Strain Gage Transducer Indicator Model: 9820 | Interface

SPECIFICATIONS. 5.13DINSTALLATION AND WIRINGINSTALLATIONThe Model 9820 enclosure is desig. ted for panel mounting in a 1/8 DIN cu. e show. 0.024, -0.000) 3.622"( 0.032, -0.000)To panel. mount the 9820, perform the following steps.Rotate the four pawl screws ( outside screws in each corner) several.


File Info : application/pdf, 21 Pages, 1.66MB

PDF preview unavailable. Download the PDF instead.

9820manlRevD21224
Strain Gage Transducer Indicator
Model: 9820
9820 Installation & User Manual

TABLE OF CONTENTS
DESCRIPTION and SPECIFICATIONS .................... 1 INSTALLATION and WIRING .........................................2 SWITCH and FUNCTION DEFINITIONS . .................... 6 OPERATING INSTRUCTIONS .....................................10 CALIBRATION INSTRUCTIONS ..................................13 PROGRAM MENU DIAGRAM ......................................17 WARRANTY..................................................................18
FIGURES
Figure 1 Model 9820 Front Panel View ................. 3 Figure 2 Model 9820 Rear Panel View .................. 3 Figure 3 Transducer Wiring ................................... 4 Figure 4 Location of Jumpers and
Adjustment Potentiometers ....................... 4 Figure 5 Location of RCAL Jumper ........................ 5

DESCRIPTION
The Model 9820 is a microprocessor-based digital indicator capable of interfacing directly to a low level strain gage load cell transducer. An internal, high gain, fully-differential amplifier and a 4 1/2 digit analog-todigital converter combine to accurately digitize the input signal. A 5V @ 60mA or 10V @ 120mA, short circuit protected, transducer excitation supply is also provided. A +/-10VDC analog output signal, with a 2kHz bandwidth, comes standard with the instrument. A front panel RCAL switch provides a convenient calibration feature. TARE and CALIBRATION are easily performed via front panel pushbuttons.

SPECIFICATIONS

Excitation voltage: Jumper selectable 5 or 10 (Default) VDC

Excitation current: 60mA @ 5VDC, 120mA @ 10VDC

Input amp gain: Jumper selectable, 40 or 80

Signal input range: +/-25mV at Input amp gain = 80

+/-50mV at Input amp gain = 40 (Default)

Max internal counts: +/- 19,999 over signal input range

Max display range: +/- 99,999

Sensitivity:

1.25uV/internal count at input amp gain = 80

Conversion rate: 2.5 samples/second

Filter:

2-pole, 10Hz low pass

Nonlinearity: +/- 0.01% +/- 1 internal count

Step Response: 500ms

CMRR:

120dB @ 50-60Hz

Input impedance: 1000 megohm (min)

Input bias current: 10nA (max)

Zero tempco:

1uV/degC (typ)

Gain tempco:

20ppm/degC (typ)

Operating range:-10 to 50 degC

Input Power: 115 or 230 VAC (solder pad jumper selectable)

Pkg dimensions: 3.78W x 1.89H x 5.13D

Interface Inc, www.interfaceforce.com

1

13-60 Rev. D 9820 Installation & User Manual

INSTALLATION AND WIRING
INSTALLATION
The Model 9820 enclosure is designated for panel mounting in a 1/8 DIN cutout. The cutout dimensions are shown below.
1.772" 45mm (+0.024, -0.000)
92 mm 3.622" (+0.032, -0.000)
To panel mount the 9820, perform the following steps.
1. Rotate the four pawl screws ( outside screws in each corner) several turns counter-clockwise to retract the pawls. Make sure the pawls retract enough to clear the back of the mounting panel. The pawls may be retracted to accomodate panel thicknesses up to 0.25 inches (6.35mm).
2. Insert the instrument into the panel cutout.
3. Position the pawls so that their elongated dimension overlaps the panel cutout, then tighten the screws. Do not over-tighten.
4. Installation complete.
WIRING
Reference Figure 3 for TRANSDUCER wiring information. Power is applied with a 3-prong AC power cord. The instrument is protected by a 250V, 250mA fast acting 5mm fuse. The fuse holder is an integral part of the input power connector. A spare fuse is provided in the fuse holder. Internal solder jumpers are provided to allow 115V AC or 230VAC operation (Reference Figure 6 on page 5).

Interface Inc, www.interfaceforce.com

2

13-60 Rev. D 9820 Installation & User Manual

Pawl Screw (1 of 4 outside screws) These 4 screws are used to mount the instrument in a panel.

