ProCon C750 Series Toroidal Conductivity Sensor User Manual

C750 Series Toroidal Conductivity Sensor

Specifications:

  • Power Supply: 9 ~ 36VDC
  • Cell Constant: 2.7
  • Measurement Range: 0 ~ 2,000mS/cm
  • Temperature Compensation: PT1000 RTD
  • Operation Temperature:
  • Output: RS485
  • Pressure Range: 232 psi | 16 Bar
  • Minimum Immersion Depth: 54mm
  • Mounting Thread: G3/4
  • Cable Length: 10m
  • Material:
    • Sensor Material: PFA | 316SS
    • Sealing Gasket: PTFE
    • O-Ring Material:

Product Usage Instructions:

1. Safety Information

DO NOT exceed maximum temperature or pressure
specifications. ALWAYS wear safety goggles or
face-shield during installation and/or service. DO
NOT
alter product construction.

Warning | Caution | Danger: Indicates a
potential hazard. Failure to follow all warnings may lead to
equipment damage, or failure, injury, or death.

Note | Technical Notes: Highlights additional
information or detailed procedure.

Basic Requirements & User Safety:

– Always depressurize and drain the pipeline before installing
or removing the sensor.

– Wear appropriate PPE (gloves, goggles, protective clothing)
when handling the sensor, especially in corrosive chemical
environments.

2. Introduction

The toroidal conductivity sensor comes with a toroidal
controller included, offering features like RS485 output,
temperature compensation, high accuracy, factory calibration, and
measurement of Conductivity, TDS, Salinity, and Resistivity.

3. Technical Specifications:

General:

  • Power Supply: 9 ~ 36VDC
  • Cell Constant: 2.7
  • Measurement Range: 0 ~ 2,000mS/cm
  • Temperature Compensation: PT1000 RTD
  • Operation Temperature:
  • Output: RS485
  • Pressure Range: 232 psi | 16 Bar
  • Minimum Immersion Depth: 54mm
  • Mounting Thread: G3/4
  • Cable Length: 10m
  • Material:
    • Sensor Material: PFA | 316SS
    • Sealing Gasket: PTFE
    • O-Ring Material:

4. Wiring:

Description Drive Receive Temp C750 — Wiring Heat-shrink Tube
Color Wire Color Red Red Brown Black Green Orange Black Clear
Yellow Green Grey Wire Description Pulse 1 Pulse 2 Receive 1
Receive 2 Receive 3 Temp + GND Temp –

5. Dimensions (mm):

– Insert dimensions diagram here –

6. Installation:

6.1 Flange Mounted:

– Install the sensor so that its center hole aligns with the
direction of water flow.

– Ensure the sensor’s submerged depth is at least 54 mm with a
minimum clearance of 30 mm from the sensor’s inner wall to the
surrounding surface.

– The pipeline’s inner diameter must be greater than 114 mm.

– Use an installation flange of DN50 or larger selected
according to site conditions.

6.1.1 Schematic Diagram of Installation:

– Insert schematic diagram here –

6.1.2 Flange Dimensions (mm):

– Insert flange dimensions diagram here –

6.2 Tee Direct Installation:

– Instructions for Tee Direct Installation –

6.3 Immersion Installation:

– Process liquid should flow upward.

Frequently Asked Questions (FAQ):

Q: Can the sensor be used in high-temperature
environments?

A: The sensor has a maximum temperature specification that
should not be exceeded. Ensure to follow all safety guidelines when
using the sensor in varying temperature conditions.

Q: How often should the sensor be calibrated?

A: The sensor comes factory calibrated; however, periodic
calibration checks are recommended to ensure accurate readings over
time.

“`

ProCon® — C750 Series
Toroidal Conductivity Sensor
Quick Start Manual

Read the user’s manual carefully before starting to use the unit. Producer reserves the right to implement changes without prior notice.

25-0865 © Icon Process Controls Ltd.

1

ProCon® — C750 Series
Toroidal Conductivity Sensor
Contents
1. Safety Information ……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………..3 2. Introduction………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………….4 3. Technical Specifications…………………………………………………………………………………………………………………………………………………………………………………………………………………………………4 4. Wiring………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………5 5. Dimensions (mm) ……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………..5 6. Installation …………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………….6 6.1 Flange Mounted……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………….6 6.1.1 Schematic Diagram of Installation ……………………………………………………………………………………………………………………………………………………………………………………………………..7 6.1.2 Flange Dimensions (mm) ……………………………………………………………………………………………………………………………………………………………………………………………………………………….8 6.2 Tee Direct Installation …………………………………………………………………………………………………………………………………………………………………………………………………………………………………..8 6.3 Immersion Installation…………………………………………………………………………………………………………………………………………………………………………………………………………………………………9 7. Model Selection ………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………….9 8. Warranty, Returns and Limitations …………………………………………………………………………………………………………………………………………………………………………………………………………10

25-0865 © Icon Process Controls Ltd.

