Answer
Aug 14, 2025 - 05:48 AM
If you encounter a situation where one of the output voltages from the ORTEC 402H Power Supply is missing, you can follow these troubleshooting steps based on the manual:
1. Check Load Removal:
- Remove the load from the output that is missing to see if the operation resumes. This step helps identify if the issue is load-related.
2. Check Connection:
- Ensure that the connection between the filter capacitor and the collector of the series pass transistor corresponding to the missing output is not interrupted or damaged.
3. Adjust Trim Potentiometer:
- Verify the correct setting of the trim potentiometer used to balance the short-circuit current sensing amplifier associated with the output. You can readjust the potentiometer by:
- Turn the potentiometer fully counterclockwise.
- Load the output to the specified current level.
- Slowly turn the potentiometer clockwise while monitoring the output voltage with a digital voltmeter. Stop when the voltage starts to drop.
4. Reset Crowbar Protection:
- If the +6-V or -6-V output is missing, turn off the power switch of the Bin and wait a few seconds before turning it back on. This step disables the crowbar protection circuit, which might have been triggered accidentally.
5. Consult the Manual:
- If the issue persists, refer to the ORTEC 402H Power Supply manual for detailed troubleshooting steps or contact ORTEC Customer Service for further assistance.
By following these steps, you can troubleshoot and potentially resolve issues related to missing output voltages from the ORTEC 402H Power Supply.