DG5000 Arbitaray Waveform Generator
“
Specifications
- Distributor: Lambda Photometrics Limited
- Address: Lambda House Batford Mill Harpenden Herts AL5 5BZ
United Kingdom - Email: info@lambdaphoto.co.uk
- Website: www.lambdaphoto.co.uk
- Telephone: +44 (0)1582 764334
- Guaranty: RIGOL approval required
- Certifications: Conforms to national and industrial standards
in China, ISO9001:2015, ISO14001:2015
Product Usage Instructions
Safety Requirement
General Safety Summary:
- Use only the exclusive power cord designed for the
instrument. - Ensure proper grounding of the instrument.
- Observe all terminal ratings.
- Use proper overvoltage protection.
- Do not operate without covers.
- Avoid inserting objects into the air outlet.
- Use the proper fuse.
- Avoid circuit or wire exposure.
- Do not operate the instrument with suspected failures.
- Provide adequate ventilation.
- Avoid operation in wet conditions.
- Avoid operation in an explosive atmosphere.
- Keep instrument surfaces clean and dry.
- Prevent electrostatic impact.
- Handle with caution.
- Use the front-panel BNC output connectors properly for signal
output only.
To Prepare for Use
Follow these steps to prepare for using the product:
- Ensure the instrument is placed on a stable surface.
- Connect the exclusive power cord to the instrument and a power
outlet. - Check that all necessary accessories are included.
To Use the Built-in Help System
Refer to the manual for detailed instructions on utilizing the
built-in help system.
To Use the Security Lock
Instructions on using the security lock can be found in the
manual. Ensure to follow them carefully for proper usage.
Parameter Setting Method
The product offers two methods for setting parameters:
- Setting parameters with the front-panel keys and knobs.
- Setting parameters using the touch screen interface.
FAQ
Q: How do I contact RIGOL for support?
A: You can reach RIGOL via email at service@rigol.com or visit
their website at www.rigol.com.
“`
Distribution in the UK & Ireland
Lambda Photometrics Limited Lambda House Batford Mill Harpenden Herts AL5 5BZ United Kingdom E: info@lambdaphoto.co.uk W: www.lambdaphoto.co.uk T: +44 (0)1582 764334 F: +44 (0)1582 712084
Guaranty and Declaration
Copyright © 2024 RIGOL TECHNOLOGIES CO., LTD. All Rights Reserved. Trademark Information RIGOL®is the trademark of RIGOL TECHNOLOGIES CO., LTD. Software Version Software upgrade might change or add product features. Please acquire the latest software version from RIGOL website or contact RIGOL to upgrade the software. Notices
· RIGOL products are covered by P.R.C. and foreign patents, issued and pending. · RIGOL reserves the right to modify or change parts of or all the specifications and pricing
policies at the company’s sole decision. · Information in this publication replaces all previously released materials. · Information in this publication is subject to change without notice. · RIGOL shall not be liable for either incidental or consequential losses in connection with the
furnishing, use, or performance of this manual, as well as any information contained. · Any part of this document is forbidden to be copied, photocopied, or rearranged without prior
written approval of RIGOL.
