SATEC PMU230 PMU PRO Phasor Measurement Unit
Specifications
- Product Name: Phasor Measurement Unit PMU PRO
- Model: BG0649 Rev. A1
- Warranty: 24-month functional warranty from date of dispatch, valid for 36 months from production date
Product Usage Instructions
- Installation:
- Ensure all incoming AC power and other power sources are turned off before starting installation.
- Check the labels on the front and side of the instrument to ensure appropriate power supply voltage, input voltages, and currents.
- Do not connect the instrument to a power source if it is damaged.
- Avoid exposing the instrument to rain or moisture to prevent fire or shock hazards.
- Operation:
- Only qualified personnel familiar with the instrument and its associated electrical equipment should perform setup procedures.
- Do not open the instrument when connected to a power source.
- Do not use the instrument for primary protection functions where failure can cause fire, injury, or death; it should only be used for secondary protection if needed.
- Hazardous voltages are present on input terminals during operation, so follow precautions to avoid serious injury or equipment damage.
- Maintenance:
- For maintenance, contact the manufacturer, local representative, or distributor for guidance.
- Do not attempt to open the unit unless by a duly authorized representative in an anti-static environment to avoid voiding the warranty.
FAQs
- Q: What should I do if the instrument malfunctions?
- A: If you experience any issues with the instrument, contact the manufacturer or your local representative for assistance as opening the unit without authorization may void the warranty.
- Q: Can the instrument be used in wet conditions?
- A: No, to prevent potential fire or shock hazards, do not expose the instrument to rain or moisture.
“`
Phasor Measurement Unit PMU PRO
Installation and Operation Manual
BG0649 Rev. A1
LIMITED WARRANTY
The manufacturer offers the customer a 24-month functional warranty on the instrument for faulty workmanship or parts from date of dispatch from the distributor. In all cases, this warranty is valid for 36 months from the date of production. This warranty is on a return to factory basis.
The manufacturer does not accept liability for any damage caused by instrument malfunction. The manufacturer accepts no responsibility for the suitability of the instrument to the application for which it was purchased.
Failure to install, set up or operate the instrument according to the instructions herein will void the warranty.
Only a duly authorized representative of the manufacturer may open your instrument. The unit should only be opened in a fully anti-static environment. Failure to do so may damage the electronic components and will void the warranty.
The greatest care has been taken to manufacture and calibrate your instrument. However, these instructions do not cover all possible contingencies that may arise during installation, operation or maintenance, and all details and variations of this equipment are not covered by these instructions.
For additional information regarding installation, operation or maintenance of this instrument, contact the manufacturer or your local representative or distributor.
WARNING
Read the instructions in this manual before performing installation, and take note of the following precautions:
Ensure that all incoming AC power and other power sources are turned off before performing any work on the instrument. Failure to do so may result in serious or even fatal injury and/or equipment damage.
Before connecting the instrument to the power source, check the labels on the front and side of the instrument to ensure that your instrument is equipped with the appropriate power supply voltage, input voltages and currents.
Under no circumstances should the instrument be connected to a power source if it is damaged.
To prevent potential fire or shock hazard, do not expose the instrument to rain or moisture.
The secondary of an external current transformer must never be allowed to be open circuit when the primary is energized. An open circuit can cause high voltages, possibly resulting in equipment damage, fire and even serious or fatal injury. Ensure that the current transformer wiring is secured using an external strain relief to reduce mechanical strain on the screw terminals, if necessary.
Only qualified personnel familiar with the instrument and its associated electrical equipment must perform setup procedures.
2
PMU PRO Phasor Measurement Unit
Do not open the instrument under any circumstances when it is connected to a power source.
Do not use the instrument for primary protection functions where failure of the device can cause fire, injury or death. The instrument can only be used for secondary protection if needed.
Read this manual thoroughly before connecting the device to the current carrying circuits. During operation of the device, hazardous voltages are present on input terminals. Failure to observe precautions can result in serious or even fatal injury or damage to equipment.
All trademarks are property of their respective owners
June 2024 Copyright 2024 © SATEC Ltd.
PMU PRO Phasor Measurement Unit
3
PMU PRO Phasor Measurement Unit
General
The PMU PRO is a compact DIN rail mounted phasor measurement unit (PMU) designed to meet the requirements of users ranging from electrical panel builders to substation operators.
The PMU PRO provides synchrophasor and frequency measurements compliant with the P and M performance classes of IEEE C37.118.1/1a and IEC/IEEE 60255-118-1:2018 standards.
PMU Features
· IEEE C37.118.1 compliant three-phase voltage, current, and positive and negative sequence phasor measurements synchronized to a common UTC time reference
· IEEE C37.118.1 synchronous frequency and rate of change of frequency (ROCOF) measurements
· Clock synchronization to a UTC time reference using an IRIG-B timecode source or an IEEE 1588 PTPv2 master clock source
· IEEE C37.118.2 commanded client-server UDP and TCP data transmission and spontaneous UDP data transmission over IP protocol
· IEEE C37.118.2 reporting rates from 1 to 200/240 frames/s
· Optional IEEE C37.118.2 data extensions with analog data (total fundamental active, reactive and apparent power and displacement power factor), and digital status data
· Streaming of phasor data over Ethernet using the IEC 61850-9-2 multicast sampled value (SV) service with IEEE C37.118.2 compliant mapping of synchrophasor data upon IEC 61850-9-2 and IEC 61850-90-5 guidelines
On-board Event and Data Recording
· 16 GB eMMC flash memory for long-term event and data recording
· Event recorder for logging internal diagnostic events, power faults and operations of the logic controller and digital I/O
PMU PRO Phasor Measurement Unit
7
1 General
· Eight data recorders with programmable datasets for data logging on a periodic basis and on any internal or external trigger
Programmable Logic Controller
· Programmable logic controller with 32 control setpoints and half-cycle scan time
· OR/AND logic, extensive triggers, programmable thresholds and delays, relay control, event-driven data recording
· 32 digital counters for counting internal events and pulses from external sources
· 16 interval timers with programmable periods from half cycle to 24 hours for periodic recording and triggering timed operations
Time and Clock
· High-accuracy real-time clock with a lithium backup battery
· Clock synchronization to a UTC time reference using an IRIG-B timecode source or an IEEE 1588 PTPv2 master clock
· Daylight saving time shift for local time indication with configurable DST start and end time
Extended Security
· 3-level password security for protecting device setups and recorded data from unauthorized changes
Display
· · · ·
High contrast 1.77″ TFT color graphics display with configurable backlight Multi-page displays; time, instrumentation and service data Menu-driven setups Multilanguage support
Built-in Digital and Analog I/O
· One optically isolated 24VDC dry contact input with programmable de-bounce time; steady-state and pulse operations; normal and inverting polarity
· One solid state relay output; unlatched, latched and pulse operations; failsafe operation for alarm notifications; programmable pulse width; direct remote relay control through communications
· One optically isolated analog input with user selectable input range of 0-1 mA, +/-1 mA, 0-20 mA or 4-20 mA
Expansion Digital I/O Options · Up to three I/O expansion modules can be daisy-chained into the side expansion connector
· 8DI module option with eight optically isolated digital inputs; dry contacts or 24V, 48V, 125V and 250V wet contacts options
8
PMU PRO Phasor Measurement Unit
1 General
· 4RO module option with four electro-mechanical or solid-state relays
· 4DI/2RO module option with four optically isolated digital inputs and two electro-mechanical or solid-state relays
Communications Options · Two 10/100Base-T IEEE 1588 Ethernet ports with packet forwarding/routing capabilities; connecting to Ethernet through one or two different network switches, or daisy chaining multiple devices to a switch using a linear or ring network topology (daisy chaining is not available for PTP-enabled devices); up to 8 simultaneous TCP connections via Modbus TCP, DNP3, IEC 60870-5-104, IEC 61850 (up to 5 client associations, GOOSE and MSV publishers) protocols; up to 5 TCP or/and UDP synchrophasor data streams via C37.118.2 protocol
· Full speed USB 2.0 type C port for local configuring and monitoring the device via Modbus RTU protocol
· RS-485 serial communication port; 2400 to 115200 bps; Modbus RTU/ASCII, DNP3 and IEC 60870-5-101 protocols (the port is not operational if the IRIG-B time synchronization source is used).
· Modbus TCP notification client · eXpertPowerTM client
Backup Power Supply · Auxiliary 88-264 VAC/125-300 VDC 50/60 Hz power supply expansion module
Upgradeable Firmware · New features can be easily added to the device by simply replacing firmware through any communication port
Software Support · PAS free configuration and data acquisition tool for remote configuring and monitoring the device
· Exporting data in Microsoft Excel workbook, CSV, or Electrotek PQDIF file format
· eXpertPowerTM SATEC proprietary Internet services
PMU PRO Phasor Measurement Unit
9
1 General
Product documentation
BG0650 PMU PRO Modbus Protocol Reference Guide
BG0651 PMU PRO DNP3 Protocol Reference Guide
BG0652 PMU PRO IEC 60870-5 Protocol Reference Guide
BG0653 PMU PRO IEC 61850 Protocol Reference Guide
BG0654 PMU PRO IEC 61850 Protocol Implementation Conformance Statement (PICS) BG0655 PMU PRO IEEE C37.118.2 Protocol Reference Guide
Reference documents
IEEE Std C37.118.1-2011, IEEE Standard for Synchrophasor Measurements for Power Systems
IEEE Std C37.118.1a-2014, IEEE Standard for Synchrophasor Measurements for Power Systems, Amendment 1: Modification of Selected Performance Requirements
IEEE Std C37.118.2-2011, IEEE Standard for Synchrophasor Data Transfer for Power Systems IEC/IEEE 60255-118-1:2018, Measuring relays and protection equipment Part 118-1: Synchrophasor for power systems Measurements IEC/TR 61850-90-5:2012, Communication networks and systems for power utility automation Part 90-5: Use of IEC 61850 to transmit synchrophasor information according to IEEE C37.118
IEEE Std 1588-2008, IEEE Standard for a Precision Clock Synchronization Protocol for Networked Measurement and Control Systems
IEEE Std C37.238-2017, IEEE Standard Profile for Use of IEEE 1588 Precision Time Protocol in Power System Applications
10
PMU PRO Phasor Measurement Unit
Installation
2.1 Mechanical Installation
Mechanical Installation
Figure 2-1 Device Dimensions and Mounting
PMU PRO Phasor Measurement Unit
11
2 Installation
2.2 Mounting Expansion Modules
Mounting Expansion Modules
REM OV E COVE R T O CONNE CT M ODULE
Module connector cover Remove Module
connector cover before assembling module
Figure 2-2 Expansion Module Dimensions and Mounting
8DI RDY
Figure 2-3 Expansion Module (front view)
12
PMU PRO Phasor Measurement Unit
2 Installation
Electrical Installation
2.3 Electrical Installation
Before installation ensure that all incoming power sources are turned off. Failure to
observe this practice can result in serious or even fatal injury and damage to equipment.
Connections to the PMU PRO are made via front terminals as shown in Figure 2-5.
Please pay attention to the following notes before installation:
· Under no circumstances should the instrument be connected to a power source if it is damaged.
· Before connecting the instrument to the power source, check the labels on the front and side of the instrument to ensure that your instrument is equipped with the appropriate power supply, voltage inputs and current inputs.
· The secondary of an external current transformer must never be allowed to be open circuit when the primary is energized. An open circuit can cause high voltages, possibly resulting in equipment damage, fire and even serious or fatal injury. Ensure that the current transformer wiring is secured using an external strain relief to reduce mechanical strain on the screw terminals, if necessary.
· When mounting the instrument on a DIN rail inside a switchgear, the metal plate of the switchgear must be grounded for safety purposes.
· Use a 2A low-voltage circuit breaker before connecting the voltage inputs to mains.
· Use non-insulated ferrules, such as Panduit Pan-Term® F77-8-M termination.
for wire
· Provide adequate segregation or barrier separation between the different classes of circuits: (a) Class 1 field and factory wiring (CT output leads, voltage measurement leads, mains input power), terminals, and bare live parts, and (b) Class 2 and Class 3 field and factory wiring, terminals, and non-insulated live parts.
Figure 2-4 Device Terminals (front view, with and without plug-in terminals)
All connection terminals except of current input terminals are detachable plug-in terminals. Measurement terminals and digital and analog I/O terminals are protected with sealable covers against unauthorized access or unwanted human contact.
