
User Manual for TiePie engineering models including: HS4 DIFF Differential USB Oscilloscope, HS4, DIFF Differential USB Oscilloscope, Differential USB Oscilloscope, USB Oscilloscope, Oscilloscope
File Info : application/pdf, 32 Pages, 359.21KB
DocumentDocumentHandyscope HS4 User manual TiePie engineering ATTENTION! Measuring directly on the line voltage can be very dangerous. The outside of the BNC connectors at the Handyscope HS4 are connected with the ground of the computer. Use a good isolation transformer or a differential probe when measuring at the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS4 is connected to a positive voltage. This short-circuit current can damage both the Handyscope HS4 and the computer. Copyright ©2024 TiePie engineering. All rights reserved. Revision 2.45, February 2024 This information is subject to change without notice. Despite the care taken for the compilation of this user manual, TiePie engineering can not be held responsible for any damage resulting from errors that may appear in this manual. Contents 1 Safety 1 2 Declaration of conformity 3 3 Introduction 5 3.1 Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 3.2 Sampling rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 3.2.1 Aliasing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 3.3 Digitizing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 3.4 Signal coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 3.5 Probe compensation . . . . . . . . . . . . . . . . . . . . . . . . . 9 4 Driver installation 11 4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 4.2 Computers running Windows 10 . . . . . . . . . . . . . . . . . . 11 4.3 Computers running Windows 8 or older . . . . . . . . . . . . . . 11 4.3.1 Where to find the driver setup . . . . . . . . . . . . . . . 11 4.3.2 Executing the installation utility . . . . . . . . . . . . . . 11 5 Hardware installation 15 5.1 Power the instrument . . . . . . . . . . . . . . . . . . . . . . . . 15 5.1.1 External power . . . . . . . . . . . . . . . . . . . . . . . 15 5.2 Connect the instrument to the computer . . . . . . . . . . . . . 15 5.3 Plug into a different USB port . . . . . . . . . . . . . . . . . . . . 15 6 Front panel 17 6.1 Channel input connectors . . . . . . . . . . . . . . . . . . . . . . 17 6.2 Power indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 7 Rear panel 19 7.1 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 7.1.1 USB power cable . . . . . . . . . . . . . . . . . . . . . . 19 7.1.2 Power adapter . . . . . . . . . . . . . . . . . . . . . . . . 20 Contents I 7.2 USB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 7.3 Extension Connector . . . . . . . . . . . . . . . . . . . . . . . . . 21 8 Specifications 23 8.1 Acquisition system . . . . . . . . . . . . . . . . . . . . . . . . . . 23 8.2 Trigger system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 8.3 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 8.4 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 8.5 Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 8.6 I/O connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 8.7 System requirements . . . . . . . . . . . . . . . . . . . . . . . . . 24 8.8 Environmental conditions . . . . . . . . . . . . . . . . . . . . . . 25 8.9 Certifications and Compliances . . . . . . . . . . . . . . . . . . . 25 8.10 Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 8.11 Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 II Safety 1 When working with electricity, no instrument can guarantee complete safety. It is the responsibility of the person who works with the instrument to operate it in a safe way. Maximum security is achieved by selecting the proper instruments and following safe working procedures. Safe working tips are given below: · Always work according (local) regulations. · Work on installations with voltages higher than 25 VAC or 60 VDC should only be performed by qualified personnel. · Avoid working alone. · Observe all indications on the Handyscope HS4 before connecting any wiring · Check the probes/test leads for damages. Do not use them if they are damaged · Take care when measuring at voltages higher than 25 VAC or 60 VDC. · Do not operate the equipment in an explosive atmosphere or in the pres- ence of flammable gases or fumes. · Do not use the equipment if it does not operate properly. Have the equipment inspected by qualified service personal. If necessary, return