dynamic-BIOSENSORS-logo

dynamic BIOSENSORS AS-2-Gc-lfs Adapter Strand 2

dynamic-BIOSENSORS-AS-2-Gc-lfs-Adapter-Strand-2-product

Specifications

  • Product Name: heliX+ ADAPTER STRAND 2 with green dye Gc
  • Manufacturer: Dynamic Biosensors GmbH & Inc.
  • Order Number: AS-2-Gc-lfs
  • Version: v5.1

Preparation and Usage Instructions

  1. Mix Adapter strand 1 – Gc (400 nM) and conjugated Ligand strand (500 nM) at a 1:1 ratio (v/v).
  2. Incubate the solution from step 1 at room temperature at 600 rpm for 30 minutes for complete hybridization.
  3. Mix the solution from step 2 with Adapter strand 2 – Gc – lfs (200 nM) at a 1:1 ratio (v/v).
  4. The solution is now ready for biochip functionalization.
  5. Note: The stability of the solution is dependent on the stability of the ligand molecule.

FAQ

  • What is the shelf life of Adapter strand 2 – Gc- lfs?
  • The product has a limited shelf life. Please refer to the expiry date on the label for specific information.
  • Can Adapter strand 2 – Gc- lfs be used for purposes other than research?
  • No, this product is intended for research use only.

Key Features

  • Adapter strand 2 for functionalization of heliX® Adapter Chip Spot 2.
  • Compatible with heliX® Adapter Chip.
  • Includes Adapter strands for 50 regenerations.
  • Ideal for MIX&RUN sample preparation.
  • Adapter strand 2 carries a hydrophobic green dye (Gc) with a neutral net charge.

heliX® Adapter Chip Overview

2 spots with 2 different anchor sequences for DNA-encoded addressing.

dynamic-BIOSENSORS-AS-2-Gc-lfs-Adapter-Strand-2-fig-1Product Description

Order Number: AS-2-Gc-lfs
Table 1. Contents and Storage Information

Material Cap Concentration Amount Buffer Storage
Adapter strand 2 – Gc- lfs White 200/250 nM 5 x 200 µL TE40 [1] -20°C

or research use only. This product has a limited shelf life, please see the expiry date on the label. To avoid many freeze-thaw cycles please aliquot the nanolever.

Preparation | MIX&RUN

In-solution hybridization of adapter and ligand strands:

  1. Mix Adapter strand 1 – Gc (400 nM) and conjugated Ligand strand (500 nM) at 1:1 ratio (v/v).
  2. Incubate the solution of step 1 at RT at 600 rpm for 30 min to ensure complete hybridization.
  3. Mix the solution of step 2 and Adapter strand 2 – Gc – lfs (200 nM) at 1:1 ratio (v/v).

dynamic-BIOSENSORS-AS-2-Gc-lfs-Adapter-Strand-2-fig-2

The solution is ready to use for biochip functionalization.  Stability of the solution is related to the stability of the ligand molecule.

Table 2. Additional material for functionalization of spot 1 and reference spot 2.

Material Concentration Buffer Related Product Name Order No
Adapter strand 1 – Gc 400 nM TE40 [1] Adapter strand 1 with green dye Gc AS-1-Gc
Ligand strand

carrying the conjugated ligand

500 nM P2] E40 [ heliX® Amine Coupling Kit 1 HK-NHS-1

Example
Required volume for 3 functionalizations: 100 µL with a final concentration of 100 nM.

Vial 1 Vial 2
Adapter strand 1 – Gc

(400 nM)

Conjugated Ligand strand

(500 nM)

Adapter strand 2 – Gc – lfs

(200/250 nM)

25 µL 25 µL 50 µL

After incubation time, mix vial 1 and vial 2 to obtain 100 μL of ready-to-use DNA solution.

Contact

Dynamic Biosensors GmbH

  • Perchtinger Str. 8/10 81379 Munich
  • Germany

Dynamic Biosensors, Inc

©2024 Dynamic Biosensors GmbH | Dynamic Biosensors, Inc. All rights reserved.[1] TE40: 10 mM Tris, 40 mM NaCl, 0.05 % Tween20, 50 μM EDTA, 50 μM EGTA [2] If the protein is not stable in PE40 (TE40, HE40), please check buffer compatibility with the switchSENSE® compatibility sheet.