PMU PRO Phasor Measurement Unit
13
2 Installation
Electrical Installation
Figure 2-5 PMU PRO Connections
2.3.1 Power Supply Connection
The power supply can be dedicated-fused, or from a monitored voltage if it is within the instrument power supply range. For AC power supply, connect line to terminal L/+ and neutral to terminal N/-. For DC power supply, connect positive wire to terminal L/+ and negative to terminal N/-
Use an external circuit breaker or switch for safety purposes.
2.3.2 Voltage Inputs Connection
Connect the mains voltage wires to terminals V1, V2, V2 and the neutral wire to terminal N. Voltages in the range of up to 277/480 VAC can be connected directly to the device voltage terminals. Voltages above 277/480 VAC must be connected via potential transformers. See Figure 2-6 and Figure 2-7 for connection diagram.
14
PMU PRO Phasor Measurement Unit
2 Installation
Electrical Installation
Figure 2-6 4-wire WYE Direct Connection
Figure 2-7 4-wire WYE PT Connection
2.3.3 Current Inputs Connection
Connect input current wires to terminals I1+/I1-, I2+/I2-, and I3+/I3-. See Figure 2-6 and Figure 2-7 for connection diagram.
2.3.4 Built-in I/O Connection
Figure 2-8 PMU PRO Built-in I/O Terminals
PMU PRO Phasor Measurement Unit
15
2 Installation
Electrical Installation
Relay Output
One built-in solid-state SPST Form A relay is provided, rated at 100 mA/250 VAC/VDC. Connect the relay load between terminals SO+ and SO-. See Figure 2-8 for relay connections.
Digital Input
One built-in optically isolated self-powered 24VDC dry contact input is provided for monitoring external dry contacts. Connect the external contacts between terminals DI1 and CM. See Figure 2-8 for digital input connections.
Analog Input
One built-in optically isolated analog input is provided with a user selectable input range of 0-1 mA, +/-1 mA, 0-20 mA or 4-20 mA. See Figure 2-8 for analog input connections.
IRIG-B Port
The IRIG-B port uses an unmodulated (pulse width coded) time code signal (5V unbalanced level) in accordance with the IRIG 200-04 standard. It supports B004 and B005 timecode formats with the timecode extensions specified for the IRIG-B profile in Annex D of IEEE C37.118.1-2011.
The IRIG-B port shares a common terminal with digital input DI1. See Figure 2-8 for IRIG-B connections.
NOTE: The IRIG-B port shares input circuitry with the RS-485 port, so only one of them can be used at a time.
2.3.5 Auxiliary Power Supply Connection
The auxiliary power supply module is designed to increase the power ability of the device. It is connected in series with the I/O expansion modules on the side expansion connector. Allowable external voltage range is 88-264 VAC or 125-300 VDC.
Figure 2-9 Auxiliary Power Supply Connection
2.3.6 Expansion I/O Modules Connection
Up to 2 expansion modules can be normally connected to the side expansion terminal without overloading the device power supply. Connecting the third expansion module requires the auxiliary power supply to be used. For I/O ratings, see the expansion I/O specifications in Appendix A .
8DI Expansion Module The 8DI expansion module provides additional 8 optically isolated dry contact or wet contact digital inputs. See the following figures for module connections.
16
PMU PRO Phasor Measurement Unit
2 Installation
Electrical Installation
Figure 2-10 8DI Dry Contact Module Connection
Figure 2-11 8DI Wet Contact Module Connection
Dry contact modules use the internal 24VDC power supply. For wet contact modules, use an external power supply with an operating voltage of 24V, 48V, 125V or 250V, which can be connected in any polarity.
4RO Expansion Module The 4RO expansion module provides four electro-mechanical or solid-state relays.
Figure 2-12 4RO Module Connection
4DI/2RO Expansion Module
The combined 4DI/2RO expansion module provides four dry contact or wet contact optically isolated digital inputs and two electro-mechanical or solid-state relays.
Dry contact inputs have an internal 24VDC power supply. For wet contact inputs, use an external power supply with an operating voltage of 24V, 48V, 125V or 250V, which can be connected in any polarity.
Figure 2-13 Dry Contact Inputs Connection
Figure 2-14 Wet Contact Inputs Connection
PMU PRO Phasor Measurement Unit
17
2 Installation
Electrical Installation
2.3.7 Communication Connections
See Figure 2-15 for the location of the PMU PRO communication ports.
Figure 2-15 Communication Ports Location
RS-485 Serial Port
The built-in RS-485 port with user-configurable response delay and inter-message break timeout can operate at speeds from 2400 to 115200 bps with 7E, 8N and 8E data formats. The port is designated as COM1. See Figure 2-16 for recommended port connections.
NOTE: The RS-485 port shares input circuitry with the IRIG-B port, so only one of them can be used at a time.
The RS-485 port supports Modbus/RTU, Modbus/ASCII, DNP3 and IEC 60870-101 serial communication protocols.
Figure 2-16 RS-485 Port Connection
18
PMU PRO Phasor Measurement Unit
2 Installation
Electrical Installation
USB Port
The PMU PRO USB port is a full speed USB 2.0 port equipped with a Type C receptacle. It is operated over Modbus RTU protocol on Windows 7, 10 and 11 using a proprietary device driver.
Figure 2-17 USB Port Connection
Use a USB Type C cable to connect your PC to the PMU PRO USB port. To prevent potential differences between the PMU PRO and PC USB ports, it is recommended to use a galvanically isolated USB adaptor or a battery powered PC.
Ethernet Ports
Two 10/100Base-T IEEE 1588 Ethernet ports with packet forwarding/routing capabilities allow direct connecting to Ethernet through one or two different network switches, or daisy chaining multiple devices to a switch using a linear or ring network topology. The ports are equipped with RJ-45 connectors.
Figure 2-18 Direct Ethernet Connection (separated port mode)
NOTE Daisy chaining is not available for PTP-enabled devices. Only one of them, directly connected to the 1588 switch, can have PTP synchronization enabled, and only using a linear topology.
PMU PRO Phasor Measurement Unit
19
2 Installation
Electrical Installation
Figure 2-19 Daisy Chain Connection with Linear Topology (daisy chain port mode)
Figure 2-20 Daisy Chain Connection with Ring Topology (daisy chain port mode)
20
PMU PRO Phasor Measurement Unit
3 Operating the PMU PRO
Device Security
3 Operating the PMU PRO
3.1 Device Security
The PMU PRO provides 3-level password security with 8-digit user passwords to protect device configuration settings and stored data from unauthorized changes. Access to particular setup and control items is granted depending on the security level of the password you entered as shown in the following table.
Password Password 1
Security level Low
Password 2 Medium
Password 3 High (administration level)
Access rights Device diagnostics Clock update Display setup Reset of pulse counters Communication setup I/O operation setup and control Memory and recorders setup Telnet Password setup Basic device setup Reset of log files
You will be prompted to enter the password each time you enter the device’s setup menus. The device is shipped with a factory-set password of 9. See 6.1 “Configuring User Passwords” for information on how to configure passwords in your device.
Additionally, mechanical means are used to secure device connection terminals. Measurement terminals and digital and analog I/O terminals are protected with sealable covers against unauthorized access or unwanted human contact.
3.2 Device Diagnostics
Device diagnostic messages may appear as a result of the PMU PRO built-in diagnostic tests performed during start-up and device operation.
See 4.1.2 “Status Indicators” for diagnostic indications on the front display.
All diagnostic events with time stamps are recorded in the device event log and can be inspected through the front display (see 4.3.6 “Events Menu”) and through PAS (see 7.5.3 “Viewing the Event Log”). See 6.5 “Viewing and Clearing Device Diagnostics” for how to clear device diagnostics in your device.
See Appendix E “Diagnostic Messages” for the list of diagnostic messages and troubleshooting device issues.
3.3 Monitoring the Battery
A backup lithium battery keeps the device clock running when the power is removed from the device.
You can see the status of the battery in the display status bar (see 4.1.2 “Status Indicators”) and via the device diagnostics.
PMU PRO Phasor Measurement Unit
21
Operating the PMU PRO
Clock and Time Synchronization
3.4 Clock and Time Synchronization
Time synchronization provides a common time basis for synchrophasor measurements and data recorders.
The PMU PRO clock must be synchronized to UTC time using an external IRIG-B timecode source or from an IEEE 1588 PTPv2 master clock.
You can see the clock status in the status bar on the front display and check the time quality on the Monitor/Clock Status screen (see “Device Clock Status” in 4.3.3).
You can also check the clock status and time accuracy via Telnet using the “time” command as shown in the following picture.
3.4.1 IRIG-B Time Synchronization
To enable the IRIG-B port as a UTC source, select the IRIG-B time synchronization option in the device Local Settings setup.
The IRIG-B port uses an unmodulated (pulse-width coded) timecode signal (unbalanced 5V level) according to the IRIG 200-04 standard. It supports B004/B005 timecode formats with time code extensions specified for the IRIG-B profile in Annex D of IEEE Std C37.118.1-2011.
The IRIG-B port can synchronize to the IRIG-B timecode source with sub-microsecond accuracy.
3.4.2 IEEE 1588 PTP Time Synchronization
The Ethernet port must be connected to an IEEE 1588 grandmaster clock through a 1588 PTPv2 peer-to-peer transparent clock switch compliant with IEEE Std 1588-2008 and IEEE C37.238 power profile.
To enable 1588 Ethernet ports as a UTC source, select the IEEE 1588/PTP time synchronization option in the device Local Settings setup.
Both Ethernet ports can be connected to the same or two different 1588 switches. The device will obtain UTC time from one of the available sources and automatically switch to the other port whenever the signal is lost for any reason.
Each port implements an ordinary PTPv2 slave clock, complying with the IEEE C37.238 power profile:
22
PMU PRO Phasor Measurement Unit
3 Operating the PMU PRO
Synchrophasor Measurements
· PTP version 2. · PTP transport over Layer 2 Ethernet (IEEE 802.3), IEEE 1588-2008, Annex F. · Multicast addressing. · Peer-to-peer (P2P) path delay measurement mechanism. · Capable of working with one-step and two-step master clocks. · Responds to peer path delay requests with two-step peer delay responses
(Pdelay_Resp, Pdelay_Resp_Follow_Up).
The 1588 ports can synchronize to the PTP grandmaster clock with sub-microsecond accuracy, which can be degraded by network topology, 1588 switch accuracy, or transmission media asymmetry.
To troubleshoot connectivity issues with 1588 switches, you can monitor the PTP clock status and time accuracy via Telnet using the “ptp” command, as shown in the following picture.
3.5 Synchrophasor Measurements
3.5.1 Synchrophasor Estimation
The PMU PRO calculates and reports synchrophasor estimates as defined and described in IEEE C37.118.1 Clause 4.
Phasor estimates include three-phase voltage and current phasors, as well as positive and negative sequence voltage and current phasors, calculated from input waveforms synchronized to an absolute time reference.
All measurements are made on a common time base and related to the system nominal frequency, so the phase angle measurements are directly comparable. Differences in the actual frequency are included in the phase angle estimation
PMU PRO Phasor Measurement Unit
23
3 Operating the PMU PRO
Instrumentation Measurements
3.5.2 Frequency and ROCOF Estimation
The frequency is computed as the first derivative of the synchrophasor phase angle, and ROCOF is computed as the second derivative of the phase angle.
The frequency and ROCOF estimates are based on positive sequence synchrophasors calculated using the symmetrical components transformation. Since phase angle changes relative to the difference between the actual frequency and the nominal frequency, this approach yields the offset from nominal.
3.5.3 Reporting Rates
The PMU supports data reporting at sub-multiples or multiples of the nominal system frequency. The actual reporting rate is user selectable. Reporting rates at 50 Hz system frequency:
Frames/s 1 2 5 10 25 50 100 200
Reporting rates at 60 Hz system frequency:
Frames/s 1 2 3 4 5 6 10 12 15 20 30 60 120 240
3.6 Instrumentation Measurements
Instrumentation measurements provide real-time measurements and one-second aggregation of voltage, current, per-phase and total power and displacement power factor, and voltage and current sequence components including voltage and current unbalance.
All electrical quantities are fundamental components and are based on voltage and current synchrophasor measurements synchronized to the device’s reporting rate.
See Appendix C for a full list of the measurements provided by the PMU PRO.
3.7 Instrument Transformer Correction
Ratio and phase angle error correction can be applied to external CTs and PTs to achieve overall metering installation accuracy, or be used in any installation to optimize the accuracy of the measurement data.