Product Certification RIGOL guarantees that this product conforms to the national and industrial standards in China as well as the ISO9001:2015 standard and the ISO14001:2015 standard. Other international standard conformance certifications are in progress. Contact Us If you have any problem or requirement when using our products or this manual, please contact RIGOL. E-mail: service@rigol.com Website: http://www.rigol.com
Section
Description
Page
1 Safety Requirement ……………………………………………………………………………………………1
1.1 General Safety Summary …………………………………………………………………………………………………… 1 1.2 Safety Notices and Symbols ……………………………………………………………………………………………… 1 1.3 Measurement Category ………………………………………………………………………………………………………2 1.4 Ventilation Requirement ……………………………………………………………………………………………………. 3 1.5 Working Environment ………………………………………………………………………………………………………… 3 1.6 Care and Cleaning ………………………………………………………………………………………………………………. 4 1.7 Environmental Considerations …………………………………………………………………………………………..5
2 Document Overview …………………………………………………………………………………………. 6 3 General Inspection ……………………………………………………………………………………………..7 4 Product Overview ……………………………………………………………………………………………….8
4.1 Appearance …………………………………………………………………………………………………………………………..8 4.2 Dimensions ………………………………………………………………………………………………………………………….. 9 4.3 Front Panel Overview ………………………………………………………………………………………………………..10 4.4 Rear Panel Overview ………………………………………………………………………………………………………… 11 4.5 User Interface Overview ……………………………………………………………………………………………………12
5 To Prepare for Use …………………………………………………………………………………………… 13
5.1 To Adjust the Supporting Legs ………………………………………………………………………………………..13 5.2 To Connect to Power …………………………………………………………………………………………………………13 5.3 Turn-on Checkout …………………………………………………………………………………………………………….. 14 5.4 To Set the System Language …………………………………………………………………………………………… 15
6 To Use the Built-in Help System ……………………………………………………………………. 16 7 To Use the Security Lock ………………………………………………………………………………… 17 8 Parameter Setting Method ……………………………………………………………………………..18
8.1 To Set Parameters with the Front-panel Keys and Knobs …………………………………………… 18 8.2 To Set Parameters with the Touch Screen ………………………………………………………………………20
9 To Use the Battery Holder (Option) ……………………………………………………………… 22 10 To Replace the Fuse ………………………………………………………………………………………… 23 11 Remote Control ……………………………………………………………………………………………….. 24 12 Troubleshooting ………………………………………………………………………………………………. 25 13 More Product Information …………………………………………………………………………….. 27
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide I
DG5000 Pro Quick Guide II
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Safety Requirement
1
Safety Requirement
1.1
General Safety Summary
Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly.
1 Only the exclusive power cord designed for the instrument and authorized for use within the destination country could be used.
2 Ensure that the instrument is safely grounded.
3 Observe all terminal ratings. 4 Use proper overvoltage
protection. 5 Do not operate without covers.
6 Do not insert objects into the air outlet.
7 Use the proper fuse. 8 Avoid circuit or wire exposure.
9 Do not operate the instrument with suspected failures.
10 Provide adequate ventilation.
11 Do not operate in wet conditions. 12 Do not operate in an explosive
atmosphere. 13 Keep instrument surfaces clean
and dry. 14 Prevent electrostatic impact.
15 Handle with caution. 16 Please use the front-panel BNC
output connectors properly. They only allow signal output.
WARNING
Equipment meeting Class A requirements may not offer adequate protection to broadcast services within residential environment.
1.2
Safety Notices and Symbols
Safety Notices in this Manual:
WARNING Indicates a potentially hazardous situation or practice which, if not avoided, will result in serious injury or death.
CAUTION Indicates a potentially hazardous situation or practice which, if not avoided, could result in damage to the product or loss of important data.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 1
Safety Requirement
Safety Notices on the Product:
· DANGER It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately.
· WARNING It calls attention to an operation, if not correctly performed, could result in potential injury or hazard.
· CAUTION It calls attention to an operation, if not correctly performed, could result in damage to the product or other devices connected to the product.
Safety Symbols on the Product:
Hazardous Safety Warning Protective Earth Chassis Ground Test Ground
Voltage
Terminal
1.3
Measurement Category
Measurement Category
This instrument can make measurements in Measurement Category I.
WARNING
This instrument can only be used for measurements within its specified measurement categories.
Measurement Category Definitions
· Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable. Thus, you must know the transient withstand capability of the equipment.
· Measurement category II is for measurements performed on circuits directly connected to low voltage installation. Examples are measurements on household appliances, portable tools and similar equipment.
· Measurement category III is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit-breakers, wiring (including cables, bus-bars, junction boxes, switches and socket-outlets) in the fixed installation, and equipment for industrial use and some other
DG5000 Pro Quick Guide 2
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Safety Requirement
1.4 1.5
equipment. For example, stationary motors with permanent connection to a fixed installation. · Measurement category IV is for measurements performed at the source of a low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
Ventilation Requirement
This instrument uses a fan to force cooling. Please make sure that the air inlet and outlet areas are free from obstructions and have free air. When using the instrument in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
CAUTION Inadequate ventilation may cause an increase of temperature in the instrument, which would cause damage to the instrument. So please keep the instrument well ventilated and inspect the air outlet and the fan regularly.
Working Environment
Temperature Operating: 0 to +40 Non-operating: -20 to +60 Humidity
· Operating: 0 to +40: 80% RH (without condensation)
· Non-operating: -20 to +40: 90% RH (without condensation) Below +60: 80% RH (without condensation)
WARNING To avoid short circuit inside the instrument or electric shock, never operate the instrument in a humid environment.
