CR-22P Microplate Centrifuge

User Manual

1. Introduction

The CR-22P is a microprocessor-controlled digital 96/384 well micro-plate centrifuge designed for sample separation and preparation in various applications such as molecular biology, genetic engineering, microbiology, genomics, immunology, and in-vitro diagnostics. It features a maintenance-free Brushless DC motor and an imbalance detection safety feature. The centrifuge is compatible with different types of standard-sized microplates.

2. Intended Use

This benchtop, two-place microplate centrifuge is intended for use with PCR plates, ELISA plates, cytokine assay plates, sample quantifications, and short spin-downs in molecular biology, genomics, proteomics, diagnostic labs, forensic laboratories, and hospitals for routine plate centrifuging applications.

Note: Please read this user manual carefully before using the instrument. This manual assists with operation and care, not repair. For repairs, contact the supplier.

3. Features

4. Standard Accessories

5. Technical Specification

Parameter Specification
Motor Type Brushless Motor
Maximum Volume 2 micro plates (96 & 384 well)
Speed Setting 500 to 2500 RPM
Maximum RCF 462 g (at 2500 RPM)
Speed Accuracy ± 100 RPM
Run Time 1 to 99 mins & infinite mode
Acceleration Time 30 + 5 seconds
Deceleration Time 20 + 5 seconds
Dimensions (W x D x H) 330 x 265 x 161 mm
Weight 4 Kg (without rotor)
Operating Humidity ≤ 80%
Input Voltage 24V ⟷ 1.5A
Power Consumption 24 W

6. Safety Precautions

Note: Insert both microplates for balancing the rotor and reducing vibrations.

7. Installation

The centrifuge is shipped in a packaging box. Remove the upper packaging and lift the centrifuge from the bottom. Ensure all packaging accessories are removed from the rotor chamber before first use. The rotor is pre-installed; confirm it is firmly tightened. Keep all packaging for at least 2 years for warranty purposes.

Note: Instructions for transportation after delivery are to be handled by the customer.

Installation Requirements:

Connecting Power Adaptor:

  1. Connect one end of the power adaptor to the rear of the centrifuge and the other to the power supply.
  2. Rotate the adaptor nut clockwise to tighten it to the centrifuge. Ensure the power switch is OFF.
  3. Match the rib on the plug with the slot on the centrifuge's adaptor port before inserting.
  4. Once matched, gently push the plug in. Do not force it.
  5. Turn the upper ring clockwise to secure the fitment.

8. Standard Parts Listing

Front View:

Rear View:

Bottom View:

9. User Interface & Display

Item Name Function
1 RPM Press to select RPM value. Use +/- to set RPM.
2 RCF Long press 'RPM/RCF' to toggle between RPM/RCF. Select RCF value for centrifugation.
3 START Press to start centrifugation after setting RPM and TIME. Works only if the lid is closed.
4 STOP/OPEN Press to stop operation. The lid opens automatically after run completion.
5 TIME Press to select Time value. Use +/- to set Time.
6 P1 Press to set program mode P1.
7 P2 Press to set program mode P2.
8 +/- Use '+' to increase and '-' to decrease values.

10. Rotor Installation

Rotor Removing and Replacing Process:

The rotor comes pre-installed. To remove or replace it:

Removing Rotor:

  1. Do not attempt to remove or loosen the rotor with bare hands.
  2. Use the T-Allen Key to loosen the rotor nut by turning it anti-clockwise.

Replacing Rotor:

  1. Load the rotor vertically onto the motor shaft.
  2. Place the rotor nut in the center hole of the rotor onto the motor shaft.
  3. Insert the T-Allen Key into the rotor nut and turn clockwise to tighten or anti-clockwise to loosen.

Note: Ensure the rotor is firmly tightened before running.

11. Balancing the Rotor

  1. Always balance the rotor before operation. Load centrifuge plates symmetrically.
  2. Ensure samples in the plates have equal volumes.
  3. Symmetrical loading is crucial to prevent vibration or imbalance, which can damage the centrifuge and harm the user.
  4. If plates are not loaded symmetrically, the imbalance detector will stop the centrifuge (Err 55). Load plates symmetrically and restart.
  5. Incorrect plate loading can lead to accidents.

Note: 'Plates Holder' are part of the Rotor.

12. Operating the Centrifuge

Switching On:

After connecting the power adaptor, switch ON the main power supply and the switch at the rear of the centrifuge. Ensure the rotor is securely fitted before use. The centrifuge will not operate with the lid open.

Note: Maintain a 3-second gap between switching OFF and ON again. Do not switch OFF and ON instantly.