the equipment to TiePie engineering for service and repair to ensure that safety features are maintained. · Measuring directly on the line voltage can be very dangerous. The outside of the BNC connectors at the Handyscope HS4 are connected with the ground of the computer. Use a good isolation transformer or a differential probe when measuring at the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS4 is connected to a positive voltage. This short-circuit current can damage both the Handyscope HS4 and the computer. Safety 1 Declaration of conformity TiePie engineering Koperslagersstraat 37 8601 WL Sneek The Netherlands EC Declaration of conformity We declare, on our own responsibility, that the product Handyscope HS4-5MHz Handyscope HS4-10MHz Handyscope HS4-25MHz Handyscope HS4-50MHz 2 for which this declaration is valid, is in compliance with EC directive 2011/65/EU (the RoHS directive) including up to amendment 2021/1980, EC regulation 1907/2006 (REACH) including up to amendment 2021/2045, and with EN 55011:2016/A1:2017 EN 55022:2011/C1:2011 IEC 61000-6-1:2019 EN IEC 61000-6-3:2007/A1:2011/C11:2012 EN according the conditions of the EMC standard 2004/108/EC, also with Canada: ICES-001:2004 Australia/New Zealand: AS/NZS CISPR 11:2011 and IEC 61010-1:2010/A1:2019 USA: UL 61010-1, Edition 3 and is categorized as 30 Vrms, 42 Vpk, 60 Vdc Sneek, 1-9-2022 ir. A.P.W.M. Poelsma Declaration of conformity 3 Environmental considerations This section provides information about the environmental impact of the Handyscope HS4. End-of-life handling Production of the Handyscope HS4 required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the Handyscope HS4's end of life. In order to avoid release of such substances into the environment and to reduce the use of natural resources, recycle the Handyscope HS4 in an appropriate system that will ensure that most of the materials are reused or recycled appropriately. The shown symbol indicates that the Handyscope HS4 complies with the European Union's requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). 4 Chapter 2 Introduction 3 Before using the Handyscope HS4 first read chapter 1 about safety. Many technicians investigate electrical signals. Though the measurement may not be electrical, the physical variable is often converted to an electrical signal, with a special transducer. Common transducers are accelerometers, pressure probes, current clamps and temperature probes. The advantages of converting the physical parameters to electrical signals are large, since many instruments for examining electrical signals are available. The Handyscope HS4 is a portable four channel measuring instrument. The Handyscope HS4 is available in several models with different maximum sampling rates. The native resolution is 12 bits, but user selectable resolutions of 14 and 16 bits are available too, with reduced maximum sampling rate: resolution 12 bit 14 bit 16 bit Model 50 50 MSa/s 3.125 MSa/s 195 kSa/s Model 25 25 MSa/s 3.125 MSa/s 195 kSa/s Model 10 10 MSa/s 3.125 MSa/s 195 kSa/s Model 5 5 MSa/s 3.125 MSa/s 195 kSa/s Table 3.1: Maximum sampling rates The Handyscope HS4 supports high speed continuous streaming measurements. The maximum streaming rates are: resolution 12 bit 14 bit 16 bit Model 50 500 kSa/s 480 kSa/s 195 kSa/s Model 25 250 kSa/s 250 kSa/s 195 kSa/s Model 10 100 kSa/s 99 kSa/s 97 kSa/s Model 5 50 kSa/s 50 kSa/s 48 kSa/s Table 3.2: Maximum streaming rates With the accompanying software the Handyscope HS4 can be used as an oscilloscope, a spectrum analyzer, a true RMS voltmeter or a transient recorder. All instruments measure by sampling the input signals, digitizing the values, process them, save them and display them. 3.1 Sampling When sampling the input signal, samples are taken at fixed intervals. At these intervals, the size of the input signal is converted to a number. The accuracy of this number depends on the resolution of the instrument. The higher the resolution, Introduction 5 the smaller the voltage steps in which the input range of the instrument is divided. The acquired numbers can be used for various purposes, e.g. to create a graph. Figure 3.1: Sampling The sine wave in figure 3.1 is sampled at the dot positions. By connecting the adjacent samples, the original signal can be reconstructed from the samples. You can see the result in figure 3.2. Figure 3.2: "connecting" the samples 3.2 Sampling rate The rate at which the samples are taken is called the sampling rate, the number of samples per second. A higher sampling rate corresponds to a shorter interval between the samples. As is visible in figure 3.3, with a higher sampling rate, the original signal can be reconstructed much better from the measured samples. 