The user can configure the ratio correction factor and phase angle error covering the typical transformer operating range. See 5.2.2 “Transformer Correction” for information on how to set up ratio and phase angle error correction in your device.
3.8 Operating Logic Controller
The PMU PRO logic controller can perform various actions in response to user-defined internal and external events. Unlike a conventional PLC, the PMU PRO uses a simplified setpoint-based programming technique that allows you to define a logical expression based on measured analog or digital values and perform the actions you want when the operate conditions are met.
The logic controller provides 32 setpoints with programmable operate and release delays, each of which evaluates a conditional expression with up to 8 arguments. Whenever the expression evaluates to true, the logic controller can perform up to 4
24
PMU PRO Phasor Measurement Unit
3 Operating the PMU PRO
Operating Device I/O
user-programmable actions, including data recording on a periodic basis at programmable rate and on any internal or external trigger.
See 5.2.10 “Configuring the Logic Controller” to learn how to configure the logic controller for your application.
3.9 Operating Device I/O
3.9.1 Operating Digital Inputs
The PMU PRO can monitor up to 25 digital inputs including one built-in digital input and others located on 3 expansion modules.
See 2.3 “Electrical Installation” for the connection diagrams and location of the digital input terminals.
All digital inputs have a programmable de-bounce time and can operate in static and pulse mode using normal (normally open) or inverting (normally closed) polarity. All inputs are sampled at a 1 ms rate.
See 5.2.5 “Configuring Digital Inputs” for information on how to configure the operation of the digital inputs in your device.
3.9.2 Operating Relay Outputs
Up to 13 relay outputs including one built-in relay and others located on 3 expansion modules can be provided with the PMU PRO for control and alarm indication.
See 2.3 “Electrical Installation” for the connection diagrams and location of the relay terminals. See Technical Specifications in Appendix A for relay ratings.
The relay outputs are updated at a half-cycle rate. Each relay is independently programmable and can operate in latched, unlatched, pulse or KYZ mode, and relay operations can be inverted so that the relay is energized when inactive and de-energized when activated.
The relays can be operated either locally via the logic controller or by a remote command sent through communications.
See 5.2.6 “Configuring Relay Outputs” for how to configure relay outputs in your device.
3.9.3 Operating Analog Input
The PMU PRO is equipped with one built-in optically isolated analog input with userselectable options for 0-20 mA, 4-20 mA, 0-1 mA, and 1 mA. The 0-1mA and 1mA inputs provide double overload up 2 mA.
See 2.3.4 “Built-in I/O Connection” for the connection diagram and location of the analog input terminals. See Technical Specifications in Appendix A for analog input ratings.
Updating the analog input is synchronized with the PMU reporting rate. The device automatically converts the raw analog input readings received from the A/D converter to a user-defined engineering scale and reports the input values in engineering units with configurable resolution.
See 5.2.7 “Configuring Analog Input” for information on how to set up the analog input in your device.
PMU PRO Phasor Measurement Unit
25
3 Operating the PMU PRO
Operating Recorders
3.10 Operating Recorders
The PMU PRO recorders provide recording events and data to the bult-in eMMC flash memory. Each log file can contain up to 65535 records. The behavior of a file when it is full can be chosen from two options: recording stops until the file is cleared (non-wrap mode), or recording continues over old records (wrap-around mode).
See 5.4.1 “Configuring Log Files” for information on configuring file size and behavior.
3.10.1 Operating the Event Recorder
The event recorder automatically records time-tagged self-supervision events related to configuration changes, resets, and device diagnostics.
It can also record logic controller operations configured for triggering the event recorder.
See 7.4 “Retrieving and Storing Files” for information on how to retrieve the event log file from your device.
3.10.2 Operating the Data Recorder
The PMU PRO provides eight programmable data recorders for recording data on a periodic basis and on any internal or external trigger. The recording is triggered via the logic controller.
Each data log file can be configured to record up to 16 selectable parameters per record and up to 65535 records per file.
See 5.4.2 “Configuring the Data Recorder” for information on setting up data recorders in the PMU PRO.
See 7.4 “Retrieving and Storing Files” for information on how to retrieve data files from your device.
3.11 Communicating with the PMU PRO
See 2.3.7 “Communication Connections” for connection diagrams. See 5.3 “Configuring Communication” to learn how to change the factory-set communication settings in your device.
3.11.1 Internet Communications
The PMU PRO provides Internet communications for a wide range of communication protocols used in industrial networks for real-time distributed control.
Assigning IP Addresses
You can assign IP addresses to your device’s Ethernet ports manually or allow IP addresses to be assigned dynamically by the network’s DHCP server on lease.
Use DHCP assigned addresses with caution because the leased IP address you are using may no longer be valid if the device has been disconnected from the network for a long time. If PAS is unable to reconnect to your device, check the device’s current IP address either through the front display or through PAS using a different communication port
26
PMU PRO Phasor Measurement Unit
3 Operating the PMU PRO
Communicating with the PMU PRO
Port Numbers
The PMU PRO uses the standard TCP and UDP port numbers assigned to the corresponding protocols by the Internet Assigned Numbers Authority (IANA) by default. For some protocols, such as DNP3 and IEEE C37.118.2, you can change the default port number through the protocol configuration.
TCP Communications
Keepalive Probes
The PMU PRO uses TCP keepalive probes to detect dead connections and prevent resource leaks. If the connection is idle longer than configured, the device sends a keepalive request to check if the connection is alive. If no response is received after 5 successive keepalive retransmissions, the connection is considered dead and will be closed.
The TCP keepalive idle time can be configured from 1 to 60 seconds or set to 0 to disable keepalive probes. It is set in the device to 20 seconds by default.
Client Connection Timeout
The client connection idle timeout can be used to terminate a connection if it has been inactive for too long. It can be configured from 1 to 300 seconds or set to 0 to disable the client idle timeout. It is set in the device to 120 seconds by default.
If the idle timeout is enabled, then the master station should periodically send probe requests to the device to maintain some kind of activity on the connection socket if it wants to keep the connection open.
When sending unsolicited messages via IEC 60870-5-104 or streaming data via IEEE C37.118.2, the device disables the connection timeout and uses keepalives regardless of port settings.
NOTE: To prevent resource leaks, the PMU PRO does not allow both the keepalive probes and connection timeout to be disabled. At least one of the dead connection preventers will always be active.
UDP Communications
Unlike TCP connections, streaming data via UDP does not provide tools for probing whether a client who initiated transmission is still alive and listening to the data stream.
Abnormal termination of client operations without explicitly disabling data transfer may leave the stream socket active and thus cause a resource leak. To avoid this, the device provides, where necessary, the appropriate means to manually stop active UDP streams and release the associated resources.
3.11.2 Telnet Communications
The PMU PRO comes with a Telnet server that can be used to monitor or check the current status and operations of the device. Run a “telnet” command with your device’s IP address, log into the device with your user password and enter a command from the suggested list.
PMU PRO Phasor Measurement Unit
27
3 Operating the PMU PRO
Communicating with the PMU PRO
If Windows does not recognize the “telnet” command, go to the Start menu, select “Settings”, click “System”, and then select “Optional features”. Click “More Windows features” in the top right, check the Telnet client checkbox, and click OK.
NOTE
Telnet access requires at least level 2 password authorization. After three unsuccessful login attempts, the Telent server will be shut down for one minute.
3.11.3 Modbus Communications
For detailed information about Modbus protocol implementation, register map and protocol parameter settings, see the PMU PRO Modbus Reference Guide.
Modbus protocol is used as a default communication protocol for communicating with the PMU PRO via PAS, remote configuring and monitoring the device, and retrieving event and data log files.
The Modbus TCP server uses the standard port number 502 assigned by IANA.
Modbus TCP Notification Client
The TCP notification client can establish connections with a remote Modbus/TCP server and send notification messages either on events, or periodically on a time basis.
Client connections are triggered via the logic controller.
See 5.3.3 “Configuring the TCP Notification Client” for how to configure the TCP notification client for your application.
eXpertPowerTM Client The eXpertPowerTM client provides communication with the eXpertPowerTM server SATEC’s proprietary Internet services. Connections to the eXpertPowerTM server are carried out via the Modbus/TCP protocol on a periodic basis managed by the eXpertPowerTM server.
See 5.3.4 “Configuring the eXpertPower Client” for information on configuring the eXpertPowerTM client in your device.
28
PMU PRO Phasor Measurement Unit
3 Operating the PMU PRO
Communicating with the PMU PRO
3.11.4 DNP3 Communications
For information about DNP3 protocol implementation, object reference and protocol parameter settings, see the PMU PRO DNP3 Reference Guide.
The DNP3 TCP port number is configurable via the DNP3 setup. By default, the DNP3 TCP server uses the standard port number 20000 assigned by IANA.
The PMU PRO implements Level 2 of the DNP3 communication protocol. The device does not support unsolicited requests or hardware collision avoidance.
The PMU PRO DNP3 implementation supports a wide variety of messages. The most common method of getting static object information from the device via DNP3 is issuing a read Class 0 request. The device allows free configuring the Class 0 response by assigning ranges of static points to be polled via Class 0 requests.
Any static object point is allowed to be linked to an associated event object point for generating data change events to be polled via event Classes 1, 2 or 3.
Up to 64 configurable event object points are available for generating events. The event scan period is half the power frequency cycle time for binary input points, and 200 ms for analog input and binary counter points.
3.11.5 IEEE C37.118.2 Communications
For information about IEEE C37.118.2 synchrophasor data transfer protocol implementation, communication features and protocol parameter settings, see the
PMU PRO IEEE C37.118.2 Reference Guide.
IEEE C37.118.2 TCP and UDP servers listen on standard ports 4712 and 4713, respectively, which can be changed in the protocol configuration.
The PMU PRO provides commanded client-server UDP and TCP data transmission and spontaneous UDP data transmission over IP protocol.
Synchrophasor data is transmitted at the device’s measurement/reporting rate. It is configurable through the basic device setup or through the IEEE C37.118.2 or IEC 61850 SV publisher settings.
Data Frames
Synchrophasor data frames transmit a time stamped set of measurements that include phasor estimates, frequency deviation from the nominal power line frequency or actual frequency, and the rate of change of frequency. In addition, the data frame can be expanded to contain analog data (total active, reactive and apparent power, and power factor) and digital input status information.
Phasor components can include three voltage and three current phasors, positive sequence voltage and current phasors, negative sequence voltage and current phasors, or all of them in one frame.
Complex phasor values can be sent in a rectangular coordinates format (real and imaginary) or in polar coordinates (magnitude and angle). Phasor and frequency data can be represented in 32-bit IEEE floating-point format or as 16-bit scaled integer numbers. When using the integer format, the data conversion factors are provided in IEEE C37.118.2 configuration frames.
PMU PRO Phasor Measurement Unit
29
3 Operating the PMU PRO
Communicating with the PMU PRO
Configuration Frames
IEEE C37.118.2 configuration frames provide information about the synchrophasor data stream in binary format. The PMU PRO supports CFG-1, CFG-2 and CFG-3 configuration frames.
In spontaneous UDP data transmission mode, CFG-2 or CFG-3 configuration frames can be sent spontaneously without an explicit user command. When enabled, a configuration frame is sent periodically every 30 seconds.
Commanded Unicast UDP and TCP Data Transmission
Commanded unicast data transmission is enabled and disabled by commands sent by the IEEE C37.118.2 client over a UDP or TCP connection.
The PMU PRO provides 5 data streaming slots for continuous streaming synchrophasor data over unicast UDP or/and TCP connections.
Although the UDP server responds to commands from unlimited number of UDP clients and the TCP server can support more client connections, the commands to turn on data transmission on new connections after all 5 data streams are running will be ignored.
To avoid resource leaks due to dead UDP connections, the PMU PRO provides a command the user can send to the device through the IEEE C37.118.2 configuration dialog that stops all active commanded UDP streams and releases the associated data stream slots.
You can check which data streams are currently active via Telnet using the “188” command as shown in the following picture.
Spontaneous UDP Data Transmission
Spontaneous data transmission is enabled and disabled in the device by the user via the IEEE C37.118.2 settings. The destination UDP port and IP address are user-configurable.
Spontaneous UDP data transmission operates in unicast, multicast or broadcast mode depending on the destination IP address you specified.
30
PMU PRO Phasor Measurement Unit
3 Operating the PMU PRO
Communicating with the PMU PRO
3.11.6 IEC 61850 Communications
For information about IEC 61850 protocol implementation, communication and protocol parameter settings, see the PMU PRO IEC 61850 Reference Guide and PMU PRO IEC 61850 Conformance Statement (PICS, PIXIT, MICS, TICS) documents.
