Command format: VSET<X>:<NR2>

  1. VSET: Command header
  2. X: output channel
  3. separator
  4. NR2: parameter

Command Details:

  1. ISET<X>:<NR2>
    Description: Sets the output current.
    Example: ISET1:2.225
    Sets the CH1 output current to 2.225A
  2. ISET<X>?
    Description: Returns the output current setting.
    Example: ISET1?
    Returns the CH1 output current setting
  3. VSET<X>:<NR2>
    Description: Sets the output voltage.
    Example: VSET1:20.50
    Sets the CH1 voltage to 20.50V
  4. VSET<X>?
    Description: Returns the output voltage setting.
    Example: VSET1?
    Returns the CH1 voltage setting.
  5. I0UT<X>?
    Description: Returns the actual output current.
    Example: IOUT1?
    Returns the CH1 output current
  6. VOUT<X>?
    Description: Returns the actual output voltage.
    Example: VOUT1?
    Returns the CH1 output voltage
  7. BEEP<Boolean>
    Description: Turns on or off the beep. Boolean: boolean logic.
    Example: BEEP1 Turns on the beep.
  8. OUT<Boolean>
    Description: Turns on off the output.
    Boolean: 0 OFF,1 ON
    Example: OUT1 Turns on the output
  9. STATUS?
    Description: reading the status of the power supply returning
    Contents: 8 bits follow the below formats
    Korad KA Series Laboratory Power Supply and Programmer - bits follow the below formats
  10. *IDN?
    Description: Returns the KA3005P identification.
    Example *IDN?
    Contents KORAD KA3005P V4.0 (Manufacturer, model name,).
  11. RCL<NR1>
    Description: Recalls a panel setting.
    NR1 1-5: Memory number 1 to 5
    Example RCL1 Recalls the panel setting stored in memory number1
  12. SAV<NR1>
    Description: Stores the panel setting.
    NR1 1-5: Memory number 1to 5
    Example SAV1 Stores the panel setting stored in memory number1
  13. OCP<Boolean>
    Description: Stores the panel setting.
    Boolean: 0 OFF,1 ON
    Example: OCP1 Turns on the OCP
  14. OVP<Boolean>
    Description: Turns on the OVP.
    Boolean: 0 OFF, 1 ON
    Example: OVP1 Turns on the OVP