Case Screw (1 of 4 inside screws) These 4 screws hold the front panel to the instrument case.

interface
ADVANCED FORCE MEASUREMENT

Mode or Menu Selection Switch S1 (Single Function Switch)

MODE

TARE

CAL

STEP

RCAL

Increment or Tare Selection Switch S2 (Dual Function Switch)

Decrement or Calibrate Selection Switch S3 (Dual Function Switch)

Status LEDS (See Status Definition)
CAL TARE H1 H2
L1 RCAL Switch S5 L2 (Single Function Switch)
Step or Min/Max Recall Selection Switch S4 (Dual Function Switch)

Figure 1. Model 9820 Front Panel View

Fuse Holder (250V, 250mA, 5mm fuse)

1 2 3 4 5 6 7 8 J1
TRANSDUCER

!

Line

50/60 Hz

115V .10A

230v .05A

FUSE 220/240V T250V 250mA

Transducer, Analog Output and RCAL Connector (8-pin Male 1.5mm Terminal Strip Connector)
Figure 2. Model 9820 Rear Panel View

AC Power Connector (3-prong grounded IEC connector)

Interface Inc, www.interfaceforce.com

3

13-60 Rev. D 9820 Installation & User Manual

P1

|

|

|

Transducer

|

|

|

|

|

|

|

|

External RCAL

| |

Analog Output

|

+/-10V(Standard), 4-20mA(Optional) |

Analog Output Return

|

|

Remote Tare

|

|

|

J1

|

7

| |

|

|

4|

|

|

|

6|

3|

5| |

2|

|

1|

|

8|

|

|

Figure 3. Transducer Wiring

Vexc(+)
Sig (+)
Vexc(-) Sig (-) RCAL Analog Out Gnd Remote Tare

Front Panel S3 CAL Switch Enable
1 2
Front Panel S1 MODE Switch Enable

Transformer 1 2 3 4

10V Analog Out Zero Adjust

5V Exc Gain 40 Gain 80 Spare
RCAL

Vexc Adjust
10V Analog Out FS Adjust
Figure 4. Location of Jumpers and Adjustment Potentiometers (Top View)

Interface Inc, www.interfaceforce.com

4

13-60 Rev. D 9820 Installation & User Manual

5 4 back

For RCAL to generate a Positive Output install solder jumper 5 as shown below. For RCAL to generate a Negative Output remove solder jumper 5 and install solder jumper 4.
NOTE: DO NOT INSTALL SOLDER JUMPERS 4 AND 5 AT THE SAME TIME. THIS WILL SHORT OUT Vexc.
Solder jumpers shown (2) for 115VAC operation. For 230VAC operation remove both 115VAC solder jumpers and add (1) solder jumper here.
Note: For use with an external RCAL resistor, remove internal RCAL resistor (see Figure 5).
Figure 5. Location of Jumper (Bottom View)

220 115

Interface Inc, www.interfaceforce.com

5

13-60 Rev. D 9820 Installation & User Manual

SWITCH AND FUNCTION DEFINITIONS

SWITCH DEFINITIONS
interface
ADVANCED FORCE MEASUREMENT

MODE

TARE

CAL

STEP

RCAL

CAL TARE H1
H2 L1 L2

S1 S2 S3 S4
Mode Selection Switch (S1)

S1 is used to cycle through the various setpoints, hysteresis, and calibration values and decimal point locations. The sequence is as follows. The designators, in parenthesis, identify text that will be displayed, momentarily, followed by the value.

Set Point High 1 (SPH1)

Set Point Low 1 (SPL1)

Set Point High 2 (SPH2)

Set Point Low 2 (SPL2)

Hysteresis High (HH)

Hysteresis Low (HL)

Auto Cal

(CAL)

Decimal Point (dP)

Exit

While in the Mode Selection sequence, the indicator is no longer monitoring the input signal. To exit, depress S1, once, after the Decimal Point selection is completed.

Increment or Tare Switch (S2)

This switch serves two functions, depending on the present status of S1. If the instrument is in the Mode Selection sequence, this switch increments the flashing digit.

If the instrument is in its normal operating mode, S2 becomes the TARE switch. If the TARE LED, located on the front panel, is not illuminated, depressing S2 once will automatically zero the readout and illuminate the TARE LED. Depressing S2 again will UNTARE the readout and extinquish the TARE LED. TARING is accomplished by storing the reading prior to TARING and subtracting this value from all subsequent readings.