The PMU PRO IEC 61850 server supports IEC 61850-8-1 client-server and GOOSE communications and IEC 61850-9-2 sampled values communications, compliant with the IEC 61850 Ed2 set of standards.
PAS provides an IED configuration tool for customizing your device’s settings and exporting IED configuration description files (CID/IID) for use with IEC 61850 client applications.
IED Configuration Description File
The IEC 61850 device data model is described in the PMU PRO IED configuration description file “PMU PRO _2007B_RevXX.icd” provided with your device. Use this as a basic template to create custom CID and IID files for use in your installation.
Client-server Communications
By default, the IEC 61850 TCP server uses the standard port number 102 assigned by IANA.
The TCP server supports up to 5 client-server associations simultaneously.
The IEC 61850 TCP server by default uses TCP keepalives to detect dead connections. The client connection idle timeout can optionally be used to terminate an idle connection, and is normally disabled.
GOOSE Communications
The PMU PRO provides a GOOSE publisher for fast distribution of measurement data over Ethernet for real-time distributed control within a substation. The GOOSE publisher settings and GOOSE dataset can be configured using PAS.
GOOSE messages are continuously retransmitted across the multicast association using a configurable multicast MAC address. The fastest retransmission time is half-cycle for the first 5 messages after an event occurred, then geometric with a time multiplier of 2 until the maximum configured retransmission time.
Sampled Values Communications
The PMU PRO provides a multicast sampled values (MSV) publisher that is responsible for streaming synchrophasor data over Ethernet. The MSV publisher settings and sampled values dataset are configurable via PAS.
Synchrophasor data transfer in the context of IEC 61850 uses IEEE C37.118 to IEC 61850 mapping mechanism defined in IEC 61850-90-5.
PMU is modeled as a dedicated logical device within the PMU PRO IED. The detailed PMU data model is shown as part of the ICD definition file provided with your device.
IEEE C37.118.2 synchrophasor data stream is mapped to IEC 61850-9-2 sampled values APDU and the data rate is synchronized with the C37.118.2 synchrophasor data rate.
PMU PRO Phasor Measurement Unit
31
3 Operating the PMU PRO
Communicating with the PMU PRO
The sampled values messages are sent at a fixed rate over a multicast association using a configurable multicast MAC address.
3.11.7 IEC 60870 Communications
For information about IEC 60870-5-101 and IEC 60870-5-104 protocol implementation, communication options, object reference and protocol parameter settings, see the
PMU PRO IEC 60870-5-101/104 Reference Guide.
The PMU PRO supports all standard ASDU types for data interrogation, event reporting and control.
TCP IEC 60870-5-104 Connections
The IEC 60870-5-104 TCP server uses the standard IANA assigned port number 2404.
Each Ethernet port can be configured for two redundant logic connections with the controlling station using two different IP addresses for spontaneous/cyclic transmission. Selecting the same IP address for both connections disables redundancy mode. In any case, both IP addresses must be defined to enable spontaneous/cyclic transmission.
Outstanding Messages
The controlling station must acknowledge receipt of information messages. The maximum number of outstanding (unacknowledged) messages allowed can be configured from 1 to 32. The default value is 12. When the number of unacknowledged messages reaches the maximum, the device pauses transmission until the controlling station sends an acknowledgement or restarts data transfer.
Interrogation
The PMU PRO supports general and group interrogation commands for binary and analog objects. Address ranges for general and group interrogation are configurable.
Up to 15 groups can be arranged for group interrogation. Any compatible ASDU data type can be separately selected for each range of points regardless of the configured default object type.
Cyclic Data Transmission
The object address range and transmission period for periodic/cyclic data transmission are configurable. The cyclic transmission period can be set from 100 to 30,000 ms, or set to 0 (default) to disable transmission.
Configured cyclic data is transmitted periodically over the active connection after the controlling station confirms start of data transfer.
Event Reporting
The PMU PRO provides up to 64 configurable setpoints for reporting events when the measured value exceeds a specified threshold or changes by a certain percentage, or the status of a binary point changes. The event scan period is half the power frequency cycle time for binary objects and 200 ms for analog objects.
Collected events are transmitted spontaneously over the active connection after the controlling station confirms start of data transfer.
32
PMU PRO Phasor Measurement Unit
Using the Front Display
Controls and Indicators
Figure 4-1 PMU PRO Controls
4.1 Controls and Indicators
4.1.1 Status LEDs
Device Status LED
The green device status LED flashes at 2 second intervals to indicate the operational status of the device as described in the following table.
Status Flashes once Flashes twice
Flashes 3 times
Description
The device is operational and works normally
Device is in boot mode. This is normal when updating the device firmware. In other cases, it indicates missing or corrupted application firmware and requires user intervention to resolve the issue.
Indicates a critical error, such as corruption of basic operating settings or a malfunction of the device clock, that requires user intervention and correction (see 3.2 “Device Diagnostics”)
Clock Sync LED
The warning LED flashes red ones a second when the PMU clock is not synchronized with the UTC time source. The LED turns off once the clock is synced to UTC time.
PMU PRO Phasor Measurement Unit
33
4 Using the Front Display
Controls and Indicators
4.1.2 Status Indicators
The graphic icons and messages in the display status bar provide additional indication about the status of the device.
Diagnostic Icon
A diagnostic icon appears in the status bar whenever diagnostic messages occur, which can be checked through the Events/Diagnostics display.
The icon will turn off once diagnostic messages are cleared, either from the display or via communication (see 3.2 “Device Diagnostics”).
Clock Sync Indicators
Clock synchronization status is indicated in the status bar with a colored message, as shown in the following table.
Message Testing
Locked Unlocked
Failure
Description
The device is trying to detect a signal from a UTC time source. When using a 1588 PTP time source, it can take up to a minute to calibrate and synchronize the device clock with the PTP master clock.
The device clock is locked to UTC time
The UTC time source previously synchronized to a GPS primary reference is now free-running based on its own internal oscillator. It is said to be in the holdover mode, as long as it is within its accuracy requirements.
The device has lost a signal from a UTC time source or the UTC master clock has lost a GPS signal
Battery Indicator
A black battery icon outlined by orange indicates that the lithium battery voltage is below the operating limit. The battery should be checked and replaced.
The icon will automatically return to its normal gray color once the battery voltage returns to normal.
Communication Activity Indicator
The communication activity indicator provides information about the activity of the serial and Ethernet ports and transmission direction. It is displayed in the status bar as two opposite horizontal arrows: orange when there is any data transfer occurring, and gray when there is no activity on the communication ports.
34
PMU PRO Phasor Measurement Unit
4 Using the Front Display
Operating the Display
Scrollbar
In a multi-page display, a scrollbar appears on the right as a line of vertical dots and shows the number of pages available. The current page position is highlighted in orange.
4.2 Operating the Display
The PMU PRO is equipped with a high contrast 1.77″ TFT color graphics display for local data read outs, device setup and servicing. The display is normally updated once per second.
4.2.1 Display Features
The PMU PRO display has configurable features that can be disabled, enabled or adjusted via the device Display Setup.
Brightness
The display brightness can be adjusted from 20 to 100%. By default, brightness is set to 50%.
Backlight
The duration of the display backlight can be adjusted from 1 to 10 minutes or it can be turned on continuously. By default, the backlight duration is set to 5 minutes.
If the display is dark, briefly press any button to turn the backlight on.
Auto-Return
When the auto-return option is enabled and no button is pressed for a configurable amount of time, the display automatically returns to the default screen from any other screen or menu.
The auto-return timeout can be adjusted from 1 to 30 minutes. The default value is 5 minutes.
Measurement Resolution
The display resolution for basic electrical quantities such as voltage, current and power can be adjusted from 1 to 3 decimal places for voltage and power, and from 1 to 2 decimal places for current. All measured data is displayed in primary units.
Multilanguage Support
The user can select one of the languages supported by the device. After changing the language, the device reboots.
Custom Name/ID
The user can optionally add a device name, username or ID, which will be displayed in the status bar. It can contain up to 9 alphanumeric characters.
PMU PRO Phasor Measurement Unit
35
4 Using the Front Display
Operating the Display
4.2.2 Navigation Buttons
The PMU PRO has four multi-function navigation buttons with functionality described in the following table.
Button
/
Function ESCAPE
UP
DOWN OK/ENTER
Operation
a) In the menu: returns to the top-level menu b) In an editable field: rejects the changes made and returns to the menu c) In the measurement display: exits the currently active screen and returns to the menu
a) In the menu: highlights a menu item to select from the top or left b) In an editable field: highlights a button on the on-screen keyboard from the top or left c) In the measurement display: scrolls forward through the screens a) In the menu: highlights a menu item to select from the bottom or right b) In an editable field: highlights a button on the on-screen keyboard from the bottom or right c) In the measurement display: scrolls backward through the screens a) In the menu: selects the highlighted menu item to display or edit b) In an editable field: confirms the changes made c) In the measurement display: adds the screen to favorites or removes it from the favorites
4.2.3 Menu Navigation
The device display is controlled through a multi-level menu with a hierarchical layout. Menu navigation is intuitive and does not require user experience. Whenever a nonstandard action is required, a text prompt appears on the display.
Use the arrow buttons and to navigate to the desired menu item or entry. The selected item is highlighted or surrounded in orange. If a submenu is marked with the icon on the left, it has a drop-down menu. Press the OK button to open the drop-down list.
Press the OK button to confirm your selection or enter the submenu. Press the ESCAPE button / to return to the top-level menu.
36
PMU PRO Phasor Measurement Unit
4 Using the Front Display
Operating the Display
In a multi-page display, a scrollbar appears on the right as a line of vertical dots and shows the number of pages available. The current page position is highlighted in orange. Use the arrow buttons and to move between pages.
4.2.4 Entering Parameters
The device provides two on-screen keyboards, numeric and alphanumeric, which allow you to change device configuration parameters in the setup menus.
Entering Numeric Values
The picture below shows what the on-screen numeric keyboard looks like.
When entering a number using the on-screen keyboard, the entered character is placed at the current cursor position, which is the first blank position after the rightmost displayed character.
Shaded buttons on the keyboard indicate buttons that are currently unavailable. If the parameter can be a negative number, then the minus sign becomes available when the cursor is in the leftmost position.
To clear the rightmost digit or the entire number, use the arrow buttons and
to highlight the “Backspace” button
on the keyboard, then press the OK button
as many times as necessary.
To enter a number, use the arrow buttons to highlight the digit you want to enter, and then press the OK button . Enter the following digits in the same way.
Once the complete number has been entered, highlight the “APPLY” button on the keyboard and press the OK button to confirm the changes.
Your input is checked for consistency and will not be accepted if the value is outside the
acceptable range. If pressing
has no effect, then the entered value is incorrect. See
“Setting Up the PMU” in Chapter 5 to determine the valid range of the parameter you
want to change.
To discard your changes and return to the menu, press the ESCAPE button / .
Entering Text Strings
The picture below shows what the on-screen alphanumeric keyboard looks like.
Text strings are entered in the same way as numbers (see above), with a choice of uppercase and lowercase letters.
To enter a text string, use the arrow buttons and to highlight the letter button on the keyboard that contains the letter you want to enter, and then press the OK button as many times as necessary until the letter you want appears in the correct case. Enter the following letters in the same way.
PMU PRO Phasor Measurement Unit
37
4 Using the Front Display
Display Menus
To enter a numeric character, change the on-screen keyboard to a numeric one by highlighting the “123” button and pressing the OK button . To return to the alphanumeric keyboard, highlight the “ABC” button and press the OK button .
Once the complete text has been entered, highlight the “APPLY” button on the keyboard and press the OK button to confirm the changes. To discard your changes and return to the menu, press the ESCAPE button / .
Entering Enumerated Values
If the selected parameter is an enumeration, it appears in a drop-down list, as shown in the following picture.
Use the arrow buttons and to highlight the parameter and press the OK button to open the drop-down list. Select the value you want using the arrow buttons, and then press the OK button to confirm your selection.
4.3 Display Menus
4.3.1 Main Menu
The main device menu is shown in the following picture. If you are in another screen, press the ESCAPE button / as many times as necessary until you return to the main menu.
38
PMU PRO Phasor Measurement Unit
4 Using the Front Display
Display Menus
The following table explains the contents of the menus.