Altitude
· Operating: below 3 km
· Non-operating: below 12 km
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 3
Safety Requirement
1.6
Protection Level Against Electric Shock
ESD ±8kV
Installation (Overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
WARNING Ensure that no overvoltage (such as that caused by a bolt of lightning) can reach the product. Otherwise, the operator might be exposed to the danger of an electric shock.
Installation (Overvoltage) Category Definitions
Installation (overvoltage) category I refers to signal level which is applicable to equipment measurement terminals connected to the source circuit. Among these terminals, precautions are done to limit the transient voltage to a low level. Installation (overvoltage) category II refers to the local power distribution level which is applicable to equipment connected to the AC line (AC power).
Pollution Degree
Pollution Degree 2
Pollution Degree Definition
· Pollution Degree 1: No pollution or only dry, nonconductive pollution occurs. The pollution has no effect. For example, a clean room or air-conditioned office environment.
· Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected. For example, indoor environment.
· Pollution Degree 3: Conductive pollution or dry nonconductive pollution that becomes conductive due to condensation occurs. For example, sheltered outdoor environment.
· Pollution Degree 4: The pollution generates persistent conductivity caused by conductive dust, rain, or snow. For example, outdoor areas.
Safety Class
Class 1 Grounded Product
Care and Cleaning
Care
Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time.
DG5000 Pro Quick Guide 4
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Safety Requirement
1.7
Cleaning Clean the instrument regularly according to its operating conditions. 1. Disconnect the instrument from all power sources. 2. Clean the external surfaces of the instrument with a soft cloth dampened with mild
detergent or water. Avoid having any water or other objects into the chassis via the heat dissipation hole. When cleaning the LCD, take care to avoid scarifying it. CAUTION To avoid damage to the instrument, do not expose it to caustic liquids.
WARNING To avoid short-circuit resulting from moisture or personal injuries, ensure that the instrument is completely dry before connecting it to the power supply.
Environmental Considerations
The following symbol indicates that this product complies with the WEEE Directive 2012/19/EU.
The equipment may contain substances that could be harmful to the environment or human health. To avoid the release of such substances into the environment and avoid harm to human health, we recommend you to recycle this product appropriately to ensure that most materials are reused or recycled properly. Please contact your local authorities for disposal or recycling information.
You can click on the following link https://int.rigol.com/services/services/declaration to download the latest version of the RoHS&WEEE certification file.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 5
Document Overview
2
Document Overview
This manual gives you a quick overview of the front and rear panels, user interface as well as basic operation methods of the DG5000 Pro Series Function/Arbitrary Waveform Generator.
TIP For the latest version of this manual, download it from RIGOL official website (www.rigol.com).
Publication Number QGB18100-1110 Format Conventions in this Manual 1. Key
The front panel key is denoted by the menu key icon. For example, indicates the “Default” key. 2. Menu
The menu item is denoted by the format of “Menu Name (Bold) + Character Shading” in the manual. For example, Setup. 3. Operation Procedures
The next step of the operation is denoted by “>” in the manual. For example,
> Utility indicates first clicking or tapping and then clicking or tapping Utility.
Content Conventions in this Manual
DG5000 Pro Series Function/Arbitrary Waveform Generator includes the following models. Unless otherwise specified, this manual takes DG5502 Pro as an example to illustrate the operation methods of DG5000 Pro series.
Model
No. of Channels Sample Rate
Max. Output Frequency
DG5252 Pro
2
2.5 GSa/s
250 MHz
DG5352 Pro
2
2.5 GSa/s
350 MHz
DG5502 Pro
2
2.5 GSa/s
500 MHz
DG5000 Pro Quick Guide 6
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
General Inspection
3
General Inspection
1. Inspect the packaging
If the packaging has been damaged, do not dispose the damaged packaging or cushioning materials until the shipment has been checked for completeness and has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument. 2. Inspect the instrument
In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative. 3. Check the accessories
Please check the accessories according to the packing lists. If the accessories are damaged or incomplete, please contact your RIGOL sales representative.