Interface Inc, www.interfaceforce.com

6

13-60 Rev. D 9820 Installation & User Manual

SWITCH DEFINITIONS (CONT)
Decrement or Calibrate Switch (S3)
This switch serves two functions, depending on the present status of S1. If the instrument is in the Mode Selection sequence, this switch will decrement the flashing digit.
If the instrument is in its operating mode, S3 becomes an AUTO CAL switch. CAUTION: Depressing this switch will cause the instrument to recalibrate its full scale reading. Be sure to read the CALIBRATION INSTRUCTIONS before using this switch.
Step Switch (S4)
This switch also serves two functions, depending on the present status of S1. If the instrument is in the Mode Selection sequence, this switch allows the user to cycle through the digits. Used in conjunction with S2 and S3, it allows rapid updating of the Set Point, Hysteresis and Cal values. If the instrument is in the operating mode, S4 recalls the max (Hi) and min (Lo) values.
RCAL Switch (S5)
This switch activates a reed relay, which places a fixed resistor across Sig(-) and Vexc(-) for a positive RCAL reading, or across Sig(+) and Vexc(-) for a negative RCAL reading. Solder jumpers 4 and 5 determine which RCAL reading is generated (Reference Figure 6).
Front Panel Switch Disable Jumper (Reference Figure 5)
Jumper 1, when removed, disables the CAL Switch (S3). Jumper 2, when removed, disables the MODE Switch (S1).

Interface Inc, www.interfaceforce.com

7

13-60 Rev. D 9820 Installation & User Manual

FUNCTION DEFINITIONS
Set Point High 1 (SPH1)
SPH1 is the present value being constantly compared with the displayed reading. If the magnitude and sign of the reading exceeds the present SPH1 value, the front panel LED (H1) will illuminate. If the displayed reading equals the SPH1 value, nothing will happen. If SPH1 is exceeded, the readout must then drop below SPH1 less the Hysteresis High (HH) value before the H1 LED turns off.
Set Point Low 1 (SPL1)
SPL1 is another preset value being constantly compared with the displayed reading. If the magnitude and sign of the displayed reading is less than the SPL1 present value, the front panel LED (L1) will illuminate. If the displayed reading equals the SPL1 value, nothing will happen. If the displayed reading drops below SPL1, it must then exceed SPL1 plus the Hysteresis Low (HL) value before the L1 LED turns off.
Set Point High 2 (SPH2)
SPH2 is independent and functions identically to SPH1. SPH2 has its own front panel LED (H2).
Set Point Low 2 (SPL2)
SPL2 is independent and functions identically to SPL1. SPL2 has its own front panel LED (L2).
Hysteresis High (HH)
HH is the hysteresis value for SPH1 and SPH2. The hysteresis value determines the number of counts the displayed reading must fall below SPH1 and SPH2 values before deactivating their respective LEDs. The maximum value of hysteresis is 99 counts while the minimum is 00 counts.

Interface Inc, www.interfaceforce.com

8

13-60 Rev. D 9820 Installation & User Manual

FUNCTION DEFINITIONS (CONT)
Hysteresis Low (HL)
HL is the hysteresis value for SPL1 and SPL2. The hysteresis value determines the number of counts the displayed reading must exceed SPL1 and SPL2 values before deactivating their respective LEDs. The maximum value of hysteresis is 99 counts while the minimum is 00 counts.
Auto Cal (CAL)
This input allows the user to calibrate the full scale readout to any desired engineering units. The CAL number is entered in the same way as the Set Point and Hysteresis values. This input must be entered prior to performing the calibration. The CAL number may be any number from 1 to 99,999. Zero is an invalid CAL number. For best stability and performance, keep the maximum CAL number below 19,999 counts.
Decimal Point (dP)
dP allows selection of decimal point locations for display. The locations selectable are x.xxxx, xx.xxx, xxx.xx, xxxx.x and xxxxx.
Min/Max Recall (HI/LO)
The HI and LO values are the maximum and minimum values, respectively, of the displayed readings. Sign and magnitude are constantly compared with the displayed readings at the instrument update rate. The maximum value is initialized to -99,999 and the minimum value to 99.999 each time the TARE switch (S2) is depressed, whether to TARE or UNTARE the readout.

Interface Inc, www.interfaceforce.com

9

13-60 Rev. D 9820 Installation & User Manual

OPERATING INSTRUCTIONS

TURN-ON INSTRUCTIONS

Reference Figures 3 and 4 for proper input and output signal wiring.

