User Manual for Newman Medical models including: SimpleABI Cuff Link Systems, Cuff Link Systems, Link Systems, Systems
simpleABI Step-by-Step Tutorials | Newman Medical
File Info : application/pdf, 1 Pages, 431.72KB
DocumentDocumentHow to properly use and hold a Doppler Doppler techniques: Hold the probe like you would a pencil, close to the end. Move the probe back and forth laterally over the artery to obtain the best signal. Support the probe with your hand resting on the patient so that the probe does not move as the cuff is inflated and deflated. One of the keys to a successful exam is being able to keep the probe in place as you inflate and deflate the cuffs. If the probe moves, you may not be able to hear the Doppler sounds return and may have to repeat the inflation. Doppler held 90° to vessel = bad angle Skin Doppler Strong Doppler signal, but "lousy" Doppler shift. Distance to intersect artery Skin ARTERY Doppler ARTERY Good Doppler frequency shift (due to the small angle). Weak Doppler signal due to the distance the signal must travel. FLOW DIRECTION Transmit frequency Transmit frequency Returning freq. Returning freq. flow away from Doppler beam, returning frequency is lower. flow towards Doppler beam, returning frequency is higher. Newman Medical | Toll-Free: 800-267-5549 | www.newman-medical.com | info@newman-medical.com