Dynamic Biosensors Adapter Strand Package
Key Features
- Adapter strand 1 – Ga and Adapter strand 2 – Ga – lfs (ligand free strand) for functionalization of heliX ® Adapter Chip on Spot 1 and Spot 2, respectively.
- Compatible with heliX ® Adapter Chip.
- Includes Adapter strands for 50 regenerations.
- Ideal for MIX&RUN sample preparation.
- Adapter strands 1 and 2 carry a moderately hydrophilic green dye (Ga) with a single negative net charge.
Adapter Chip Overview
2 spots with 2 different anchor sequences for DNA-encoded addressing.
Product Description
Order Number: ASP-1-Ga
Table 1. Contents and Storage Information
Material | Cap | Concentration | Amount | Buffer | Storage |
Adapter strand 1 – Ga | Black | 400 nM | 5 x 100 µL | TE40 [1] | -20°C |
Adapter strand 2 – Ga – lfs | White | 200/250 nM | 5 x 200 µL | TE40 [1] | -20°C |
For research use only.
his product has a limited shelf life, please see expiry date on label.
To avoid many freeze thaw cycles please aliquot the nano lever.
Preparation | Mix&run
In-solution hybridization of adapter and ligand strands:
- Mix Adapter strand 1 – Ga (400 nM) and conjugated Ligand strand (500 nM) at 1:1 ratio (v/v).
- Incubate the solution of step 1 at RT at 600 rpm for 30 min to ensure complete hybridization.
- Mix solution of step 2 and Adapter strand 2 – Ga – lfs (200 nM) at 1:1 ratio (v/v).
Solution is ready to use for biochip functionalization.
Stability of the solution is related to the stability of the ligand molecule.
Table 2. Additional material for functionalization of spot 1 and reference spot 2.
Material | Concentration | Buffer | Related Product Name | Order No |
Ligand strand carrying the conjugated ligand | 500 nM | PE40 [2] | heliX ® Amine Coupling Kit 1 | HK-NHS-1 |
Example
Required volume for 3 functionalizations: 100 µL with a final concentration of 100 nM.
Vial 1 | Vial 2 | |
Adapter strand 1 – Ga (400 nM) | Conjugated Ligand strand (500 nM) | Adapter strand 2 – Ga – lfs (200/250 nM) |
25 µL | 25 µL | 50 µL |
After incubation time, mix vial 1 and vial 2 to obtain 100 µL of ready-to-use DNA solution.
Customer Support
Contact
Dynamic Biosensors GmbH
Perchtinger Str. 8/10
81379 Munich
Germany
Bruker Scientific LLC
40 Manning Road, Manning Park
Billerica, MA 01821
USA
Order Information order@dynamic-biosensors.com
Technical Support support@dynamic-biosensors.com
www.dynamic-biosensors.com
Instruments and chips are engineered and manufactured in Germany.
©2025 Dynamic Biosensors GmbH
For Research Use Only. Not for use in clinical diagnostic procedures.
[1] TE40: 10 mM Tris, 40 mM NaCl, 0.05 % Tween20, 50 μM EDTA, 50 μM EGTA