Midmark v. 11.0 Digital Spirometer

Using Instruction
- In IQmanager®, search for a patient from the Opening Screen and double-click the patient name, or select New Patient. Ensure the patient’s race, date of birth, sex, height and weight are entered..
- Select the New Test button.
- Enter the names for the Technician, Physician and Indication. Click the Play icon next to the spirometer..
- Prepare the patient for the spirometry test.
NOTE: Proper patient prep results in fewer repeat manoeuvres:- a. Explain the purpose of the spirometry procedure to determine how hard and fast air can be exhaled.
- b. Stress proper posture, including shoulders back and chin up (not leaning forward during exhalation).
- c. Demonstrate the procedure.
- Insert a new mouthpiece into the spirometer.
- Apply the nose clip. Instruct the patient to hold the spirometer up and to the side of their face..
- Click on Start Pre-Test. Wait for the device to zero. If performing a Post test, click Start Post. If applicable, click Start after choosing the incentive.
- Begin the test:
- a. Coach the patient to inhale maximally, then quickly place the mouthpiece in the mouth and seal the lips.
- b. Instruct the patient to BLAST out the air upon exhalation (for 6 seconds). Adjust posture if necessary.
NOTE: Obtain three good manoeuvres (with two matching). Do not exceed eight manoeuvres in one session.
- Click the Review button to display the review screen. Select Save after reviewing the report to save the test session.
- Eject the mouthpiece into the trash receptacle after all manoeuvres are completed.
Examples of Normal, Restrictive and Obstructive Flow-Volume curves
NORMAL
- Normal FEV1
- Normal PEF
- Normal FVC
- Normal FEV1 / FVC % (~75%)
RESTRICTION

- Normal FEV1
- Normal PEF
- Low FVC
- Normal FEV1 / FVC % (>80%)
BORDERLINE TO MODERATE OBSTRUCTION

- Low FEV1
- Low PEF
- Normal FVC
- Low FEV1 / FVC % (<70%)
SEVERE OBSTRUCTION

- Low FEV1
- Low PEF
- Normal FVC
- Low FEV1 / FVC % (<50%)
Midmark Corporation
- 60 Vista Drive, Versailles, OH 45380 USA
- Support Services 844.856.1230 #2
- midmark.com
FAQ
- Q: How many manoeuvres should be performed in one session?
- A: Obtain three good manoeuvres with two matching results. Do not exceed eight manoeuvres in one session.
- Q: What should be done before starting the test?
- A: Before starting the test, ensure that the technician and physician names are entered, proper patient preparation is done, and a new mouthpiece is inserted into the spirometer.