Display Menu Menu Icon
Display Contents
Favorites Monitor
Predefined user favorite screens
Electrical measurements, device identification, communication settings, device clock and I/O status
Phasor
Real-time phasor graph
Trend Events
Real-time voltage, current, power and frequency trend graph
Device event log and diagnostic events
Setup
Device configuration setup and control
4.3.2 Favorites Menu
The Favorites menu lists the user-selected screens to display. If the Favorites menu is empty, the following message appears when you enter the Favorites menu.
The Favorites menu can contain up to 5 entries with screens that can be selected from any other menu.
To add a new screen to your Favorites menu, navigate to the screen that you want to place in the menu and that is not already listed in the menu, press the OK button , and then confirm your choice, as shown in the following picture.
PMU PRO Phasor Measurement Unit
39
4 Using the Front Display
Display Menus
To remove a screen from the Favorites menu, navigate to the screen listed in the menu, press the OK button , and then confirm the changes.
4.3.3 Monitor Menus
The monitor menu lists screens that display current electrical measurements and device status information.
All electrical measurements are one-second averages derived from the voltage and current synchrophasor vectors and synchronized with the UTC second rollover.
The following table shows the monitor menu layout.
Main Menu
Submenu Custom Screen
Voltage & Currents
Power
Monitor
Device I/O
Device Counters Communication Clock Status Device Info
Drop-down Menu
Voltage L-N Voltage L-L Currents Total Phase 1 Phase 2 Phase 3 Digital Inputs Relay Outputs Analog Input
The following sections describe the monitor screens and show what they look like.
Default Screen
The default screen is the screen that opens when you turn on the device.
The display also automatically returns to the default screen from any other screen or menu when the auto-return option is enabled and no button is pressed for a configurable amount of time.
By default, this is the screen that displays line-to-neutral voltages. You can choose any other screen as your default screen.
The example below shows how to select the line-to-line voltage screen as your default screen. See Navigating in Setup Menus for information on navigating in setup menus.
40
PMU PRO Phasor Measurement Unit
4 Using the Front Display
Display Menus
Custom Screen
The PMU PRO allows you to build your own customized screen that will be listed in the Monitor menu and will appear like any other screen. You can also set it as your default screen.
The custom screen can contain up to 4 voltage and current readings. By default, the custom screen is blank.
The example below shows how to organize your own screen. Navigate to the Custom Screen in the Display Setup menu, press the OK button twice, and then navigate to the parameters you want to display on the custom screen and check them by pressing the OK button . Then press the ESCAPE button / to return to the menu.
See Navigating in Setup Menus for more information on navigating in setup menus.
Voltage and Current Screens
PMU PRO Phasor Measurement Unit
41
4 Using the Front Display Power Screens
Display Menus
Device I/O Status
I/O status screens display the current status of the device’s digital and analog I/O. The analog input value is indicated in engineering units according to the user-configured analog input scale.
42
PMU PRO Phasor Measurement Unit
4 Using the Front Display Device Counters
Display Menus
Communication Status The communication status screen shows the activity of the communication ports. The two dots on the right flash green when data is being transferred to or from the device.
Device Clock Status This screen provides more detailed information about the device clock synchronization and current time quality. For the IRIG-B time source, no time error is reported.
Device Information These screens provide information about the device serial number and firmware versions, as well as detailed information about communication port settings.
PMU PRO Phasor Measurement Unit
43
4 Using the Front Display
Display Menus
4.3.4 Phasor Diagram
The phasor diagram provides a graphical representation of the relationship between voltage and current phasors. It can be used when installing the device to check whether the phase voltage and current wires are connected to the input terminals in the correct order.
The numbers on the right show the maximum measured three-phase voltage and current.
4.3.5 Trend Graph
The real-time trend graph represents the latest 10-minute measurements plotted at one-second intervals.
The device displays trends of voltage, current, total power and frequency, as shown in the following picture. The numbers below the graph indicate last measured values.
4.3.6 Events Menu
The Events menu allows you to view the device’s event log and device diagnostic messages.
Event records are listed in order from newest to oldest.
44
PMU PRO Phasor Measurement Unit
4 Using the Front Display
Display Menus
4.3.7 Setup Menus
Navigating in Setup Menus
Setup menus allow you to view or change the device configuration settings and perform certain management functions in your device.
The following table shows the setup menu layout.
Main Menu
Submenu General Setup
Communication Setup
Setup
Device I/O Setup Display Setup
Date & Time Reset
Drop-down Menu Basic Setup Transformer Correction COM1 Network 1 Network 2 Switch Mode Digital Inputs Relay Outputs Analog Input Custom Name/ID Default Screen Custom Screen Adjustment Language Auto-return Resolution Clock Localization DST Counters Diagnostics Logs
PMU PRO Phasor Measurement Unit
45
4 Using the Front Display
Display Menus
Password authorization is required to access the setup menus. Some setups and controls will prevent you from making any changes to the device if the security level of the password you entered does not give you the appropriate access rights.
Entering the Password
When prompted for a password, enter your password as a regular numeric value (see Entering Parameters).
Once the password is set to the desired value, highlight the “APPLY” button and press the OK button .
Viewing and Changing Parameters
Configuring the device parameters is demonstrated below using an example of changing the primary current of the input current transformers.
Navigate to the General Setup menu with the arrow buttons and , press the OK button to open the drop-down menu and select the Basic Setup entry. The selected item is surrounded in orange. Press the OK button to confirm your selection.
See 4.2.4 “Entering Parameters” for information on how to change a parameter value using the on-screen keyboard.
Once the desired value has been set and confirmed, press the ESCAPE button / to return to the top-level menu.
46
PMU PRO Phasor Measurement Unit
4 Using the Front Display
Display Menus
PMU PRO Phasor Measurement Unit
47
5 Setting Up the PMU PRO
Authorization
Setting Up the PMU PRO
See 4.3.7 “Setup Menus” for information on how to view and modify device settings via the front display. See 7.1.4 “Configuring the Device” to learn how to work with PAS.
5.1 Authorization
Password authorization is required to access the setup menus. See 3.1 “Device Security” for information on the device password security.
Using the Front Display See “Entering the Password” in 4.3.7 for information on entering the user password using the front display.
Using PAS You are prompted for the password every time you try to send new configuration data or a command to the device.
Enter the password and click OK. If your authorization was successful, you are not prompted for the password again until you close the dialog window.
5.2 General Device Settings
5.2.1 Basic Setup
These parameters provide your device with basic information about the electrical connections and device measurement options.
Using the Front Display Navigate to the General Setup menu, press the OK button to open the drop-down menu, and then select Basic Setup. Use UP and DOWN buttons to scroll through the device settings.
See the table below for the available options and acceptable parameter ranges.
48
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
General Device Settings
Using PAS Select General Setup from the Meter Setup menu and click the Basic Setup tab.
The following table lists available options.
Parameter PT Ratio
PT Secondary (L-L), V CT Primary Current Nominal Frequency Data Rate, frames/s
Phase Order L1 Current Direction L2 Current Direction L3 Current Direction Volts Scale, V
Amps Scale, A
Options 1.0-6500.0
Default 1.0
50.0-500.0
120.0 V
1-30,000 A
5 A
50, 60 Hz
60 Hz
1,2,3,4,5,6,10,12, 10 15,20,25,30,50,60, 100,120,200,240
ABC, CBA
ABC
Regular, Reverse Regular
Regular, Reverse Regular
Regular, Reverse Regular
60-600 V
144 V
1.0-20.0 A
10 A
Description The ratio of the primary and secondary windings of potential transformers (PT) Nominal secondary line-to-line voltage Primary rating of current transformers Nominal line frequency Data measurement rate and data reporting and transmission rate in IEEE C37.118.2 and IEC 61850 SV publisher Voltage phase sequence L1 current wiring direction L2 current wiring direction L3 current wiring direction Maximum voltage scale, in secondary volts (see Appendix D ) Maximum current scale, in secondary amps (see Appendix D )
PMU PRO Phasor Measurement Unit
49
5 Setting Up the PMU PRO
General Device Settings
5.2.2 Transformer Correction
Transformer correction allows you to compensate ratio and phase angle inaccuracies of the voltage and current instrument transformers.
Using the Front Display
Navigate to the General Setup menu, press the OK button to open the drop-down menu, and then select Transformer Correction.
Using PAS
Select General Setup from the Meter Setup menu and click the Transformer Correction tab.
50
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
General Device Settings
The following table lists available options.
Parameter Ratio Correction Factor Phase Angle Error
Options 0.900-1.100
+/-600 minutes
Default
Description
1.000
The ratio of the true transformer ratio to the marked ratio
0
The phase offset between the primary and
secondary values in arc minutes. The phase
angle is positive when the secondary value
leads the primary value.
5.2.3 Display Setup
Using the Front Display
Navigate to the Display Setup menu, press the OK button to open the drop-down menu, and then select the entry you want.
Using PAS
Select General Setup from the Meter Setup menu, and then click the Display Setup tab.
The following table lists available options.
Parameter Brightness Backlight time, min
Language
Volts resolution Current resolution Power resolution Auto-return, min
Custom name/ID
Options 20-100 % Continuous, 1-10 min English, Spanish, Russian, Chinese 1-3 1-2 1-3 Disabled, 1-5, 10, 15, 20, 25, 30 min
Default 50 % 5 min
English
1 1 1 5 min
Description Display brightness Backlight duration (see Backlight in 4.2.1) Display language
Number of decimal places Number of decimal places Number of decimal places Auto-return timeout (see Auto-Return in 4.2.1) Custom name or identifier shown in the display status bar (up to 9 letters)
PMU PRO Phasor Measurement Unit
51
5 Setting Up the PMU PRO
General Device Settings
5.2.4 Device Localization
Device localization allows you to adjust your time zone’s settings, including time zone offset and daylight-saving time options, and select an external time synchronization source.
Using the Front Display
Navigate to the Date & Time menu, press the OK button to open the drop-down menu, and then select Localization. Use UP and DOWN buttons to scroll through the settings.
Using PAS Select General Setup from the Meter Setup menu, and then click the Local Settings tab.
52
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
General Device Settings
The following table lists available options.
Parameter Country
Time zone offset Daylight saving time (DST)
DST Start Month DST Start Week
DST Start Day DST Start Hour DST End Month DST End Week
DST End Day DST End Hour Time synchronization input
Options Country name
-720 to 720 min Disabled Enabled
January-December First, Second, Third, Fourth, Last Sunday-Saturday 1-6 January-December First, Second, Third, Fourth, Last Sunday-Saturday 1-6 IRIG-B, IEEE 1588/PTP
Default Default -300 min Disabled
March Second
Description Indicates the location of the device. The default setting stands for the U.S.A. Local time zone offset from UTC Disabled: the clock operates in standard time. Enabled: the clock is automatically updated on daylight saving time dates Month when daylight saving time begins Week when daylight saving time begins
Sunday Day when daylight saving time begins
2
Hour when daylight saving time begins
November Month when daylight saving time ends
First
Week when daylight saving time ends
Sunday 2 IRIG-B
Day when daylight saving time ends Hour when daylight saving time ends Port receiving time sync signal
PMU PRO Phasor Measurement Unit
53
5 Setting Up the PMU PRO
General Device Settings
5.2.5 Configuring Digital Inputs
Using the Front Display
Navigate to the Device I/O Setup menu, press the OK button to open the drop-down menu, and then select Digital Inputs.
To view or configure an expansion digital input, first enter the input number.
Using PAS Select General Setup from the Meter Setup menu and click the Digital Inputs tab. Digital input numbers are assigned by the device in the order in which expansion modules are connected to the device. The built-in digital input is designated DI1. Digital inputs that are not present in the device do not appear in the setup dialog.
54
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
General Device Settings
The following table lists available options.
Parameter Input Polarity Pulse Input Mode
Pulse Polarity
Debounce Time
Options NORMAL INVERTING PULSE MODE KYZ MODE
NORMAL (N.O.) INVERTING (N.C.)
1-1000 ms
Default NORMAL PULSE MODE NORMAL
10 ms
Description Input polarity for static input
PULSE: one of the edges, leading or trailing, is recognized as a pulse KYZ: both edges are recognized as a pulse Input polarity for pulse input NORMAL: the transition from open to closed is considered a pulse INVERTING: the transition from closed to open is considered a pulse. N/A in KYZ mode The time during which the state of a digital input must not change to be recognized as a new state
5.2.6 Configuring Relay Outputs
Using the Front Display Navigate to the Device I/O Setup menu, press the OK button to open the drop-down menu, and then select Relay Outputs.