Recommended Calibration Interval
RIGOL suggests that the instrument should be calibrated every 12 months.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 7
Product Overview
4
4.1
Product Overview
With up to 2.5 GSa/s sample rate and 64 Mpts/CH arbitrary waveform length (128 Mpts/CH optional), the DG5000 Pro Series Function/Arbitrary Waveform Generator is an all-in-one generator that integrates Function Generator, Arbitrary Waveform Generator, Noise Generator, Pulse Generator, Harmonics Generator, and Analog/ Digital Modulator. It is a multi-functional and cost-effective dual-channel function/ arbitrary waveform generator.
Appearance
Figure 4.1 Front Panel
DG5000 Pro Quick Guide 8
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Product Overview
4.2
Dimensions
Figure 4.2 Rear Panel
358 mm
215 mm
Figure 4.3 Front View
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 9
Product Overview
122 mm
Figure 4.4 Side View
4.3
Front Panel Overview
1
23
4
5 6 7
8 9
15 14
13 12 11 10
Figure 4.5 Front Panel
1
10.1-inch Touch Screen
9
CH2 Sync Output Connector,
Modulation Input Connector
2
Output Mode Selection Area
10 CH2 Output Connector
3
Frequency/Period and Phase/Delay 11 CH1 Sync Output Connector,
Toggle Keys
Modulation Input Connector
4
Basic Waveform Selection Area 12 CH1 Output Connector
DG5000 Pro Quick Guide 10
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Product Overview
4.4
5
Amplitude/High Level and
13 Help Key, Manual Trigger Key, Align
Offset/Low Level Toggle Keys
Phase Key
6
Parameter Input Area (Knob, Arrow 14 USB HOST Interface
Keys,and Numeric Keypad)
7
Quick Operation Key
15 Power Key
8
Channel Output Control Area
Rear Panel Overview
12
3
4 5 6 7
8 9
12 11
10
Figure 4.6 Rear Panel
1 10 MHz Reference Output
7 Security Lock Hole
Connector
2 10 MHz Reference Input Connector 8 AC Power Socket
3 Mounting Screw Holes (VESA 100 x 9 Fuse
100)
4 HDMI Interface
10 Battery Holder Connector
5 USB DEVICE Interface
11 Auxiliary Output Connector
6 LAN Interface
12 External Trigger Input Connector
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 11
Product Overview
4.5
User Interface Overview
Figure 4.7 User Interface
1
Basic Waveform Drop-down Button 8
Channel Copy Function Key
2
Output Mode Drop-down Button 9
Channel Tab
3
Parameter Configuration Area
10 Channel Labels
4
Notification Area
11 Function Navigation Icon
5
Screen Capture Key
12 Waveform Tab
6
Store/Recall Key
13 Channel Identifier
7
Align Phase Function Key
DG5000 Pro Quick Guide 12
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
To Prepare for Use
5
5.1
To Prepare for Use
To Adjust the Supporting Legs
Adjust the supporting legs properly to use them as stands to tilt the instrument upwards for stable placement of the instrument as well as better operation and observation. You can also fold the supporting legs when the instrument is not in use for easier storage or shipment, as shown in the figure below.
5.2
Supporting Legs (a) Unfold the supporting legs
(b) Fold the supporting legs
Figure 5.1 Adjust the Supporting Legs
To Connect to Power
The power requirements of this signal source are 100 V to 240 V, 47 Hz to 63 Hz or 115 V, 360 Hz to 440 Hz. Please use the power cord provided in the accessories to connect the instrument to the AC power source, as shown in the figure below.
Power Connector Figure 5.2 Connect to Power
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 13
To Prepare for Use
5.3
After the power is connected, the backlight of the front-panel power key can have the following three states:
· Breathing Red: indicates that the instrument is connected to a compliant power adaptor and is in standby state. At this time, you can press the power key to power on the instrument.
· Solid Yellow: indicates that the instrument does not detect a compliant power adaptor and cannot be powered on.
· Solid Green: indicates that the instrument is running normally.
WARNING To avoid electric shock, please ensure that the instrument is correctly grounded.
NOTE
If this signal generator is equipped with a battery holder, the AC power source will power both the signal source and the battery holder.
Turn-on Checkout
After the instrument is connected to the power source, press at the lower-left corner of the front panel to power on the instrument. During the start-up process, the instrument performs a series of self-tests. After the self-test, the splash screen is
displayed. You can also click or tap > Utility > Setup to set the “Power Set” to “Auto”. The instrument powers on once connected to power.
TIP You can shut down the instrument in the following ways.
