Elitech LogoGSP-8
User Manual

Overview

GSP-8 data logger s used to record the temperature/humidity of foods, medicines, chemicals and other goods during storage, transportation and i each stage of the cold chain including cooler bags, cooling cabinets, refrigerators, medicine cabinets, laboratories, reefer containers,and trucks, ‘The logger features a arge LCD screen, one button for operation, an internal temperature sensor, and an external temperature-humidity-combined probe (optional). It supports audible-visual alarm, two storage modes, and auto shortens logging interval function during alarms. also for wall mounting through the holes at the back of logger.

Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - OverView

  1. DLEDIndicator
  2. LCD Screen
  3. IsulatorStrip
  4. Button
  5. Short External Temperature-humidity Probe (optional)
  6. Long External Temperature-humidity Probe (2m, optional)

Specifications

Sensor  Digital temperature humidity sensor 
Temperature
Measurement Range
Internal sensor: -20∼60°C (-4°F~140°F);
External probe: -40∼85°C (-40°F~185°F) *
Humidity
Measurement Range
External probe: 1~99%RH
Temperature Accuracy ± 03’C/± 0.6°F (-20∼40°C) , ±  0.5°C/+0.9°F (others)
Humidity Accuracy  ±3%RH (25°C)
±5%RH (others) M183
Operating
Temperature Range
-20∼60°C (-4°F~140°F)
Ambient Temperature during power charging 0~40°C (32°F~104°F)
Resolution 01°C/’F; 01%RH
Memory Maximum 100,000 points
Logging Interval 10 seconds ta 24 hours
LCD ON 15/30/45/60 seconds; always on
Backlight ON 15/30/45/60 seconds; always on
Start Mode Press button; auto start; timed start
Stop Mode Press button; use software; auto stop
Alarm Thresholds
Data Interface
Configurable, temperatue: up to 3 high limits and 5 1o, imits; humidity:

1 high limit and 1 low limit

Alarm Type Single; cumulative
Alarm Delay 10 seconds to 24 hours
Data Interface USB
Report Format A0 POF et

PDF/EXCEL/TXT** by ElitechLog software

Battery Rechargeable 18650 lithium battery
battery life Store and use for 2 years

(@25°C and 15 minutes logging interval)

Dimensions 94x90x29 mm
Weight 173

* At ultralow temperoture, the LCD is slow but does rot offect normal Iogging. ft will be back to normol after the temperature rises.
“*TXT for Windows ONLY.

Operation

  1. Activate Logger
    Pull out the insulator strip from the side of the logger to activate the battery.
  2. Install Software
    Please download and install the free ElitechLog software (mac0S and Windows) via www.elitechlog.com.
  3. Configure Parameters
    First, connect the data logger to the computer via USB cable, wait until the Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 2 icon shows on the LCD, then configure via:
    ElitechLog software:
    – If you do not need to change the default parameters (in Appendix); please click Quick Reset under Summary menu to synchronize local time before usage:
    – If you need to change the parameters, please click the Parameter menu, enter your preferred values, and click the Save Parameter button to complete the configuration.
    Warning! For first time user or after battery replacement: To avoid time or time zone errors, pleose make sure you click Quick Reset or Save Parameter before usage to sync and configure your local time into the logger.
  4. Start Logging
    Press Button:
    Press and hold the button for 5 seconds until the Right icon shows on the LCD, indicating the logger starts logging Auto Start:
    Immediate Start: The logger starts logging after remove it from the computer.
    Timed Start: The logger starts counting after being removed from the computer; it will start logeing automatically after the set date/time.
    Note: If the Right  icon keeps floshing, it means the logger configured with stort delay; it willstart logging after the set delay time elapses
  5. Mark Events
    Double click the button to mark current temperature and time, up to 10 groups of data. After marked, it will be indicated by Log X and SUC on the LCD screen (X means the marked group)
  6. Stop Logging
    Press Button®: Press and hold the button for 5 seconds until the Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon icon shows on the LCD, indicating the logger stops logging.
    Auto Stop**: When the logging points reach the maximum memory, the Iogger will stop automatically.
    Use Software: Open Elitechlog software, click Summary menu, and Stop Logging button
    Note: “Default stop method is Press Button. If you disabled it, the button operation would be invalid; please open the ElitechLog software. and click Stop Logging button to stop it **The Auto Stop function will be disabled automatically if you enabled the  Circular Logging
  7. Download Data
    Connect the data logger to the computer via USB cable, wait until 2 shows on the LCD screen, then download data:
    Without ElitechLog Software: Simply find and open the removable storage device ElitechLog, save the auto generated PDF report to your computer for viewing With ElitechLog software: After the logger auto-upload its data to the ElitechLog software, click Export and select your preferred file format to export. If data failed to auto-upload, please manually click Download and then repeat above operation.
  8. Reuse the Logger
    To reuse a logger, please stopped it first; then connect it to your computer and use ElitechLog software to save or export the data.
    Next, reconfigure the logger by repeating the operations in 3. Configure Parameters*. After finished, follow 4. Start Logging to restart the logger for new logging.
    Warning! *To make space for the new loggings, All previous logging data inside the logger will be deleted ofter re-configuration, f you forgot to save/export dato, please try to locate the logger in the History meru of ElitechLog software.
  9.  Repeat Start:
    If you have enabled Repeat Start function”, simply press and hold the button for § seconds to quickly restart the stopped logger without any configuration.
    Note: * Repeat Start i disabled by defoult to prevent dota loss coused by misoperation. Pleose backup data before Repeat Start by repeating 7.Download Data.

