pyroscience MM33 Micro Manipulator User Manual

pyroscience logoMicromanipulator MM33
Manual Micromanipulator
USER MANUAL

pyroscience MM33 Micro ManipulatorDocument Version 1.14

INTRODUCTION

With the manual micromanipulator MM33 it is possible to measure oxygen micro profiles in semi-solid samples by moving a micrometer screw stepwise manually. Due to its precise and smooth operation, the micromanipulator MM33 is a standard tool for microsensor studies. For best operation conditions, the micromanipulator MM33 should be mounted on the heavy stand HS1 from PyroScience.
The three control knobs for the three axes are positioned very compact and user-friendly. An additional micrometer screw for the z-axis allows positioning down to 10 µm resolution.

pyroscience MM33 Micro Manipulator - Fig1

SAFETY GUIDELINES

PLEASE READ THESE INSTRUCTIONS CAREFULLY BEFORE STARTING TO WORK WITH THIS DEVICE

  • Ensure that the device is only operated under the climatic conditions specified in the chapter “Technical Specifications” in this instruction manual.
  • Maintenance, modifications and repair work must only be carried out by a suitable, qualified technician, trained by PyroScience.
  • If there is any reason to assume that the instrument can no longer be employed without a risk, it must be set aside and appropriately marked to prevent any further use, e.g. if the instrument a) is visibly damaged, b) no longer operates as specified, c) has been stored under adverse conditions for a longer period.
  • If there is any problem, contact PyroScience and send the micromanipulator back for repair.
  • The user has to ensure the following laws and guidelines:
    • EEC directives for protective labor legislation
    • National protective labor legislation
    • Safety regulations for accident prevention
  • This device is not protected against water spray
  • This device is not water-proof
  • This device must not be used under environmental conditions which cause water condensation
  • This device must not be opened

This device is only intended for use in the laboratory by qualified personal according to this instruction manual and these safety guidelines!
This device is not intended for medical or military purposes! Keep away from moving parts of the device! Keep this device out of the reach of children!

Installation and Operation

PRECAUTION: The Micromanipulator MM33 comprises high-precision mechanics which can easily be damaged by improper or harsh treatment. Handle with care!
Please follow these steps for mounting the Micromanipulator MM33:

  1. Release the tilting screws and open the tilting mechanism. Screw the mounting screw carefully into the mounting nut of the Heavy Stand HS1 (if the spring-loaded mounting nut is not exactly centred within the groove, move it back into position with your fingers).pyroscience MM33 Micro Manipulator - Fig2
  2. Fasten the mounting screw.pyroscience MM33 Micro Manipulator - Fig3
  3. Fasten the tilting screws.pyroscience MM33 Micro Manipulator - Fig4
  4. By releasing the tilting screws and the mounting screw, the mounting angle and the position of the micromanipulator can be adjusted. Ensure to fasten the screws carefully, if the desired position is reached.pyroscience MM33 Micro Manipulator - Fig5
  5. Microsensors with diameters from 6.5 mm to 12 mm can be mounted directly to the sensor mount.pyroscience MM33 Micro Manipulator - Fig6
  6. For mounting microsensors with diameters from 0.5 mm up to 3 mm use the included adapter for thin microsensors. The adapter can hold two thin microsensors simultaneously.pyroscience MM33 Micro Manipulator - Fig7

PRECAUTION: Keep any parts of your body (especially fingers and hairs!) out of the mechanics!

MAINTENANCE

PRECAUTION
The performance of the MM33 depends strongly on the conditions of its precision mechanics. Ensure that its parts are kept free from dirt, dust, liquids and other contaminants!
Be aware, that aerated aquariums will deposit a fine spray of water drops in the surrounding. Ensure that the Micromanipulator MM33 is shielded from this spray! Especially seawater is very aggressive, causing rapid corrosion of metallic parts!
The mechanical performance of the Micromanipulator MM33 depends strongly on the condition of its precision mechanics. Periodically you might carefully clean the gear wheels and gear racks. For cleaning, first remove contaminants and old grease with ethanol. Then apply a small amount of silicone-based grease to the gear wheels, gear racks, and the lead screw. In order to protect all other parts of the MM33 against contaminants, you might coat them thinly with a Teflon spray.

SPECIFICATIONS

Feature Specification
Dimensions 65 mm x 200 mm x 160 mm
Weight 0.65 kg
Travel range / resolution x-axis (manual) 20 mm / 0.1 mm
Travel range / resolution y-axis (manual) 25 mm / 0.1 mm
Travel range / resolution z-axis (manual) 37 mm + 10 mm fine adjustment / 0.01 mm
Operating temperature 0 to +50°C
Max. relative humidity up to 95% (non-condensing conditions)
Mounting adapter M6 screw, 10 mm length
Sensor mount Clamp for microsensors from 6.5 mm to 12 mm
Adapter for thin sensors Clamp for two microsensors from 0.5 mm to 3 mm

CONTACT 

PyroScience GmbH
Kackertstraße 11
52072 Aachen
Deutschland
Tel.: +49 (0)241 5183 2210
Fax: +49 (0)241 5183 2299
info@pyroscience.com
www.pyroscience.com

pyroscience logoThe MM33 is provided by:
PyroScience GmbH
Kackertstrasse 11
52072 Aachen
Germany
Phone: +49 (0)241 5183 2210
Fax: +49 (0)241 5183 2299
Email: info@pyroscience.com
Web: www.pyroscience.com
Registered: Aachen HRB 17329, Germany
© PyroScience GmbH

Documents / Resources

PDF thumbnailMM33 Micro Manipulator
User Manual · MM33 Micro Manipulator, MM33, Micro Manipulator, Manipulator

References

Ask a Question

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

Ask a Question

Ask about setup, compatibility, troubleshooting, or anything missing from this manual. Name and email are optional.