1. Apply power to the instrument.

2. Verify the instrument readout blanks momentarily, then displays the value of the input signal.

PROGRAM INSTRUCTIONS

Four front panel switches (S1, S2, S3 and S4) allow the user to program Set Point, Hysteresis, Calibration Number (CAL) and Decimal Points and to monitor minimum, maximum or real time values.

interface
ADVANCED FORCE MEASUREMENT

MODE

TARE

CAL

STEP

RCAL

CAL TARE H1 H2
L1 L2

S1 S2 S3 S4
1. Depress MODE (S1) switch once.
The text SPH1 followed by the current value of Set Point High 1 (SPH1) will be displayed. The system will idle in this condition until S2, S3 or S4 is depressed. If S2 or S3 is depressed, the most significant digit will be incremented (S2) or decremented (S3). Incrementing the most significant digit will display numbers from -9 to 9.
Depressing S4 will select the next least significant digit position for updating. The digit being updated will be flashing. Using S2, S3, and S4, set Set Point High 1 to the desired value.

Interface Inc, www.interfaceforce.com

10

13-60 Rev. D 9820 Installation & User Manual

PROGRAM INSTRUCTIONS (CONT)
2. Depress MODE (S1) switch once.
The text SPL1 followed by the current value of Set Point Low 1 will be displayed. Repeat Step 1 to set Set Point Low 1 to the desired value.
3. Depress MODE (S1) switch once.
The text SPH2 followed by the current value of Set Point High 2 will be displayed. Repeat Step 1 to set Set Point High 2 to the desired value.
4. Depress MODE (S1) switch once.
The text SPL2 followed by the current value of Set Point Low 2 will be displayed. Repeat Step 1 to set Set Point Low 2 to the desired value.
5. Depress MODE (S1) switch once.
The text HH followed by the current value of Hysteresis High will be displayed. Repeat Step 1 to set Hysteresis High to the desired value. The maximum value for HH is 99 counts.
6. Depress MODE (S1) switch once.
The text HL followed by the current value of Hysteresis Low will be displayed. Repeat Step 1 to set Hysteresis Low to the desired value. The maximum value for HL is 99 counts.
7. Depress MODE (S1) switch once.
The text CAL followed by the current CAL number will be displayed. Repeat Step 1 to set the CAL number. The maximum CAL number is 99,999.

Interface Inc, www.interfaceforce.com

11

13-60 Rev. D 9820 Installation & User Manual

PROGRAM INSTRUCTIONS (CONT)
8. Depress MODE (S1) switch once.
The text dP followed by the current decimal position will be shown flashing. Increment (S2) or decrement (S3) the decimal point to the desired location.
9. Depress MODE (S1) switch once.
The instrument will exit the Mode Selection Sequence and enter the normal operating mode (i.e. monitoring the input signal) This is the end of the PROGRAM INSTRUCTIONS.

Interface Inc, www.interfaceforce.com

12

13-60 Rev. D 9820 Installation & User Manual

CALIBRATION INSTRUCTIONS
The Interface Model 9820 can be calibrated using (2) different methods.
1. The first method uses actual or simulated ZERO and FULL SCALE (FS) inputs. Actual inputs may come from a strain gage transducer loaded with calibrated weights. Simulated inputs may come from a strain gage bridge simulator.
2. The second method utilizes an RCAL resistor to simulate a load on a strain gage bridge. The RCAL resistor is activated when the RCAL (S5) switch is depressed. Reference Figure 3 for location of S5.
ACTUAL OR SIMULATED LOAD METHOD
1. Connect the strain gage transducer to the instrument using Figure 3 as a reference. Do not apply any load to the transducer.
2. Apply power to the instrument.
NOTE: If the instrument was previously calibrated and no new calibration is required, do not proceed any further. All previous calibration information is stored in non-volatile EEPROM and is ready for use when power to the instrument is reapplied.
3. Calculate the CAL NUMBER by verifying the load that will be used to calibrate the full scale of the Model 9820.
Example: If a 500 pound load is used to calibrate full scale and the desired readout is 500.0, the CAL NUMBER will be 5000. The decimal point can be selected later. If the desired readout is 500 then the CAL NUMBER will be 500.
4. Program in the CAL NUMBER. Reference PROGRAM INSTRUCTIONS, Step 7. If the CAL NUMBER was previously programmed, skip this step.

