Inogen

Inogen One G3 Replacement Columns

Inogen-One-G3-Replacement-Columns

To change the columns on your Inogen One G4, first:

  1. Turn off the Inogen One® G4 concentrator by pressing the power button to shut down the device.
  2. Remove the Inogen One® G4 concentrator from the carrying case (if attached).
  3. Remove the battery from the Inogen One® G4 concentrator.Inogen-One-G3-Replacement-Columns-1
  4. Place the Inogen One® G4 concentrator on its side so that the underside is visible. The metal column assembly can be seen on one side of the device.Inogen-One-G3-Replacement-Columns-2
  5. Unlock the column assembly by pushing the latch button away from the columns.Inogen-One-G3-Replacement-Columns-3
  6. While holding the button open, slide the column assembly out of the device by pulling on the column pull handle.Inogen-One-G3-Replacement-Columns-4
  7. Remove the columns completely from the Inogen One® G4. Both columns are removed as one piece.Inogen-One-G3-Replacement-Columns-5
  8. Remove dust caps of new coulmn assembly. Make sure there is no dust or debris where the dust caps were located.Inogen-One-G3-Replacement-Columns-6
  9. Insert column assembly into the Inogen One® G4 concentrator. Do not leave the column ends exposed; column assembly should be inserted into the Inogen One G4 as soon as the dust caps have been removed.Inogen-One-G3-Replacement-Columns-7
  10. Push the column assembly into the device such that the columns are fully seated into the Inogen One® G4 concentrator. The spring loaded latch button should fully return to the closed position.Inogen-One-G3-Replacement-Columns-8
  11. Connect the AC power supply cord to the Inogen One® G4 and plug the power supply AC cord into an electrical outlet. Do Not Power on the Inogen One® G4 concentrator. Inogen-One-G3-Replacement-Columns-9
  12. Press and hold the plus (+) and (-) minus buttons for 5 seconds. The screen will display “sieve reset” message. Release buttons once message is displayed on screen.Inogen-One-G3-Replacement-Columns-10
  13. Press the alert/bell button once and screen will display “sieve reset success.” Press the power button to turn on the Inogen One® G4, and use it normally.Inogen-One-G3-Replacement-Columns-11
  14. If you are using Inogen Connect App, navigate to the Adv anced screen, then to Additional Information screen and click on the Column Reset button.

Note: Column change instructions are only to be used when maintenance is required and are not intended for practice purpose.