MEDTECH-logo

MEDTECH TMP 11 Digital Thermometer

MEDTECH-TMP-11-Digital-Thermometer-product

 

Warning : A Read instructions thoroughly before using digital thermometer.

  • Choking Hazard: Thermometer cap and battery may be fatal if swallowed. Do not allow children to use this device without parental supervision.
    A Do not use thermometer in ear. Designed use is for oral, rectal, and armpit (axilla) readings only.
  • Do not place thermometer battery near extreme heat as it may explode.
  • Remove battery from the device when not in operation for a long time.
  • The use of temperature readings for self-diagnosis is dangerous. Consult your doctor for the interpretation of results. Self-diagnosis may lead to the worsening of existing disease conditions.
  • Do not attempt measurements when the thermometer is wet as inaccurate readings may result.
  • Do not bite the thermometer. Doing so may lead to breakage and/or injury.
  • Do not attempt to disassemble or repair the thermometer. Doing so may result in inaccurate readings. * After each use, disinfect the thermometer especially in case the device is used by more than one person. A Do not force the thermometer into the rectum. Stop insertion and abort the measurement when pain is present. Failure to do so may lead to injury. A Do not use thermometer orally after being used rectally.
  • For children who are two years old or younger, please do not use the devices orally. A If the unit has been stored at temperatures over 5°C ~40°C(41°F ~ 104°F) , leave it in 5°C ~40°C (41°F ~104°F) ambient temperature for about 15 minutes before using it.

PLEASE READ CAREFULLY BEFORE USING
This digital thermometer provides a quick and highly accurate reading of an individual’s body temperature. To better understand its functions and to provide years of dependable results, please read all instructions first.

CONTENTS

1 Thermometer, I Owner’s Manual, I Storage Case

PRODUCT ILLUSTRATION

MEDTECH-TMP-11-Digital-Thermometer- (1)

PRECAUTION

  • performance of the device may be degraded should one or more of the following occur:
    • Operation outside the manufacturer’s stated temperature and humidity range.
    • Storage outside the manufacturer’s stated temperature and humidity range.
    • Mechanical shock (for example, drop test) or degraded sensor.
    • Patient temperature is below ambient temperature.
  • Portable and mobile RF communications can affect the device. The device needs special pre-cautions regarding EMC according to the EMC information provided in the accompany documents.

SPECIFICATIONS

MEDTECH-TMP-11-Digital-Thermometer-01

°C, °F SWITCHABLE
Temperature readings are available in the Celsius or Fahrenheit scale (*C/°F; located in the upper right corner of LCD.) With the unit off, press and hold the On/Off Button for approximately 2 seconds to change the current setting.

DIRECTIONS

  1. Press the On/Off Button next to LCD display. A tone will sound as the screen shows 188.8.%, followed by last recorded temperature. After showing the self-test temperature, the thermometer is now in the testing mode.
  2. Position thermometer in desired location (mouth, rectum, or armpit.)
    • Oral Use: Place thermometer under tongue as indicated by ” y” position shown in Figure 2. Close your mouth and breathe evenly through the nose to prevent the measurement from being influenced by inhaled/exhaled air.
    • Rectal Use: Lubricate silver probe tip with petroleum jelly for easy insertion. Gently insert sensor approximately 1em (less than 1/2 “) into rectum.
    • Armpit Use: Wipe armpit dry. Place probe in armpit and keep arm pressed firmly at side. From a medical viewpoint, this method will always provide inaccurate readings, and should not be used if precise measurements are requiredMEDTECH-TMP-11-Digital-Thermometer- (2)
  3. The degree sign flashes throughout the testing process. When flashing stops an alarm will beep for approximately 10 seconds. The measured reading will appear on the LCD simultaneously. The minimum measurement time until the signaling tone (beep) must be maintained without exception. The measurement continues even after the buzzer notification. So that in order to achieve better body temperature measurement result, recommend to keep the probe in mouth and rectum about 2 minutes, or in armpit about 5 minutes regardless of the beep sound and at least 30 seconds measurement interval should be maintained.
    *Note: Normally the buzzes are ” Bi-Bi-Bi-Bi-” ; Alarm beeps more rapidly when temperature reaches 37.8 C (100 F) or higher ,and the buzzes are “Bi-Bi-Bi-.. Bi-Bi-Bi-______ Bi-Bi-Bi”
  4. To prolong battery life, press the On/Off Button to turn unit off after testing is complete. If no action is taken, the unit will automatically shut off after around 10 minutes.

