mantracourt logoPSDS-HSBK Quick Start Guide
mantracourt.com
mantracourt PSDS-HSBK Portable Sensor Display Strain BridgePSDS-HSBK
Portable Sensor Display – Strain Bridge

Introduction/overview

The PSDS-HSBK allows simple display of strain bridge based measurements such as load cells and pressure gauges with sensitivity up to +/-480 mV/V.
Up to six calibration ranges are available allowing for different loading modes (tension and compression) or different sensors. Each calibration range will remember settings that contribute to the user experience such as selected units and tare values.
TEDS devices using templates 33, 40 and 41 can be connected and will update the viewed calibration. The last twenty TEDS devices connected will be remembered and recognized when connected again reverting to the last user experience settings for that device. TEDS can be disabled and the internal calibration ranges become available again.
There are also up to six display modes available. These determine what is visible on the LCD display and what actions are available from the set of three soft keys.
Full configuration is available with a free PC based toolkit. Some simple configuration such as two point calibration is available from within the handheld using the menu system.

Getting started

This document is designed to give a very quick overview of the PSDS-HSBK and its general, standard functionality.
For more details, see the full manual.

Hardware overview

Front

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Hardware overview

Back 

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Hardware overview1

Top 

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Hardware overview Top

The Load Cell connector fitted to PSD Standard Product is BINDER 770-8. The cable fitted mating connector is a BINDER 771 8-way male connector. There are three versions available with different field cable diameters.

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Hardware overview Top1

View from solder connector side of the connector

Cable pair Suggested Colour Connector Pin Function
1 White 1 Loadcell Reference +ve
Black 2 Loadcell Reference -ve
2 Green 3 Loadcell Signal +ve
Black 4 Loadcell Signal -ve
3 Red 5 Loadcell Excitation +ve
Black 6 Loadcell Excitation -ve
4 Blue 7 TEDS
Black 8 Ground
Screen Grey Cable screen should only be connected to chassis of the sensor.
If this cannot be achieved, then it should be connected to Excitation -ve.

Setting up without the Toolkit

Quick, limited setup is possible using just the handheld and no toolkit. To access all the configuration features, you will need to use the toolkit.
Calibration
saci pumps OPTIMA Swimming Pool Pump and Filter - icon1 Please note that this must be done in the correct order for the calibration to work correctly. Changing the Sensitivity after inputting mV/V readings will cause the calibration to be void. Always make sure you finish with Apply

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Calibration mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Calibration1

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Calibration2

Change units 

In all standard display modes,mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - icon scrolls through the available units for the selected calibration.
Change decimal places
In all standard display modes, a 2 second pressmantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - icon1 shifts the decimal place position for the selected units.

Overload/underload alarm

saci pumps OPTIMA Swimming Pool Pump and Filter - icon1 Please note, the values saved for the thresholds are applied in the calibrated units of the currently selected range. This means that different ranges will trigger at different loads if the calibrated unit is different.

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - underload alarm mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - underload alarm1

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - underload alarm2

Using TEDS

Plug in a TEDS enabled load cell.

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - underload alarm3 mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - underload alarm4

Toolkit

Toolkit overview
More complex setup is possible using the supplied toolkit software.
Install the toolkit onto your windows PC and connect the USB to the PSDS-HSBK.
Then open the toolkit:

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Toolkit overview

Navigation is achieved using the icons along the top of the toolkit.

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - Toolkit overview1

For further information, please see the PSDS-HSBK User Manual.

Document Title: PSDS-HSBK Quick Start
Applies To: PSD (Portable Sensor Display)
Part Number: 517-952
Issue Number: 01.00
Dated: 23rd May 2022

In the interests of continued product development, Mantra court Electronics Limited reserves the right to alter product specifications without prior notice.

mantracourt logowww.mantracourt.com
Mantra court Electronics Limited
PSDS HSBK Quick Start Guide
mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge - icon2

Documents / Resources

mantracourt PSDS-HSBK Portable Sensor Display Strain Bridge [pdf] User Guide
PSDS-HSBK Portable Sensor Display Strain Bridge, PSDS-HSBK, Portable Sensor Display Strain Bridge, Sensor Display Strain Bridge, Display Strain Bridge, Strain Bridge, Bridge

References

Leave a comment