2

ProCon® — C750 Series
Toroidal Conductivity Sensor

1. Safety Information
DO NOT exceed maximum temperature or pressure specifications. ALWAYS wear safety goggles or face-shield during installation and/or service. DO NOT alter product construction.

Warning | Caution | Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, or failure, injury, or death.

Note | Technical Notes
Highlights additional information or detailed procedure.

Basic Requirements & User Safety
Always depressurize and drain the pipeline before installing or removing the sensor. Wear appropriate PPE (gloves, goggles, protective clothing) when handling the sensor, especially in corrosive
chemical environments.
Ensure the mounting flange and fittings are tightened properly to avoid leaks. Do not exceed the sensor’s rated pressure or temperature limits to prevent damage or accidents. Avoid installation in locations with heavy vibration or strong electrical interference. Keep the sensor’s cable and connectors dry and away from corrosive chemicals. Do not modify or disassemble the sensor; service should be done only by qualified personnel. Disconnect power before performing any maintenance or inspection.

25-0865 © Icon Process Controls Ltd.

3

ProCon® — C750 Series
Toroidal Conductivity Sensor

2. Introduction
TOROIDAL CONTROLLER
INCLUDED

G3/4″ Connection

? Eliminate Costly Conductivity Module ? RS485 Output ? Temperature Compensated ? High Accuracy ? Factory Calibrated ? Toroidal Measurement ? Conductivity | TDS | Salinity | Resistivity

Direct 8-Wire PFA Electrode

Introducing a toroidal conductivity sensor for when you need extreme range.
The ProCon® C750 series toroidal (inductive) conductivity sensors provide a broad measuring range and reliable toroidal technology spanning the range 0-2000 mS/cm.
The C750 series is built for a trouble-free and extended service life, and is resistant to corrosion, coatings, and fouling that are typical of contacting conductivity sensors. The standard PFA Teflon® body material and Viton o-ring offers excellent chemical resistance.
The C750 series models are equipped with G ¾” thread for submersion installation. Built into the sensor is a PT1000 RTD temperature element.

3. Technical SpecificPaFAtiToenflosn

designed for difficult industrial applications and to

General

Power Supply

9 ~ 36VDC

Cell Constant

2.7

Measurement Range

0 ~ 2,000mS/cm

Temperature Compensation PT1000 RTD

Operation Temperature

230 Psi -4 ­ 266°F | -20 ­ 130°C

Output

RS485 | 4-20mA (via Procon® controller)

Pressure Range

232 psi | 16 Bar

Minimum Immersion Depth 54mm

Mounting Thread

G ¾”

Cable Length

10m

Material

Sensor Material

PFA | 316SS

Sealing Gasket

PTFE

O-Ring Material

FEP | Viton®

25-0865 © Icon Process Controls Ltd.

4

order to avoid air within pipe)
ProCon® — C750 Series
Toroidal Conductivity Sensor

4. Wiring

Description Drive
Receive Temp

C750 — Wiring

Heat-shrink Tube Color

Wire Color

Red

Red Brown

Black Green

Orange Black Clear Yellow Green Grey

Wire Description
Pulse 1 Pulse 2
Receive 1 Receive 2 Receive 3
Temp + GND Temp –

5. Dimensions (mm)
30.00
PFA
R27.00

187.00

RS485

2.7

60.50 46.00
/cm

G3/4″

14.00 27.00 40.00 30.00

30.00

25-0865 © Icon Process Controls Ltd.

5

ProCon® — C750 Series
Toroidal Conductivity Sensor
6. Installation 6.1 Flange Mounted

L>54mm T>30mm

T>30mm

T>30mm

Notes:
? Install the sensor so that its center hole aligns with the direction of water flow (see illustration on the left). ? Ensure the sensor’s submerged depth is at least 54 mm, with a minimum clearance of 30 mm from the sensor’s inner wall to
the surrounding surface. The pipeline’s inner diameter must be greater than 114 mm. ? Use an installation flange of DN50 or larger, selected according to site conditions.
Recommended:

25-0865 © Icon Process Controls Ltd.

6

ProCon® — C750 Series
Toroidal Conductivity Sensor
Not Recommended:

6.1.1 Schematic Diagram of Installation
6
5
4 3 2
1

Part Number 1 2 3 4 5 6

Item C750 Sensor Teflon Gasket FEP Seal Ring Flange Gasket DN50 Flange G3/4″ Hex Nut

25-0865 © Icon Process Controls Ltd.

