08-06 Page 13 of 13
Documents / Resources
![]() |
linshang LS171 Provide Color Comparison [pdf] User Manual LS171 Provide Color Comparison, LS171, Provide Color Comparison, Color Comparison, Comparison |