6 Chapter 3 Figure 3.3: The effect of the sampling rate 3.2.1 The sampling rate must be higher than 2 times the highest frequency in the input signal. This is called the Nyquist frequency. Theoretically it is possible to reconstruct the input signal with more than 2 samples per period. In practice, 10 to 20 samples per period are recommended to be able to examine the signal thoroughly. Aliasing When sampling an analog signal with a certain sampling rate, signals appear in the output with frequencies equal to the sum and difference of the signal frequency and multiples of the sampling rate. For example, when the sampling rate is 1000 Sa/s and the signal frequency is 1250 Hz, the following signal frequencies will be present in the output data: Multiple of sampling rate ... -1000 0 1000 2000 ... 1250 Hz signal -1000 + 1250 = 250 0 + 1250 = 1250 1000 + 1250 = 2250 2000 + 1250 = 3250 -1250 Hz signal -1000 - 1250 = -2250 0 - 1250 = -1250 1000 - 1250 = -250 2000 - 1250 = 750 Table 3.3: Aliasing As stated before, when sampling a signal, only frequencies lower than half the sampling rate can be reconstructed. In this case the sampling rate is 1000 Sa/s, so we can we only observe signals with a frequency ranging from 0 to 500 Hz. This means that from the resulting frequencies in the table, we can only see the 250 Hz signal in the sampled data. This signal is called an alias of the original signal. If the sampling rate is lower than twice the frequency of the input signal, aliasing will occur. The following illustration shows what happens. Introduction 7 Figure 3.4: Aliasing In figure 3.4, the green input signal (top) is a triangular signal with a frequency of 1.25 kHz. The signal is sampled with a rate of 1 kSa/s. The corresponding sampling interval is 1/1000Hz = 1ms. The positions at which the signal is sampled are depicted with the blue dots. The red dotted signal (bottom) is the result of the reconstruction. The period time of this triangular signal appears to be 4 ms, which corresponds to an apparent frequency (alias) of 250 Hz (1.25 kHz - 1 kHz). To avoid aliasing, always start measuring at the highest sampling rate and lower the sampling rate if required. 3.3 Digitizing When digitizing the samples, the voltage at each sample time is converted to a number. This is done by comparing the voltage with a number of levels. The resulting number is the number corresponding to the level that is closest to the voltage. The number of levels is determined by the resolution, according to the following relation: LevelCount = 2Resolution. The higher the resolution, the more levels are available and the more accurate the input signal can be reconstructed. In figure 3.5, the same signal is digitized, using two different amounts of levels: 16 (4-bit) and 64 (6-bit). 8 Chapter 3 Figure 3.5: The effect of the resolution The Handyscope HS4 measures at e.g. 12 bit resolution (212=4096 levels). The smallest detectable voltage step depends on the input range. This voltage can be calculated as: V oltageStep = F ullInputRange/LevelCount For example, the 200 mV range ranges from -200 mV to +200 mV, therefore the full range is 400 mV. This results in a smallest detectable voltage step of 0.400 V / 4096 = 97.65 µV. 3.4 Signal coupling The Handyscope HS4 has two different settings for the signal coupling: AC and DC. In the setting DC, the signal is directly coupled to the input circuit. All signal components available in the input signal will arrive at the input circuit and will be measured. In the setting AC, a capacitor will be placed between the input connector and the input circuit. This capacitor will block all DC components of the input signal and let all AC components pass through. This can be used to remove a large DC component of the input signal, to be able to measure a small AC component at high resolution. When measuring DC signals, make sure to set the signal coupling of the input to DC. 3.5 Probe compensation The Handyscope HS4 is shipped with a probe for each input channel. These are 1x/10x selectable passive probes. This means that the input signal is passed through directly or 10 times attenuated. Introduction 9 When using an oscilloscope