7

ProCon® — C750 Series
Toroidal Conductivity Sensor
6.1.2 Flange Dimensions (mm)
Ø165

18
30.2 125 165

18

6.2 Tee Direct Installation

2″ Screw Thread

3 2
1

25-0865 © Icon Process Controls Ltd.

Part Number 1 2 3

Item C750 Sensor 2″ – 3/4″ Tee Adapter G3/4″ Hex Nut

8

ProCon® — C750 Series
Toroidal Conductivity Sensor
6.3 Immersion Installation

5

Process liquid should flow upwar4d.
(for downward flow ensure order to avoid air within pipe)
3
2

Part Number 1 2 3 4 5

1
Pulse 1 Pulse 2

Item CG7re5y0 Sensor

1 2 3 Temp + GND Temp –

1″-3/4″ Immersion Mounting Adapter

O-Ring

Protective Tube

LP100 Junction Box

7. Model Selection

Part Number C750-T-D-1-F

Material PFA

Output RS485

Cell Constant

Range

Connection

2.7

0-2000mS/cm Flying Lead

25-0865 © Icon Process Controls Ltd.

9

ProCon® — C750 Series
Toroidal Conductivity Sensor
8. Warranty, Returns and Limitations
Warranty
Icon Process Controls Ltd warrants to the original purchaser of its products that such products will be free from defects in material and workmanship under normal use and service in accordance with instructions furnished by Icon Process Controls Ltd for a period of one year from the date of sale of such products. Icon Process Controls Ltd obligation under this warranty is solely and exclusively limited to the repair or replacement, at Icon Process Controls Ltd option, of the products or components, which Icon Process Controls Ltd examination determines to its satisfaction to be defective in material or workmanship within the warranty period. Icon Process Controls Ltd must be notified pursuant to the instructions below of any claim under this warranty within thirty (30) days of any claimed lack of conformity of the product. Any product repaired under this warranty will be warranted only for the remainder of the original warranty period. Any product provided as a replacement under this warranty will be warranted for the one year from the date of replacement.
Returns
Products cannot be returned to Icon Process Controls Ltd without prior authorization. To return a product that is thought to be defective, go to www.iconprocon.com, and submit a customer return (MRA) request form and follow the instructions therein. All warranty and non-warranty product returns to Icon Process Controls Ltd must be shipped prepaid and insured. Icon Process Controls Ltd will not be responsible for any products lost or damaged in shipment.
Limitations
This warranty does not apply to products which: 1) are beyond the warranty period or are products for which the original purchaser does not follow the warranty procedures outlined above; 2) have been subjected to electrical, mechanical or chemical damage due to improper, accidental or negligent use; 3) have been modified or altered; 4) anyone other than service personnel authorized by Icon Process Controls Ltd have attempted to repair; 5) have been involved in accidents or natural disasters; or 6) are damaged during return shipment to Icon Process Controls Ltd reserves the right to unilaterally waive this warranty and dispose of any product returned to Icon Process Controls Ltd where: 1) there is evidence of a potentially hazardous material present with the product; or 2) the product has remained unclaimed at Icon Process Controls Ltd for more than 30 days after Icon Process Controls Ltd has dutifully requested disposition. This warranty contains the sole express warranty made by Icon Process Controls Ltd in connection with its products. ALL IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY DISCLAIMED. The remedies of repair or replacement as stated above are the exclusive remedies for the breach of this warranty. IN NO EVENT SHALL Icon Process Controls Ltd BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND INCLUDING PERSONAL OR REAL PROPERTY OR FOR INJURY TO ANY PERSON. THIS WARRANTY CONSTITUTES THE FINAL, COMPLETE AND EXCLUSIVE STATEMENT OF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES OR REPRESENTATIONS ON BEHALF OF Icon Process Controls Ltd. This warranty will be interpreted pursuant to the laws of the province of Ontario, Canada.
If any portion of this warranty is held to be invalid or unenforceable for any reason, such finding will not invalidate any other provision of this warranty.
For additional product documentation and technical support visit:
www.iconprocon.com | e-mail: sales@iconprocon.com or support@iconprocon.com | Ph: 905.469.9283

by

Phone: 905.469.9283 · Sales: sales@iconprocon.com · Support: support@iconprocon.com

25-0865 © Icon Process Controls Ltd.

10

Documents / Resources

ProCon C750 Series Toroidal Conductivity Sensor [pdf] User Manual
25-0865, c750, C750 Series Toroidal Conductivity Sensor, C750 Series, Toroidal Conductivity Sensor, Conductivity Sensor, Sensor

References

Leave a comment

Your email address will not be published. Required fields are marked *