Interface Inc, www.interfaceforce.com

13

13-60 Rev. D 9820 Installation & User Manual

CALIBRATION INSTRUCTIONS (CONT.)
5. The TARE and CAL LED's on the front panel should be illuminated.
6. Depress the TARE(S2) switch once. The TARE LED should extinquish. The instrument should now display the UNTARED reading.
7. Depress the CAL (S3) switch once. Hold for 3 seconds. The CAL LED should extinquish. The instrument should now display the UNCALIBRATED reading.
8. With no load on the transducer, verify the instrument reading is low.
9. Depress the TARE(S2) switch once. The TARE LED should illuminate and the readout should be ZERO +/- 1 digit.
10. Apply the FULL SCALE load to the transducer. If a strain gage simulator is used, set it to the the desired mV/V setting. The instrument should display a value greater than 01000. For optimum performance the readout should be closer to 19000. Ignore decimal point.
11. Depress the CAL (S3) switch once. Hold for 3 seconds. The CAL LED should turn on and the readout should display the CAL NUMBER +/- 1 digit.
12. The CALIBRATION sequence is now complete.
NOTE: Calibration values (tare and slope values) are maintained in nonvolatile memory and are updated every time CALIBRATION Steps 9 and 11 are performed. TARING the display does not change the scale factor (slope value) calculated when Step 11 is completed.
S3 has a 3 second delay feature to prevent inadvertent loss of calibration.

Interface Inc, www.interfaceforce.com

14

13-60 Rev. D 9820 Installation & User Manual

CALIBRATION INSTRUCTIONS (CONT.)

RCAL METHOD

1. Connect the strain gage transducer to the instrument using Figure 3 as a reference. Do not apply any load to the transducer.

2. Verify that either an internal or external RCAL resistor is provided

with the instrument. If an internal RCAL resistor is provided, no

installation is required. Proceed to Step 3. If an external RCAL

resistor is provided, install it across pins 5 and 6 (Positive

Reading) or pins 5 and 7 (Negative Reading). Verify internal

RCAL resistor is removed. P1

|

|

|

|

|

Transducer

|

|

|

|

|

|

| | | | | | |

|

|

External RCAL

J1

|

|

7

| |

|

|

|

|

4| |

|

|

|

|
6|

|

|

|

5| |

Vexc(+)
Sig (+)
Vexc(-) Sig (-) RCAL

3. Apply power to the instrument.
NOTE: If the instrument was previously calibrated and no new calibration is required, do not proceed any further. All previous calibration information is stored in non-volatile EEPROM and is ready for use when power to the instrument is reapplied.
4. Set the CAL NUMBER to the full scale value provided with the RCAL resistor.
Example: If the RCAL resistor supplied with the instrument creates an output equivalent to 80% of a full scale 500 pound load, set the CAL NUMBER to 400.00, 0400.0 or 00400.

Interface Inc, www.interfaceforce.com

15

13-60 Rev. D 9820 Installation & User Manual

CALIBRATION INSTRUCTIONS (CONT.)
5. The TARE and CAL LED's on the front panel should be illuminated.
6. Depress the TARE (S2) switch once. The TARE LED should extinquish. The instrument should now display the UNTARED reading.
7. Depress the CAL(S3) switch once. (Hold for 3 seconds) The CAL LED should extinquish. The instrument should now display the UNCALIBRATED reading.
8. With no load on the transducer, verify the readout is approximately zero.
9. Depress the TARE(S2) switch once. The TARE LED should illuminate and the readout should be ZERO +/- 1 digit.
10. Depress and hold the RCAL switch. The display should read a value greater than 01000. For optimum performance the readout should be closer to 19000. Ignore decimal point.
While holding down the RCAL switch, depress the CAL (S3) switch once. The CAL LED should turn on and the readout should display the CAL NUMBER +/- 1 digit.
11. The RCAL CALIBRATION sequence is now complete.
NOTES: Calibration values (tare and slope values) are maintained in non-volatile memory and are updated every time CALIBRATION Steps 9 and 10 are performed. TARING the display does not change the scale factor (slope value) calculated when Step 10 is completed.
S3 has a 3 second delay feature to prevent the inadvertent loss of calibration.

