KIMO CLASS 50 LUX Meter
Specifications
- Model: Class 50 Portables
- Parameters Measured: Temperature, Relative Humidity, CO, Pressure, Air Velocity, Light
- Brands: Kimo, a Sauermann brand
- Website: sauermanngroup.com
Quok Start Guide
Class 50 portables
These portable instruments are used to measure the following parameters: temperature, relative humidity, CO, pressure, air velocity and light.
Operating temperature, protection of the instrument, and information about storage
- Conditions of use (OC/%RH/m): from 0 to +50 oc. In non-condensing conditions. From 0 to 2000 m.
- Protection: IP54
- Storage temperature: HD 50, LX 50: from -20 to +70 oc. CT 50, CO 50, HM 50, LV 50, MP 50, MP 51, MP 55, VT 50: from -20 to +80 oc. If the instrument is stored outside the operating temperature (for example in a van, a warehouse…), please wait for 1 0 minutes in its operating temperature before start using it.
Keys
- Select
- On/Off/Hold
Take a measurement
WARNING: VT 50 Before using the instrument with the hotwire probe, please lower the protection tube (1) of the sensing element.
WARNING: VT 50: Always use the hotwire probe with the red point in front of the flow.
Switch on the instrument by pressing briefly on “On/Off/Hold”.
The instrument displays its name, then the measurement.
- CO 50: the instrument displays its name then a 5 second countdown. This count corresponds to the preheating time of the sensor. At the end of the countdown, the measured value of CO is displayed.
- MP 50, MP 51: the instrument displays its name then an autozero is performed. At the end of the autozero, the measured value displays.
Press briefly on “Select” to switch measurement parameter display. MP 50, MP 51, MP 55:Presss briefly and successively on “Select” to change the unit.
- Available units for MP 50: Pa, mmH20, inWg, daPa.
- Available units for MP 51: mmH20, kPa, in Wg, mbar, mmHg, daPa.
- Available units for MP 55: mbar, mmHg, hPa.
HM 50: Press briefly on “Select” to select a type of material. Press the spikes in the material to be measured.
When a measurement is over-range, above the maximum range, the value flashes. Below the minimum range, the value is no longer measurable, and the device indicates that it means that for:
- Wood type 1: the measurement is < than 10%
- Wood type 2: the measurement is < than 9%
Plaster, brick, and concrete: the measurement is < than 0.1 %
Carry out a manual autozero (MP 50, MP 51)
During measurement: Remove pressure tubes. Carry out a long press on “Select”. The instrument makes an autozero then returns to the measurement display.
Hold the measurement
During measurement, press briefly on “On/Off/Hold” to hold measurement:
“Hold” appears on the screen, and measurements are fixed. CO 50, HD 50, HM 50, LX 50, MP 55, VT 50: Press briefly on “Select” to move from one parameter to the other.Press again briefly on “On/Off/Hold” to exit the Hold function.
Change the unit
The instrument is on.
- CO 50, CT 50, HD 50, LX 50:
- Carry out a long press on “Select” to change the unit.
- Press briefly on “Select” to select the required unit of measurement.
- Carry out a long press on “Select” to confirm the unit and return to the normal measuring mode.
VT 50
- Change the air velocity unit:
- Press briefly on “Select to display the air velocity.
- Carry out a long press on “Select” to change the air velocity unit.
- Press briefly on “Select” to select the required unit of measurement.
- Carry out a long press on “Select” to confirm the unit and return to the normal measuring mode.
- Change the temperature unit:
- Press briefly on “Select” to display the temperature.
- Carry out a long press on “Select” to change the temperature unit.
- Press briefly on “Select” to select the required unit of measurement.
- Carry out a long press on “Select” to confirm the unit and return to the normal measuring mode.
Select the velocity type: contact or optical (CT 50)
The instrument is on.
- Carry out a long press on “Select”: “Cont” or “Optic” flashes.
- Press briefly on “Select” to select contact velocity (“Cont”) or optical velocity (“Optic”).
- Carry out a long press on “Select” to confirm.
Perform a self-test (HM 50)
Turn the instrument on. Remove the cap and turn the cap. Press “Select” to select the type of material to verify. Press the 2 pins of the instrument into the contact plugs of the cap. Read the value indicated by the instrument.
- If the indicated value matches the given value in the table below, the instrument runs properly.
- If the indicated value does not match with the given value in the table below, clean the tips and carry out a new self-test. If the value is still wrong, change the pins. If it does not work, please contact the after sales service.
Material | |||||
Resistance | Wood type 1 | Wood type 2 | Plaster | Brick and concrete | |
1 MΩ | 30.8% ±1% | 27.5% ±1% | 4.6% ±1% | 5.2% ±1% |
Materials humidity
Wood
Moisture content | State of the wood |
< 15% | Dry |
From 15 to 30% | Moist |
Plaster
Moisture content | State of the plaster |
< 2% | Dry |
From 2 to 4% | Moist |
Brick and concrete
Moisture content | State of the brick/concrete |
< 2% | Dry |
From 2 to 4% | Moist |
Batteries replacement
The instrument is OFF. Remove the front part from the back of the instrument. Replace used batteries with new batteries, AA LR03 1.5 V. Replace the front.
WARNING: To ensure safety, replace batteries with the same type. Not doing so might lead to explosion. Dispose of used batteries according to instructions.
Auto shut-off
The instrument switches off if it is not used for 20 minutes.
FAQs
Q: How do I register my product online for warranty?
A: Visit sauermanngroup.com to register your product online and confirm your warranty.
Documents / Resources
![]() |
KIMO CLASS 50 LUX Meter [pdf] User Guide CO 50, CT 50, HD 50, HM 50, LV 50, LX 50, MP 50, MP 51, MP 55, VT 50, CLASS 50 LUX Meter, CLASS 50, LUX Meter |