probe in 1:1 the setting, the bandwidth of the probe is only 6 MHz. The full bandwidth of the probe is only obtained in the 1:10 setting The x10 attenuation is achieved by means of an attenuation network. This attenuation network has to be adjusted to the oscilloscope input circuitry, to guarantee frequency independency. This is called the low frequency compensation. Each time a probe is used on an other channel or an other oscilloscope, the probe must be adjusted. Therefore the probe is equiped with a setscrew, with which the parallel capacity of the attenuation network can be altered. To adjust the probe, switch the probe to the x10 and attach the probe to a 1 kHz square wave signal. Then adjust the probe for a square front corner on the square wave displayed. See also the following illustrations. Figure 3.6: correct Figure 3.7: under compensated 10 Chapter 3 Figure 3.8: over compensated Driver installation 4 Before connecting the Handyscope HS4 to the computer, the drivers need to be installed. 4.1 4.2 4.3 4.3.1 4.3.2 Introduction To operate a Handyscope HS4, a driver is required to interface between the measurement software and the instrument. This driver takes care of the low level communication between the computer and the instrument, through USB. When the driver is not installed, or an old, no longer compatible version of the driver is installed, the software will not be able to operate the Handyscope HS4 properly or even detect it at all. Computers running Windows 10 When the Handyscope HS4 is plugged in into a USB port of the computer, Windows will detect the instrument and will download the required driver from Windows Update. When the download is finished, the driver will be installed automatically. Computers running Windows 8 or older The installation of the USB driver is done in a few steps. Firstly, the driver has to be pre-installed by the driver setup program. This makes sure that all required files are located where Windows can find them. When the instrument is plugged in, Windows will detect new hardware and install the required drivers. Where to find the driver setup The driver setup program and measurement software can be found in the download section on TiePie engineering's website. It is recommended to install the latest version of the software and USB driver from the website. This will guarantee the latest features are included. Executing the installation utility To start the driver installation, execute the downloaded driver setup program. The driver install utility can be used for a first time installation of a driver on a system and also to update an existing driver. The screen shots in this description may differ from the ones displayed on your computer, depending on the Windows version. Driver installation 11 Figure 4.1: Driver install: step 1 When drivers were already installed, the install utility will remove them before installing the new driver. To remove the old driver successfully, it is essential that the Handyscope HS4 is disconnected from the computer prior to starting the driver install utility. When the Handyscope HS4 is used with an external power supply, this must be disconnected too. Clicking "Install" will remove existing drivers and install the new driver. A remove entry for the new driver is added to the software applet in the Windows control panel. Figure 4.2: Driver install: Copying files 12 Chapter 4 Figure 4.3: Driver install: Finished Driver installation 13 14 Chapter 4 Hardware installation 5 Drivers have to be installed before the Handyscope HS4 is connected to the computer for the first time. See chapter 4 for more information. 5.1 5.1.1 5.2 5.3 Power the instrument The Handyscope HS4 is powered by the USB, no external power supply is required. Only connect the Handyscope HS4 to a bus powered USB port, otherwise it may not get enough power to operate properly. External power In certain cases, the Handyscope HS4 cannot get enough power from the USB port. When a Handyscope HS4 is connected to a USB port, powering the hardware will result in an inrush current higher than the nominal current. After the inrush current, the current will stabilize at the nominal current. USB ports have a maximum limit for both the inrush current peak and the nominal current. When either of them is exceeded, the USB port will be switched off. As a result, the connection to the Handyscope HS4 will be lost. Most USB ports can supply enough current for the Handyscope