APEX MEDICAL CORP.
XT Auto
Notice! Please remove the guide before giving the device to the patient. Only medical professionals could adjust pressure settings.
Provider Setup Navigation
- Press “Menu”
to any
Menu (See Operation Flow on the right) while in the Standby screen.
- Hold down both the “UP”
& ”Down”
buttons, and then simultaneously press the “Menu”
for 1 second. When the screen starts blinking, the Provider Mode would be unlocked.
- Press “UP”
and ”Down”
to adjust the settings.
- Press “Menu”
to confirm the settings.
- Press the “Start/Standby”
to go back to the standby screen.
Provider Mode Quick Guide
Re-set the Compliance Meter
- In the standby screen, press “Menu”
to select [CM xxxx.x hr] menu.
- Accessing the Provider Mode first (Press
), then the screen starts blinking.
- Press “UP”
or ”Down”
, then the screen will show [Clear].
- Press “Menu”
to confirm the settings, and the Compliance Meter record will be erased. The screen will show [Clear OK] when finishing the process.
Engineering Mode Quick Guide
Accessing the Engineering Mode
1. Plug the power supply into the iCH device.
2. Before it shows [Standby] screen, hold down both the “Starts/Standby” and “Menu”
buttons for 1 second, the
Engineering Mode would be unlocked.
Re-set the Total Meter
- Accessing the Engineering Mode first, then the screen will show [TM xxxx.x hr] menu.
- Hold down both the “UP”
& ”Down”
buttons, and simultaneously press the “Menu”
, and then the screen starts blinking.
- Press “UP”
or ”Down”
, then the screen will show [Clear].
- Press “Menu”
to confirm the settings, and the screen will show [TM CLr Clear],and the Total Meter record will be erased.
- The screen will show [TM End Clear OK] when finishing the process.