rotronic CP11 Handheld Instrument for CO2 Humidity and Temperature Instruction Manual
GENERAL DESCRIPTION
The CP11 is a precise, portable instrument that can display and save relative humidity, temperature and CO2 content and calculate parameters such as dew point or wet bulb temperature.
PROGRAMMING
Most settings, such as clearing the memory, CO2 alarm, the units (OC /OF), sampling rate and real time clock can be changed with the function keys.
POWER SUPPLY
The CP11 is battery powered and requires 4 AA alkaline batteries or a 5 VDC adapter (optional accessory). To insert or change the batteries, use the enclosed screwdriver. Unscrew the two screws and insert the batteries correctly. Ensure contact with the battery poles.
CONNECTION TO PC OR LAPTOP
Connect the CP11 to a PC or laptop using a mini USB cable. The stored data can then be downloaded easily with the HW4 software. The software can be downloaded free of charge from www.rotronic.com.
FUNCTION KEYS
DISPLAY
- Top LCD line: Humidity/Temperature/CO2 measurement
- Bottom LCD line: Real time clock
- CO2: Carbon dioxide measurement
- HOLD: The measurements are halted unchanged
- MIN/MAX: Minimum and maximum measurements
- AVG: Average measurements
: Low battery
- DP: Dew point temperature
- TEMP: Ball temperature
- WBT: Wet bulb temperature
- %RH: Unit of relative humidity
- °C/°F: Celsius/Fahrenheit (temperature)
- CAL: In calibration mode
- REC: In manual/automatic logging state
- RECALL: Recall mode in manual recording
USE
DATE/TIME SETTINGS
When the instrument is on, press for more than 2 seconds to enter setting mode. Press
or
to select the program P80 (rtc) and press
to start the program.
First select the time format by pressing or
. You can select between 12h and 24h mode and confirm this by pressing the key
. Then select the date format by pressing
or
. You can display the date either as month-day-year or as day-month-year. Press
to confirm your selection. Then set the year, month, day, hours, minutes and seconds. Press
or
to set the values and press
to confirm them. Press
to return to normal mode.
SETTING OF THE UNITS
When the instrument is in operation, press for more than 2 seconds to enter setting mode. Press
or
to select the program P50 (Unit) and press
to start the program. Press
or
to select °C or °F and press
to confirm your selection. Press
to return to normal mode.
TAKING MEASUREMENTS
The instrument starts measuring when it is on and refreshes the measurements every second.
If the environment changes, it takes 30 seconds before the CO2 sensor reacts. NOTE: Keep the instrument away from faces as exhalation can change the CO2 value.
HUMIDITY
Press to view the humidity. The bottom line of the display shows the real time clock.
TEMPERATURE, DEW POINT, WEB BULB TEMPERATURE
Press to view the temperature. The bottom line of the display shows the real time clock.
CO2 (CARBON DIOXIDE)
Press to view the CO2 content. Users see the CO2 measurement in ppm on the main display. The bottom line of the display shows the real time clock.
DATA HOLD
When you are in the normal display, press to stop the measurements. The “HOLD” symbol flashes at the top left of the LCD screen. All measurements taken so far remain unchanged. Press
again to cancel the “HOLD“ function. NOTE: The hold function is not available in Min/Max/ AVG mode.
BACKLIGHT
The backlight is activated by pressing any key for 10 seconds.
MIN, MAX, AVG, CURRENT VALUES
This instrument enables you to check the minimum, maximum, average and current values as soon as you press . By pressing
in normal mode, the MIN, MAX, AVG and current values are shown one after the other. Press
to return to normal mode. In this mode, the corresponding measurements and time that this mode has been active are shown. This mode can be active for up to 18 hours.
You can change the parameter shown by pressing .
ALARM
The instrument sounds an audible alarm when the CO2 limit is exceeded. (You can set the respective limit values in the instrument’s settings.) It beeps (approx. 80 dB) when the CO2 content exceeds the set limit, and only stops again when the measurements drop below the set value. It beeps again when the value exceeds the limit again.
AUTO POWER OFF
The instrument switches itself off automatically when it has not been used for 20 minutes. To override this function, first switch the instrument off. Then press and
hold and for 2 seconds and switch on the instrument until an “n“ appears.
MANUAL RECORDING
The instrument has a 99 data point memory. In normal mode or hold mode, press to start recording. The
symbol and main display light up three times. Up to 99 measured values can be stored for later viewing directly on the instrument’s display. The log data contains all information on the parameters, not only of the one being shown at the moment. When the 99 data point memory is full, the message “FULL” appears on the screen. If you feel that the readings are changing too quickly, you can press
to stop the current recordings and then record the data manually. The function is not available in min/max mode.
99 SETPOINT MEMORY
Press and hold in normal mode or in hold mode for more than 2 seconds until the
symbol appears. Press
or
to scroll through the contents. The log sequence number is shown on the main LCD display first, then the readings. The time in memory mode shows how long readings were recorded.