HS4 to work without an external power supply, but this is not always the case. Some (battery operated) portable computers or (bus powered) USB hubs do not supply enough current. The exact value at which the power is switched off, varies per USB controller, so it is possible that the Handyscope HS4 functions properly on one computer, but does not on another. In order to power the Handyscope HS4 externally, an external power input is provided for. It is located at the rear of the Handyscope HS4. Refer to paragraph 7.1 for specifications of the external power intput. Connect the instrument to the computer After the new driver has been pre-installed (see chapter 4), the Handyscope HS4 can be connected to the computer. When the Handyscope HS4 is connected to a USB port of the computer, Windows will detect new hardware. Depending on the Windows version, a notification can be shown that new hardware is found and that drivers will be installed. Once ready, Windows will report that the driver is installed. When the driver is installed, the measurement software can be installed and the Handyscope HS4 can be used. Plug into a different USB port When the Handyscope HS4 is plugged into a different USB port, some Windows versions will treat the Handyscope HS4 as different hardware and will install the drivers again for that port. This is controlled by Microsoft Windows and is not caused by TiePie engineering. Hardware installation 15 16 Chapter 5 Front panel 6 Figure 6.1: Front panel 6.1 Channel input connectors The CH1 CH4 BNC connectors are the main inputs of the acquisition system. The outside of all four BNC connectors is connected to the ground of the Handyscope HS4. Connecting the outside of the BNC connector to a potential other than ground will result in a short circuit that may damage the device under test, the Handyscope HS4 and the computer. 6.2 Power indicator A power indicator is situated at the top cover of the instrument. It is lit when the Handyscope HS4 is powered. Front panel 17 18 Chapter 6 Rear panel 7 Figure 7.1: Rear panel 7.1 Power The Handyscope HS4 is powered through the USB. If the USB cannot deliver enough power, it is possible to power the instrument externally. The Handyscope HS4 has two external power inputs located at the rear of the instrument: the dedicated power input and a pin of the extension connector. The specifications of the dedicated power connector are: Pin Center pin Outside bushing Dimension Ø1.3 mm Ø3.5 mm Description ground positive Figure 7.2: Power connector Besides the external power input, it is also possible to power the instrument through the extension connector, the 25 pin D-sub connector at the rear of the instrument. The power has to be applied to pin 3 of the extension connector. Pin 4 can be used as ground. The following minimum and maximum voltages apply to both power inputs: Minimum 4.5 VDC Maximum 14 VDC Table 7.1: Maximum voltages 7.1.1 Note that the externally applied voltage should be higher than the USB voltage to relieve the USB port. USB power cable The Handyscope HS4 is delivered with a special USB external power cable. Rear panel 19 Figure 7.3: USB power cable One end of this cable can be connected to a second USB port on the computer, the other end can be plugged in the external power input at the rear of the instrument. The power for the instrument will be taken from two USB ports of the computer. The outside of the external power connector is connected to +5 V. In order to avoid shortage, first connect the cable to the Handyscope HS4 and then to the USB port. 7.1.2 Power adapter In case a second USB port is not available, or the computer still can't provide enough power for the instrument, an external power adapter can be used. When using an external power adapter, make sure that: · the polarity is set correctly · the voltage is set to a valid value for the instrument and higher than the USB voltage · the adapter can supply enough current (preferably >1 A) · the plug has the correct dimensions for the external power input of the in- strument 7.2 USB The Handyscope HS4 is equipped with a USB 2.0 High speed (480 Mbit/s) interface with a fixed cable with type A plug. It will also work on a computer with a USB 1.1 interface, but will then operate at 12 Mbit/s. 20 Chapter 7 7.3 Extension Connector Figure 7.4: Extension connector To connect to the Handyscope HS4 a 25 pin female D-sub connector is available, containing the following signals: