MSP-450 series
450W Single Output Medical Type
MSP-450 series Single Output Medical Type
https://www.meanwell.com/Upload/PDF/Enclosed_Type_EN.pdf
GTIN CODE
https://www.meanwell.com/serviceGTIN.aspx
Features:
- Universal AC input / Full range
- Built-in active PFC function, PF>0.95
- High efficiency up to 89.5%
- Withstand 300VAC surge input for 5 seconds
- Protections: Short circuiit / Overload / Over voltage / Over temperature
- Built-in constant current limiting circuit
- Medical safety approved (MOOP level)
- Built-in cooling Fan ON-OFF control
- Built-in DC OK signal
- Built-in remote ON-OFF control
- Stand by 5V@0.3A
- Built-in remote sense function
- No load power consumption<0.6W (Note.6)
- 5 years warranty
SPECIFICATION
MODEL | MSP-450-3.3 | MSP-450-5 | MSP-450-7.5 | MSP-450-12 | MSP-450-15 | MSP-450-24 | MSP-450-36 | MSP-450-48 | |
OUTPUT | DC VOLTAGE | 3.3V | 5V | 7.5V | 12V | 15V | 24V | 36V | 48V |
RATED CURRENT | 90A | 90A | 60A | 37.5A | 30A | 18.8A | 12.5A | 9.5A | |
CURRENT RANGE | 0 ~ 90A | 0 ~ 90A | 0 ~ 60A | 0 ~ 37.5A | 0 ~ 30A | 0 ~ 18.8A | 0 ~ 12.5A | 0 ~ 9.5A | |
RATED POWER | 297W | 450W | 450W | 450W | 450W | 451.2W | 450W | 456W | |
RIPPLE & NOISE (max.) Note.2 | 80mVp-p | 80mVp-p | 100mVp-p | 120mVp-p | 150mVp-p | 150mVp-p | 240mVp-p | 240mVp-p | |
VOLTAGE ADJ. RANGE | 2.8 ~ 3.8V | 4.3 ~ 5.8V | 6.8 ~ 9V | 10.2 ~ 13.8V | 13.5 ~ 18V | 21.6 ~ 28.8V | 28.8 ~ 39.6V | 40.8 ~ 55.2V | |
VOLTAGE TOLERANCE Note.3 | ±2.0% | ±2.0% | ±2.0% | ±1.0% | ±1.0% | ±1.0% | ±1.0% | ±1.0% | |
LINE REGULATION | ±0.5% | ±0.5% | ±0.5% | ±0.3% | ±0.3% | ±0.2% | ±0.2% | ±0.2% | |
LOAD REGULATION | ±1.0% | ±1.0% | ±1.0% | ±0.5% | ±0.5% | ±0.5% | ±0.5% | ±0.5% | |
SETUP, RISE TIME | 1000ms, 100ms/230VAC 2500ms, 100ms/115VAC at full load | ||||||||
HOLD UP TIME (Typ.) | 16ms/230VAC 16ms/115VAC at full load | ||||||||
INPUT | VOLTAGE RANGE Note.4 | 85 ~ 264VAC 120 ~ 370VDC | |||||||
FREQUENCY RANGE | 47 ~ 63Hz | ||||||||
POWER FACTOR (Typ.) | PF>0.95/230VAC PF>0.99/115VAC at full load | ||||||||
EFFICIENCY (Typ.) | 80% | 83% | 86.5% | 88% | 89% | 88% | 89% | 89.5% | |
AC CURRENT (Typ.) | 5A/115VAC 2.4A/230VAC | ||||||||
INRUSH CURRENT (Typ.) | 35A/115VAC 70A/230VAC | ||||||||
LEAKAGE CURRENT | Earth leakage current < 300μA/264VAC , Touch leakage current < 100μA/264VAC | ||||||||
PROTECTION | OVERLOAD | 105 ~ 135% rated output power | |||||||
Protection type : Constant current limiting, recovers automatically after fault condition is removed | |||||||||
OVER VOLTAGE | 3.96 ~ 4.62V | 6 ~ 7V | 9.4 ~ 10.9V | 14.4 ~ 16.8V | 18.8 ~ 21.8V | 30 ~ 34.8V | 41.4 ~ 48.6V | 57.6 ~ 67.2V | |
Protection type : Shut down o/p voltage, re-power on to recover | |||||||||
OVER TEMPERATURE | Shut down o/p voltage, recovers automatically after temperature goes down | ||||||||
FUNCTION | 5V STANDBY | 5VSB : 5V@0.3A ; tolerance±5%, ripple : 50mVp-p(max.) | |||||||
DC OK SIGNAL | PSU turn on : 3.3 ~ 5.6V ; PSU turn off : 0 ~ 1V | ||||||||
REMOTE CONTROL | RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off | ||||||||
FAN CONTROL (Typ.) | Load 20±10% or RTH2≧50℃ Fan on | ||||||||
ENVIRONMENT | WORKING TEMP. | -40 ~ +70℃ (Refer to “Derating Curve”) | |||||||
WORKING HUMIDITY | 20 ~ 90% RH non-condensing | ||||||||
STORAGE TEMP., HUMIDITY | -40 ~ +85℃, 10 ~ 95% RH non-condensing | ||||||||
TEMP. COEFFICIENT | ±0.03%/℃ (0 ~ 50℃) | ||||||||
VIBRATION | 10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes | ||||||||
SAFETY & EMC (Note 8) |