To view or configure an expansion relay output, first enter the output number. See the table below for the available options and acceptable parameter ranges.
Using PAS Select General Setup from the Meter Setup menu and then click the Relay Outputs tab. Relay numbers are assigned by the device in the order in which expansion modules are connected to the device. The built-in relay output is designated RO1. Relays that are not present in the device do not appear in the setup dialog.
PMU PRO Phasor Measurement Unit
55
5 Setting Up the PMU PRO
General Device Settings
The following table lists available options.
Parameter Operation mode
Polarity
Options
Default
UNLATHCED, UNLATCHED
LATCHED, PULSE, KYZ
NORMAL (N.O.), NORMAL
INVERTING (N.C.)
Retentive mode NO, YES
NO
Pulse width
10-3000 ms
100 ms
Description Relay operating mode (see below)
NORMAL: the relay is de-energized when inactive and energized when activated INVERTING: the relay is energized when inactive and de-energized when activated Applicable for latched relays only. NO: the relay always returns to the inactive state upon power up YES: the relay state is restored to the state it was before power loss The actual pulse width is a multiple of the half-cycle time rounded to the nearest higher value. The pause between pulses is equal to the pulse width.
56
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
General Device Settings
The relay can operate in the following modes:
· UNLATCHED: the relay goes into the active state when the control setpoint is in the active (operated) state, and returns to the inactive state when the control setpoint is released
· LATCHED: the relay goes into the active state when the control setpoint goes to the active state and remains active until it is returned to the inactive state by a local or remote command
· PULSE: the relay goes into the active state for a specified time, goes into the inactive state for the specified time and remains in the inactive state
· KYZ: the relay output state changes with each command and remains in this state until the next command
5.2.7 Configuring Analog Input
Using the Front Display
Navigate to the Device I/O Setup menu, press the OK button to open the drop-down menu, and then select Analog Input.
Using PAS Select General Setup from the Meter Setup menu and click the Analog Inputs tab. The following table lists available options.
Parameter AI type Zero scale
Full scale
Dec. Places Value label
Options
±1 mA, 0-1 mA, 0-20 mA, 4-20 mA
Up to 9 digits including decimal places (±)
Up to 9 digits including decimal places (±)
0-3
Default
Description
±1 mA User-selectable AI type.
0
Low engineering scale (in primary units)
corresponding to the minimum input
current (0 or 4 mA)
0
High engineering scale (in primary units)
corresponding to the maximum input
current (2 or 20 mA)
0
Number of decimal places in a fractional
part of the scaled value
An arbitrary name that you can give the analog input quantity
PMU PRO Phasor Measurement Unit
57
5 Setting Up the PMU PRO
General Device Settings
NOTES
1. The ±1 mA and 1 mA inputs provide double overload and are calibrated for ±2 mA and 2 mA input currents respectively, so you should set the high engineering scale for the 2 mA input current.
2. The engineering scale for the ±1 mA bidirectional analog input is symmetrical around zero. Only the high engineering scale needs to be configured.
3. Analog input labels are not stored in the device. Always save the analog inputs setup to the site database in order to keep your labels.
5.2.8 Configuring Counters
The PMU PRO has 32 nine-digit general purpose counters intended to count input pulses and events with a programmable scale factor. They can be used by the logic controller to perform repeated operations or as temporary storage.
To set up counters, select General Setup from the Meter Setup menu, and then click the Counters tab.
58
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
General Device Settings
If you are connected to a device, the left pane displays the current contents of the counters. You can preset the counters to any value you want: enter a number in the “Counter Value” field and click Send.
To link a digital input as a pulse source to a counter, select the digital input in the “Pulse Source” field, enter the pulse multiplier, then select the counter you want to link the input to and click Send. Several digital inputs can be connected to the same counter.
5.2.9 Configuring Periodic Timers
The PMU PRO provides 16 interval timers with programmable time periods from halfcycle to 24 hours. They are intended to be used by the logic controller for periodic recording and triggering time-based operations. Seven timers from #10 to #16 are factory preset, and others can be configured by the user.
To set up periodic timers, select General Setup from the Meter Setup menu and click the Periodic Timers tab.
To run a periodic timer, select the time units in which you define the period and specify a non-zero timer period. To stop the timer, set the timer period to zero.
PMU PRO Phasor Measurement Unit
59
5 Setting Up the PMU PRO
General Device Settings
5.2.10 Configuring the Logic Controller
The logic controller provides 32 setpoints with programmable operate and release delays, each of which evaluates a conditional expression with up to 8 arguments.
Whenever the expression evaluates to true, the logic controller can perform up to 4 user-programmable actions, including data recording on a periodic basis at programmable rate and on any internal or external trigger.
The scan time for all setpoints is half-cycle time
To set up the logic controller, select General Setup from the Meter Setup menu and then click the Control/Alarm Setpoints tab.
60
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
General Device Settings
The setpoint options are described in the following table.
Parameter OR/AND Input group Trigger parameter Relation
Operate limit
Options OR, AND See Appendix C See Appendix C
<=, >=, =, <>, ON, OFF, Delta, Delta+, Delta-, rDelta, rDelta+, rDeltaTrigger specific
Release limit Trigger specific
Action
See below
Target
Operate delay 0-1000.000 sec
Release delay 0-1000.000 sec
Description The logical operator for the trigger The trigger parameter group The trigger parameter used as an argument in the conditional expression The relational operator used in the conditional expression (see explanation below on using Delta operators)
The threshold (in primary units) at which the expression will evaluate to true. Not applicable for digital triggers. The threshold (in primary units) at which the expression will evaluate to false. Defines the hysteresis for analog triggers. Not applicable for digital triggers. Action taken when the setpoint expression evaluates to true The optional action target Time delay before activation of the setpoint when the operate conditions are met Time delay before the setpoint is deactivated when the release conditions are met
PMU PRO Phasor Measurement Unit
61
5 Setting Up the PMU PRO
General Device Settings
Delta operators evaluate a logical expression to true if the following conditions are met:
· Delta: the absolute difference between the last reported value and the current value exceeds the specified threshold
· Delta+ (incremental delta): the positive difference between the current value and the last reported value exceeds the specified threshold
· Delta- (decremental delta): the positive difference between the last reported value and the current value exceeds the specified threshold
· rDelta (relative delta): the absolute difference between the last tested value and the current value exceeds the specified threshold
· rDelta+ (incremental relative delta): the positive difference between the current value and the last tested value exceeds the specified threshold
· rDelta- (decremental relative delta): the positive difference between the last tested value and the current value exceeds the specified threshold
The available setpoint actions are listed in the following table.
Action SET EVENT FLAG CLEAR EVENT FLAG OPERATE RELAY RELEASE RELAY INCREMENT COUNTER DECREMENT COUNTER CLEAR COUNTER EVENT LOG DATA LOG SEND NOTIFICATION
Target 1-32 1-32 1-13 1-13 1-32 1-32 1-32 OPER, RELS, ANY 1-8 –
Description None (no action) Raise an event flag Drop an event flag Operate relay Release latched relay Increment counter Decrement counter Clear counter Event log on setpoint operation, release or both Data recording to log file Send a notification to a remote TCP server
Using Logical Operators
Logical operators OR/AND are treated in a simplified manner. They have no specific priority or precedence rules.
Any trigger condition that is ORed in a logical expression and evaluates to true overrides any preceding condition evaluated to false. Likewise, any trigger condition that is ANDed in a logical expression and evaluates to false overrides any condition evaluated to true before it.
To avoid confusion, it is recommended not to interleave different logical operators in one expression. Instead, combine all conditions that use the same logical operator on one side of the expression, and the others on the opposite side.
To explicitly override all preceding conditions with a critical trigger, place it at the end of the logical expression using an OR operator if you want the setpoint to be operated anyway when the trigger condition evaluates to true, and with an AND operator, if the setpoint should not be operated while the critical trigger is not asserted.
62
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
General Device Settings
Using Numeric Triggers
For numeric (analog) triggers, a setpoint allows you to specify two thresholds for each trigger to provide hysteresis (dead band) for setpoint operations. The operate limit defines the operating threshold, and the second release limit defines the release threshold for the trigger. Trigger thresholds are always specified in primary units.
If you do not want to use hysteresis, set the release limit to the same as the operate limit.
Using Binary Triggers
Binary (digital) triggers, as digital inputs, relays, or event flags, are tested for ON (closed/set) or OFF (open/cleared) conditions.
Binary triggers are divided into two types: static and pulsed. Static events are levelsensitive and are asserted as long as the corresponding condition exists, such as event flags or static digital inputs.
Pulsed events are edge-sensitive events with auto-reset, such as pulse inputs (transition pulses on digital inputs) and events generated by interval timers. The logic controller clears pulsed events at the end of each scan, so that triggers that used these events are prevented from being triggered once again.
Using Interval Timers
Interval timers are not synchronized with the clock. When the setpoint is activated, the timer is restarted and then generates the next timer pulse when the timer interval expires.
Using Time Triggers
Unlike interval timers, time triggers are synchronized with the device clock and can be used to program time-synchronized actions, such as synchronous recording interval data.
Delaying Setpoint Operations
Two optional delays can be added to each setpoint to extend monitoring setpoint for a longer time before making a decision on whether the expected event has occurred or not. If a delay is specified, the logic controller changes the setpoint state only if all conditions are asserted for a period at least equal to the delay time.
Setpoint Actions
When a setpoint state changes, that is, the setpoint event is either activated or deactivated, the following happens in your device:
· The new setpoint status is logged to the setpoint status register that can be monitored via communications in order to give an indication of the expected event.
· The status of the activated setpoint is latched in the setpoint alarm register, which is also accessible remotely. The register retains the last setpoint alarm state until it is explicitly cleared.
PMU PRO Phasor Measurement Unit
63
5 Setting Up the PMU PRO
Configuring Communication
· Up to four programmable actions can be performed when the setpoint is activated. Typically, setpoint actions are performed independently for each setpoint and can be repeated several times for the same target.
The exceptions are relay operations and data recording that are shared among all setpoints using an OR scheme for each target. The relay output is activated when one of the setpoints linked to the relay is activated and remains energized until all those setpoints are released.
5.3 Configuring Communication
5.3.1 Network Setup
Using the Front Display
Navigate to the Communication Setup menu, press the OK button to open the dropdown menu, and then select Network 1 or Network 2 for Ethernet port 1 and 2, respectively. Scroll through network settings using the UP and DOWN arrow buttons.
Using PAS
Select Communication Setup from the Meter Setup menu, click the Network Setup tab, and then select the network port you want to configure.
The top panel shows the currently active port settings. If you are using DHCP assigned addresses, they may differ from manually configured static network settings.
The following table lists available options.
Parameter Device IP Address Network Subnet Mask
Options
Default 192.168.0.203 255.255.255.0
Description Static port IP address Static port subnet mask
Network Default Gateway
Use DHCP
NO, YES
192.168.0.1 NO
Connection Idle Timeout, s
0 = disabled, 120 s 30-300 s
TCP Keepalive Time, s Switch Mode
0 = disabled, 1-60 Daisy Chain, Separated
20 s Separated
Static default network gateway
Enables dynamic IP address assignment by a network DHCP server Client connection idle timeout (see “Client Connection Timeout” in 3.11.1) TCP keepalive idle time (see “Keepalive Probes” in 3.11.1) Network topology
64
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
Configuring Communication
5.3.2 Serial Communications
Using the Front Display Navigate to the Communication Setup menu, press the OK button to open the dropdown menu, and then select COM 1. Scroll through serial port settings using the UP and DOWN arrow buttons.
See the table below for available communication options. Using PAS
Select Communication Setup from the Meter Setup menu and click the Serial Ports tab.
PMU PRO Phasor Measurement Unit
65
5 Setting Up the PMU PRO
Configuring Communication
To change the port settings in your device, select desired port parameters and click Send.
The following table lists available communication options.
Parameter Protocol
Interface
Options
Default
Modbus RTU, Modbus ASCII, DNP3, IEC 60870-5-101
Modbus RTU
RS485
RS485
Description Communication protocol
Fixed for the port
Device address Baud rate
Data format Response delay, ms
1-247
2400, 4800, 9600, 19200, 38400, 57600, 115200 bps
8N, 8E
0-99 ms
1 19.2 kbps
8N 5 ms
Character Timeout, ms 0-99 ms
5 ms
Device communication address Port baud rate
Data format and parity Minimum delay before sending a response message Inter-frame idle time to terminate frame reception
66
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO
Configuring Communication
5.3.3 Configuring the TCP Notification Client
The TCP notification client can establish connections with a remote Modbus/TCP server and send notification messages either on events or periodically. See the PMU PRO Modbus Reference Guide for more information on the client operation and the notification message structure.