Interface Inc, www.interfaceforce.com

16

13-60 Rev. D 9820 Installation & User Manual

MODEL 9820 PROGRAM MENU DIAGRAM

SETUP PROCEDURE
START

CALIBRATION PROCEDURE USING ACTUAL OR SIMULATED LOADS
START

Depress MODE

SPH1

Set SPH1

Depress MODE

SPL1

Set SPL1

Depress MODE

SPH2

Set SPH2

Depress MODE

SPL2

Set SPL2

Depress MODE

HH

Set HH

Depress MODE

HL

Set HL

Depress MODE

CAL

Set CAL

Depress MODE

dP

Set dP

Depress MODE

Depress
Depress
Depress STEP Depress
Depress
Depress STEP Depress
Depress
Depress STEP Depress
Depress
Depress STEP Depress
Depress
Depress STEP Depress
Depress
Depress STEP Depress
Depress
Depress STEP Depress
Depress
Depress STEP

Increments Flashing Digit
Decrements Flashing Digit
Steps Flashing Digit to Right
Increments Flashing Digit
Decrements Flashing Digit
Steps Flashing Digit to Right
Increments Flashing Digit
Decrements Flashing Digit
Steps Flashing Digit to Right
Increments Flashing Digit
Decrements Flashing Digit
Steps Flashing Digit to Right
Increments Flashing Digit
Decrements Flashing Digit
Steps Flashing Digit to Right
Increments Flashing Digit
Decrements Flashing Digit
Steps Flashing Digit to Right
Increments Flashing Digit
Decrements Flashing Digit
Steps Flashing Digit to Right
Shifts Decimal Pt to the Right
Shifts Decimal Pt to the Left
No Change

Is TARE LED on?
NO

YES

Depress

TARE Led turns off

Is CAL LED on?
NO

YES

Depress

CAL Led turns off

Apply input signal for zero reading on display

Depress

Readout displays 0000

Apply full scale input signal

TARE Led turns on

Depress

Readout displays CAL

CAL Led turns on

CALIBRATION PROCEDURE USING RCAL RESISTOR
START

Is TARE LED on?
NO

YES

Depress

TARE Led turns off

Is CAL LED on?
NO

YES

Depress

CAL Led turns off

Apply no load to the strain gage transducer

Depress

Readout displays 0000

Depress RCAL

&

switch

Readout displays CAL

TARE Led turns on
CAL Led turns on

OPERATE

Interface Inc, www.interfaceforce.com

17

13-60 Rev. D 9820 Installation & User Manual

Warranty and Repair Policy
1. Interface warrants that its instruments shall be free from defects in material and workmanship for one year under normal and proper use when correctly installed. 2. Any Interface product which proves defective in material or in workmanship within one year from date of shipment by Interface, will be repaired or replaced free of charge provided that (1) buyer provides Interface with satisfactory proof of the defect and that the product was properly installed, maintained, and operated within the limits of rated and normal usage; (2) buyer obtains from Interface authorization to return the product; and (3) products claimed to be defective must be returned with transportation charges prepaid. Product will be returned to Buyer with transportation charges collect unless the item is found to be defective, in which case Interface will pay the return transportation charges. 3. The remedy set forth herein does not apply to damage to or defects in any product caused by the Buyer's misuse or neglect, nor does it apply to any product which has been repaired or disassembled, which in the sole judgement of Interface affects the performance of the product. 4. Interface makes no warranty concerning components not manufactured by it. However, in the event of the failure of any component or accessory not manufactured by Interface, reasonable assistance will be given to Buyer in obtaining from the respective manufacturer whatever adjustment is reasonable based on the manufacturer's own warranty. Interface expressly disclaims any liability to its customers, dealers, and representatives, and to users of its products, and to any other person for special or consequential damages of any kind and from any cause whatsoever arising out of or in any way connected with the manufacture, sale, handling, repair, maintenance, or replacement arising out of or in anyway connected with the use of said products. 5. Representations and warranties made by any person, including dealers and representatives of Interface, which are inconsistent or in conflict with the terms of this warranty (including but not limited to the limitations of the liability of Interface, as set forth above), shall not be binding upon Interface unless reduced to writing and approved by an officer of Interface, Inc.
THIS EXPRESS WARRANTY SUPERCEDES ANY AND ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE.
OBTAINING SERVICE UNDER WARRANTY Advance authorization is required before any product is returned to Interface. Prior to the return of any product, write or call the Repair Department at Interface advising them of; (1) a part number; (2) a serial number of the defective product; (3) a technical description of the defect including specific test data, written observations on the failure and specific corrective action required; (4) a no-charge purchase order number (so the product can be returned to sender correctly); and (5) ship and bill addresses. Non-verified problems or defects may be subject to an evaluation charge. Please return the original calibration data with the unit.
REPAIR WARRANTY All repairs of Interface products are warranted for a period of 90 days from date of shipment. This warranty applies only to those items which were found defective and repaired. It does not apply to products in which no defect was found and returned as is or merely recalibrated. Out of warranty products may not be capable of being