Speed Setting:

After closing the lid, press the 'RPM/RCF' button to select speed setting (RPM or RCF). Use the '+' and '-' buttons to adjust the value. The speed range is 500 to 2500 RPM. The set value is stored automatically after approximately 1 second of inactivity. Speed can be changed during operation by pressing 'RPM/RCF' and using the +/- buttons. Long press 'RPM/RCF' to toggle to RCF setting.

Time Setting:

Operating time can be set from 1 minute to 99 minutes or in continuous mode (' ][: ][ '). Press the 'TIME' button and use the '+' and '-' buttons to adjust the value. The timer is a countdown timer displaying 'Min/Sec'.

Start and Stop Operation:

Press the 'START' button to begin operation. Press the 'STOP/OPEN' button to stop the ongoing operation. The lid opens automatically after the operation stops.

Program Mode & Line Setting:

Press 'P1' or 'P2' once to enter program mode. Use 'RPM/RCF' and '+/-' to adjust speed/RCF, and 'TIME' to adjust time. Set values are saved after 5 blinks. Press 'START' to run the program. Long pressing 'P1' or 'P2' returns the centrifuge to normal mode.

Changing Settings During a Run:

Settings can be changed during operation. Pressing a function button (TIME, RPM, RCF) allows modification of the current value. New values are activated after entry and are stored upon confirmation (5 blinks).

Opening Centrifuge Lid in Power Failure:

Disconnect the centrifuge from the main supply. Wait for the rotor to stop completely. Pull out the emergency lid release pin to open the lid. After opening, push the thread back into the hole and refit the emergency release cap.

13. Troubleshooting

The centrifuge has a self-diagnostic function. Error/warning codes displayed on the screen indicate malfunctions:

ERROR PROBLEM SOLUTION
No display No main power connection Check power and plug connections.
Power failure Check the main fuse.
Improper connection Connect adaptor properly.
Err 1 Latch jammed or limit switch damaged. Contact service representative.
Err 52 Rotor stuck or incorrect operating voltage. Turn off, check rotor fitment, apply correct voltage (230 VAC ± 10 VAC), and restart.
Err 55 Rotor not loaded symmetrically. Load rotor symmetrically and restart.
Centrifuge lid cannot be opened Rotor is still spinning. Wait for the rotor to stop.
Power failure Use emergency lid release after rotor stops.
Centrifuge shakes during acceleration & exceptional running noise Rotor not loaded symmetrically. Load rotor symmetrically and restart.
Broken plates, rotor, or motor damage. Replace broken plates. Contact service for rotor/motor damage.
Rotor is damaged. Remove and change rotor.
Display Error Loose display connection. Contact service representative.
Power tripping Cable not fitted properly. Remove and connect cable properly.

Notes:

14. Warranty Statement

This product is warranted against defects in material and workmanship for two (2) years from the date of purchase. The warranty covers repair upon prompt notification under the following conditions:

For your reference:

Serial No.: _______________

Purchase Date: _______________

Supplier: _______________

15. Product Disposal

Observe relevant legal regulations when disposing of the product. Electrical and electronic devices are regulated by national regulations based on EU Directive 2012/19/EU (WEEE). Devices supplied after 13.06.05 in the business-to-business sphere, to which this product is assigned, may not be disposed of in municipal or domestic waste. They are marked with the following symbol: ♺.

Disposal regulations vary within the EU; contact your supplier if necessary.

Models: CR-22P CAPPRondo Microplate Centrifuge, CR-22P, CAPPRondo Microplate Centrifuge, Microplate Centrifuge, Centrifuge, CR-22P CAPPRondo Microplate Centrifuge

PDF preview unavailable. Download the PDF instead.

CR-22P Microplate-Centrifuge UG Corel PDF Engine Version 17.0.0.491

Related Documents

Preview CAPP CRPI-412X Thermo Shaker User Manual - Operation and Specifications
Comprehensive user manual for the CAPP CRPI-412X Thermo Shaker, covering features, technical specifications, installation, operation, safety, troubleshooting, and warranty from AHN Biotechnologie GmbH.
Preview CAPPRondo Hot Plate Stirrer CRS-22H User Manual
Comprehensive user manual for the Capp CAPPRondo Hot Plate Stirrer, model CRS-22H. Learn about its features, specifications, safe operation, maintenance, and troubleshooting for laboratory applications.
Preview CAPP CR-MI1 Mini Incubator User Manual
User manual for the CAPP CR-MI1 mini incubator, detailing its features, technical specifications, installation, operation, safety precautions, cleaning, warranty, and disposal. This 22L incubator offers digital temperature control from ambient -15°C to 60°C.
Preview CAPP CR3-08X CAPPRondo 3D Shaker User Manual
Comprehensive user manual for the CAPP CR3-08X CAPPRondo 3D Shaker, detailing its features, operation, safety instructions, technical specifications, and maintenance for laboratory use.