To set up the notification client, select Communication Setup from the Meter Setup menu and click the TCP Notification Client tab.
To enable the notification client, enter the server’s IP address, change the Client Enabled status to YES, and click Send.
Client connections are initiated through the logic controller. To send event notifications, configure the setpoint to respond to the desired triggers and add the “Send notification” action to the end of the setpoint’s action list.
5.3.4 Configuring the eXpertPower Client
The eXpertPowerTM client provides communication with the eXpertPowerTM server the SATEC proprietary Internet services. Connections to the eXpertPowerTM server are handled on a periodic basis and are managed by the server.
PMU PRO Phasor Measurement Unit
67
5 Setting Up the PMU PRO
Configuring Recorders
To set up the eXpertPowerTM client, select Communication Setup from the Meter Setup menu and click the ExpertPower Client tab.
To enable the eXpertPowerTM client, change the XPW Client Enabled status to YES, and click Send. Other client attributes are set to default.
5.4 Configuring Recorders
5.4.1 Configuring Log Files
To view device file settings, click the “Memory/Log setup” icon on the PAS toolbar, or select Memory/Log from the Meter Setup menu and click the Log Memory tab.
The file size is defined for each file statically when you configure files and will not change unless you reorganize your files.
To change file properties or create a new file:
1. Double-click the file you want to configure, or select the file and click “Configure File…”.
2. Select the number of records, the number of parameters per record and the file type.
68
PMU PRO Phasor Measurement Unit
5 Setting Up the PMU PRO 3. Click OK.
Configuring Recorders
For your reference, the record size in bytes and the number of available records are shown at the top. To delete an existing file: 1. Click Delete. 2. Click OK. Save your new setup to the device database, and send it to the device.
PMU PRO Phasor Measurement Unit
69
5 Setting Up the PMU PRO
Configuring Recorders
5.4.2 Configuring the Data Recorder
The Data recorder is programmable for recording up to 16 data parameters per record in each of the data log files. The list of parameters to be recorded is configurable individually for each file.
To define the contents of the file:
1. On the Log Memory tab, select the data log file you want to configure and click “Setup Recorder”, or click the “Data Recorder” tab and select the corresponding log number.
2. Configure the list of parameters that will be written to the log file. You are not allowed to select more parameters than you specified when configuring the file. Refer to Appendix C for a list of available parameters.
3. For your convenience, PAS follows your selection and helps you configure a number of adjacent parameters: when you open the “Group” box for the next parameter, PAS highlights the same group as the previous selection. If you select this group again, PAS automatically updates the “Parameter” box with the next parameter in the group.
4. Add a name for the data log file in the “Name” box. It will appear in the data log reports and in the Logs menu to the right of the data log number.
5. Save the new setup to the device site database, and send it to the device.
70
PMU PRO Phasor Measurement Unit
Device Management
Configuring User Passwords
6 Device Management
This chapter describes how to perform management and control functions on your device using the front display and PAS. To access device management options from PAS, you must be connected to the device.
6.1 Configuring User Passwords
Changing user passwords in the device requires high-level password authorization. From the Monitor menu in PAS, select Administration -> Change Password and click the password you want to change.
Enter a new password, repeat it in the following “Confirm new password” field, and then click Send to update the password in the device.
6.2 Setting Up Device Clock
Using the Front Display Navigate to the Date & Time menu, press the OK button to open the drop-down menu, and then select Clock. Use the UP and DOWN arrow buttons to move through the clock parameters.
Using PAS
Click the Real-time Clock icon on the PAS toolbar, or select RTC from the Monitor menu. The RTC dialog displays the current PC time and the time of your device.
PMU PRO Phasor Measurement Unit
71
6 Device Management
Manual Relay Control
Click Set to synchronize the device clock with your PC’s clock. NOTE
When the device is connected to an external UTC source, your setting will be overridden by UTC time adjusted to your local time using the time zone offset. If the time reading looks out of order, check your local time zone offset in the device’s Local Settings.
6.3 Manual Relay Control
Manual relay control allows you to manually operate (connect or disconnect) output relays by sending commands though communications.
To send a remote command to a relay, select Device Control from the Monitor menu in PAS and click the Remote Relay Control tab.
Select the relay command in the Remote Command box, OPERATE to activate the relay, or RELEASE to remove the command, and then click Send.
72
PMU PRO Phasor Measurement Unit
6 Device Management
Manual Event Flag Control
6.4 Manual Event Flag Control
Event flags are normally used as temporary event storage in control expressions during logic controller operations. You can change the state of event flags using PAS to manually control the operation of the logic controller.
Select Device Control from the Monitor menu in PAS and click the Event Flags tab.
Select the flag status you want in the Status box for the corresponding event flag and click Send.
6.5 Viewing and Clearing Device Diagnostics
Device diagnostics can be cleared using the front display or communications.
Using the Front Display See 4.3.6 “Events Menu” for how to view the device diagnostic on the display. To clear the diagnostic messages, navigate to the Reset menu, press the OK button to open the drop-down menu, and then select Diagnostics. Press the OK button and then confirm your command.
PMU PRO Phasor Measurement Unit
73
6 Device Management
Clearing Counters
Using PAS Select Device Control from the Monitor menu and click the Device Diagnostics tab.
To clear the device diagnostics events, click Clear. See Appendix E “Diagnostic Messages” for the list of diagnostic messages and troubleshooting device issues.
6.6 Clearing Counters
Using the Front Display
Navigate to the Reset menu, press the OK button to open the drop-down menu, select Counters, and then press the OK button. To clear all counters, select “Reset all counters” and press the OK button. To clear a specific counter, select “Reset selected counter”, select the counter you want to reset, and then press the OK button.
74
PMU PRO Phasor Measurement Unit
6 Device Management
Clearing Files
Using PAS Select Reset from the Monitor menu and click Reset Counters.
Click “Check All” to clear all counters, or check specific counters you want to reset. Click OK, and then confirm your command.
6.7 Clearing Files
Using the Front Display You can only clear the device’s event log using the front display. Navigate to the Reset menu, press the OK button to open the drop-down menu, and then select Logs. Press the OK button and then confirm your command.
PMU PRO Phasor Measurement Unit
75
6 Device Management
Firmware Update
Using PAS
To clear all files: Select Reset from the Monitor menu, click Clear All Files, and then confirm your command.
To clear the device’s event log file: Select Reset from the Monitor menu, click Clear Event Log, and then confirm your command.
To clear the device’s data log: Select Reset from the Monitor menu and click Clear Data Log.
Click “Check All” to clear all data log files, or check specific files you want to clear. Click OK, and then confirm your command.
6.8 Firmware Update
You can update device firmware using any communication port installed in your device. Firmware update is only supported via Modbus RTU/ASCII serial protocol and Modbus/TCP protocol. If you are connected to the device via a serial port, make sure it is in Modbus mode. To download a firmware file to your device: 1. Make sure the On-line icon on the PAS toolbar is checked, select Flash Downloader
from the Monitor menu, and confirm changes. 2. Point to the firmware update file, click Open, and then confirm the device update.
76
PMU PRO Phasor Measurement Unit
6 Device Management
Firmware Update
3. When asked for the password, enter your user password, and click OK.
4. Wait until download completes and the device reboots before any further manipulating with your device.
NOTE After rebooting, your Internet connection may be temporarily lost. You may have to wait a while before PAS reestablishes a connection with your device.
PMU PRO Phasor Measurement Unit
77
7 Using PAS
Working with PAS
7 Using PAS
The free PAS software is a configuration and data acquisition tool that allows you to configure the PMU PRO and monitor your devices online through any communication port.
7.1 Working with PAS
7.1.1 Software Installation
Use PAS V1.5.5.41 or higher to access all PMU PRO features. Use the installation package included with the device or download it from www.satec-global.com.
PAS 1. Log into Windows as an administrator or make sure you have administrator rights.
2. Make sure there is no a SATEC USB device connected to your computer.
3. Open File Explorer, locate the PAS installation package, click setup.exe with the right mouse button and select Run as administrator (your administrator privileges do not yet grant the programs you run the permission to make changes in the system folders).
4. Follow installation instructions on the screen.
PAS is installed to the C:Pas folder by default, but you can change the installation folder during installation. Do not install PAS to the C:Program Files folder to avoid possible ODBC security issues.
Once the installation is complete, a PAS icon appears on your Desktop. Double-click the PAS icon to launch PAS.
7.1.2 Creating a Site for your Device
PAS stores all communication and configuration data for your device in a configuration database called a site database. During setup, store all setup data in the site database so that PAS recognizes device properties regardless of whether you are connected to the device or work offline.
To create a new site database for your device:
1. Click the Configure Device icon on the PAS toolbar or select Configuration from the Tools menu.
2. In the Site field, select the site that you want to use as the template for your new site’s database. For the first time it will be “DefaultSite”.
3. Click Sites on the right.
4. In the “Look in” field, select the folder where the new database will be saved. By default, this is the “Sites” folder. Enter the site name for your device in the “File name” field. Click New to create an empty database, or click Duplicate to create a copy of the template database, and click OK.
5. In the “Model” field, select “PMU PRO” as the device type of your PMU PRO. PAS automatically configures the appropriate settings for your device.
6. If you want to add any comments, type them into the “Comment” field. 7. Click OK.
78
PMU PRO Phasor Measurement Unit
7 Using PAS
Working with PAS
Once created, your new site appears in the site list in the PAS toolbar and in the PAS site tree. Check the “View/Hide Site Tree” icon on the PAS toolbar to open the site tree pane on the left. Uncheck it if you want to hide the site tree.
You can organize the site tree so it is convenient for you to work with it. All sites are organized into groups. You can rename existing groups or create new groups and drag sites from one group to another with the mouse.
Click the plus sigh to the left of the tree items to expand the tree and view device configuration information.
PMU PRO Phasor Measurement Unit
79
7 Using PAS
Working with PAS
7.1.3 Setting up Communications
To configure communications with the PMU PRO:
1. Click the Configure Device icon on the PAS toolbar or select Configuration from the Tools menu. In the Communication group, select the type of connection to your device.
2. Set the device address you assigned to the PMU PRO port.
3. In the “Sampling Rate” field, select the rate at which PAS will update data on your screen when continuously polling the device in the PAS Data Monitor.
The communication protocol and port settings must match the settings you have made in your device.
Communicating via the Internet
To communicate through the device’s Ethernet ports, you should provide the IP address of your device on the IP network.
1. On the Instrument Setup tab, select Internet Site and click the Connection tab.
2. Click “IP address” and enter the IP address of your device. The device’s factory set IP addresses are 192.168.0.203 for Ethernet 1 and 192.168.0.204 for Ethernet 2 ports.
3. In the “Protocol” field, select the communication protocol for the TCP port. The port number is configured automatically using the standard numbers assigned by IANA.
4. In the “Wait for answer” field, configure the amount of time PAS waits for a connection before reporting an error and the number of retries to receive a response from the device if communication fails.
80
PMU PRO Phasor Measurement Unit
7 Using PAS
Working with PAS
Communicating via USB On the Instrument Setup tab, click USB Port, and then click OK.
Serial Communications To configure your PC’s serial port: 1. On the Instrument Setup tab, select Serial Port/Modem Site and click the Connection tab. 2. In the Device field, select the PC’s COM port to which your device is connected and click Configure.
3. Select the baud rate and data format that match your device settings and click OK. To configure the communication protocol: 1. On the Connection tab, click Protocol.
PMU PRO Phasor Measurement Unit
81
7 Using PAS
Working with PAS
2. In the “Protocol” field, select the serial protocol that matches your device’s port settings.
3. In the “Response Timeout” box, define the maximum time PAS should wait for the device to response before declaring a failure.
4. In the “Break Timeout” box, define the maximum amount of idle time between frames that PAS must wait after receiving the last character to complete receiving the frame. If you have many applications running on your computer, PAS may not be able to respond quickly enough to received characters and may drop the communication while the device is still transmitting. If you frequently receive the “Communication error” message, try increasing the Break Timeout. This time is added to the message transmission time, and increasing it too much can slow down communications.
5. In the “Retries” box, define the number of retries PAS should use to obtain a response from the device if communication fails.
7.1.4 Configuring the Device
Configuring Setups
You can prepare configuration data for your device offline without having to connect it to your computer.
