mPower Electronics logomPower Electronics MP100 HF Calibrator MonitorHF Calibrator
For MP100 HF Monitor
PN 006-0003-M01
Quick Start Guide

MP100 HF Calibrator Monitor

mPower Electronics MP100 HF Calibrator Monitor - overview

Specifications

* The unit may be used for bump testing for up to one year, but after 6 months may no longer be accurate enough for calibrations.

Concentrations 1-15 ppm
Accuracy ±15% (new)
Operating Life 200 calibrations or 6 months,
whichever is first*
Temperature 15-30°C (59-86°F)
Storage Temp. 10-30°C (50-86°F)
Pressure 1±0.1 atm
Humidity 15-90% non-condensing
Weight 220 g
Height 15.7 cm (6.2 in.)
Diameter 11.4 cm (4.5 in.)
Power 3 AAA batteries or 5V/2A
charger with USB C fitting
Safety Data Sheet https://www.mpowerinc.com/materials-safety-data-sheets/ 

* The unit may be used for bump testing for up to one year, but after 6 months may no longer be accurate enough for calibrations.mPower Electronics MP100 HF Calibrator Monitor - figure

Overview

The HF Generator has a reservoir that contains a sorbent material onto which HF is impregnated. It creates a constant HF gas concentration in the 1-15 ppm range, depending on the temperature. It is designed for bump test and calibration of diffusion gas monitors and cannot be used for pumped instruments because there is no flowing gas. The outlet fitting is specifically made to fit mPower UNI gas monitors.

Power Connection

Power the unit with 3 AAA alkaline batteries.
These typically last past the 6-month life of the unit.
Or use a 5V/2A charger with USB C connector.

Operation

  1. Allow the unit at least at least 30 minutes at ambient temperature for the concentration reading to stabilize.
  2. Adjust the UNI Cal Set value to match the concentration reading on the calibrator.
  3. Perform a zero calibration on the UNI.
  4. Place a block or piece of foam about 1 cm (3/8”) thick on the battery compartment.
  5. Unscrew the calibrator cap and set aside.
  6. Start the UNI calibration timing and quickly place the UNI face down with the sensor head onto the HF gas opening and the display supported by the 1cm block/foam, and open the valve.
  7. Start a stopwatch to measure when the calibration time is complete before removing the UNI. Add one minute if it is desired to see the final span reading, to allow the UNI display to automatically revert to the gas reading. Read immediately, as this value will drop quickly once the UNI is removed from the gas source.
  8. Close the valve and screw the cap back on.

warning CAUTION warning
Handle the unit carefully!
DO NOT SHAKE OR TIP UPSIDE DOWN as this may cause damage and will not help speed up the concentration equilibration process. mPower Electronics MP100 HF Calibrator Monitor - figure 1

mPower Electronics logomPower Electronics Inc.
2910 Scott Blvd. Santa Clara, CA 95054
www.mpowerinc.com
info@mpowerinc.com
v1.3 May 2025