TROUBLESHOOTING

MEDTECH-TMP-11-Digital-Thermometer- (4)

BATTERY REPLACEMENT

  1. Replace battery when ” MEDTECH-TMP-11-Digital-Thermometer- (5)” appears in the lower right corner of LCD display.
  2. Pull battery cover off as shown in Figure 3.
  3. Gently pull out plastic circuit board with battery chamber approximately 1 cm (slightly less than 1/2″.) (See Figure 4)
  4. Use pointed object such as a pen to remove old battery. Discard battery lawfully. Replace with new 1.5V DC button type LR41 or SR41,UCC392, or equivalent. Be sure battery is installed with “+” polarity facing up. (See Figure 5) 5. Slide battery chamber back into place and attach cover.

MEDTECH-TMP-11-Digital-Thermometer- (6)

CALIBRATION
The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to the use instruction, periodic readjustment is not required. However, we recommend checking calibration every two years or whenever clinical accuracy of the thermometer is in question. Turn on the thermometer and insert into the water bath and then check the laboratory accuracy of thermometer. Please send the complete device to the dealers or manufacturer. The above recommendations do not supersede the legal requirements. The user must always comply with legal requirements for the control of the measurement, functionality, and accuracy of the device which are required by the scope of relevant laws, directives or ordinances where the device is used.

CLEANING AND DISINFECTION

  1. Immerse the thermometer probe in distilled water for at least 1 minute;
  2. Using a clean, soft cloth to wipe the thermometer down to remove any residue;
  3. Repeat step 1 and 2 for three times until no soil is seen with visual inspection after cleaning;
  4. For thoroughly clean and disinfection, please use method A or B: Method A(High level disinfection): immerse the thermometer probe in 0.55% OPA(O-Phthaldehyde), such as CIDEX OPA, for at least 12 minutes under temperature at 20°C ; Method B(Low level disinfection): Using a clean soft cloth dipped in 70% medical alcohol, wipe the probe 3 times, at least one minute for each time.
  5. Repeat step 1 to 3 to remove OPA residuals; Note: Rectal use is not recommended for home use as OPA will not be readily available outside of a hospital. If rectal measurement is necessary, we strongly recommend high level disinfection.
    Note2: Please operate according to the manual of OPA for reference. To prevent damage to the thermometer please note and observe the following:
    • Do not use benzene, paint thinner, gasoline or other strong solvents to clean the thermometer.
    • Do not attempt to disinfect the sensing probe (tip) of the thermometer by immersing in alcohol, OPA or in hot water (water over 122 F (50 °C) for long time.
    • Do not use ultrasonic washing to clean the thermometer.

LIMITED WARRANTY

The thermometer is guaranteed for one year from the date of purchase. If the thermometer does not function properly due to defective components or poor workmanship, we will repair or replace it free of charge. All components are covered by this warranty excluding the battery. The warranty does not cover damages to your thermometer due to improper handling. To obtain warranty service, an original or copy of the sales receipt from the original retailer is required.

Electromagnetic Compatibility Information

The device satisfies the EMC requirements of the international standard IEC 60601-1-2. The requirements are satisfied under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use. Portable and mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the electromagnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.

Table 1

MEDTECH-TMP-11-Digital-Thermometer- (7)

Table 2

MEDTECH-TMP-11-Digital-Thermometer- (8)

Table 3

MEDTECH-TMP-11-Digital-Thermometer- (9)

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

  • The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
  • The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,7 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges.
  • Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
  • Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Table 4

MEDTECH-TMP-11-Digital-Thermometer- (10)

Table  5MEDTECH-TMP-11-Digital-Thermometer- (11)WARNINGS!

  • This device should not be used in the vicinity or on the top of other electronic equipment such as cell phone, transceiver or radio control products. If you have to do so, the device should be observed to verify normal operation.
  • The use of accessories and power cord other than those specified, with the exception of cables sold by the manufacturer of the equipment or system as replacement parts for internal components, may result in increased emissions or decreased immunity of the equipment or system.

Documents / Resources

MEDTECH TMP 11 Digital Thermometer [pdf] Instruction Manual
TMP 11, TMP 11 Digital Thermometer, TMP 11, Digital Thermometer, Thermometer

References

Leave a comment

Your email address will not be published. Required fields are marked *