NOTE
If the On-line icon on the PAS toolbar is checked, PAS will attempt to connect to the device and obtain the current configuration settings as soon as you open the settings tab. If you are not connected to a device, uncheck the On-line icon before entering the setup.
To prepare the setup for your device:
1. Select a device site from the list box in the PAS toolbar or from the PAS site tree.
2. Select the desired setup group from the Meter Setup menu and click the setup tab you want to create or modify. Alternatively, click the plus sigh to the left of the site name in the site tree to expand the tree and click the desired setup.
3. Fill in the fields with new configuration data.
82
PMU PRO Phasor Measurement Unit
7 Using PAS
Working with PAS
4. Click the “Save as…” button to save the new data to the device site database and click OK.
Always configure and save Basic Setup data in the site database first. PAS uses this data as a reference when configuring other setups. To save your setup to another site’s database: 1. Click the “Save as…” button. 2. Select the target database and click OK. You can also reuse a setting from another site by copying it to the current site’s database. To copy a setup from another site’s database: 1. Click Open.
2. Select the source site database and click OK.
3. The open setup is copied to your dialog window. 4. Click the “Save as…” button. 5. Select the target database and click OK. To copy all settings from one site’s database to another site’s database: 1. In the site list on the PAS toolbar, select the source device site from which you want
to copy settings. 2. Select “Copy to…” from the Meter Setup menu. 3. Select the target site database where you want to copy the settings and click OK.
Downloading Setups to the Device
You can update each setting in your device individually, or download all settings together from the site’s database. Individual Download
To update a specific setup in your device:
1. Make sure the On-line icon on the PAS toolbar is checked. 2. Select the device site from the site list on the PAS toolbar.
3. Select the desired setup group from the Meter Setup menu. Click the tab of the setup you want to download to the device. When the setup dialog opens, PAS retrieves and displays the device’s current settings.
4. If you want to download the settings saved in the site database, click Open and then click OK, otherwise fill in the fields with new configuration data before sending to the device, and then click Send.
Batch Download
To download all settings to the device at once:
1. Make sure the On-line icon on the PAS toolbar is checked. 2. Select the device site from the site list on the PAS toolbar.
PMU PRO Phasor Measurement Unit
83
7 Using PAS
Monitoring Devices
3. Select Download Setups from the Meter Setup menu.
Uploading Setups to PC Individual Upload To get a specific setup from your device:
1. Make sure the On-line icon on the PAS toolbar is checked. 2. Select a device site from the site list on the PAS toolbar, and then select the desired
setup group from the Meter Setup menu. 3. Click on the tab of the setup you want to read from the device. When the dialog
opens, PAS retrieves and displays the current setup data from the device. Click Receive if you want to get the device settings again. 4. Click “Save As” to save the settings to the device site database, and then click OK. Batch Upload To upload all settings from the device to the device site database at once:
1. Check the On-line icon on the PAS toolbar. 2. Select the device site from the site list on the PAS toolbar. 3. Select Upload Setups from the Meter Setup menu.
7.2 Monitoring Devices
7.2.1 Viewing Device Properties
To view device properties, make sure the On-line icon on the PAS toolbar is checked, check the site tree icon on the PAS toolbar to open the site tree pane, right-click the device’s site name, and select Properties.
The Properties window provides information about the device’s factory settings and device firmware.
84
PMU PRO Phasor Measurement Unit
7 Using PAS
Monitoring Devices
7.2.2 Viewing Real-time Data
To get real-time data from your device:
1. Make sure the On-line icon on the PAS toolbar is checked.
2. Select the device site from the site list on the PAS toolbar.
3. Click the Data Monitor icon on the PAS toolbar to select the default dataset for your device. To select a different dataset, click the down arrow on the Data Monitor icon to open the drop-down list, and then select the dataset you want to view. Alternatively, point to “RT Data Monitor” in the Monitor menu and select a dataset from the list.
Real-time data is updated on the screen at the rate you set in the Instrument Setup.
Organizing Datasets
RT Data Monitor supports 32 programmable datasets, each containing up to 40 data parameters. Set #0 is intended for devices with a limited number of low-resolution scaled parameters and is not recommended for use with the PMU PRO.
PAS comes with a number of preconfigured datasets and others are blank. To reorganize datasets, select RT Data Sets from the Monitor menu, or open the dataset you want to modify and click the Data Set icon on the window toolbar.
The example below shows a user-configured dataset for monitoring synchrophasor data.
PMU PRO Phasor Measurement Unit
85
7 Using PAS
Monitoring Devices
Polling Devices
To start polling data, click the Poll icon or Continuous Poll icon on the window toolbar. Click the Stop icon to stop polling.
You can open as many monitor windows as you like, either for different sites, or for the same site using different datasets. The open monitor window is tied to the selected device site and does not change when you select another site in the PAS site list.
The acquired data can be viewed in a tabular form or graphed as a data trend. The picture below shows a typical data monitor window.
Polling Options
To change polling options, click the Options icon on the window toolbar, or right-click on the data monitor window and select Options.
If you check “Do not stop on errors”, polling resumes automatically when a communication error occurs, otherwise polling stops until you manually restart it.
Viewing Data Table
See “Working with Tables” in 7.5.2 to learn about the options available in data tables.
PAS can display a single data record or multiple records in consecutive rows. To change the view, right-click on the data monitor window and select either “Wrap” to see a single record, or “UnWrap” to go to the multi-record view.
The pictures below show what the data monitor windows look like in non-wrapped and wrapped views.
86
PMU PRO Phasor Measurement Unit
7 Using PAS
Monitoring Devices
The number of records in the multi-record window can be selected on the Options tab: click the Options icon on the window toolbar, or right-click on the data monitor window and select Options. When the number of records received exceeds the number of rows in the window, the window scrolls up and older records are erased.
PMU PRO Phasor Measurement Unit
87
7 Using PAS
Viewing Event and Data Files Online
Viewing Data Trend
To view a data trend, click the Data Trend icon on the window toolbar. See “Working with Graphic Windows” in 7.5.2 for the options available in graphic windows.
Saving Data to a File
To save the acquired data to a file, click the Save icon on the PAS toolbar, select a database or enter a name for the new database, and then click Save.
To avoid confusion, do not store data files into the “Sites” folder where site databases are located.
Printing Data
To see how the report will look when printed, select Print Preview from the File menu.
To print the data table or trend graph, click the Print icon on the PAS toolbar, select a printer, and click OK.
7.2.3 Real-time Data Logging
PAS can automatically save data records to a database while updating data on the screen.
To set up real-time logging:
1. Make sure the On-line icon on the PAS toolbar is checked.
2. Open the Data Monitor window.
3. Click the “RT Logging On/Off” icon on the window toolbar, or select “RT Logging Options” from the Monitor menu.
4. Select a database or enter a name for the new database and select the folder where you want to save it.
5. Select the number of tables and the number of records in each table that you want to record.
6. Set the file update rate for automatic recording. It must be a multiple of the sampling rate defined in the Instrument Setup dialog.
7. Click Save.
When you start real-time data polling, PAS automatically saves the acquired records to the database at the rate you specified.
In order for data to be recorded continuously, the “RT Logging On/Off” icon on the window toolbar should be checked at all times. You can pause logging by unchecking this icon, and then resume logging by checking it again.
7.3 Viewing Event and Data Files On-line
It may be useful to retrieve and view a portion of the recorded data from your device online while you wait for new logged data to appear.
88
PMU PRO Phasor Measurement Unit
7 Using PAS
Retrieving and Storing Files
PAS can retrieve recorded data from a specific log file without storing it to a file on your PC. The data is only displayed in a window on your screen. You can then manually save it to the database if you wish.
To view recorded data online, open the Logs menu and select the file you want to view. Only new records that have not yet been read through this communication port will be retrieved. Click the Poll icon on the window toolbar to read more new records.
If you want to view the entire log file from the beginning, click the Restore log icon , and then click the Poll icon .
7.4 Retrieving and Storing Files
PAS can retrieve recorded event and data files from your devices and save them to files on your PC in Microsoft Access database format on demand or periodically using the Upload Scheduler that can automatically retrieve data on a predefined schedule.
Unless you change the target database location, new data is added to the same database, so you can store long-term data profiles in the same database regardless of the upload schedule you selected.
7.4.1 Retrieving Files on Demand
To retrieve recorded files from your device: 1. Select a device site from the list box on the PAS toolbar.
2. Make sure the On-line icon on the PAS toolbar is checked.
3. Click the Upload Log Files icon on the PAS toolbar or select Upload Logs from the Logs menu.
4. Select a database or enter a name for the new database, and select the folder where you want to save it.
5. Click “Select Logs” and check the boxes for the files you want to retrieve from the device.
PMU PRO Phasor Measurement Unit
89
7 Using PAS
Retrieving and Storing Files
6. To retrieve data starting with a known date, check the “From” box and select the start date for retrieving data.
7. To retrieve data recorded before a known date, check the “To” box and select the latest date to retrieve data.
8. Click OK.
7.4.2 Using the Upload Scheduler
To set up the Upload Scheduler:
1. Click the Upload Scheduler icon on the PAS toolbar or select Upload Scheduler from the Logs menu.
2. Click Add Site, specify the site database for which you want to organize the schedule, and click OK.
3. Click Browse and select a database to store the retrieved data, or enter a name for the new database, select the folder where you want to save it, and click OK.
4. Click Configure or double-click the site row.
5. Select a daily, weekly or monthly schedule, and set the start time. If you want to upload data periodically at predefined intervals, click “Periodic” and specify the time period in hours and minutes.
6. Select the number of attempts to upload data in case of temporary connection problems or when your device is unavailable, as well as the delay between attempts in minutes and seconds.
90
PMU PRO Phasor Measurement Unit
7 Using PAS
Retrieving and Storing Files
7. Click Select Logs, check the boxes for the files you want to upload on a schedule, and click OK.
8. Check the Enabled box on the left to activate the schedule and click Close to save it.
For the scheduler to work, the On-line icon on the PAS toolbar must always be checked. If you uncheck it, the scheduler stops running. This does not result in any data loss, because the scheduler will resume as soon as you check this icon again.
Pausing the Scheduler
To pause the Upload Scheduler, check the Suspend Scheduler box on the right. To activate the Upload Scheduler, leave this box unchecked.
Running the Scheduler on Demand
You can start the scheduler at any time outside of your schedule by checking the Start Now box on the right. Once the upload is complete, the box is automatically unchecked.
Reviewing Upload Issues
If the Upload Scheduler fails to retrieve data, some data is missing, or another problem occurs, it writes an error message to the system log file. Select System Log from the View menu to review this file.
7.4.3 Exporting Files
PAS can convert data log files stored in mdb databases into Microsoft Excel workbook, CSV, or Electrotek PQDIF file format.
The first and second rows of the Excel and CSV tables list the data names and data codes that identify the recorded data and are intended for use in automated table processing. Each table row is tagged with a device ID, which you can specify in the Instrument Setup tab when creating the site database for your device.
PQDIF file names contain the site name followed by the timestamp of the first event recorded in the file, and may look like SITE-NAME_20240125T133038.pqd.
Manual Converting Files To manually convert a data log file: 1. Click the Export icon on the PAS toolbar. 2. Open the database and data table you want to export. 3. Select the folder where you want to store the exported files, enter a file name that identifies your files, select the desired output file format, and click Save. PQDIF files are usually written in a compressed format. If you do not want your files to be compressed, uncheck the Compress box before saving the file.
Automatic Converting Files To automatically convert data files using the Upload Scheduler: 1. Open the Upload Scheduler. 2. Select the site row with the left mouse button and click Export.
PMU PRO Phasor Measurement Unit
91
7 Using PAS
Retrieving and Storing Files
3. Check the Enabled box for the data log file that you want to automatically convert during upload.
4. Highlight the “Record to…” row for the selected table and click Browse. 5. Select the folder where you want to save the converted files, enter a file name for
the converted files, select the output file format, and click Save.
6. Repeat the same for all data log files you want to convert.
7. Click OK.
7.4.4 Archiving Files
Microsoft Access databases tend to grow quickly. Files larger than 0.5 Gigabytes can drastically slow down database performance
Documents / Resources
![]() |
SATEC PMU230 PMU PRO Phasor Measurement Unit [pdf] Instruction Manual PMU230, PMU230 PMU PRO Phasor Measurement Unit, PMU PRO Phasor Measurement Unit, Phasor Measurement Unit, Measurement Unit |