FC/FE
TEMPERATURE SWITCHES
INSTALLATION DRAWING
RANGE SELECTION TABLE
Range Code | Range °C (°F) | Differential* °C (°F) | Maximum Working Temperature °C (°F) |
Approximate Maximum for “Al” microswitch | |||
T1H | 25 to 90 (77 to 194) |
15.0 (59) |
150.0 (302) |
T2H | 70 to 150 (158 to 302) |
20.0 (68) |
200.0 (392) |
T3H | 120 to 215 (248 to 419) |
30.0 (86) |
300.0 (572) |
Note:
- The minimum differential increases with the setpoint. The differential values mentioned in the above table are approximate maximum for FSR. The differential value will vary according to the pressure range selected and microswitch type. For actual values of differential please contact sales office.
- When using 2SPDT switching arrangement, both microswitches may not actuate and/or deactuate at the same point. A small stage gap, normally upto +/- 5% FSR (depending on range code) may be observed. The On-Off differential (hysterisis) typically tends to be atleast double of those published for 1SPDT pressure switches.
If actuation and/or deactuation at same point is critical part of operation, then it can be achieved by using a separate DPDT relay. This relay will need a separate power supply for it’s coil.
HOW TO ORDER FLAMEPROOF TEMPERATURE SWITCHES
Group 1 | Group 2 | Group 3 | Group 4 | Group 5 | Group 6 | Group 7 | Group 8 |
Non standard allocation | Model | Cable Entry Size | Switch Type | Range Code (values in Deg. Cen.) | Microswitch Type | Temp. Bulb Material / Size |
Capillary Material / Size |
Reserved for Non-standard Options not covered in Catalogue. Will Be given by Manufacturer, Only after Agreement of Supply details With customer. |
FC = IP66 Flameproof pressure switch, ATEx, IECEx & PESO approved FE = IP66 Flameproof pressure switch, PESO approved |
1 = Al. head ½” NPT threads 2 = Al. head ¾” NPT threads 3 = Al. head M20 x 1.5 threads *4 =Grey CI head ½” NPT threads *5 = Grey CI head ¾” NPT threads *6 = Grey CI head M20 x 1.5 threads 7 = SS head ½” NPT threads 8 = SS head ¾” NPT threads 9 = SS head M20 x 1.5 threads *Not available in FE models |
T1 = Temperature Switch, fixed differential without scale T2 = Temperature Switch, fixed differential with scale in °C For adjustable differential |
T1H = 25 – 90 T2H = 70 – 150 T3H = 120 – 215 |
A1 = General purpose microswitch rated at 15 A; 250 VAC *A6 = elements with adjustable deadband *A7 = 2SPDT switching elements A8 = General purpose microswitch rated at 5 A, 250 VAC; 5 A, 28 VDC *B7 = 2SPDT Switching Elements C6 = 1SPDT Adjustable Differential Please refer to page nos. 120 & 121 for options and specifications of microswitches * Please contact sales office for additional information |
B1 = Brass /Dia. 9.5 mm, 123 mm length, with 3/8″BSP (M) thermowell connection B2 = Brass /Dia. 9.5 mm, 123 mm length, with 3/8″NPT (M) thermowell connection B3 = Brass /Dia. 9.5 mm, 123 mm length, with 1/2″ NPT (M) thermowell connection S1 = SS316 /¼” BSP(F) S2 = SS316 / ¼” NPT(F) S3 = SS316 /1″ BSP(M) Please contact sales office for additional information |
2 = SS316 /2.0 meter |
E.g. A Flameproof Temperature switch, with 1/2”NPT cable entry in aluminum housing as 1 SPDT, fixed differential without scale, having 25˚C to 90˚C temperature range, with 15 Amp. microswitch, with Brass 9.5 mm diameter bulb, having length 123 mm with 3/8”BSP(M),with 2.0 meter SS316 capillary length shall be specified by
Group 1 |
Group 2 | Group 3 | Group 4 | Group 5 | Group 6 | Group 7 | Group 8 |
|
FC | 1 | T1 | T1H | A1 | B1 |
2 |
Please specify full model number to avoid ambiguity.
Documents / Resources
![]() |
Orion FC/FE Temperature Switches [pdf] Instructions FC FE Temperature Switches, FC, FE Temperature Switches, Temperature Switches, Switches |
![]() |
Orion FC/FE Temperature Switches [pdf] Installation Guide FC FE Temperature Switches, FC FE, Temperature Switches, Switches |