DATA LOGGING
The instrument can record the humidity, temperature and CO2 content automatically and monitor the environment over the long term. The memory has a capacity of 6000 points for every parameter. Users can set the sampling rate from 1 second to 4 hours, 59 minutes and 59 seconds. The factory defaults are 30 seconds.
To set the sampling rate while the instrument is on, press and hold for more than 2 second to enter setting mode. Press
or
to select the program P60 (rAtE) and press
to start the program.
Press or
to select the hours, minutes and seconds of the sampling rate and press
to confirm your selection. Press
to return to normal mode. When you have set the sampling rate, press
and hold for 2 seconds in normal mode to start the logging program. The
symbol appears to show the logging status. The main LCD display shows the real time value. The bottom line of the display shows the real time clock.
Repeat the above to create another log. When the log memory is full, the message “FULL” appears on the screen. Press and hold for 2 seconds to close the data log. The
symbol goes out. Press
to enter the mode of your choice. Except for the mode selection function, auto power off, min/max, manual recording, and the hold and memory function are not available during logging.
INSTRUMENT SETTINGS
When the instrument is on, press and hold for more than 2 seconds to enter setting mode.
Press to return to normal mode. Press
or
to select the program and press
to start it. The programmable settings light up on the display. Press
or
to select the settings and press
to confirm your selection. If you wish to return to the main menu without saving your selection, press
.
P 10, clear 99 data point memory |
|
Note: “![]() |
P 11, select yes or no |
|
|
P20, delete auto log data |
|
Note: “ ![]() |
P21, select yes or no |
|
|
P30, CO2 alarm settings |
|
Note:
|
P31, select between 1000 and 9900 |
|
|
P40, CO2 ABC settings |
|
Note:
|
P41, ON or OFF |
|
|
P70, pressure equalization |
|
Note:
|
P71, select between 700 and 1990 hpa |
|
BATTERY INDICATOR
The battery indicator lights up when the battery level is low. Please replace the batteries to ensure accurate measurement. For long-term measurements and data records, use of an adapter is recommended (optional accessory: AC1214).
HUMIDITY CALIBRATION
The instrument can be calibrated either with a 35 % or 80 % salt solution or a humidity generator. It is recommended that the ambient conditions are at 25 °C. Place the measurement probe in a 35 % salt solution while the instrument is off. Press and hold ,
,
simultaneously for 3 seconds to enter calibration mode. Press
or
to select the 35.0 % calibration and press
to stop it. “CAL“ and the calibration value light up on the LCD screen.
WARNING: Do not calibrate the humidity without the standard calibration salt because otherwise serious damage could result. For the calibration salt or other services, contact your dealer.
CO2 CALIBRATION
The sensor has an automatic calibration function to ensure long-term accuracy. If you nevertheless wish to calibrate the sensor, press and hold ,
,
simultaneously for 3 seconds in off-state to enter calibration mode. Press
or
to select 400 ppm calibration (outdoor air calibration).
Press to enter the mode. “CAL“ and the CO2 value light up on the LCD screen. Wait for about 10 minutes until the display stops flashing and calibration is finished.
TROUBLESHOOTING
Error |
Message |
Solution |
E01 |
The CO2 sensor is not working | Switch the instrument off and back on again |
E33 |
The CO2 sensor is not working | Repeat the CO2 calibration |
E02 |
The measured value is too low | Place the device in a normal environment |
E03 |
The measured value is too high | Place the device in a normal environment |
E04 |
The dew point and wet bulb temperature cannot be determined because the necessary values for calculation were not measured correctly. (DP, WB) | Correct the error |
E11 |
RH calibration error | Repeat the humidity calibration |
E31 |
Temperature sensor or AD damaged | Send the instrument in for repair |
E32 |
Memory IC damaged | Send the instrument in for repair |
E33 |
RH sensor or circuit damaged | Send the instrument in for repair |
TECHNICAL DATA
Humidity / Accuracy | @23 °C (±0.5 K) ±2 %rh (10…90 %rh), others ±5 %rh |
Temperature / Accuracy | @23 °C (±0.5 K) ±0.3 K |
CO2 accuracy | 0…9999 ppm/+-(30 ppm+5 % of the measurements) @0~5000 ppm |
Storage and transmission | -20…60 °C/10…90 %RH, non-condensing |
Operating limit of the electronics | 0…50 °C for CO2, -20…60 °C for the other parameters, non-condensing |
Memory | 99 manual readings, 18000 readings in auto logging |
Size, mm | 234 (L) x 77 (W) x 42 (H) |
Weight | 200 g |
Battery | 4 AA alkaline batteries |
Power supply | 5 VDC, 250 mA (available as optional extra AC1214) |
PC connection | Mini USB port |
OPTIONAL ACCESSORIES
Order code | |
AC1214 |
5 VDC universal adapter |
DELIVERY PACKAGE
- CP11
- Mini USB cable
- Screwdriver
Website: www.rotronic.com
Documents / Resources
![]() |
rotronic CP11 Handheld Instrument for CO2 Humidity and Temperature [pdf] Instruction Manual CP11, Handheld Instrument for CO2 Humidity and Temperature, Instrument for CO2 Humidity and Temperature, CO2 Humidity and Temperature Instrument |