Status Indication

  1. Buttons
    Operations Function
    Press and hold the button for S seconds Start/Stop logging
    Press and release the button Check/Switch interfaces
    Double click the button Mark events
  2. LCD ScreenElitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Screen
    1. BatteryLevel
    2. Functions
    3. Channel Selection
    4. Over Limit Alarm
    5. Connected to PC
    6. Stopped/Logging
    7. Circular Logging
    8. Buzzer (Enabled)
    9. Temperature
    10. Humidity
    11.  Logging Interval/Date or Time
  3. LCD Interface
    Not Start : Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - LED Interface 1 Maximum: Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - LED Interface 5
    Delay/Timed Start (Right flashes) Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - LED Interface 2 Minimum: Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - LED Interface 6
    Logging: Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - LED Interface 3 High Limit: Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - LED Interface 7
    Average: Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - LED Interface 4 Low Limit : Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - LED Interface 8

    Note: I configured to use the internol sensor (temp). the hurmidity will not
    disploy by defoult.

  4. LED & Buzzer Indication
    LED Indicator Buzzer* Meaning…
    Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 3 flash once Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 5 beeps once Configuration completed
    Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 3  flash twice Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 5 beeps twice Started
    Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 3 flash 3 times Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 5 beeps 3 times Stopped
    Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 3 flash 4 times Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 5 beeps 4 times Timed/Delay start
    Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 3 flash once every 15 seconds Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 5 in alarm status*
    Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 4 flashes once every 1 seconds / Charging
    Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 4 always on, and then off / Charged

Battery Charging

Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - BettaryConnect the included USB cable into a computer or poweradapter for battery charging.

Wall Mounting

Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Wall GSP-8 supports hanging-mount and stand (optional) only, please confirm the installation distance and preferred nail size before hanging the logger to wall (nail not included).

What’s Included

  • GSP-8 Temperature (Humidity) Data Logger x 1
  • Micro USB Cable x 1
  • User Manual x1
  • Certificate of Verification x 1
  • Long External Temperature-humidity Probe (2m) x 1 (Optional)
  • Certificate of Verification (2m) x 1 (Optional)
  • Short External Temperature-humidity Probe x 1 (Optional)
  • Certificate of Verification (Short) x 1 (Optional)
  • Stand x1 (Optional)

Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 6  Warning!
Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 7 Ifyou use the logger for the first time, pleose use ElitechL og software to synchronize syster time and configure parameters.
Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 7 Any parameter configuration on ElitechLog softwore will defete oll Iogged dota inside the logger. Please save data before you opply new configurations
Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 7 The LLD screen will auto off after 15 seconds of inactiviy (by default) Press the button again to turn on the screen. The backiight ON time is configurable via ElitechLog software.
Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 7 To ensure the humidity accuracy, please avoid contact with unstable chemical solvents or compounds, especially avoid fong-term storage or exposure to the environments with high concentrations of ketene, acetone, ethariol, isopropanol, toluene, etc.
Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm - Icon 7 Recommended to remove the battery from the – side (cathode).

Appendix

Default Parameters:

Logging Interval 15 minutes
High Temperature Limit 60
Low Temperatura Limit 0°C
Button Tone Enable
Audible Alarm Disable
Logging interval shortened in alarm Disable
Loon 15 seconds
Backdight ON OFF

Elitech LogoElitech Technology, INC.
wwwlitechlog.com
1551 McCarthy Bivd, Suite 112
Milpitas, CA 95035 USA

Documents / Resources

Elitech GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm [pdf] User Manual
GSP-8A, GSP-8A Temperature and Humidity Data Logger Recorder Audio Alarm, Temperature and Humidity Data Logger Recorder Audio Alarm, Humidity Data Logger Recorder Audio Alarm, Data Logger Recorder Audio Alarm, Logger Recorder Audio Alarm, Recorder Audio Alarm, Audio Alarm, Alarm

References

Leave a comment

Your email address will not be published. Required fields are marked *