· Click or tap
> Shut Down or press the front-panel and a dialog box “Do you
need to shut down the instrument?” is displayed. Click or tap Shut Down to shut down
the instrument.
· Press twice to shut down the instrument.
· Press for three seconds to shut down the instrument.
DG5000 Pro Quick Guide 14
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
To Prepare for Use
5.4
To Set the System Language
The instrument supports system languages including Chinese and English. You can
click or tap > Utility > Setup to enter the basic settings menu. Then click or tap the drop-down button of Language to set the system language to Chinese or English.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 15
To Use the Built-in Help System
6
To Use the Built-in Help System
The built-in help file provides information about the functions and menu
introductions of the instrument. Click or tap > Help to enter the help system.
In the help system, you can get its help information by clicking on or tapping the link for the specified chapter.
DG5000 Pro Quick Guide 16
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
To Use the Security Lock
7
To Use the Security Lock
If necessary, you can lock the instrument to a fixed location by using a standard laptop security lock (please purchase it by yourself), as shown in the figure below.
The method is as follows: align the lock with the lock hole and plug it into the lock hole vertically, turn the key clockwise to lock the instrument, and then pull the key out.
Security Lock Hole
Figure 7.1 Use the Security Lock
CAUTION Please do not insert other objects into the security lock hole to avoid damaging the instrument.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 17
Parameter Setting Method
8
8.1
Parameter Setting Method
This instrument allows you to use the front-panel parameter input area and the touch screen to set parameters for the instrument.
To Set Parameters with the Front-panel Keys and Knobs
You can use the front-panel parameter input area to set some parameters of the instrument. The parameter input area consists of a knob, a numeric keypad, unit selection keys, and arrow keys, as shown in the figure below.
Knob
You can rotate the knob to move the cursor and navigate through the menu items. Then perform the following operations:
· If the cursor selects a parameter input field, you can press the knob to enter the parameter editing mode. Then you can use the front-panel arrow keys to move the cursor to select the digit place to be modified. Rotate the knob clockwise to increment the value at the selected digit place or counter-clockwise to decrement the value. Press the knob again to confirm the parameter setting and disable the parameter editing mode.
· If the cursor selects a drop-down button, you can press the knob to expand the drop-down menu and then rotate the knob to select a parameter in the menu. After that, press the knob again to confirm your selection and collapse the drop-down menu.
DG5000 Pro Quick Guide 18
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Parameter Setting Method
· If the cursor selects a key, on/off switch, or tab control, pressing the knob is equivalent to tapping the corresponding key, on/off switch, or tab control using the touch screen.
Numeric Keypad
The numeric keypad is composed of numeric keys (from 0 to 9), the decimal point, and symbol keys. If the current cursor selects the input field, you can press the
numeric key to input a number; press
to input “.”; press
to input “-” or “+”.
When using the numeric keypad, you can also perform the following operations:
· Press the knob to confirm the input.
· Press to delete characters.
· Press to cancel the input.
TIP
When inputting hexadecimal characters, use
,
,
,
,
, or
to input
“A, “B, “C, “D”, “E”, or “F” respectively.
Unit Selection Keys
When setting a parameter with the numeric keypad on the front panel, you can use the keys to select the unit of the parameter.
·
: sets the parameter unit to the default one. For example, when setting the
phase, press
>
to set the phase to 1°; when setting the frequency,
press
>
to set the frequency to 1 Hz.
·
/ / : when setting the frequency, use the unit (M/k/G) before “/”;
when setting time/amplitude/offset, use the unit (/m/n) after “/”. For example,
when setting the frequency, press
>
to set the frequency to 1 kHz;
when setting the period, press
>
to set the period to 1 ms.
TIP
When the set value exceeds the limit value, the instrument automatically adjusts the parameter to meet the requirements.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 19
Parameter Setting Method
8.2
Arrow keys
· In normal mode, you can use the keys to move the cursor to select the desired menu item. It is equivalent to rotating the knob.
· In parameter editing mode, you can use the keys to select the digit place to be modified.
· When inputting parameters using the ring-type numeric keypad, is used to
delete the character while field.
is used to cancel the input and close the input
To Set Parameters with the Touch Screen
For this instrument, you can use the touch screen function to set all of its parameters. Click or tap the parameter input field and a virtual keypad will be displayed. You can use the pop-up keypad to complete your parameter setting. The method of using the virtual keypad is as follows.