SAFETY STANDARDS | ANSI/AAMI ES60601-1, IEC60601-1, EAC TP TC 004 approved; Design refer to BS EN/EN60601-1, BS EN/EN62368-1 | |||||||
ISOLATION LEVEL | Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP, Secondary-Earth: 1×MOOP | ||||||||
WITHSTAND VOLTAGE | I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC | ||||||||
ISOLATION RESISTANCE | I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH | ||||||||
EMC EMISSION | Compliance to BS EN/EN55011 (CISPR11) Class B, BS EN/EN61000-3-2,-3, EAC TP TC 020 | ||||||||
EMC IMMUNITY | Compliance to BS EN/EN61000-4-2,3,4,5,6,8,11, BS EN/EN60601-1-2, BS EN/EN55035, EAC TP TC 020 | ||||||||
OTHERS | MTBF | 1171.9K hrs min. Telcordia SR-332 (Bellcore) ; 159.4K hrs min. MIL-HDBK-217F (25℃) | |||||||
DIMENSION | 218*105*41mm (L*W*H) | ||||||||
PACKING | 1.19Kg; 12pcs/15.3Kg/0.82CUFT |
NOTE
- All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25’C of ambient temperature.
- Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0. 1uf & 47uf parallel capacitor.
- Tolerance : indudes set up tolerance, line regulation and load reguiation
- Derating may be needed under low input voltages. Please check the derating curve for more details.
- Length of set up time is measured at first cold start. Tuming ON/OFF the power supply may lead to increase of the set up time.
- No load power consumption<0.5W when RC- & RG+ (CN100 pin1,2) 0 ~ 0.8V or short
- When the input voltage is less than 40VAG, the SPS may exhibit degradation of performance. The final product manufacturers must re-confirm this deviation that does not affect basic safety or essential performance.
- The power supply is considered a componert which wil be installed into a final equipment. All the EMG tests are been exectted by mounting the unit on 2 360mm”360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that stil meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supples.”(as available on https://www.meanwel.com/Upload/PDF/EMI_statemert_en.pdf)
- The ambient temperature derating of 3.5’C/1000m with faniess models and of 5’C/1000m with fan models for operating alttude higher than 2000m(500t)
Product Liabilty Disclaimer : For detalled information, please refer to htips://www.meanwell.comiserviceDisdaimer.aspx
Mechanical Specification
Block Diagram
Derating Curve
Output Derating VS Input Voltage
Function Description of CN100
Pin No. | Function | Description |
1 | RC+ | Turns the output on and off by electrical or dry contact between pin 2 (RC-), Short: Power OFF, Open: Power ON. |
2 | RC- | Remote control ground. |
3 | AUX | Auxiliary voltage output, 4.75~5.25V, referenced to pin 4(AUXG). The maximum load current is 0.3A. This output is not controlled by the “remote ON/OFF control”. |
4 | AUXG | Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V). |
5 | DC-OK | DC-OK Signal is a TTL level signal, referenced to pin6(DC-OK GND). High when PSU turns on. |
6 | GND | This pin connects to the negative terminal(-V). Return for DC-OK signal output. |
7 | +S | Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V. |
8 | -S | Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V. |
Function Manual
1.Remote Sense
The remote sensing compensates voltage drop on the load wiring up t0 0.5V.2.DC-OK Signal
DC-OKsignal is a TTL level signal. High when PSU turns on.
Between DC-OK(pin5) and GND(pin6) | Output Status |
3.3 ~ 5.6V | ON |
0 ~ 1V | OFF |
3.Remote Control
The PSU can be turned ON/OFF by using the “Remote Control” function.
Between RC+(pin1) and RC-(pin2) | Output Status |
SW ON (Short) | OFF |
SW OFF (Open) | ON |