Pin Description 1 Ground 2 Reserved 3 External Power in DC 4 Ground 5 +5V out, 10 mA max. 6 Ext. sampling clock in (TTL) 7 Ground 8 Ext. trigger in (TTL) 9 Data OK out (TTL) 10 Ground 11 Trigger out (TTL) 12 Reserved 13 Ext. sampling clock out (TTL) Pin Description 14 Ground 15 Ground 16 Reserved 17 Ground 18 Reserved 19 Reserved 20 Reserved 21 Reserved 22 Ground 23 I2C SDA 24 I2C SCL 25 Ground Table 7.2: Pin description Extension connector All TTL signals are 3.3 V TTL signals which are 5 V tolerant, so they can be connected to 5 V TTL systems. Pins 9, 11, 12, 13 are open collector outputs. Connect a pull-up resistor of 1 kOhm to pin 5 when using one of these signals. Rear panel 21 22 Chapter 7 Specifications 8.1 Acquisition system 8 Number of input channels 4 analog CH1, CH2, CH3, CH4 BNC, female Type Single ended Resolution 12, 14, 16 bit user selectable Accuracy 0.2% of full scale ± 1 LSB Ranges (full scale) ±200 mV ±2 V ±20 V ±400 mV ±4 V ±40 V Coupling AC/DC Impedance 1 M / 30 pF Maximum voltage 200 V (DC + AC peak <10 kHz) Bandwidth (-3dB) 50 MHz AC coupling cut off frequency (-3dB)±1.5 Hz Maximum sampling rate HS4-50 HS4-25 HS4-10 HS4-5 12 bit 50 MSa/s 25 MSa/s 10 MSa/s 5 MSa/s 14 bit 3.125 MSa/s 3.125 MSa/s 3.125 MSa/s 3.125 MSa/s 16 bit 195 kSa/s 195 kSa/s 195 kSa/s 195 kSa/s Maximum streaming rate HS4-50 HS4-25 HS4-10 HS4-5 12 bit 500 kSa/s 250 kSa/s 100 kSa/s 50 kSa/s 14 bit 480 kSa/s 250 kSa/s 99 kSa/s 50 kSa/s 16 bit 195 kSa/s 195 kSa/s 97 kSa/s 48 kSa/s Sampling source internal quartz, external Internal Quartz Accuracy Stability ±0.01% ±100 ppm over -40C to +85C External On extension connector Voltage 3.3 V TTL, 5 V TTL tolerant Frequency range 95 MHz to 105 MHz Memory 128 kSamples per channel ±800 mV ±8 V ±80 V Specifications 23 8.2 Trigger system 8.3 System Source Trigger modes Level adjustment Hysteresis adjustment Resolution Pre trigger Post trigger Trigger hold-off Digital external trigger Input Range Coupling Interface digital, 2 levels CH1, CH2, CH3, CH4, digital external, AND, OR rising slope, falling slope, inside window, outside window 0 to 100% of full scale 0 to 100% of full scale 0.024 % (12 bits) 0 to 128 ksamples (0 to 100%, one sample resolution) 0 to 128 ksamples (0 to 100%, one sample resolution) 0 to 1 MSamples, 1 sample resolution extension connector 0 to 5 V (TTL) DC Interface 8.4 Power USB 2.0 High Speed (480 Mbit/s) (USB 1.1 Full Speed (12 Mbit/s) and USB 3.0 compatible) Input Consumption 8.5 Physical from USB or external input 500 mA max Instrument height Instrument length Instrument width Weight USB cord length 8.6 I/O connectors 25 mm / 1.0" 170 mm / 6.7" 140 mm / 5.2" 480 gram / 17 ounce 1.8 m / 70" CH1 .. CH4 Power Extension connector USB BNC, female 3.5 mm power socket D-sub 25 pins female Fixed cable with type A plug 8.7 System requirements PC I/O connection Operating System USB 2.0 High Speed (480 Mbit/s) (USB 1.1 Full Speed (12 Mbit/s) and USB 3.0 compatible) Windows 10, 32 and 64 bits 24 Chapter 8 8.8 Environmental conditions Operating Ambient temperature Relative humidity Storage Ambient temperature Relative humidity 0C to 55C 10 to 90% non condensing -20C to 70C 5 to 95% non condensing 8.9 Certifications and Compliances CE mark compliance Yes RoHS Yes REACH Yes EN 55011:2016/A1:2017 Yes EN 55022:2011/C1:2011 Yes IEC 61000-6-1:2019 EN Yes IEC 61000-6-3:2007/A1:2011/C11:2012 Yes ICES-001:2004 Yes AS/NZS CISPR 11:2011 Yes IEC 61010-1:2010/A1:2019 Yes UL 61010-1, Edition 3 Yes 8.10 Probes Model Bandwidth Rise time Input impedance Input capacitance Compensation range Working voltage (DC + AC peak) 8.11 Package contents HP-3250I X1 6 MHz 58 ns 1 M oscilloscope impedance 56 pF + oscilloscope capacitance - 300 V 150 V CAT II X10 250 MHz 1.4 ns 10 M incl. 1 M oscilloscope impedance 13 pF 10 to 30 pF 600 V 300 V CAT II Instrument Probes Accessories Software Drivers Software Development Kit Manual Handyscope HS4 4 x HP-3250I X1 / X10 switchable USB power cable Windows 10, 32 and 64 bits, via website Windows 10, 32 and 64 bits, via website Windows 10 and Linux, via website Instrument manual and software manual Specifications 25 26 Chapter 8 If you have any suggestions and/or remarks regarding this manual, please contact: TiePie engineering Koperslagersstraat 37 8601 WL SNEEK The Netherlands Tel.: Fax: E-mail: Site: +31 515 415 416 +31 515 418 819 support@tiepie.nl www.tiepie.com TiePie engineering Handyscope HS4 instrument manual revision 2.45, February 2024