Input a Value
When setting or modifying a function parameter, you can input an appropriate value with the numeric keypad.
Input Field
Unit Drop-down Button Back Key Clear Key
Enter Key
Unit Key
Figure 8.1 Numeric Keypad
Use the numeric keys in the numeric keypad to input a value. Then select the desired unit, and the numeric keypad is turned off automatically. This indicates that you have completed the parameter setting. The unit drop-down menu includes all available units. You can also click or tap the unit drop-down button to select the desired unit
DG5000 Pro Quick Guide 20
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Parameter Setting Method
when multiple units are available. Then click or tap “Enter” to confirm input and close the numeric keypad.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 21
To Use the Battery Holder (Option)
9
To Use the Battery Holder (Option)
The DG5000 Pro Series Function/Arbitrary Waveform Generator can be powered by the battery holder (option). After the battery holder is installed, it powers the instrument as a power supply when the instrument has not been connected to AC power. It enables the instrument to work for hours in place where an AC power source is not available (e.g. in the field). The actual operating time varies depending on the instrument model. A battery holder option makes it more flexible to use the instrument, making your test no longer limited by the test site.
For how to use the battery holder and safety precautions, refer to BatHolder138 User Guide. For the latest version of this manual, download it from RIGOL official website (www.rigol.com).
CAUTION
The battery holder does not support hot swapping. Before installing and removing the battery holder, power off the instrument and disconnect all power cables to avoid damaging the instrument or battery holder.
DG5000 Pro Quick Guide 22
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
To Replace the Fuse
10
To Replace the Fuse
If you need to replace the fuse, please use the proper fuse (AC 250 V, T3.15 A; 5.2 mm×20 mm) and follow the steps shown below (see Figure 10.1 ). 1. Power off the instrument and remove the power cord. 2. Insert a small straight screwdriver into the slot at the power socket and pry out the
fuse holder gently. 3. Remove the fuse. 4. Insert the proper fuse into the fuse holder. 5. Re-insert the fuse holder into the power socket.
Fuse
Fuse Holder
Figure 10.1 Replace the Fuse
WARNING To avoid electric shock, please make sure that the instrument is powered off and disconnected from the power before replacing the fuse. Also, please make sure the fuse is consistent with the required fuse rating.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 23
Remote Control
11
Remote Control
The following ways of remote control are supported:
· User-defined Programming Users can program and control the instrument by using the SCPI (Standard Commands for Programmable Instruments) commands. For details about the SCPI commands and programming, refer to Programming Guide of this product series.
· PC Software You can use the PC software to send SCPI commands to control the instrument remotely. RIGOL Ultra Sigma is recommended. You can download the software from RIGOL official website (http://www.rigol.com). Operation Procedures:
– Set up communication between the instrument and PC.
– Run Ultra Sigma and search for the instrument resource.
– Open the remote command control panel to send commands.
· Web Control This instrument supports Web Control. You can view the display of the real-time interface of the instrument using Web Control. Through the Web Control method, you can migrant the device control to the control terminals (e.g. PC, Mobile, iPad, and other smart terminals) to realize remote control of the instrument. Connect the instrument to the network, then input the IP address of the instrument into the address bar of the browser of your computer. When mDNS is enabled, you can also input “hostname.local” (see #unique_32) to use Web Control. You have to log in before using the Web Control to modify network settings. When you first log in to the Web Control, the user name is “admin” and password is “rigol”.
This instrument can be connected to the PC via the USB and LAN interface to set up communication and realize remote control through the PC.
This chapter will illustrate how to use the RIGOL Ultra Sigma software to remotely control the instrument via various interfaces.
CAUTION
Before connecting the communication cable, please turn off the instrument to avoid causing damage to the communication interfaces.
DG5000 Pro Quick Guide 24
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Troubleshooting
12 Troubleshooting
1. When I power on the instrument, the instrument stays black and does not display anything.
a. Check whether the power supply has been connected correctly.
b. Check whether the power key is really pressed.
c. Check whether the fuse is blown. If you need to replace the fuse, use only the specified fuse that conforms to the product.
d. Restart the instrument after finishing the above inspections.
e. If the problem still persists, please contact RIGOL.
2. The settings are correct but no waveform is generated.
a. Check whether the output cable is connected to the corresponding channel output terminal tightly.
b. Check whether the output cable works properly.
c. Check whether the output cable is connected to the test instrument tightly.
d. If the problem still persists, please contact RIGOL.
3. The USB storage device cannot be recognized.
a. Check whether the USB storage device can work normally when connected to other instruments or PC.
b. Make sure that the USB storage device is FAT32, NTFS, or exFAT type. The instrument doesn’t support hardware USB storage device.
c. After restarting the instrument, insert the USB storage device again to check whether it can work normally.
d. If the USB storage device still cannot work normally, please contact RIGOL.
4. Performance verification test is failed.
a. Check whether the generator is within calibration period (1 year).
b. Check whether the generator has been warmed up for at least 30 minutes before the test.
c. Check whether the generator is under the specified temperature.
d. Check whether the test is under strong-magnetism environment.
e. Check whether the power supplies of the generator and test system have a strong interference.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 25
Troubleshooting
f. Check whether the performance of the test device used meets the requirement.
g. Make sure that the test device used is within the calibration period.
h. Check whether the test device used meets the required conditions of the manual.
i. Check whether all the connections are tight.
j. Check whether any cable has internal damage.
k. Make sure that the operations conform to settings and processes which are required by the performance verification manual.
l. Check whether the error calculation has faults.
m. Correctly understand the definition of “typical value” for this product: the performance specification of this product under specified conditions.
5. The touch-enabled operation does not work.
a. Check whether you have locked the touch screen. If yes, unlock the touch screen.
b. Check whether the screen or your finger is stained with oil or sweat. If yes, please clean the screen or dry your hands.
c. Check whether there is a strong magnetic field around the instrument. If the instrument is close to the strong magnetic field (e.g. a magnet), please move the instrument away from the magnet field.
d. If the problem still persists, please contact RIGOL.
DG5000 Pro Quick Guide 26
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
More Product Information
13
More Product Information
1. Obtain the Device Information
Click or tap > Utility > About to obtain the information of this instrument, such as the model, serial number, and hardware version number. You can also click or tap the notification area at the lower-right corner of the screen and obtain the information by clicking or tapping About in the pop-up Utility menu.
2. View the Option Information and Install the Option
The instrument is installed with the trial versions of the options when leaving factory. When the instrument is powered on for the first time when leaving factory,
the trial time is about 2,160 minutes. Click or tap > Utility > Options to view the options currently installed in the instrument and their information. For how to install the option, see the detailed descriptions in the User Guide of this product.
For more information about the instrument, refer to the relevant manuals by logging in to the official website of RIGOL (http://www.rigol.com) to download them. · DG5000 Pro User Guide introduces the functions of the instrument and the
operation methods, remote control methods, possible failures and solutions in using the instrument, specifications, and order information. · DG5000 Pro Programming Guide provides detailed descriptions of SCPI commands and programming examples of the instrument. · DG5000 Pro Data Sheet provides the main features and technical specifications of the instrument.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DG5000 Pro Quick Guide 27
Boost Smart World and Technology Innovation
Industrial Intelligent Manufacturing
Education& Research
0
ii·
·,o,m
S
· : ,—- 11 –.,
System Integration
Semiconductors New Energy
Cellular-SG/WIFI £ UWB/RFID/ ZIGBEE ..t;.. Digital Bus/Ethernet (q>) Optical Communication
0 Digital/Analog/RF Chip
[::] Memory and MCU Chip Third-Generation Semiconductor @ Solar Photovoltaic Cells
New Energy Automobile
d: PV/lnverter 0 PowerTest
$ Automotive Electronics
Provide Testing and Measuring Products and Solutions for Industry Customers
Distribution in the UK & Ireland
Characterisation, Measurement & Analysis
Lambda Photometrics Limited Lambda House Batford Mill Harpenden Herts AL5 5BZ United Kingdom E: info@lambdaphoto.co.uk W: www.lambdaphoto.co.uk T: +44 (0)1582 764334
F: +44 (0)1582 712084
RIGOL® is the trademark of RIGOL TECHNOLOGIES CO., LTD. Product information in this document is subject to update without notice. For the latest information about RIGOL’s products, applications and services, please contact local RIGOL channel partners or access RIGOL
official website: www.rigol.com
Documents / Resources
![]() |
RIGOL DG5000 Arbitaray Waveform Generator [pdf] Instruction Manual DG5000 Arbitaray Waveform Generator, DG5000, Arbitaray Waveform Generator, Waveform Generator |